Outgas of Methane from NEG Coating
|
|
- Beverley Oliver
- 5 years ago
- Views:
Transcription
1 WS63: SURFACE PHENOMENA LIMITING PRESSURE IN VACUUM SYSTEM Outgas of Methane from NEG Coating Chia-Mu Cheng ( 鄭家沐 ) Vacuum Group NSRRC, Taiwan 18 /09/ chiamucheng@nsrrc.org.tw 1
2 Application of NEG coating for synchrotron light source NEG(non-evaporable getter) has been developed at CERN about thirty years ago. It can provide large pumping speed of H 2 and CO after activation for UHV systems. A NEG coated ID chamber for large synchrotron light source provide linear pumping and large area surface pumping efficiency. It can solve the problem of limited conductance. The Ti-Zr-V coating with lower activation temperature has been developed for the aluminum chambers will be feasible for the ID chambers. 2
3 Preparation of the Al chamber for NEG coating, Al(NEG) A6063T5 extruded pipe Chemical cleaning (CERN recipe) TIG welding (in clean room ) Leakage checking NGL cleaning (CERN) Vacuum baking (degassing) N 2 purging and Packing NEG coating (SAES Getters) 3
4 Surface characteristics for NEG film Concentration of NEG film Elmt Spect. Element Atomic Type % % Al K ED 0.13* 0.89* 10 μm Ti K ED V K ED Zr L ED Total μm 4
5 NEG coated Aluminum-SIMS (negative ion) C - CH - O - OH - F - C - 2 C 2 H - C 2 H O - 2 HO - 2 Cl - Cl - AlO There is hydrocarbon,h 2 O and contamination (F and Cl) on the surface. 5
6 NEG coated Aluminum-SIMS (positive ion) Na + Al + K + Ti + V + TiO + VO + Zr ? 92 6
7 AL(NEG) chamber was installed in TLS 19B(PSD) beam line for SR photon exposure Beam Line Test Chamber (2m) NEG-coated Elec. Feedthrough - I (PEY) RGA Safety shutter Absorber IG (P1) IG (P2) Bottle neck Conductance, C ~ 5 L/s E I θ SR beam energy (1.5GeV) e-beam current (300mA) photon span (0.0033rad) Pumping chambers - IG (Po) X-Y Slits - Define horiz. photon span of 2.42 mrad yield of photon stimulated desorption η =ΔQ/ΔN ΔQ =3.5x10 19 x C (ΔP 1 -ΔP 0 ) ΔN =8.05x10 17 x E x I x θ/2π 7
8 The typical baking and activation process for Al(NEG) chamber , 24h NEG - CH Other - CH , 8h h 5h 13h 17h 37h 40h 42h 8
9 History of photon exposure for Al(NEG) chamber at 19B beamline It motivated us to study the outgas of NEG! 9
10 To study the outgas of NEG film We found outgas of CH 4 and Kr from NEG chamber during activation and build up. The D 2 O has been exposed into NEG chamber for inspecting the surface phenomena. It has been found the outgas of CmDxHy compound molecules in the NEG chamber exposed to D 2 O. It shows D 2 O has been dissociated and recombined with C and H atoms on the surface. 10
11 Build up of AL(NEG) Chamber with D 2 O exposure, (a) before and (b) after SR exposure (a) The compound molecules of CD x, CD x H y, C m D x, and C m D x H y are observed for the AL(NEG) chamber exposed with D 2 O and after activation. (b) The compound molecules of C m D x and C m D x H y are reduced for the AL(NEG) chamber after the photon exposure. (c) The beam cleaning effect addresses the desorption of the compound molecules are from the NEG film. CD x & CD x H y C m D x & C m D x H y (a) CD x & CD x H y C m D x & C m D x H y D 2 O dissociated and outgas of C m D x, C m D x H y complex produced. SR beam cleaning (b) Ref : G. Y. Hsiung, C. M. Cheng, C. Y. Yang, C.K. Chan and J.R. Chen; AIP vol. 1099, p.17(2009) 11
12 SS(NEG) Chambers with D 2 O and H 2 O Exposure (a) D 2 O exposed on the activated SS(NEG) chamber The pressure build up curves for SS(NEG) chamber is similar with AL(NEG) chamber. The D 2 O has been dissociated and the outgases of high mass compound molecules of C m D x H y and C m D x are formed and desorbed. Outgas of D 2 O is not significant. (b) H 2 O exposed on the baked SS(NEG) chamber Hydrocarbon molecules, CH 4 and C m H y, are formed and desorbed. (a) SS(NEG) / D 2 O Exposure (b) SS(NEG) / H 2 O Exposure CH 4 C m H y Dissociation of D 2 O on the NEG film is confirmed Ref : G. Y. Hsiung, C. M. Cheng, C. Y. Yang, C.K. Chan and J.R. Chen; AIP vol. 1099, p.17(2009) 12
13 Al- and SS-chambers (non-coated) with D 2 O Exposure D 2 O exposure to the Al- and SSchambers without NEG-coating The pressure build up curves for the Al- and SS- chambers without NEG coating are similar. The outgas species contain the D 2 O and hydrocarbon (CH 4, C m H y ), except C m D x H y compounds, are measured. No evidence shows D 2 O is dissociated and combined with hydrocarbons. The outgas of hydrocarbon is existed. Only D 2 O and C m H y are measured. Ref : G. Y. Hsiung, C. M. Cheng, C. Y. Yang, C.K. Chan and J.R. Chen; AIP vol. 1099, p.17(2009) 13
