Specification for Medical Gas Pipeline s

Size: px
Start display at page:

Download "Specification for Medical Gas Pipeline s"

Transcription

1 M.S Specification for Medical Gas Pipeline s Supply, installation and commissioning of medical gas pipelines at pediatric emergency and Outborn NICU S. No. Area Beds Gas outlet Oxygen- O2 Flow meter Ward Vacuum Unit 1 Outborn NICU Neonate receiving area 3 Emergency ward 4 Emergency receiving area TOTAL MGPS Should be designed as per norms laid down in guidelines of HTM2022 /EN737 standards. The Products, wherever offered, should fully meet and comply with HTM 2022 standards. COMPONENT OF MGPS SYSTEM Medical Gas Pipeline component like Gas outlet points, Medical Gas area Line pressure alarm, Zonal Valve Box, Oxygen flow Meter, Humidifier Bottle, Vacuum Regulator, Collection Jar, High pressure antistatic tubing, Bed Head Wall Panels with railing should be from one source for easy compatibility and of good quality as per the specifications. MGPS PRE-INSTALLATION CERTIFICATION FOR MATERIAL SUPPLIED Test certificate from the manufacturer saying that products like Gas outlet points, Medical Gas Area Line pressure alarm, Zonal Valve Box, Oxygen flow Meter, Humidifier Bottle, Vacuum Regulator, Collection Jar, High pressure antistatic tubing, Bed Head Wall Panels with railing are fully tested and suitable for medical use. MGPS POST COMMISSIONING AND TESTING CERTIFICATES Testing and installation certificate from tenderer with complete details of the product. 1. The entire project has to be completed by the contractor before handover to the Hospital.

2 2. The specifications provided are basic guidelines. The contractor should collect the copy of building layout plan, visit the site and make their proposal. The details of the number of outlets and BOQ are indicative. SPECIFICATIONS OF MGPS SYSTEM Medical Gases Supply System DISTRIBUTION PIPING Copper Pipeline 1. The piped distribution system must use copper pipes manufactured from phosphorus deoxidized non arsenical copper. All pipelines used along with flow meter etc. should be ISI mark preferably size, 12 mm, 15 mm, 22 mm, and 28 mm (outer diameter) and of thickness > 0.7mm to 0.9mm,.0.9mm and 0.9mm respectively. 2. Should be drawn, seamless, deoxidized, non-arsenical, half-hard tempered and degreased materials conforming to BS-6017/1981 and manufactured as per BS-2871/the new European standard EN All copper pipes are to be certified by Authorized Test House Services for physical properties and Chemical properties. 4. Copper to copper joints shall be made on site using a silver copper phosphorus brazing alloy. Copper to brass or gun metal joints shall not be made on site. 5. Pipe lines should be supported at the intervals specified in HTM 2022 using suitable metallic, non ferrous material. Plastic supports may be used only for the drown drops to the terminal units. 6. Fitting shall be end feed type, manufactured from the same grade of copper on the pipes 7. Fittings shall be degreased suitable for oxygen use and supplied sealed in protective polythene bags. Gas Outlet/Terminal Unit-(Front Loaded-Emerald Type) 1. Outlets/valve with probe- each adapter should have suitable threads so that it can be connected to tube/ flow meter/ wvu /tsu etc. 2. All oxygen outlets and Suction outlets should be preferable Imported.

