METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL PRESSURE AND COMPOSITIONAL ANALYSIS
|
|
- Dorcas Arline Clarke
- 5 years ago
- Views:
Transcription
1 METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL PRESSURE AND COMPOSITIONAL ANALYSIS Robert E. Ellefson REVac Consulting Dayton, OH USA ERMP IND12 QMS Workshop Bled, Slovenia April 10-13, 2012
2 What do Industrial & R/D Users Need and Expect? Partial Pressures or Composition of Process Gas Expect Stable Operation to Trust the Data How do they Know it is Stable? Accurate? Process Control practices typically call for a Verification of Performance Test Stand Calibration of RGA may Qualify the MS In Situ Measurement of Reference Mixture is needed to provide Verification that the RGA is In Control
3 Logic Plan: Define a MS Measurement and Calibration Method for a Process or Vacuum System Application Vacuum: PVD- UHV -XHV P Process < 0.01 mb Molecular Flow Process Gas Species (Any Condensibles?) CVD, Etch, ALD, Degas 0.01 mb < P Process < 100 mb Vis., Trans. or Molecular Process Gas Species (Any Condensibles?) Atmospheric Processes 100 mb < P Process < 2 bar Viscous Flow Process Gas Species (Any Condensibles?) Sampling Design Molecular Flow Direct Insertion of QMS (Isolation Valve) Bakeout Provision Viscous, Transition or Molecular Flow [P Process ] Orifice / Small Tube toms (Heated Inlet and QMS) Viscous Flow Capillary Sampling with Sample Pump and Molecular Leak to MS (Heated Inlet and QMS) Select and Qualify MS Calibration Design Open Ion Source (CIS) High ev: ev High Emission: e.g. 2 ma Mass Range for Species Detector: FC & EM(?) Demo: Linearity, S vs t, A-B Std Mix: Process Species Reservoir: PV for 1 year Cal Flow Std: 1x10-4 mb-l/s Flow Mix into CIS Closed Ion Source (CIS) Low ev: e.g. 40 ev Low Emission: e.g. 0.2 ma Mass Range for Species Detector: FC & EM(?) Demo: Linearity, S vs t, A-B Std Mix: Process Species Volume: 5-10 L@ 1 Atm Capillary Sampling of Standard Mixture Quality Assurance Use Std Mix for QA Data Log Data & Plot PP i vs Time Plot X i =PP i / Sum PPi vs Time Establish Action Limits for Sens or Gain
4 RGA Conditions for Accurate Partial Pressure and Compositional Analysis begins in the Ion Source Vacuum System Base Pressure PP Measurements Requires Sensitivity OIS: ev e - ; 1-2 ma Emission Process Gas Composition allows Lower Sensitivity operation which minimizes Fragmentation CIS: ev e - ; 200 μa Emission Ion Extraction, Focus and Ion Energy Potentials affect Space Charge, Ion Residence Times and Linearity
5 Evidence of Gas Scattering Loss at Higher Pressure is seen as Ions Traverse the Mass Analyzer to the Detector. XPR Correction for Gas Scattering is I = I meas e KP ; Gives PP i s
6 The Potential Well formed by the Ionizing Electron Beam Lengthens Ion Residence Time (Longer Path) in Ion Source CIS equipotentials 1V well (SIMION) 10V 70V 75V 79V 79.9V E-Beam Well Depth For the Geometry of this CIS: V well = i e / (V e ) 1/2 V well (40eV/200uA) = V V well (70eV/2000uA) = V Note for a given i e the well gets deeper for lower V e (slower electrons; higher ρ e ) R Ellefson and M Vollero, AVS-57, 2010
7 The Same Open Ion Source can be Linear or Non-Linear depending on Operating Potentials
8 O 2 and Ar Partial Pressures and (O 2 /O 2 +Ar)*100 Ratio The Ar and O 2 sensitivity decreases in OIS with time due to O 2 Oxidation of Ion Source Grids changing surface potentials. However Ratio O 2 /[O 2 +Ar] remains Nearly Constant Conditioning RGA exposed to pure Ar for many hours Reduces the Surface Oxide: H(Diff) + OH(S) H 2 O Ar + (E-9 A) Ratio = % Exposure to Ar +3% O 2 changes the sensitivity for both Ar and O 2 by surface re-oxidation with a time constant of ~ 2hrs O 2 + (E-10 A) Practical Solution for PVD Process Control: Where PP (O 2 ) = R(O 2 )*P(CDG) R (O 2 ) = O 2 /[O 2 +Ar] Recommended Cal Reference: 3% O 2 in Ar Time, minutes Data provided by Wm Sproul
9 Examples of In Situ RGA Calibration Methods In Situ Calibration for an Open Ion Source (OIS) RGA In Situ Calibration for a Closed Ion Source (CIS) RGA System Calibration using a Fixed Reference Supply with Sampling equivalent to the Process Sampling Portable Calibration Reference Source introduced directly into the CIS
10 In Situ Calibration: A Reference Pressure & Composition is established at the IG and RGA when the Valve to the Gas Mixture is Open Pumping is provided by the Vacuum System A Calibrated Fixed-Flow Rate produces a Reproducible Pressure and Composition at IG and RGA Ionizers The Pressure is P cal = Q cal / C cal where C Cal can be Calculated from Geometry. Mixture Composition chosen for Application Vacuum System must tolerate the Q cal Flow Rate
11 For UHV and XHV Systems, a Lower Calibration Pressure can be created with a Q vs P Fill Calibration Q = A P fill 2 The Plot shows Flow Rate, Q; The Pressure generated in the example is ~ Q/10. P Fill can be adjusted without altering Composition using the Gas Pipettes. Suggested UHV Composition: 90% H 2 ; 9% CO; 1% CO 2
12 A Two-Stage Pressure Reduction with multiple probe types enables sampling at a representative point over a wide range of process pressures Process Gas Species in CIS reflect Process: Molecular Flow into CIS; Molecular Flow out Calibration Reference Mix Species are Altered by Mass: Viscous Flow into CIS; Molecular Flow out (Addressed Later)