14 PM-IRRAS Spectra for Al and Al- NEG samples without exposure (a), with D 2 O exposure (b), and with H 2 O exposure (c). (1) (2) (3) (4) (a) Peaks of (1) C-H stretching (2929, 2857), (2) C=O stretching (1741), (3) C=C bending (1650), (4) C-H bending (1460, 1406), (5) O-D stretching (2577), (6) O-H stretching (3456) are identified for the Al samples. Peaks (1) ~ (4) recognize the molecules adsorbed on the Al surfaces contains CH 4 and CO. (1) (5) (3) (2) (4) (b) Peaks (5) and (6) recognize the residence of D 2 O and H 2 O on the surface after exposure, respectively. There is not any peak measured for the Al-NEG surface. It is suspected (a) the strong binding energy on the adsorbed molecules from the NEG material or (b) D 2 O have been dissociated. (6) (1) (2) (3) (4) 14 (c)
15 NEG coating for Aluminum ID chamber at NSRRC DC sputtering coating parameters Discharge gas Kr Working pressure Torr voltage -800 V current ~ 18 ma Magnetic field (center) 200 Gauss Depositing rate 0.07 nm/s 15
16 SEM and EDS analysis of NEG film X25000 X5000 EDS analysis Ti 30%±5% V 35%±5% Zr 35%±5% 16
17 Activation for AL(NEG) chamber The ultimate pressure is 8e-11 Torr. The major outgas is hydrogen and methane. 17
18 Exposure test with ION-NEG pump Two sets of starcell Ion pump combined with Pd/NEG chamber, 100CF and 150CF, called ION-NEG provided by Varian are for the test. Pd/NEG chamber was coating Pd film on NEG coated chamber. Pd film can prevent NEG film from oxygen contamination. 200nm Pd 850nm NEG (TiVZr) Stainless steel chamber 18
19 Experimental The ION-NEG pump have been baked and activated at 200 for 8h to reach ultimate pressure. The outgassing rate of ION-NEG was measured by pressure build up with TMP isolated and ion pump turned off. 19
20 starcell IP + 100CF Pd/NEG build up for 3 min Build Before up 3min build up The major residual gas is hydrogen, H The major outgas 2 O, CO and CO is hydrogen 2. and methane. 20
21 Starcell IP + 150CF Pd/NEG build up for 3 min Before Build up build 3minup The major residual gas is hydrogen, H 2 O, CO and CO 2. The major outgas is only methane. 21
22 Compare with pressure and outgassing rate Starcell IP + 100CF Pd/NEG system Starcell IP+ 150CF Pd/NEG system Ultimate pressure (Torr) 1.3e-10 (1.4e-10) Outgassing rate (Torr.L/sec) 3.8e-9 (7e-9) Outgassing rate of H 2 (Torr.L/sec) 3.7e-9 (6.8e-9) Outgassing rate of CH 4 (Torr.L/sec) 3e-11 (8e-11) 4e-11 3e-10 5e e-10 The major residual gas is hydrogen(> 90%). A pressure bellow 1e-10 Torr has been achieved for 150CF ION-NEG after baking. The 100CF Pd/NEG chamber seems to be no working. The property of ION-NEG still not stable. 22
23 Does Pd film protect NEG film? In order to test the function of Pd film, we expose D 2 O to 150 CF Pd/NEG chamber after it been activated. To study the outgas from Pd film surface. If there is no C m H x D y compound which is like only NEG coating chamber. It was believed that Pd film can isolate NEG from D 2 O. 23
24 150CF Pd/NEG chamber before exposed D 2 O H 2 H 2 O CH 4 CO C m H x There is lots of hydrocarbon before exposing D 2 O. 24
25 150CF Pd/NEG chamber after exposed D 2 O DHO H DH DO D 2 O 2 D2 CO CO 2 C m H x There is no C m H x D y compound. Pd film can prevent NEG film from reacting with D 2 O. 25
26 Conclusion D 2 O or H 2 O are dissociated and recombined with C to produce CH 4 on NEG surface. There is no C, H and O related bonding of Al(NEG) sample on the PM-IRRAS spectra. The ION-NEG (Pd/NEG chamber with IP) can reach lower ultimate pressure Pd/NEG chamber will not produce C m D x H y outgas after exposed D 2 O. 26
27 Thank you for attention 27
IUVSTA WORKSHOP WS-63 Surface Phenomena Limiting Pressure in Vacuum Systems REPORT TO IUVSTA. prepared by JOSÉ L. DE SEGOVIA
IUVSTA WORKSHOP WS-63 Surface Phenomena Limiting Pressure in Vacuum Systems INTRODUCTION REPORT TO IUVSTA prepared by JOSÉ L. DE SEGOVIA on behalf of the SPANISH VACUUM SOCIETY Madrid,February 1, 2011
More informationA novel class of Non Evaporable Getter (NEG) alloys with lower hydrogen equilibrium pressure and reduced outgassing properties
A novel class of Non Evaporable Getter (NEG) alloys with lower hydrogen equilibrium pressure and reduced outgassing properties P. Manini, FabrizioSiviero, A. Gallitognotta, L. Caruso, T. Porcelli,G. Bongiorno,
More informationAchieving Ultra-High Vacuum without (in-situ) Bakeout
Achieving Ultra-High Vacuum without (in-situ) Bakeout Matthew Cox Diamond Light Source, UK WS 63 Ávila September 2010 1 What is Diamond Light Source? The UK national synchrotron facility 3 rd generation
More informationResist round robin ELETTRA
Resist round robin outgassing @ ELETTRA Trieste - Italy M. Bertolo, P. Lacovig, L.Rumiz, D. Giuressi, R. Sergo, A. Stolfa Experimental setup (1) 100 mm linear translator Sample manipulator Mask manipulator
More informationMASSIVE TITANIUM SUBLIMATION PUMPING IN THE CESR INTERACTION REGION *
MASSIVE TITANIUM SUBLIMATION PUMPING IN THE CESR INTERACTION REGION * N.B. Mistry, R. Kersevan and Yulin Li, Laboratory of Nuclear Studies, Cornell University, Ithaca, NY 153 USA Abstract * The residual