3 3. The socket assembly must retain a capsule assembly, a sealed with check valve and probe O ring seal 4. Outlets should be gas specific and only accept the correct medical gas probe. 5. Should be pin indexed to ensure correct gas assembly Oxygen Flow meter with Humidifier Bottle 1. It should have Humidifying bottle with whole flow meter assembly, should be of light weight material. 2. It should have backpressure compensated flow meter. 3. It should provide accurate gas flow measure by 0-15 liter/m flow meter. 4. Flow tube should be of good quality, transparent with measuring humidifying bottle. 5. Flow meter should be fitted with matching adaptor (probe) for oxygen outlet point. Ward Vacuum Unit 1. It should include vacuum pressure regular up to 760 mm Hg with Dial gauge 2. Should have re-usable collection bottle made with a wall bracelet. or unbreakable polycarbonate materials, fully auto-clavable along with safety device to prevent overflow and with a wall bracetk. 3. Should be wall mounted and back flow resistance should have 600ml capacity with play contracted jar 4. The controller should be operated by means of a 180 degree on/off switch and a bonnet adjustment. Isolation Valve (medical gas line Ball Valve- indigenous) 1. Each point should have separate isolation valve 2. Isolation/ shut off ball valves shall be of non lubricated 90 degree turn level stainless steel all valve with PTFE seat, having opening by quarter turn of handles. 3. Sizes should be appropriate of copper pipes with screw threaded ends and bran adapter. 4. All valves should be pneumatically tested for twice the working pressure degreased of medical gen service before supply.

4 5. The shut of valves should be from any of make CIM/RB/AUDKO- Italy make in ISO factories Antistatic Rubber Tube Should comply and meets requirement of ISO standards. It should be as per the specified standard Color. Bed head unit The bed head panel (horizontal) should be made from high strength extruded Aluminium with anodized finish. The length should be 5feet & 4 feet as per BOQ. The panel should not be flipped opened. The panel should be able to house 2 X Oxygen, 1 X vacuum and 1 X Air outlets, 5 nos. X 5/15 amp electrical receptor (4 nos. electrical receptors in 4-feet panel). The panel should have provision for double rail mounting system and include accessories i.e. I.V. poles with four hooks, Infusion pump mounts, monitor tray with slider and utility basket. Alarm System 1. The medical gas alarms must be capable of monitoring medical gas services by means of pressure sensors, which detect deviations from the normal operating limits of either pressure or medical vacuum. 2. Should have digital display of low pressure & high pressure valve. 3. Each gas service must be displayed by colored LEDs to show normal (green) low and high Pressure (red) conditions. Medical Vacuum systems must be displayed in normal; (green) and low vacuum (red) conditions only. Failure indicator must be displayed by flashing lights and normal indications must be steady 4. An audible warning must sound simultaneously with any failure indication and a mute facility must be provided. Following a mute selection the audible must resound after approx 15 minutes, or must operate simultaneously.a maintenance mute switch must be provided internally to the panel for use during maintenance which results in prolonged pipeline or plant shutdown. This facility must automatically reset when the gas services return to normal. 5. The alarm panel must have a test facility to prove the integrity of the internal circuits. LEDs an audible warning.each alarm must provide a green LED to indicate that electrical power is available at the panel and a red led to indicate system alarm. In the event of an electrical power failure the System alarm LED illuminate (flashing) and the audible warning.

5 Installation Testing Painting 1. Installation of piping shall be carried out with utmost cleanliness. Only pipes fitting and valves, which have been degreased and brought in polythene sealed bags, shall be used at the site. Pipe fixing clamps will be non-ferrous or non- deteriorating plastic suitable for the diameter of the pipe. 2. All pipe joint shall be made using insert gas fluxes brazing method.all joints shall be of copper to copper and shall be brazed by silver brazing filter material 3. All pipes shall be installed without printing or forcing. All pipes shall be protected against mechanical injury in a manner satisfactory to authorities having justification. After erection all the new pipelines would be cleared and purged with the help of dry nitrogen gas. Testing will be carried out a pressure of 6-kg/cm2 pressures for 24 hours to make the system leak out. All existing and proposed exposed pipe be painted with two coats of synthetic enamel paint and color codification as per prevalent hospital practice. Certification After completion of installation, contractor will certify that complete installation has been tested and all relevant tests have been carried. General terms and conditions 1. The firm should be an authorized agent/manufacturer of gas pipeline system along with accessories related to pipeline system. The medical gas specialist/ product manufacturer must be having ISO certificate (ISO/CE certified) 2. Product quoted must be installed in a Government Hospital preferably in Delhi. The firm should have satisfactory working equipment certificate from other Govt Hospital 3. The firm should have minimum 5 years of working experience in Medical Gas pipeline system along with control panel and manifold equipment in other Govt. Hospital/major hospitals having at least one manifold room with 100 oxygen and suction outlets +2 outlets of nitrous oxide.