13 Ar-40 Sensitivity (A/Torr) Sensitivity Monitoring Data shows Shifts in Sensitivity Related to Process Changes 3.4E-04 MASS(40) Sensitivity vs Days for CIS2 [W Filament; 40eV/200uA] Reference Gas: VTI Flow Std with INFICON Mix 3.2E E % 2-Sigma 2.8E-04 Event: Accidental Air Venting of CIS2 Vacuum due to Power Failure 2.6E E E % 2-Sigma 2.0E Time (Days)
14 Remote Process Monitoring: D/T/He Gas Mixture Compositional Measurements Process Room m Analytical Lab T P T V Process Capillaries D/T Pump Tritium Glove Box Standards Capillaries CDG << 1 mb Sample Pressure 3 L D/T Pump Tritium Fume Hood Magnetic Sector Mass Spectrometer D/T/He - 1 D/T/He - 2 Tritium Glove Box D/T/He - 3 D/T/He - 4 Sampling Capillaries m long Capillaries to Standard Mixtures (Similar) Capillary Sample Purged 3 5 Capillary Vol s ~ 0.1 mb sample P in 3 L volume Species deplete through Molecular Leak Record time for peak for t=0 Composition Analysis Time: ~ 20 minutes Qualify MS with Std Mix each AM + Run Time Enables < 0.5% relative accuracy for D/T/He-3
15 The D/T/He-3 mixtures were used to Assess Accuracy and do Daily Qualification of the MS s A daily routine for operators was to run an analysis of a known mixture to check performance Call Customers We are up today so they could run a Test, etc. If a MS was Out of Control, a re-run was done to note permanent shifts in calibration (due to power failures, etc)
16 Can we define a Multi-Application, Portable InSitu Calibration Gas Source? Vacuum Process 2x10-7 mbar Process Q=2x10-4 mbar-l/s 1x10-5 mbar Q=2x10-4 mbar-l/s 2x10-5 mbar 2x10-4 mbar RGA Calibration Source Fixed Flow Rate Gas Source Pumping by Process through a Fixed Conductance (~ 10 L/s) Process Monitor Calibration Gas Source Fixed Flow Rate Gas Source Pumping by Monitor Vacuum System through the Closed Ion Source (~ 1 L/s)
17 Portable Calibration Sources have been used for many years Crimped-Capillary Leak Standards (Boeckmann, et al.) VTI Positive Shut off Flow Standards (Traceable to NIST) Sandia Flow Standards [Chamberlin, et al. JVST A7 (1989)] High accuracy primary flow references (Traceable to NIST) INFICON has used their compact Reference Gas Source for InSitu Calibration of OIS and CIS products for over 10 years SS Frit Flow Element For Stable Flow Low dead volume valve to minimize turn-on bursts Measured Flow Rate stated on Label (+ 10% of Rate) Predictable Flow Rate (Depletion < 10%/Yr normal use) Gas mixture composition is constant (viscous flow) Gas fill Pressure < 2.8 bar (non-compressed gas for simple shipping) Flow Rate can be independently Certified at a Standards Lab
18 INFICON Calibration Reference Source 4-VCR Male Connection Air Operated Valve 38mm- (+VCR) x 270mm 132 cc Volume
19 Flow Through Element goes as P fill 2 RGA Calibration Pressure vs Fill Pressure of Reference 1.6E-05 RGA Calibration Pressure: Q (P) / C [mbar] 1.4E E E E E E E-06 Vacuum Process Q=2x10-4 mbar-l/s PP(Ar) = Q(P) / C Q(P) / C = A P 2 0.0E Fill Pressure (Ar) [bar]
20 P Cal [Ar-40] (mbar) RGA Calibration Pressure Depletion is Predictable with Time of Flow 2.5E-05 Expected 10% Depletion in 1 year of automated use (5 minutes of flow/day ) 2.0E-05 P Cal [Ar-40] (t) 1.5E E E-06 Model for Depletion of Calibration Reference Pressure: P Cal (t) = P cal (0) / [1+Q o t/p o V] Q o = 2.8x10-4 T-L/s Fill Parameters: P o = 2053 Torr [2.7 bar] V = L 0.0E Elapsed Flow Time (Days)
21 Ion Current (A - Corrected to Ref mix) & Ratio 40/ Composition Ratio of 50/50 He/Ar Mix does not change with 30% Depletion of Fill Gas Flow Rate Depletion: Ar-40 (Data Sensitivity Corrected) Flow Rate Depletion: He-4 (Data Sensitivity Corrected) Ratio 40/4 = t Time (Days)
22 Pcal produced from Cal Reference Source changes with Ambient Temperature by %/ o C 4.26E-09 Ion Current [Ar-40] (A) from P Cal 4.24E E E E E E E E-09 Temperature Coefficient of Flow Rate 0.15% / o C 4.08E E Average Temperature [Valve & Tank] (C)
23 INFICON Gas Mixture options allow choice of gas pumped by the process Component Argon (99.995%) Ar with 5% Impurities PVD Mix Argon % 95 % Balance H % 200 ppm He % 1000 ppm N % 50 ppm CO 2 20 ppm Kr - 1.0% 1000 ppm Xe % 1000 ppm
24 Composition of the Gas Mixture in the Ionizer is altered by the RGA s molecular flow pumping The Mixture is in viscous flow from the Calibration Reference Source. The composition entering the vacuum chamber is the stated Mixture. The partial flow rate q i of a species is q i (in) = X i (Ref) Q o (mbar-l/s) The partial flow rate out of the chamber depends on the mass of the species, M i and the pumping system conductance, C N2 at the ion source: q i (out) = PP i C i = PP i C N2 [28 / M i ] 1/2 But q i (in) = q i (out) So at the ionizer: PP i (Ion Source) = [M i / 28] 1/2 X i (Ref) Q o / C N2 From which a Sensitivity Factor can be calculated: SF i (mb/a) = PP i (Ion Source) / [ I i Interference Contributions ]
25 Composition of Mixture at the MS ionizer shows depleted light gases and enhance heavy gases CEM 1:1:1 Ar-He-N2 Mix Component Tank Mix X i -Ion Source Ar He N
26 Ion Current (A) Mass Spectrum of a PVD Mixture Calibration Reference Source 1.0E E E ppm H Ar ++ Ar + INFICON PVD Mixture Calibration Reference Source Flow RATE 1x o C Transpector CIS: 70eV/2000uA/EM 1.0E ppm He Xe ++ Kr + 1.0E E E Mass
27 Mass Scale Calibration with Calibration Reference Source H 2 & He 36 Ar & 38 Ar Kr