More informationExtreme High Vacuum: The Need, Production, and Measurement
Extreme High Vacuum: The Need, Production, and Measurement, Philip Adderley, Matt Poelker (Jefferson Lab) Polarized Electron Gun Group Newport News, Virginia Run by JSA for the US DOE What is XHV Extreme
More informationContamination Removal of EUVL masks and optics using 13.5-nm and 172-nm radiation
Contamination Removal of EUVL masks and optics using 13.5-nm and 172-nm radiation Takeo Watanabe, Kazuhiro Hamamoto and Hiroo Kinoshita University of Hyogo IEUVI Resist & Contamination TWG Oct.19, 2006
More informationGun Vacuum Performance Measurements February 2003
JLAB-TN-03-013 Gun Vacuum Performance Measurements February 2003 Marcy Stutzman Abstract. The vacuum in the polarized photoguns at Jefferson Lab is crucial to long lifetime operation of the polarized source.
More informationNumerical Tools for LCLS-II Vacuum Systems
Numerical Tools for LCLS-II Vacuum Systems G. Lanza, D. Gill, SLAC National Accelerator Laboratory AVS, November 2 nd 2017 Doc. N. SLAC-PUB-17166 Work described in this presentation is supported by the
More informationPractical approach and problems in in-situ RGA calibration
Practical approach and problems in in-situ RGA calibration Oleg Malyshev and Keith Middleman Vacuum Science Group, ASTeC Accelerator Science and Technology Centre STFC Daresbury Laboratory UK Workshop
More informationVacuum acceptance tests for particle accelerator equipment
Vacuum acceptance tests for particle accelerator equipment Giuseppe Bregliozzi TE-VSC - CERN Geneva Vacuum for Particle Accelerators Glumslöv, Sweden, 6-16 June, 2017 2 Outline 1. The CERN accelerator
More informationUltrahigh Vacuum in Superconducting Synchrotrons A.Smirnov, JINR, Dubna, Russia
Ultrahigh Vacuum in Superconducting Synchrotrons A.Smirnov, JINR, Dubna, Russia Topics: - Pumping & outgasing of hydrogen - Cryogenic pumps & non evaporable getters for UHV - Superconducting accelerators:
More informationIntroduction of Vacuum Science & Technology. Diffusion pumps used on the Calutron mass spectrometers during the Manhattan Project.
Introduction of Vacuum Science & Technology Diffusion pumps used on the Calutron mass spectrometers during the Manhattan Project. 1 What is a vacuum? 760 mm Hg Vacuum ATM A vacuum is defined as less than
More informationVacuum Science and Technology in Accelerators
Vacuum Science and Technology in Accelerators Ron Reid Consultant ASTeC Vacuum Science Group (ron.reid@stfc.ac.uk) Session 5 Materials Properties Relevant to Vacuum Aims To understand which mechanical
More informationGo/no-go test for the acceptance of components for unbaked vacuum sectors
Go/no-go test for the acceptance of components for unbaked vacuum sectors Paolo Chiggiato (Thanks to Sophie Meunier for the preparation of the presentation) Introduction CERN policy: All vacuum components
More informationThe Principles of Vacuum Technology
The Principles of Vacuum Technology Vacuum Terminology Vacuum units Vacuum regimes How to measure vacuum. Gauge designs. How to create vacuum Pump classifications and designs UHV compatibility considerations
More informationDevelopment of Cleaning and Processing Techniques for UHV/XHV at Daresbury Laboratory Keith Middleman
Development of Cleaning and Processing Techniques for UHV/XHV at Daresbury Laboratory Keith Middleman ASTeC Vacuum Science Group STFC Daresbury Laboratory Accelerators + Vacuum Particle accelerators come
More informationGraphite and C-C materials for UHV applications
Graphite and C-C materials for UHV applications Prepared by J.M. JIMENEZ AT/VAC/SL Section Graphite Outgassing Measurements for UHV Applications Main topics What did we learn from past experiences? Static
More informationIonic outgassing from photoresist compositions upon irradiation at 13.5 nm
Ionic outgassing from photoresist compositions upon irradiation at 13.5 nm Grace H. Ho, Chih- Jen Liu, Chih- Han Yen, Ming- Hsuan Ho, Shih-Yu Wu, and Yu-Hsian Shih Department of Applied Chemistry National
More informationVacuum Science Techniques and Applications Dan Dessau Adv. Lab 2007
Vacuum Science Techniques and Applications Dan Dessau Adv. Lab 2007 Vacuum increases the mean-free-path of gas molecules. Vacuum prevents chemical reaction. Vacuum removes contaminants from surfaces. Vacuum
More informationDevelopment of a high vacuum sample preparation system for helium mass spectrometer
Journal of Physics: Conference Series Development of a high vacuum sample preparation system for helium mass spectrometer To cite this article: P Kumar et al 2012 J. Phys.: Conf. Ser. 390 012056 View the
More informationLeak Detection. Kirsten Zapfe, DESY CAS Vacuum in Accelerators Platja D Aro, May 21, Special thanks to S. Holm, N. Mildner and A.