6 4. It should give warranty of minimum 5yrs for pipelines and its accessories. 5. Firm should provide good quality of Medical gas pipeline/medical gas system that should be Safe and deliver medical gas in wards with safety norms. Medical gas Specialist/product manufacturer should be responsible for preparing design, flow rate calculations, pressure drop calculations, and operation maintenance manuals. 6. The firm should be manufacturer/sole distributor of manufacturer. 7. At least one engineer of the bidder must have gone through training programme at the original manufacturing unit. 8. Bidder must have completed/awarded three projects of single order value of 10lac in a government hospital preferably in Delhi. The company should have experience of 5 years or more in Government hospital. The company to provide work satisfactory report from the same institutions. 9. Firm should come and visit the actual installation site and submit the tender along with technical details, costs and pamphlets showing equipment to be supplied by them along with all the accessories required for maintenance. 10. Performance Certificate for All indigenous item quoted must be submitted. 11. Company should provide an undertaking that they will provide AMC for next 5 years after expiry of warranty period and should quote rates of AMC for next 5 years. 12. On site repair of the system to be provided. 13. The firm should keep spare s of sales equipment for 10 years 14. The entire project has to be done on turnkey basis including internal civil and electrical works at the time of execution. 15. Should have service facility in Delhi. 16. Copper pipes must be certified by a recognized certified agency for its compliance to specific standard. 17. Any misinformation regarding the specification of equipment offered would mean technical rejection.

7 18. The tender should be submitted in double i.e. Technical & Financial bid separately 19. Demonstration of equipment in working condition is a must. After satisfactory installation the firm should be able to show proper functioning of outlet points and pressure drop test to check the leakage. 20. Payments will be made on the actual measurement basis in front of the board members. Measurements will be made by the installing agency, which will be verified by the board on actual basis. 21. The quantity may be decreased or increased as per actual requirements of installation at a given time. 22. List of all spare parts with the rates should be submitted with the tender along with reasonability certificate for rates.

Technical Specification

Technical Specification Medical Gas Terminal Units (Gas Outlet Points) Technical Specification EN 9170-1 Terminal Unit Product Description CPX Terminal Units comprise a first fix, second fix and check valve. The first fix assembly

More information

SUPPLY, DESIGN, ERECTION & COMMISIONING OF COMPLETE MEDICAL GAS PIPE LINE SYSTEM BOQ (NSR)

SUPPLY, DESIGN, ERECTION & COMMISIONING OF COMPLETE MEDICAL GAS PIPE LINE SYSTEM BOQ (NSR) Page 1 1 Expression of Interest (E.O.I.) Govt. Lady Goshen Hospital ONGC MRPL Block Phase II SUPPLY, DESIGN, ERECTION & COMMISIONING OF COMPLETE MEDICAL GAS PIPE LINE SYSTEM BOQ (NSR) Sl. Specification

More information

PRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

PRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL PRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL Company Registered Office: Crown House, Stockport, Cheshire, SK13RB No. 850 0700 67 Registered in England and Wales No. 05058855

More information

CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL Company Registered Office: Crown House, Stockport, Cheshire, SK13RB No. 850 0700 67 Registered in England and Wales No. 05058855

More information

Manifolds & pressure reducing sets

Manifolds & pressure reducing sets Manifolds & pressure reducing sets Automatic Changeover Manifold (Electronic) Providing continuous supply of medical gases Pneumatech Medical Gas Solutions (PMGS) PureGAS Manifolds provide a duty or standby