28 Electron Multiplier Gain Adjustment Adjust RGA EM (HV) to give Ion Current Level: e.g.: P Cal = 2x10-5 mbar S FC = 2x10-4 A/mbar Gain = 200 Species Abundance Ion current (EM) Ar ppm 8.0x10-7 A Ar ppm 1.5x10-7 A Ar % Off Scale 1% N2 1% 2.4x10-7 A Measure Gain of the EM at a value of High Voltage Gain = I(Ar-38: EM) I(Ar-38: FC)
29 Ion Current (A) CIS Total Pressure (mb) Sensitivity and Gain can be quickly measured with sequential FC/EM scans with Selected Ion Monitoring of Ar species. 1.0E E-06 Pressure: Calibration Reference ON Ar-40 EM (Off Scale) 1.0E E E E E-08 S(Ar-40) = 1.1E-5 A/mb 1.0E E-10 Ar-40 FC Gain = Ar-36 EM Ar-38 EM 1.0E E E-11 Ar-36 FC Gain = E E-12 Ar-38 FC 1.0E E-13 FC EM FC 00: : : : : : : : : : :26.4 Time (m:s) 1.0E-10
30 Ar-40 Sensitivity (Amp/Torr) Sensitivity Monitoring Data shows Shifts in Sensitivity Related to the Sensor MASS(40) Sensitivity Vs Time For CIS2 (W filament; 40eV/200uA) Reference Gas: Inficon/VTI Calibration Standard 2.60E E E E E-04 2-Sigma Deviation 5% of Average over 15 days 2-Sigma: 2% over 7 days 5% Shift in Average Sensitivity on Day E E E E Time (Days) 2-Sigma: 3% over 8 days
31 Composition Calculations: Accurate Compositional Analysis is the Goal for Process Monitoring QMS measures the Ion Current vs Time for the Gas being sampled Partial Pressures of Species Present are Calculated Ion Currents with Corrections for Interferences, and Sensitivities Mol-% Composition is defined as X i (mol-%) = 100 * PP i / PP j For Quality Assurance, Accuracy is measured with known mixtures: ΔX i = X i - X i (Std) [for Major Components] 8 Capillaries P process = Atm i.d. = 0.25 mm L = 1.5 m 35μ Leak CIS 100 amu QMS w FC TMP Drag Diaphragm Pump
32 Ion Currents vs Time from 2 Gas Streams How do we relate Ion Current to Partial Pressure? Calibration
33 Ionization Cross Sections(Å 2 ) differ substantially with e - Energy Ion Gauges use 150 ev INFICON OIS uses 105 ev Most OIS and CIS s use 70 ev for high Sensitivity Some OIS and CIS use 40 ev for reducing Fragmentation and Multi-Charge peaks e.g. [Ar ++ ] IG Sensitivity ratios to N2 are at best a guess for RGAs Sensitivity Measurements are Required for Species of interest for a RGA Molecule e-cross Sections at Atom e-cross Sections inferred from Wutz Handbuch Vacuumtechnik Ed , Bild 12.42
34 Partial Pressures are Calculated from Ion Current in Real Time using a Previous Calibration Partial Pressure = Sens Factor * Trans(M) * [Ion Current Interferences] PP i = SF i * M Z * [ I i - Σ k i A ik * I k ]
35 The Result of the Partial Pressure Calculation is Real Time Display of the Partial Pressures and Clearing Times when Gas Comp Changes 4m-% H2 in Ar Air 4m-% H2 in Ar
36 More informative is the Composition of the Gas Species showing Stable Composition after Conditioning the Inlet System 4m-% H2 in Ar Air 4m-% H2 in Ar 5 Min Elapsed Time Composition Table at the Time with the Yellow Cursor
37 Safe Hydrogen [4 mol-% H 2 in Ar] is a Low-Cost Reference Mixture Initial Calibration for H 2 was done with 2.0% H 2 in Synthetic Air (N 2 & O 2 ) in August 2011 Data for 4% H 2 in Ar (at right) began a month later. Data shown are values from a single scan sampled at random during the sampling of the 4 mol-% H 2 in Ar from the Trend Data over 12 days The Mean Value for H 2 over 12 days is Mol-%
38 Sampling Room Air is a Convenient Reference Gas Capillary Sampling and CIS Analysis of Room Air over 160 Days Calibration for this data done in August 2011; No Sensitivity Drift Trend Mean Value of N 2, O 2, Ar is within 1 σ Data displayed is a randomly selected, single measurement value of X i O 2 and Ar have σ ~ 2 %-RSD H 2 O and CO 2 are Biased due to Seasonal ΔRH and Breathing in Lab Air Components Altered by Seasons and Breathing
39 Now to Summarize
40 In Situ Calibration by Two Methods Provides Accurate Partial Pressures and/or Composition to a RGA / IG Process Room m Analytical Lab T P T V Process Capillaries D/T Pump Tritium Glove Box Standards Capillaries CDG 3 L << 1 mb Sample Pressure D/T Pump Tritium Fume Hood Magnetic Sector Mass Spectrometer D/T/He - 1 D/T/He - 2 Tritium Glove Box D/T/He - 3 D/T/He - 4 XHV UHV Low Pressure Processes Atmospheric Processes Calibrated Flow Reference Gas Mixtures Deliver Known PP i s and X i s to the Ionization Region Method of Introduction is Designed for a Particular Application Regular Measurements for PP i s and X i s Plotted over Time Reveals Accuracy, Drift Trends, Time for Recalibration and Data for Sensitivity or Gain Changes Flow Standards should be designed to last > 1 year or Flow Rate Re-Calculated Based on Time Measured Use A Tank of Reference Gas is Sampled by the Same Means as Process Gas Certified Gas Mixture should be Pertinent to the Process Do QA Measures and Plot Difference Results Reference Gas Supply should last > 1 Year in planned useage