Leak Detection, DESY CAS Vacuum in Accelerators Platja D Aro, May 21, 2006 Special thanks to S. Holm, N. Mildner and A. Wagner Outline Introduction Leak Rate and Leak Types Leak Detection Methods Helium
More informationPlasma Sources and Feedback Control in Pretreatment Web Coating Applications
Plasma Sources and Feedback Control in Pretreatment Web Coating Applications Joseph Brindley, Benoit Daniel, Victor Bellido-Gonzalez, Dermot Monaghan Gencoa Ltd., Physics Rd, L24 9HP Liverpool, UK (+44)
More informationThe use of remote plasma emission spectroscopy as an alternative method of gas sensing for direct monitoring of vacuum processes
Enlighten Conference and Vacuum Symposium 12 th October 2017 The use of remote plasma emission spectroscopy as an alternative method of gas sensing for direct monitoring of vacuum processes Joe Brindley,
More informationA NOVEL SENSOR USING REMOTE PLASMA EMISSION SPECTROSCOPY FOR MONITORING AND CONTROL OF VACUUM WEB COATING PROCESSES
A NOVEL SENSOR USING REMOTE PLASMA EMISSION SPECTROSCOPY FOR MONITORING AND CONTROL OF VACUUM WEB COATING PROCESSES F. Papa 1, J. Brindley 2, T. Williams 2, B. Daniel 2, V. Bellido-Gonzalez 2, Dermot Monaghan
More informationDr. José Miguel JIMENEZ CERN, Technology Department Head of the Vacuum, Surfaces and Coatings group
Vacuum Systems Outline Vacuum in Particle Accelerators Vacuum Basics Instrumentation & Pumping The LHC Case Vacuum in Particle Accelerators Why is Vacuum needed in particle accelerators? Vacuum aims to
More informationANL/APS/TB-16 ADVANCED PHOTON SOURCE ACCELERATOR ULTRAHIGH VACUUM GUIDE. Chian Liu and John Noonan
ANL/APS/TB-16 ADVANCED PHOTON SOURCE ACCELERATOR ULTRAHIGH VACUUM GUIDE Chian Liu and John Noonan CONTENTS 1. Ultrahigh Vacuum Overview..... 1 1.1 Vacuum..... 1 1.2 Sources of Residual Gas... 2 1.3 Material
More informationTotal Pressure Measurement Using Residual Gas Analyzers
Total Pressure Measurement Using Residual Gas Analyzers M. Maskell, Old Dominion University; G. Myneni, P. Adderley, Jefferson Lab; G. Brucker, Stanford Research Systems; C. Day, Forschungszentrum, Karlsruhe
More informationOutgassing, photoablation and photoionization of organic materials by the electron-impact and photon-impact methods
Outgassing, photoablation and photoionization of organic materials by the electron-impact and photon-impact methods Grace H. Ho*, Yen-H. Huang, Hung-M. Lin, Chih-H. Shao, Chen-Y. Yeh, and Jia-J. Sung Department
More informationSAES Packaging service. making innovation happen, together
SAES Packaging service MEMS Packaging service description Facility and operations Capabilities and future prospective Choose your type of service : Fast prototyping Process developments Packaging foundry
More informationLeaks, contamination and closed loop control how RGAs make coating processes more profitable
Leaks, contamination and closed loop control how RGAs make coating processes more profitable Backup: What s an RGA Residual Gas Analyzer An instrument that is measuring the gas composition in a chamber
More informationHow To Use Getters and Getter Pumps
A Journal of Practical and Useful Vacuum Technology From By Phil Danielson How To Use Getters and Getter Pumps Gettering is a vacuum pumping technology that has been with us, in many forms, for almost
More informationat NIST: ultra-low outgassing rates
Traceable outgassing measurements at NIST: ultra-low outgassing rates and new materials James Fedchak Thermodynamic Metrology Group Sensor Science Division Physical Measurement Laboratory Vacuum Project:
More informationChapter 8: Cryo-sorption pumps
Chapter 8: Cryo-sorption pumps Cryo-sorption pumps offer a clean, quiet, safe, vibration free and inexpensive way to rough pump a vacuum system. They are often used on vacuum systems that are sensitive
More informationDC gun high voltage conditioning with Krypton gas Carlos Hernandez-Garcia October 2008
DC gun high voltage conditioning with Krypton gas Carlos Hernandez-Garcia October 2008 Index Background 1 High voltage processing with Helium gas 2 High voltage processing with Krypton gas 3 Some thoughts
More informationThis educational seminar discusses creating, measuring, and troubleshooting Rough Vacuum.