More information

MANUAL GAS MANIFOLDS

MANUAL GAS MANIFOLDS MANUAL GAS MANIFOLDS Phoenix Pipeline Products Limited. Unit 8, McKenzie Industrial Park, Tel No.: 44 (0) 161 428 7200 Bird Hall Lane, Fax No.: 44 (0) 161 428 7010 Stockport, Email: info@p3-phoenix.com

More information

American Society of Sanitary Engineering PRODUCT (SEAL) LISTING PROGRAM

American Society of Sanitary Engineering PRODUCT (SEAL) LISTING PROGRAM American Society of Sanitary Engineering PRODUCT (SEAL) LISTING PROGRAM ASSE STANDARD #1048 - REVISED: 2009 Double Check Detector Fire Protection Backflow Prevention Assemblies Separate, complete laboratory

More information

AUTOMATIC GAS MANIFOLDS

AUTOMATIC GAS MANIFOLDS AUTOMATIC GAS MANIFOLDS Phoenix Pipeline Products Limited. Unit 8, McKenzie Industrial Park, Tel No.: 44 (0) 161 428 7200 Bird Hall Lane, Fax No.: 44 (0) 161 428 7010 Stockport, Email: info@p3-phoenix.com

More information

COMPACT TRIPLE WALL OUTLET STATION MODEL INSTALLATION AND OPERATING INSTRUCTIONS

COMPACT TRIPLE WALL OUTLET STATION MODEL INSTALLATION AND OPERATING INSTRUCTIONS COMPACT TRIPLE WALL OUTLET STATION MODEL 6255-1 INSTALLATION AND OPERATING INSTRUCTIONS To assure safe operation and conformation to local fire codes, all Porter Outlet Stations are designed to be used

More information

Perfect Medical Product

Perfect Medical Product +91-8048711203 Perfect Medical Product https://www.pmpmediindia.com/ We are one of the best and well known Manufacturer, Exporter, Wholesaler and Distributors of superior quality Medical Equipments like

More information

1 Exam Prep NFPA 99 Health Care Facilities Questions and Answers (Plumbing Contractor)

1 Exam Prep NFPA 99 Health Care Facilities Questions and Answers (Plumbing Contractor) 1 Exam Prep NFPA 99 Health Care Facilities Questions and Answers (Plumbing Contractor) 1. According to NFPA 99, SCFM is an acronym for. a) Surface Conditions at Fahrenheit Mercury b) Standard Conditions

More information

Instructions for Use. Plug connection for medical gas and vacuum acc. to DIN / 2004 DIN EN ISO

Instructions for Use. Plug connection for medical gas and vacuum acc. to DIN / 2004 DIN EN ISO Instructions for Use Plug connection for medical gas and vacuum acc. to DIN 13260-2 / 2004 DIN EN ISO 9170-1 Table of Contents Page 1. MANUFACTURER S DATA... 3 2. INTRODUCTION... 4 3. APPLICATION... 5

More information

Instruction Manual. CG16K Barometrically Compensated Capsule Dial Gauge. CG16K Capsule Dial Gauge, 0 to 25 mbar

Instruction Manual. CG16K Barometrically Compensated Capsule Dial Gauge. CG16K Capsule Dial Gauge, 0 to 25 mbar Instruction Manual D356-10-880 Issue G Original CG16K Barometrically Compensated Capsule Dial Gauge Description CG16K Capsule Dial Gauge, 0 to 1040 mbar CG16K Capsule Dial Gauge, 0 to 125 mbar CG16K Capsule

More information

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet: OXY Integral INTERCON ENTERPRISES INC Tel: 800 665 6655 Fax: 604 946 5340 E-Mail: sales@intercononline.com Internet: www.intercononline.com Manual Integral 2006 1 INDEX 2-3 PREFACE 4 INTRODUCTION 5 Principle

More information

Nitrous Oxide / Oxygen FLOOR MOUNTED OUTLET STATION Installation and Instructions