41 Summary: Operation of RGAs for Analytical Measurements Choose Ion Source potentials that produce linear stable operation Electron Energy, Emission Current, Extraction/Focus Voltages and Ion Energy Choose a Calibration Reference Gas for Calibration Verification (QA) For UHV/XHV: Consider a Viscous Leak with Mix Pressure defining Flow Rate. Needs Flow Rate vs P mix Calibration + Conductance (Calc/Measure); or Total RGA For UHV: Fixed Flow Rate Calibration Reference Mix (INFICON or VTI) + OIS 0.01 mb<p process <100 mbar: Calibration Reference Mix into CIS; Conductance of CIS In each application, calculate Composition in Ion Source given Viscous in Molecular Out For Atm Processes: Tank with Gas Mix representing the Process; Sample with Capillary [In this case, the composition in Ion Source is the same as Tank due to Mol. In - Mol. Out] Calculate Sensitivities from Partial Pressures at Ion Source from Cal Gases Initial Calibration requires known gases in the list of Species to be analyzed. Determine Fragmentation Factors during these initial calibrations If a gas species not available, estimates of the Sens can come from Gauge Constants or σ(e) Choosing a model for Ion Transmission, e.g. T(M) = M z where 0 < Z < 1 to make SF σ(e) For High Ion Source Pressure, X i -mol-% more stable than PP i for species QA: Plot differences, X i X i (Std) vs Time to monitor when to Re-Calibrate
42 Acknowledgements INFICON RGA Engineers (Developed the Calibration Reference Source) Tim Karandy, Michael Vollero, Peter Schubert, Ken Rosys, Lou Frees Mound Technical Solutions, Inc (MTS) Doug McClelland, Steve Huff AVS/IUVSTA Colleagues EMRP IND12 Chairs for the Invitation to Speak Janez Šetina and Karl Jousten And to All of You for Listening: Hvala (Thank You)
Vacuum. Methods for in situ QMS calibration for partial pressure and composition analysis. Robert E. Ellefson * abstract
Vacuum 101 (2014) 42e42 Contents lists available at ScienceDirect Vacuum journal homepage: www.elsevier.com/locate/vacuum Methods for in situ QMS calibration for partial pressure and composition analysis
More informationBEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION
BEST KNOWN METHODS Transpector XPR3 Gas Analysis System DESCRIPTION The Transpector XPR3 is a third-generation, quadrupole-based residual gas analyzer that operates at PVD process pressures and is the
More informationPractical approach and problems in in-situ RGA calibration
Practical approach and problems in in-situ RGA calibration Oleg Malyshev and Keith Middleman Vacuum Science Group, ASTeC Accelerator Science and Technology Centre STFC Daresbury Laboratory UK Workshop
More informationTotal Pressure Measurement Using Residual Gas Analyzers
Total Pressure Measurement Using Residual Gas Analyzers M. Maskell, Old Dominion University; G. Myneni, P. Adderley, Jefferson Lab; G. Brucker, Stanford Research Systems; C. Day, Forschungszentrum, Karlsruhe
More informationRecent steps toward traceability for partial pressure and outgassing measurements Karl Jousten, PTB, Berlin
Recent steps toward traceability for partial pressure and outgassing measurements Karl Jousten, PTB, Berlin 1. Problems of calibrating QMS 3. Toward traceability in EMRP IND12 2. Toward standardization
More informationApplications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples
Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples Outline of Presentation Introduction Process Gas Analysis Magnetic Sector Mass Spectrometer Standard
More informationChoosing a Quadrupole Gas Analyzer Application Note #9
Choosing a Quadrupole Gas Analyzer Application Note #9 Modern-day contamination control requirements for gas phase processes are constantly pushing the limits of performance of quadrupole gas analyzers.
More informationat NIST: ultra-low outgassing rates
Traceable outgassing measurements at NIST: ultra-low outgassing rates and new materials James Fedchak Thermodynamic Metrology Group Sensor Science Division Physical Measurement Laboratory Vacuum Project:
More informationResist round robin ELETTRA
Resist round robin outgassing @ ELETTRA Trieste - Italy M. Bertolo, P. Lacovig, L.Rumiz, D. Giuressi, R. Sergo, A. Stolfa Experimental setup (1) 100 mm linear translator Sample manipulator Mask manipulator
More informationISO/TC 112/WG 2 N 49. ISO TS QMS WD v6. Date of document Expected action Comment Due Date Background
ISO/TC 112/WG 2 N 49 ISO/TC 112/WG 2 Vacuum instrumentation E-mail of Secretary: juergen.eisenreich_vdma.org@livelink.din.de Secretariat: DIN ISO TS 20175 QMS WD v6 Date of document 2016-05-02 Expected
More informationThis educational seminar discusses creating, measuring, and troubleshooting Rough Vacuum.
This educational seminar discusses creating, measuring, and troubleshooting Rough Vacuum. Specifically, today s talk will cover: Brief review of Vacuum Fundamentals Applications Using Rough Vacuum Rough
More informationQuadruple mass spectrometers (transducer type)
Quadruple mass spectrometers that meet the need for sensitive analysis and low outgassing which were specifically designed to be controlled by PCs Quadruple mass spectrometers (transducer type) M-070QA-TDF,
More informationThe new PTB standard for dynamic vacuum pressures
The new PTB standard for dynamic vacuum pressures Karl Jousten, Physikalisch-Technische Bundesanstalt, Berlin, Germany 1. Idea and design of the standard 2. Development 3. Pressure approximation and experiments
More informationLow-Power Atmospheric Gas Sampling System Based on ART MS Sensor and NEG-Ion Pump
8 th Harsh Environment Mass Spectrometry Wokshop Low-Power Atmospheric Gas Sampling System Based on ART MS Sensor and NEG-Ion Pump Sept 22, 2011 Gerardo A. Brucker and G. Jeffery Rathbone Granville-Phillips
More informationGun Vacuum Performance Measurements February 2003
JLAB-TN-03-013 Gun Vacuum Performance Measurements February 2003 Marcy Stutzman Abstract. The vacuum in the polarized photoguns at Jefferson Lab is crucial to long lifetime operation of the polarized source.