This educational seminar discusses creating, measuring, and troubleshooting Rough Vacuum. Specifically, today s talk will cover: Brief review of Vacuum Fundamentals Applications Using Rough Vacuum Rough
More informationFURNACE N 2 DOPING TREATMENTS AT FERMILAB
Proceedings of SRF215, Whistler, BC, Canada FURNACE DOPING TREATMENTS AT FERMILAB M. Merio, A.C. Crawford, A. Grassellino, A. Rowe, M. Wong, Fermilab, Batavia, IL, USA M. Checchin, M. Martinello, Fermilab,
More informationLow-Power Atmospheric Gas Sampling System Based on ART MS Sensor and NEG-Ion Pump
8 th Harsh Environment Mass Spectrometry Wokshop Low-Power Atmospheric Gas Sampling System Based on ART MS Sensor and NEG-Ion Pump Sept 22, 2011 Gerardo A. Brucker and G. Jeffery Rathbone Granville-Phillips
More informationTechnical Specifications of Hydrogen Isotope Handling and Recovery System
SECTION - C TECHNICAL SPECIFICATIONS OF STORES AND DRAWINGS. Technical Specifications of Hydrogen Isotope Handling and Recovery System INSTITUTE FOR PLASMA RESEARCH GANDHINAGAR, GUJARAT 382428 ANNEXURE-I
More informationChoosing a Quadrupole Gas Analyzer Application Note #9
Choosing a Quadrupole Gas Analyzer Application Note #9 Modern-day contamination control requirements for gas phase processes are constantly pushing the limits of performance of quadrupole gas analyzers.
More informationThis educational seminar discusses creating, measuring, and troubleshooting Ultra-High Vacuum ( Torr) systems.
This educational seminar discusses creating, measuring, and troubleshooting UltraHigh Vacuum ( 10 8 10 12 Torr) systems. Specifically, today s talk will cover: Brief review of High Vacuum (Characteristics,
More informationUser s Manual Differential Pump
User s Manual Differential Pump Model DP-03 XIA LLC 31057 Genstar Road Hayward, CA 94544 USA Tel: (510)-401-5760; Fax: (510)-401-5761 http://www.xia.com/ Information furnished by XIA LLC (XIA) is believed
More informationDetermination of the Vacuum System for R2R by using a New Pumping Simulation Program
Determination of the Vacuum System for R2R by using a New Pumping Simulation Program AIMCAL WCHC Conference Memphis Oct 10 th, Peter Klingner, Dr. Stefan Lausberg, Dr. Marlis Sydow, Dr. Derek Corcoran
More informationVacuum Simulations of the KATRIN Experiment
Vacuum Simulations of the KATRIN Experiment Marcel Krause (KATRIN collaboration) 31.03.2014 KIT University of the State of Baden-Wuerttemberg and National Research Center of the Helmholtz Association www.kit.edu
More informationIssues Confronting Vacuum System Design for e+e- Storage Rings
1991 IEEE. Personal use of this material is permitted. However, permission to reprint/republish this material for advertising or promotional purposes or for creating new collective works for resale or
More informationShort Introduction to Cryo-Pumps and Refrigerators. Dr Graham Rogers Leybold UK Ltd.
Short Introduction to Cryo-Pumps and Refrigerators Dr Graham Rogers Leybold UK Ltd. Cryo Pumps Topics Physical principles of cryo-pumping Design and control of modern refrigerator cryo-pumps UHV and XHV
More informationCanadian Light Source Inc. Vacuum Component Leak Test Technical Procedure
Canadian Light Source Inc. Vacuum Component Leak Test Technical Procedure CLSI Document 8.7.33.2 Rev. 2 Date: 2006-03-08 Copyright 2006, Canadian Light Source Inc. This document is the property of Canadian
More informationA Road Map to Extreme High Vacuum
A Road Map to Extreme High Vacuum Philip Adderley, Ganapati Myneni Jefferson Lab 12000 Jefferson Ave, Newport News, VA; 23606 adderley@jlab.org, rao@jlab.org Abstract. Ultimate pressure of a well-designed
More informationTests for High Voltage & Vacuum of the 2 nd 500 kv DC Gun at KEK
Tests for High Voltage & Vacuum of the 2 nd 500 kv DC Gun at KEK OLAV IV @NSRRC 3 rd April 2014 Masahiro Yamamoto KEK 2 nd DC-gun Development Team T. Miyajima, Y. Honda, T. Uchiyama, M. Kobayashi, Y. Saito,
More informationBEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION
BEST KNOWN METHODS Transpector XPR3 Gas Analysis System DESCRIPTION The Transpector XPR3 is a third-generation, quadrupole-based residual gas analyzer that operates at PVD process pressures and is the
More informationQuadruple mass spectrometers (transducer type)
Quadruple mass spectrometers that meet the need for sensitive analysis and low outgassing which were specifically designed to be controlled by PCs Quadruple mass spectrometers (transducer type) M-070QA-TDF,
More informationResidual Gas Analysis Systems for Industry
HIDEN RC SYSTEMS QUADRUPOLE MASS SPECTROMETERS FOR RGA, GAS ANALYSIS AND PROCESS MONITORING The HAL RC systems are designed for RGA, gas analysis and process monitoring applications including leak detection,