Nitrous Oxide / Oxygen FLOOR MOUNTED OUTLET STATION Installation and Instructions FM-171 Rev. B 8/00 Nitrous Oxide / Oxygen FLOOR MOUNTED OUTLET STATION Installation and Instructions MODEL 6300-1 PORTER INSTRUMENT COMPANY, INC. 245 TOWNSHIP LINE RD. P.O. BOX 907 HATFIELD, PA 19440-0907

More information

Installation and Maintenance Manual. Compact Medical Gas Outlets

Installation and Maintenance Manual. Compact Medical Gas Outlets Installation and Maintenance Manual Compact Medical Gas Outlets Contents Product Description 3 Cleaning and Lubricating 4 Inspection and Testing 4 Installation and Dimensions 5-6 Compact Outlet Indexing

More information

ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS

ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS Annex 3, page 2 ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS The text of existing chapter 15 is replaced by the following: "1 Application This

More information

AREA VALVE SERVICE UNITS

AREA VALVE SERVICE UNITS AREA VALVE SERVICE UNITS. Unit 8, McKenzie Industrial Park, Tel No.: 44 (0) 161 428 7200 Bird Hall Lane, Fax No.: 44 (0) 161 428 7010 Stockport, Email: info@p3-phoenix.com U.K., SK3 0SB www.p3-phoenix.com

More information

Technical Datasheet. Rigid & Retractable Pendants. Rigid Pendant

Technical Datasheet. Rigid & Retractable Pendants. Rigid Pendant Rigid & Retractable Pendants Rigid Pendant Description Rigid pendants are intended for use operating theatres, anaesthetic rooms and similar medical locations. They are an ergonomic solution for the provision

More information

M03 WUNDATRADE. Heat Pump Manifold. Before you start: Check the contents. Optional. ...a division of WUNDA GROUP PLC

M03 WUNDATRADE. Heat Pump Manifold. Before you start:  Check the contents. Optional. ...a division of WUNDA GROUP PLC Before you start: Check the manifold box contents against the list below. Check the contents Flow gauges 1. Bar assembly with manual return valves 2. 2 x Auto air vent & drain hose attachments Optional

More information

Emergency Mobile Unit Medicinal gases everywhere, always on time

Emergency Mobile Unit Medicinal gases everywhere, always on time Emergency Mobile Unit Medicinal gases everywhere, always on time EMU Emergency Mobile Unit service consider a full management of medicinal gases emergency situations inside hospitals in different steps:

More information

MEDICAL EQUIPMENT CATALOG

MEDICAL EQUIPMENT CATALOG THE MOST TRUSTED MEDICAL EQUIPMENT IN NORTH AMERICA MEDICAL EQUIPMENT CATALOG Total Components Including Bulk Oxygen Supply, Back up and Delivery Equipment Designs for High Pressure, Liquid or Bulk Applications

More information

ASSE International Product (Seal) Listing Program

ASSE International Product (Seal) Listing Program ASSE International Product (Seal) Listing Program ASSE 1070-2015 / ASME A112.1070-2015 / CSA B125.70-15 Performance Requirements for Water Temperature Limiting Devices Manufacturer: Contact Person: E-mail:

More information

American Society of Sanitary Engineering PRODUCT (SEAL) LISTING PROGRAM

American Society of Sanitary Engineering PRODUCT (SEAL) LISTING PROGRAM American Society of Sanitary Engineering PRODUCT (SEAL) LISTING PROGRAM ASSE STANDARD #1003 - REVISED: 2009 Water Pressure Reducing Valves for Domestic Water Distribution Systems MANUFACTURER: CONTACT

More information

Specialists in HTM Medical Gas Pipeline Equipment AVSU/Module Operating and Maintenance Instructions

Specialists in HTM Medical Gas Pipeline Equipment AVSU/Module Operating and Maintenance Instructions The Area Value Service Units and Module format (AVSU) provide a local gas isolation facility for use during normal installation and maintenance or in the event of an emergency. They are built in accordance

More information

SATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products

SATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products Saturn Evo SATURN EVO COLOR A N A E S T H E S I A S Y S T E M i n t e l l i g e n t s i m p l i c i t y Manufacturer of high quality anaesthesia products SATURN EVO COLOR PATIENT SAFETY AND USER PATIENT