More informationThe Helium Leak Detector
The Helium Leak Detector Helium Leak Detector Main Components The main components of a helium leak detector are: 1. The analyzed, which enables to separate the tracer gas from other gases inside leak detector.
More informationHigh Performance Mass Spectrometers
High Performance Mass Spectrometers For Vacuum and Semiconductor Process Monitoring Detailed product information / introduction MOCVD semiconductor processing Applications: XXCVD/MOCVD XXPECVD XXALD XXSputtering
More informationOMNISTAR / THERMOSTAR. The efficient solution for gas analysis. Intelligent software. Wide range of applications.
OMNISTAR / THERMOSTAR The efficient solution for gas analysis. Intelligent software. Wide range of applications. OMNISTAR /THERMOSTAR The efficient solution for gas analysis. Intelligent software. Wide
More informationThe Principles of Vacuum Technology
The Principles of Vacuum Technology Vacuum Terminology Vacuum units Vacuum regimes How to measure vacuum. Gauge designs. How to create vacuum Pump classifications and designs UHV compatibility considerations
More informationGPP model GPPT model GPC model CLP model
CHOOSING A CALIBRATOR FOR HELIUM LEAK DETECTORS Unique Vacuum Solutions deals with four models of Calibrated Leak Standards for calibrating Helium Leak Detectors and Leak Testing Systems so that one is
More informationVIC offers a variety of Calibrated Gas Leaks
CALIBRATED GAS LEAKS VIC offers a high quality line of calibrated gas leaks traceable to National Institute of Standards Technology (NIST). Calibrated gas leaks are used to provide a controlled delivery
More informationThe Broadest Gauging Line in the Industry
The Broadest Gauging Line in the Industry Vacuum Measurement Varian s broad line of vacuum gauge controllers and gauge tubes are the most reliable, accurate, and economical means of measuring, monitoring,
More informationLeaks, contamination and closed loop control how RGAs make coating processes more profitable
Leaks, contamination and closed loop control how RGAs make coating processes more profitable Backup: What s an RGA Residual Gas Analyzer An instrument that is measuring the gas composition in a chamber
More informationIntroductory Lab: Vacuum Methods
Introductory Lab: Vacuum Methods Experiments in Modern Physics (P451) In this lab you will become familiar with the various components of the lab vacuum system. There are many books on this topic one of
More informationCommercial Practice Test Method Internal Vapor Analysis of Hermetic Devices
Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices Oneida Research Services, Inc. 8282 Halsey Road Whitesboro, NY 13492 Phone: (315) 736-5480 FAX: (315) 736-9321 1.0 Purpose The
More informationA Mass Spectrometry Method for Water Vapour Transmission Rate (WVTR) testing to 10-6 gm/m 2 /day, utilising Deuterium Oxide
A Mass Spectrometry Method for Water Vapour Transmission Rate (WVTR) testing to 10-6 gm/m 2 /day, utilising Deuterium Oxide Stephen Swann Development Scientist www.vgs-plasticelectronics.com WVTR Illustration
More informationMeasuring mass flows in hermetically sealed MEMs & MOEMs to ensure device reliability
Measuring mass flows in hermetically sealed MEMs & MOEMs to ensure device reliability R. C. Kullberg*, D. J. Rossiter** Oneida Research Services, Inc., 8282 Halsey Road, Whitesboro, NY13492 Abstract Many
More informationOXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:
OXY Integral INTERCON ENTERPRISES INC Tel: 800 665 6655 Fax: 604 946 5340 E-Mail: sales@intercononline.com Internet: www.intercononline.com Manual Integral 2006 1 INDEX 2-3 PREFACE 4 INTRODUCTION 5 Principle
More informationLeak Detection. Kirsten Zapfe, DESY CAS Vacuum in Accelerators Platja D Aro, May 21, Special thanks to S. Holm, N. Mildner and A.