More informationAPPLICATION OF A CHANNEL PLATE IN FIELD-ION MICROSCOPY
R 718 Philips Res. Repts 25, 87-94,1970 APPLICATION OF A CHANNEL PLATE IN FIELD-ION MICROSCOPY by A. van OOSTROM Abstract Further progress has been made in applying a channel-plate image intensifier in
More informationHigh Performance Mass Spectrometers
High Performance Mass Spectrometers For Vacuum and Semiconductor Process Monitoring Detailed product information / introduction MOCVD semiconductor processing Applications: XXCVD/MOCVD XXPECVD XXALD XXSputtering
More informationApplication Note. ASTRON Remote Plasma Source Ignition Best Practices PROBLEM
PROBLEM The ASTRON family of remote plasma sources produce reactive gas species for semiconductor device fabrication applications. Reliable ignition of the plasma in the ASTRON depends strongly on the
More informationVacuum System at IUAC
Vacuum System at IUAC A. Mandal Inter-University Accelerator Centre, Aruna Asaf Ali Marg, Post Box - 10502, New Delhi 110067, Fax:011-26893666. email: mandal@iuac.res.in Abstract: Vacuum technology is
More informationLASER INTERFEROMETER GRAVITATIONAL WAVE OBSERVATORY - LIGO - CALIFORNIA INSTITUTE OF TECHNOLOGY MASSACHUSETTS INSTITUTE OF TECHNOLOGY
LASER INTERFEROMETER GRAVITATIONAL WAVE OBSERVATORY - LIGO - CALIFORNIA INSTITUTE OF TECHNOLOGY MASSACHUSETTS INSTITUTE OF TECHNOLOGY Document Type LIGO-T980008-01- W 2/18/98 Bake Oven Requirements and
More informationOutgassing. the hows and whys of measurement. Ron Reid ASTeC Vacuum Science Group STFC Daresbury Laboratory UK
Outgassing the hows and whys of measurement Ron Reid ASTeC Vacuum Science Group STFC Daresbury Laboratory UK Vacuum System Performance Vacuum Equation P = Q S Where P is the base pressure in the system
More informationINSTRUCTIONS. VP-100 Ion Pump. Model / R001
INSTRUCTIONS 87-900-130-01 (A) March 3, 2011 VP-100 Ion Pump Model 819-1490 / 819-1490R001 GENERAL INFORMATION This equipment is destined for use by professionals. The user should read this instruction
More informationBoth physical and chemical effects come into play either separately or in combination.
A Journal of Practical and Useful Vacuum Technology By Phil Danielson WHY CREATE A VACUUM? From Creating a vacuum has become a fairly common and often routine undertaking, but the commonness of the activity
More informationPlasma Cleaner. Yamato Scientific America. Contents. Innovating Science for Over 125 Years. Gas Plasma Dry Cleaner PDC200/210/510 PDC610G.
Yamato Scientific America Innovating Science for Over 125 Years Plasma Cleaner Contents Gas Plasma Dry Cleaner PDC200/210/510 PDC610G Gas Plasma Reactor 145 146 147 149 144 Gas Plasma Dry Cleaner Plasma
More informationThe Broadest Gauging Line in the Industry
The Broadest Gauging Line in the Industry Vacuum Measurement Varian s broad line of vacuum gauge controllers and gauge tubes are the most reliable, accurate, and economical means of measuring, monitoring,
More informationSHMS. Downstream Beamline. Large SHMS/HMS Angles (Config-1) Large SHMS / Min HMS Angle (Config-2) Dan Young Ext /26/2015 1
SHMS Dan Young Ext 5255 Downstream Beamline Large SHMS/HMS Angles (Config-1) Large SHMS / Min HMS Angle (Config-2) 1/26/2015 1 10 14.5 SHMS Beam Line Large SHMS/HMS Angles (Configuration-1) SHMS minimum
More informationThermo K-Alpha XPS Standard Operating Procedure
Thermo K-Alpha XPS Standard Operating Procedure Quick Guide Draft v.0.1 Procedure overview 1. Vent the loadlock 2. Secure your sample to the stage using clips, check the height of the final assembly. 3.
More informationSingle and Canted Undulator Front-Ends
Single and Canted Undulator Front-Ends 3470 Mainway, Burlington JUV Front End - Installation stage photo (ID11-1/ Duplex FE at CLS) Johnsen Ultravac Front Ends Technical Specifications Page 1 of 7 FRONT
More informationRemote Plasma Cleaning from a TEM Sample Holder with an Evactron De-Contaminator ABSTRACT
Remote Plasma Cleaning from a TEM Sample Holder with an Evactron De-Contaminator Christopher G. Morgan, David Varley, and Ronald Vane, XEI Scientific, Inc., 1755 E. Bayshore Rd., Suite 17, Redwood City,
More informationLCLS Heavy Met Outgassing Tests * K. I. Kishiyama, D. M. Behne Lawrence Livermore National Laboratory
UCRL-TR-223160 LCLS-TN-06-12 July 12, 2006 LCLS Heavy Met Outgassing Tests * K. I. Kishiyama, D. M. Behne Lawrence Livermore National Laboratory Abstract A Heavy Met that is 95% tungsten, 3% nickel and
More informationOOC RGA. Ion Pump 400 l/s. 8" ports with baked viton o-ring seals. Welsh rotary roughing pump 650 l/s manual isolation valve.