More information

Zeus Vertical Headwall

Zeus Vertical Headwall Zeus Vertical Headwall Medical Supply Units Installation Instructions Published by Pneumatech Medical Gas Solutions All possible care has been taken in the preparation of this publication, but Pneumatech

More information

MEDICAL GAS OUTLETS & PROBES

MEDICAL GAS OUTLETS & PROBES MEDICAL GAS OUTLETS & PROBES Quick-connect medical gas outlets for operative connection of the medical equipment to medical gas supply systems TECHNICAL DATA Maximum inlet pressure to 10 bar Vacuum rarefaction

More information

ACCU-PULSE Installation and Operation Instructions

ACCU-PULSE Installation and Operation Instructions ACCU-PULSE Installation and Operation Instructions Pump Discharge Installation: Chargeable Models Step 1: Mounting Position Mount ACCU-PULSE as close to the pump discharge as possible to absorb the pulse

More information

PRS(TC)4,8 USER MANUAL. Read the complete manual before installing and using the regulator.

PRS(TC)4,8 USER MANUAL. Read the complete manual before installing and using the regulator. PRS(TC)4,8 USER MANUAL Read the complete manual before installing and using the regulator. WARNING INCORRECT OR IMPROPER USE OF THIS PRODUCT CAN CAUSE SERIOUS PERSONAL INJURY AND PROPERTY DAMAGE. Due to

More information

HYQUEST SOLUTIONS PTY LTD INSTRUCTION MANUAL TIPPING BUCKET RAIN GAUGE MODEL TB6

HYQUEST SOLUTIONS PTY LTD INSTRUCTION MANUAL TIPPING BUCKET RAIN GAUGE MODEL TB6 INSTRUCTION MANUAL TIPPING BUCKET RAIN GAUGE MODEL TB6 QUALITY SYSTEM ISO:9001 CERTIFIED HYQUEST SOLUTIONS PTY LTD PO BOX 332, LIVERPOOL B.C NSW 1871, AUSTRALIA Phone:(Int.) 612 9601 2022 Fax: :(Int.)

More information

Installation and Maintenance Manual. Dual Medical Gas Outlets

Installation and Maintenance Manual. Dual Medical Gas Outlets Installation and Maintenance Manual Dual Medical Gas Outlets Contents Product Description 3 Cleaning and Lubricating 4 Inspection and Testing 4 Installation and Dimensions 5-6 Model Numbers 7 Service 8

More information

USM21 Sealed Bimetallic Steam Trap for use with Pipeline Connectors Installation and Maintenance Instructions

USM21 Sealed Bimetallic Steam Trap for use with Pipeline Connectors Installation and Maintenance Instructions 6250250/1 IM-P625-03 ST Issue 1 USM21 Sealed Bimetallic Steam Trap for use with Pipeline Connectors Installation and Maintenance Instructions 1. General safety information 2. General product information

More information

Section GAS CHLORINATION SYSTEM

Section GAS CHLORINATION SYSTEM PART 1 GENERAL 1.01 SUMMARY Section 11260 This Section includes the furnishing and subsequent installation, complete and in place, of a functioning gas chlorination disinfection system for water as shown

More information

A Rationale for Pressure Relief Device(s) Qualification Requirements (LH2)

A Rationale for Pressure Relief Device(s) Qualification Requirements (LH2) UN-GTR PART A INSERTION Stand 09.02.2011, 17:20 Uhr A.3.3 HYDROGEN STORAGE SYSTEM The hydrogen storage system consists of all components that form the primary pressure boundary of the stored hydrogen in

More information

RS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator.