Leak Detection, DESY CAS Vacuum in Accelerators Platja D Aro, May 21, 2006 Special thanks to S. Holm, N. Mildner and A. Wagner Outline Introduction Leak Rate and Leak Types Leak Detection Methods Helium
More informationTechnical Specifications of Hydrogen Isotope Handling and Recovery System
SECTION - C TECHNICAL SPECIFICATIONS OF STORES AND DRAWINGS. Technical Specifications of Hydrogen Isotope Handling and Recovery System INSTITUTE FOR PLASMA RESEARCH GANDHINAGAR, GUJARAT 382428 ANNEXURE-I
More informationDevelopment of a high vacuum sample preparation system for helium mass spectrometer
Journal of Physics: Conference Series Development of a high vacuum sample preparation system for helium mass spectrometer To cite this article: P Kumar et al 2012 J. Phys.: Conf. Ser. 390 012056 View the
More informationThe use of remote plasma emission spectroscopy as an alternative method of gas sensing for direct monitoring of vacuum processes
Enlighten Conference and Vacuum Symposium 12 th October 2017 The use of remote plasma emission spectroscopy as an alternative method of gas sensing for direct monitoring of vacuum processes Joe Brindley,
More informationVacuum Science Techniques and Applications Dan Dessau Adv. Lab 2007
Vacuum Science Techniques and Applications Dan Dessau Adv. Lab 2007 Vacuum increases the mean-free-path of gas molecules. Vacuum prevents chemical reaction. Vacuum removes contaminants from surfaces. Vacuum
More informationA NOVEL SENSOR USING REMOTE PLASMA EMISSION SPECTROSCOPY FOR MONITORING AND CONTROL OF VACUUM WEB COATING PROCESSES
A NOVEL SENSOR USING REMOTE PLASMA EMISSION SPECTROSCOPY FOR MONITORING AND CONTROL OF VACUUM WEB COATING PROCESSES F. Papa 1, J. Brindley 2, T. Williams 2, B. Daniel 2, V. Bellido-Gonzalez 2, Dermot Monaghan
More informationWhite Paper. Chemical Sensor vs NDIR - Overview: NDIR Technology:
Title: Comparison of Chemical Sensor and NDIR Technologies TSN Number: 25 File:\\MII- SRV1\Metron\Bridge_Analyzers\Customer_Service_Documentation\White_Papers\25_G en EGA NDIR vs Chemical Sensor.docx Created
More informationApplication Note AN-107
SPEC Sensor TM Characterization & Calibration Considerations Scope This document is provided to describe the considerations needed to characterize, calibrate, verify and validate the measurement performance
More informationHiden QIC Series Gas Analysers Real-time gas analysis systems
Hiden QIC Series Gas Analysers Real-time gas analysis systems vacuum analysis surface science plasma diagnostics gas analysis Application Focussed Options Multistream valves 20, 40 and 80 way versions
More informationResidual Gas Analysis Systems for Industry
HIDEN RC SYSTEMS QUADRUPOLE MASS SPECTROMETERS FOR RGA, GAS ANALYSIS AND PROCESS MONITORING The HAL RC systems are designed for RGA, gas analysis and process monitoring applications including leak detection,
More informationGenerating Calibration Gas Standards
Technical Note 1001 Metronics Inc. Generating Calibration Gas Standards with Dynacal Permeation Devices Permeation devices provide an excellent method of producing known gas concentrations in the PPM and
More informationAGILENT TURBOMOLECULAR PUMPS MEASUREMENT METHODOLOGY
AGILENT TURBOMOLECULAR PUMPS MEASUREMENT METHODOLOGY Agilent Vacuum Products: Our Committment to Quality Agilent s commitment to quality and Measurement s philosophy Reference Standards (PNEUROP, ISO)
More informationSupporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial
Supporting Information for Micro-Collection of Gases in a Capillary Tube: Preservation of Spatial and Temporal Resolution AUTHOR INFORMATION Kristin A. Herrmann Favela, 1 * Pieter Tans, 2 Thomas Jaeckle,
More informationLINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER
LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER Michael T Tang, Ph.D. Grace Feng Greg Merideth Rui Huang Matheson Gas Applied Lab
More informationChapter 6 Leak detection
Chapter 6 Leak detection 1 2 1. Leak rate q L = V p t mbar l s -1 A leak rate of 1 mbar l s -1 exists in a closed vessel having a volume of 1 liter when the pressure increases by 1 mbar within one second,
More informationMass Spec will not Autotune
Mass Spec will not Autotune Applies to 5973A/N MSD What could be the problem? There could be several things that would cause your Mass Spec not to Autotune. The most common, easily corrected Autotune problems
More informationHIGH ACCURACY MULTI-GAS MONITORING USING AUTOMATED SELF- CALIBRATION
HIGH ACCURACY MULTI-GAS MONITORING USING AUTOMATED SELF- CALIBRATION SCENTROID is the global leader in sensory technology providing intelligent environmental solutions across a wide range of industries
More informationLCLS Heavy Met Outgassing Tests * K. I. Kishiyama, D. M. Behne Lawrence Livermore National Laboratory
UCRL-TR-223160 LCLS-TN-06-12 July 12, 2006 LCLS Heavy Met Outgassing Tests * K. I. Kishiyama, D. M. Behne Lawrence Livermore National Laboratory Abstract A Heavy Met that is 95% tungsten, 3% nickel and
More informationImproving the Measurement Accuracy of Water Partial Pressure Using the Major Constituent Analyzer
Improving the Measurement Accuracy of Water Partial Pressure Using the Major Constituent Analyzer Ben D. Gardner, Phillip M. Erwin, Wai Tak Lee, Amber M. Tissandier, Souzan M. Thoresen Hamilton Sundstrand
More informationGas Leak Measurement of TRT Barrel Module
DUKHEP01-07-02 Gas Leak Measurement of TRT Barrel Module Shannon Watson, Seog Oh Department of Physics Duke University Durham, NC, 27708, USA 1. Introduction One of the requirements for the TRT module
More informationAPPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT
APPLICATION NOTE Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT The INFICON 3000 Micro GC provides two fast and accurate solutions for the analysis of coal mine gas components.
More informationDEVICES FOR FIELD DETERMINATION OF WATER VAPOR IN NATURAL GAS Betsy Murphy MNM Enterprises 801 N. Riverside Drive Fort Worth, Texas 76111
INTRODUCTION Water vapor in natural gas has more than a substantial effect on the quality of the gas stream. Without quality measurement of water vapor the gas is basically not saleable. Contracts are
More informationCanadian Light Source Inc. Vacuum Component Leak Test Technical Procedure
Canadian Light Source Inc. Vacuum Component Leak Test Technical Procedure CLSI Document 8.7.33.2 Rev. 2 Date: 2006-03-08 Copyright 2006, Canadian Light Source Inc. This document is the property of Canadian
More information1)! DO NOT PROCEED BEYOND THIS MARK
Operating Instructions for X-ray Photoelectron Spectrometer: Physical Electronics Model 555 XPS/AES (John H. Thomas, III, Ph.D., Electron Spectroscopy) Sample Insertion: figure 1. Sample insertion rod
More informationLOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary
ADH 1/7/014 LOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary This experiment will introduce you to the kinetic properties of low-pressure gases. You will make observations on the
More informationContinuous Gas Analysis In situ laser gas analyzers TÜV and MCERTS add-on for LDS 6 operating instructions Compact Operating Instructions
Continuous Gas Analysis In situ laser gas analyzers Compact Operating Instructions 1 LDS 6 TÜV/MCERTS certified analyzers for NH3/H2O, NH3 and H2O 1.1 Operation requirements Please observe the following:
More informationMajor Design Features
Major Design Features 3Flex Features Design Benefits Discussion 316 SS fittings and pneumatically actuated, hard-seal valves Provides virtually leak-free gas management with the lowest outgassing rate
More informationTitle: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)
Procedure No: SOP-026 Revision No: 1.0 January 24, 2011 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting and to meet
More informationOperating Procedures for Metal Evaporator I
Operating Procedures for Metal Evaporator I Metal Evaporator I is intended as a tool and a training device. Understanding the operation of this equipment should give you a basic knowledge of vacuum and
More informationVacuum System at IUAC
Vacuum System at IUAC A. Mandal Inter-University Accelerator Centre, Aruna Asaf Ali Marg, Post Box - 10502, New Delhi 110067, Fax:011-26893666. email: mandal@iuac.res.in Abstract: Vacuum technology is
More informationThis test shall be carried out on all vehicles equipped with open type traction batteries.