LASER INTERFEROMETER GRAVITATIONAL WAVE OBSERVATORY - LIGO - CALIFORNIA INSTITUTE OF TECHNOLOGY MASSACHUSETTS INSTITUTE OF TECHNOLOGY Technical Note LIGO-T000126-00- R 11/20/2000 Vacuum pressure at the
More informationGas Release from Solids, Materials for Vacuum and Leak Detector
VACUUM AND PLASMA SCIENCE AND TECHNOLOGY W9 Gas Release from Solids, Materials for Vacuum and Leak Detector Ming-Show Wong Nov. 2015 Gas Release from Solids Vaporization Thermal Desorption First Order
More informationOperating Procedures for Metal Evaporator I
Operating Procedures for Metal Evaporator I Metal Evaporator I is intended as a tool and a training device. Understanding the operation of this equipment should give you a basic knowledge of vacuum and
More informationVacuum and Surface Contamination Problems in Experiments with Ultracold Neutrons
Vacuum and Surface Contamination Problems in Experiments with Ultracold Neutrons Reinhold Henneck Paul-Scherrer-Institut, CH-5232 Villigen, Switzerland reinhold.henneck@psi.ch Ultracold neutrons (UCN)
More informationThe Transition Radiation Detector of the AMS-02 Experiment
The Transition Radiation Detector of the AMS-02 Experiment Simonetta Gentile Università di Roma La Sapienza, INFN (Italy) for: RWTH Aachen, MIT, KNU Daegu, IEKP Karlsruhe, Universita La Sapienza, & INFN
More informationKodial Zero Length Viewports
Kodial Zero Length Viewports Specification Seal Type Induction Maximum Temperature 350 C (KF versions 150 C) Kodial Zero Length Viewports Minimum Temperature minus 20 C Maximum Rate of Temperature Change
More informationAbstract The physics and practical consequences of thermal outgassing in vacuum chambers are described in this article.
111 THERMAL OUTGAING Karl Jousten Physikalisch-Technische Bundesanstalt, Berlin, Germany Abstract The physics and practical consequences of thermal outgassing in vacuum chambers are described in this article.
More informationVacuum Systems Operation EiC Training and Refresher Course 26 th April 2007
Vacuum Systems Operation EiC Training and Refresher Course 26 th April 2007 Contents 1. Introduction 2. Pumps and pumping on JET Primary Volumes Primary Pumps Secondary Pumps Vacuum Diagnostics Vessel
More informationUsage Policies Notebook for Xenon Difluoride (XeF 2 ) Isotropic Si Etch
Usage Policies Notebook for Xenon Difluoride (XeF 2 ) Isotropic Si Etch Revision date September 2014 2 Emergency Plan for XeF 2 Si Etcher Standard Operating Procedures for Emergencies Contact information
More informationMatching Vacuum Pump to Process
A Journal of Practical and Useful Vacuum Technology From By Phil Danielson Matching Vacuum Pump to Process What s the best vacuum pump? is one of those maddening questions that can only be answered with
More informationParticle Control under Wall Saturation in Long-pulse High-density H-mode Plasmas of JT-60U
Particle Control under Wall Saturation in Long-pulse High-density H-mode Plasmas of JT-6U H. Kubo, T. Nakano, N. Asakura, H. Takenaga, K. Tsuzuki, N. Oyama, H. Kawashima, K. Shimizu, H. Urano, K. Fujimoto,
More information[Hussain*, 4.(7): July, 2015] ISSN: (I2OR), Publication Impact Factor: 3.785
IJESRT INTERNATIONAL JOURNAL OF ENGINEERING SCIENCES & RESEARCH TECHNOLOGY DESIGN OF ULTRA HIGH VACUUM (UHV) SYSTEM TO EVALUATE OUTGASSING RATE OF THE MATERIALS TO ASSES THEIR UHV COMPATIBILITY Ibrahim
More informationCommercial Practice Test Method Internal Vapor Analysis of Hermetic Devices
Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices Oneida Research Services, Inc. 8282 Halsey Road Whitesboro, NY 13492 Phone: (315) 736-5480 FAX: (315) 736-9321 1.0 Purpose The
More informationLEAK-TIGHT WELDING EXPERIENCE FROM THE INDUSTRIAL ASSEMBLY OF THE LHC CRYOSTATS AT CERN
EUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH European Laboratory for Particle Physics Large Hadron Collider Project LHC Project Report 1072 LEAK-TIGHT WELDING EXPERIENCE FROM THE INDUSTRIAL ASSEMBLY OF THE
More informationA Mass Spectrometry Method for Water Vapour Transmission Rate (WVTR) testing to 10-6 gm/m 2 /day, utilising Deuterium Oxide
A Mass Spectrometry Method for Water Vapour Transmission Rate (WVTR) testing to 10-6 gm/m 2 /day, utilising Deuterium Oxide Stephen Swann Development Scientist www.vgs-plasticelectronics.com WVTR Illustration
More informationUHP - UHV APPLICATIONS
UHP - UHV APPLICATIONS High Performance Seals and Sealing Systems FOREWARD Today s sealing requirements for Ultra-High Purity and Ultra-High Vacuum are more demanding than ever before: Ultra-low Leak Rates
More informationE SC 412 Nanotechnology: Materials, Infrastructure, and Safety Wook Jun Nam The Pennsylvania State University
E SC 412 Nanotechnology: Materials, Infrastructure, and Safety Wook Jun Nam Unit 2 Infrastructure Lecture 6 More on Vacuum Systems, Part III Unit Outline Infrastructure * What is it? * Why do we need it