5.4. Determination of hydrogen emissions page 1 RESS-6-15 5.4.1. This test shall be carried out on all vehicles equipped with open type traction batteries. 5.4.2. The test shall be conducted following
More informationLOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12
LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/ This experiment will introduce you to the kinetic properties of low-pressure gases. You will make observations on the rates with which selected
More informationLogin to ilab Kiosk. Revised 05/22/2018. Load your sample:
Login to ilab Kiosk Load your sample: 1. Check: The analysis chamber pressure is
More informationOutgas of Methane from NEG Coating
WS63: SURFACE PHENOMENA LIMITING PRESSURE IN VACUUM SYSTEM Outgas of Methane from NEG Coating Chia-Mu Cheng ( 鄭家沐 ) Vacuum Group NSRRC, Taiwan 18 /09/2010 e-mail : chiamucheng@nsrrc.org.tw 1 Application
More informationD R A F T. Operating Procedures for the NPDGamma Liquid Hydrogen Target at the BL 13. Version 1.00
D R A F T Operating Procedures for the NPDGamma Liquid Hydrogen Target at the BL 13 Version 1.00 October 06, 2010 Operating Procedures for the NPDGamma Liquid Hydrogen Target V1.00 6/10/10 2 Table of Content
More informationTitle: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator
Procedure No: SOP-034 Revision No: 1.0 Revised Dec. 29, 2010 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE This procedure is intended to describe the operations of the Dasibi model 5008 calibrator. The Dasibi
More informationLaboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.
Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions
More informationContamination Removal of EUVL masks and optics using 13.5-nm and 172-nm radiation
Contamination Removal of EUVL masks and optics using 13.5-nm and 172-nm radiation Takeo Watanabe, Kazuhiro Hamamoto and Hiroo Kinoshita University of Hyogo IEUVI Resist & Contamination TWG Oct.19, 2006
More informationZebron Gas Management
Zebron www.phenomenex.com/gasmanagement What are the key benefits of Zebron filters and traps? The high adsorption capacity of the Zebron Filters removes contaminants from mobile phase gases, reducing
More informationCharacterizing a Cu/Mn Alloy for Extracting Oxygen from Inert Gas. Streams
Characterizing a Cu/Mn Alloy for Extracting Oxygen from Inert Gas Streams Grace Lenhard Prattsburgh Central School LLE advisors: Walter Shmayda and Matthew Sharpe Laboratory for Laser Energetics University
More informationTitle: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator
Procedure No: SOP-029 Revision No: 1.1 (December 29, 2010) Page No.: 1 of 7 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting, air quality
More informationChemistry Chapter 11 Test Review
Chemistry Chapter 11 Test Review Multiple Choice Identify the choice that best completes the statement or answers the question. 1. Pressure is the force per unit a. volume. c. length. b. surface area.
More informationAPPLICATION NOTE. GC Integrated Permeation Device
GC Integrated Permeation Device GC-integrated Design PPB to PPM Level Calibration No Cylinders or Regulators Required Option for Dual Ovens Cost Effective, Safe, Clean, Flexible, NIST Traceable Keywords:
More informationA Journal of Practical and Useful Vacuum Technology. By Phil Danielson
A Journal of Practical and Useful Vacuum Technology From By Phil Danielson Thermal Conductivity Gauges Thermal conductivity pressure gauges are extremely common in vacuum technology, but an understanding
More informationAppendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation
Page 1 of 19 : SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor Description and Principle of Operation The photoacoustic multi-gas monitor (INNOVA 1412, Innova AirTech Instruments, Denmark) is a highly
More informationPURE SUBSTANCE. Nitrogen and gaseous air are pure substances.
CLASS Third Units PURE SUBSTANCE Pure substance: A substance that has a fixed chemical composition throughout. Air is a mixture of several gases, but it is considered to be a pure substance. Nitrogen and
More informationMETHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER
1250 METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER 1.0 Scope and Application. 1.1 Analytes. Analyte CAS No. Sensitivity Total Organic Compounds N/A
More informationPressure & Vacuum Measurement. Series 370. Solutions. Stabil-Ion Vacuum gauge and Controller
w w w. m k s i n s t. c o m Pressure & Vacuum Measurement Solutions Series 370 Stabil-Ion Vacuum gauge and Controller Features & Benefits All-metal, rack-mount controller for Stabil-Ion and Convectron
More informationThree Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor
Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor Anwar Sutan, Metco Services Ltd. Charles Johnson, Metco Services Ltd. Jason Laidlaw,
More informationThe Gas Attenuator of FLASH
INTRODUCTION The Gas Attenuator of FLASH K. Tiedtke, N. von Bargen, M. Hesse, U. Jastrow, U. Hahn The experimental hall of the FLASH user facility is located approximately 30m behind the last dipole magnet
More informationA novel class of Non Evaporable Getter (NEG) alloys with lower hydrogen equilibrium pressure and reduced outgassing properties
A novel class of Non Evaporable Getter (NEG) alloys with lower hydrogen equilibrium pressure and reduced outgassing properties P. Manini, FabrizioSiviero, A. Gallitognotta, L. Caruso, T. Porcelli,G. Bongiorno,
More informationGranville-Phillips Series 370 Stabil-Ion Vacuum Gauge Controller
Granville-Phillips Series 370 Stabil-Ion Vacuum Gauge Controller All-metal, rack-mount controller for Stabil-Ion and Convectron vacuum gauges is noise-immune and CE compliant Accurate vacuum pressure measurement
More information29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies
IMPROVED CALIBRATION AND OPERATION OF A STABLE-GAS QUANTIFICATION MANIFOLD AS PART OF A RADIOXENON COLLECTION SYSTEM ABSTRACT Richard M. Williams, James C. Hayes, and Randy R. Kirkham Pacific Northwest
More informationA Reliable and Tracer Gas Independent Leak Detector for Food Packages
19 th World Conference on Non-Destructive Testing 2016 A Reliable and Tracer Gas Independent Leak Detector for Food Packages Silvio DECKER 1 1 INFICON GmbH, Köln, Germany Contact e-mail: Silvio.Decker@inficon.com-
More informationLab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow
Georgia Institute of Technology School of Earth and Atmospheric Sciences EAS 4641, Spring 2008 Lab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow Purpose of Lab 3: 1) To gain a
More informationGFM. Typical Aluminum GFM Mass Flow Meter NIST MASS FLOW METERS. Principles of Operation. Design Features
Design Features Rigid metallic construction. Maximum pressure of 1000 psig (70 bars). Leak integrity 1 x 10-7 of helium. NIST traceable certification. Built-in tiltable LCD readout. 0-5 Vdc and 4-20 ma
More informationLaboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.
Laboratory Hardware Custom Gas Chromatography Solutions Custom solutions for your analytical needs. Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions for advanced chromatography
More informationChapter 10: Gases. Characteristics of Gases
Chapter 10: Gases Learning Outcomes: Calculate pressure and convert between pressure units with an emphasis on torr and atmospheres. Calculate P, V, n, or T using the ideal-gas equation. Explain how the
More informationReal-time Analysis of Industrial Gases with Online Near- Infrared Spectroscopy
Real-time Analysis of Industrial Gases with Online Near- Infrared Branch 5 Petrochemical, Biofuels Keywords Process, NIR, Gases, Gas Cell, Ethylene, Carbon Dioxide, Propane, Butane, Acetylene Summary Near-infrared
More informationFlare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1
Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost HERMAN HOLM SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 4C CONFERENCE APRIL 2018 SAN ANTONIO, TX SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC
More informationRevision 2013 Vacuum Technology 1-3 day Good Vacuum Practice 1 Day Course Outline
Revision 2013 Vacuum Technology 1-3 day Good Vacuum Practice 1 Day Course Outline This training course outline is intended to cover the following: Introduction to vacuum Measurement Lubricated rotary pumps
More informationInstruction Manual. Model DSP-Ex Portable Dewpoint Meter. Alpha Moisture Systems Alpha House 96 City Road Bradford BD8 8ES England
Model DSP-Ex Portable Dewpoint Meter Instruction Manual Alpha Moisture Systems Alpha House 96 City Road Bradford BD8 8ES England Tel: +44 1274 733100 Fax: +44 1274 733200 Email: info@amsytems.co.uk Web:
More informationDevelopment of Cleaning and Processing Techniques for UHV/XHV at Daresbury Laboratory Keith Middleman
Development of Cleaning and Processing Techniques for UHV/XHV at Daresbury Laboratory Keith Middleman ASTeC Vacuum Science Group STFC Daresbury Laboratory Accelerators + Vacuum Particle accelerators come
More informationProudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Varian 310 310 LC/MS Manual System INCLUDES: 310-MS LC/MS triple quadrople
More informationThe Experts in Vacuum Solutions
By Woodrow Farrow - Reprinted with permission from Specialty Gas Report Vacuum: a space that is relatively empty of matter, especially when that void has been created through artificial means. The earth
More informationName /74. MULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question.
Ch 11 Gases STUDY GUIDE Accelerated Chemistry SCANTRON Name /74 MULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question. 1) Which of the following statements
More informationValidation of Custody Transfer Metering Skid at Site After Laboratory Proving
flotek.g 2017- Innovative Solutions in Flow Measurement and Control - Oil, Water and Gas August 28-30, 2017, FCRI, Palakkad, Kerala, India Validation of Custody Transfer Metering Skid at Site After Laboratory
More informationGA / 3.02 TL 4 TL 6. Cat. No Calibrated leaks GA / 3.02 TL 4 TL 6. Cat. No
O P E R A T I N G I N S T R U C T I O N S GA 10.18 / 3.02 TL 4 TL 6 Cat. No. 155 65 155 66 Calibrated leaks O P E R A T I N G I N S T R U C T I O N S GA 10.18 / 3.02 TL 4 TL 6 Cat. No. 155 65 155 66 Calibrated
More informationVacuum Simulations of the KATRIN Experiment
Vacuum Simulations of the KATRIN Experiment Marcel Krause (KATRIN collaboration) 31.03.2014 KIT University of the State of Baden-Wuerttemberg and National Research Center of the Helmholtz Association www.kit.edu
More informationVOC Emissions From the Drying of Red Alder Lumber
VOC Emissions From the Drying of Red Alder Lumber Report to Cascade Hardwoods Report by Michael R. Milota Department of Wood Science and Engineering Oregon State University Corvallis, OR 97331 February
More informationPHOENIX L300i Best-in-class Helium Leak Detectors
180.85.02 Best-in-class Helium Leak Detectors The perfect Helium leak detector matching your demands The helium leak detector opens up new dimensions of productivity and reliability for your applications.
More informationIntroduction of Vacuum Science & Technology. Diffusion pumps used on the Calutron mass spectrometers during the Manhattan Project.
Introduction of Vacuum Science & Technology Diffusion pumps used on the Calutron mass spectrometers during the Manhattan Project. 1 What is a vacuum? 760 mm Hg Vacuum ATM A vacuum is defined as less than
More information