More informationThe Gas Attenuator of FLASH
INTRODUCTION The Gas Attenuator of FLASH K. Tiedtke, N. von Bargen, M. Hesse, U. Jastrow, U. Hahn The experimental hall of the FLASH user facility is located approximately 30m behind the last dipole magnet
More informationQuantification of Contamination Using Quartz Crystal Thickness Monitors
Quantification of Contamination Using Quartz Crystal Thickness Monitors Christopher G. Morgan, Mark M. Gleason and Ronald Vane XEI Scientific, Inc. 1755 E. Bayshore Rd., Suite 17 Redwood City, CA 9463
More informationMaterials and Properties IV Outgassing
Outline Materials and Properties IV Outgassing Paolo Chiggiato CERN Accelerator School (CAS) on Vacuum for Particle Accelerators June 18 th 2017 - Gas source and main features. - Order of magnitude. -
More informationCharacteristics and Production of Gas Sealed, Radiation Hard, Small Ionization Chambers
Characteristics and Production of Gas Sealed, Radiation Hard, Small Ionization Chambers Mayda Velasco 1 Gianni Tassotto 3 on behalf of the SIC team P. Ball 2, A.Dabrowsky 1, G. Graham 2, C. Kendziora 3,
More informationSPINE workshop Uppsala, Sweden, Jan 17-19, 2011 Spacecraft outgassing: Rosetta results
SPINE workshop Uppsala, Sweden, Jan 17-19, 2011 Spacecraft outgassing: Rosetta results B. Schläppi, K. Altwegg, H. Balsiger, M. Hässig, A. Jäckel, P. Wurz, B. Fiethe, M. Rubin, S. A. Fuselier, J. J. Berthelier,
More informationCombination of a getter pump with turbomolecular pumps in UHV applications
Vacuum 55 (1999) 27}37 Combination of a getter pump with turbomolecular pumps in UHV applications Roberto Giannantonio*, Magda Bovisio, Andrea Conte SAES Getters S.p.A., Viale Italia, 77-20020-Lainate
More informationPlasma-Therm PECVD. Operating Characteristics. Operating Instructions. Typical Processes. I. Loading. II. Operating
Plasma-Therm PECVD A PECVD (plasma enhanced chemical vapor deposition) reacts gases in a RF (radio frequency) induced plasma to deposit materials such as silicon dioxide and silicon nitride. This PECVD
More informationACCELERATOR VACUUM PROBLEMS: IN-LINE OIL TRAPPING BETWEgN THE STANFORD LINEAR ACCELERATOR AND THE BEAM SWITCHYARD
slac-pub-75 March 1965 ACCELERATOR VACUUM PROBLEMS: IN-LINE OIL TRAPPING BETWEgN THE STANFORD LINEAR ACCELERATOR AND THE BEAM SWITCHYARD E. L. Hoyt and E. L. Garwin Stanford Linear Accelerator Center Stanford
More informationDevelopment of o-carborane boronization for MST
Development of o-carborane boronization for MST J. Ko, D. J. Den Hartog, J. A. Goetz, P. J. Weix, and S. T. Limbach University of Wisconsin Madison, WI, USA 52nd APS DPP Meeting Chicago, Illinois, USA
More informationDEVICES FOR FIELD DETERMINATION OF WATER VAPOR IN NATURAL GAS Betsy Murphy MNM Enterprises 801 N. Riverside Drive Fort Worth, Texas 76111
INTRODUCTION Water vapor in natural gas has more than a substantial effect on the quality of the gas stream. Without quality measurement of water vapor the gas is basically not saleable. Contracts are
More information-6- THE DIII-D MAGNETIC FUSION TOKAMAK OUTGASSING TESTS ON MATERIALS USED IN. KaLm HOLTROP, M. HANSINK, and AmGm KELLMAN DECEMBER 1998 ATOMICS GENE-L
-6- OUTGASSNG TESTS ON MATERALS USED N THE D-D MAGNETC FUSON TOKAMAK by KaLm HOLTROP, M. HANSNK, and AmGm KELLMAN DECEMBER 1998 GENE-L ATOMCS DSCLAMER Portions of this document may be illegible in electronic
More informationDemystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.
Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards. Jamie Brown R&D Scientist Supelco 595 North Harrison RD. Bellefonte, PA 16823 National
More informationInformation Sheet. About this information sheet. Floor and wall loadings. Attachment of process module to floor. PlasmaPro Estrelas100
Information Sheet PlasmaPro Estrelas100 About this information sheet This data sheet provides basic information about the installation of a PlasmaPro Estrelas100 tool. Some of the information in this sheet
More informationREPRODUCIBILITY IN OPTICAL THIN FILM PROCESSING PART 1, THE VACUUM AND PUMPING
ABSTRACT REPRODUCIBILITY IN OPTICAL THIN FILM PROCESSING PART 1, THE VACUUM AND PUMPING Ronald R. Willey Willey Optical, Consultants, 13039 Cedar Street, Charlevoix, MI 49720, USA Ph 231-237-9392, ron@willeyoptical.com
More informationMETHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL PRESSURE AND COMPOSITIONAL ANALYSIS
METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL PRESSURE AND COMPOSITIONAL ANALYSIS Robert E. Ellefson REVac Consulting Dayton, OH 45459 USA robert.ellefson@sbcglobal.net ERMP IND12 QMS Workshop Bled,
More informationCharacterizing a Cu/Mn Alloy for Extracting Oxygen from Inert Gas. Streams
Characterizing a Cu/Mn Alloy for Extracting Oxygen from Inert Gas Streams Grace Lenhard Prattsburgh Central School LLE advisors: Walter Shmayda and Matthew Sharpe Laboratory for Laser Energetics University
More information