RS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator. RS(H)10,15 USER MANUAL Read the complete manual before installing and using the regulator. WARNING INCORRECT OR IMPROPER USE OF THIS PRODUCT CAN CAUSE SERIOUS PERSONAL INJURY AND PROPERTY DAMAGE. Due to

More information

Piped Air SCBA Refilling System Standard

Piped Air SCBA Refilling System Standard WALT WHITE Fire Chief 5770 Freeport Blvd., Suite 200 Sacramento, CA 95822-3516 Ph: (916) 808-1300 Fax: (916) 808-1629 www.sacfire.org Piped Air SCBA Refilling System Standard SCOPE: This specification

More information

OIL SUPPLY SYSTEMS ABOVE 45kW OUTPUT 4.1 Oil Supply

OIL SUPPLY SYSTEMS ABOVE 45kW OUTPUT 4.1 Oil Supply OIL SUPPLY SYSTEMS ABOVE 45kW OUTPUT 4.1 Oil Supply 4.1.1 General The primary function of a system for handling fuel oil is to transfer oil from the storage tank to the oil burner at specified conditions

More information

HOSPITAL STRUCTURAL PRODUCTS 12. MEDICAL GAS SYSTEMS LUMINAIRES AND OTHER EQUIPMENTS WALL CORNER PROTECTION SYSTEMS A. B. C.

HOSPITAL STRUCTURAL PRODUCTS 12. MEDICAL GAS SYSTEMS LUMINAIRES AND OTHER EQUIPMENTS WALL CORNER PROTECTION SYSTEMS A. B. C. 244 // HOSPITAILOR 12. HOSPITAL STRUCTURAL PRODUCTS MEDICAL GAS SYSTEMS LUMINAIRES AND OTHER EQUIPMENTS WALL CORNER PROTECTION SYSTEMS A. B. C. HOSPITAILOR // 245 A. MEDICAL GAS SYSTEMS PATIENT BED HEAD

More information

TECHNICAL DATA. Q = C v P S

TECHNICAL DATA. Q = C v P S Page 1 of 13 1. DESCRIPTION The Viking 6 Model G-6000 Dry Valve Riser Assembly consists of a small profile, light weight, pilot operated valve that is used to separate the water supply from the dry sprinkler

More information

TECHNICAL DATA OBSOLETE

TECHNICAL DATA OBSOLETE April 9, 2009 Preaction 326a 1. DESCRIPTION Viking supervised Double-Interlocked Electric/Pneumatic Release Preaction Systems utilizing the Viking G-4000P Deluge Valve. The small profi le, lightweight,

More information

DUPLEX PRESSURE REDUCING SET

DUPLEX PRESSURE REDUCING SET DUPLEX PRESSURE REDUCING SET Phoenix Pipeline Products Limited. Unit 8, McKenzie Industrial Park, Tel No.: 44 (0) 161 428 7200 Bird Hall Lane, Fax No.: 44 (0) 161 428 7010 Stockport, Email: info@p3-phoenix.com

More information

TECHNICAL DATA 3 MODEL G-3000 DRY VALVE RISER ASSEMBLY

TECHNICAL DATA 3 MODEL G-3000 DRY VALVE RISER ASSEMBLY Page 1 of 13 1. DESCRIPTION The Viking 3 Model G-3000 Dry Valve Riser Assembly is equipped with a small profile, light weight, pilot operated valve that is used to separate the water supply from the dry

More information

MEDAP-MOBILE GAS SUPPLY

MEDAP-MOBILE GAS SUPPLY MEDAP-MOBILE GAS SUPPLY This document is intended to provide information to an international audience outside of the US. MOBILE GAS SUPPLY 3 MOBILE GAS SUPPLY FOR TRANSPORTATION AND WARD PRESSURE REGULATORS,

More information

SECTION BUTTERFLY VALVES

SECTION BUTTERFLY VALVES SECTION 15112 BUTTERFLY VALVES PART 1 GENERAL 1.01 SUMMARY A. All butterfly valves shall be of the tight closing, rubber seated type and fully comply with the latest revision of AWWA Standard C504, Class

More information

TECHNICAL DATA. Q= Cv S

TECHNICAL DATA. Q= Cv S Page 1 of 13 1. DESCRIPTION The Viking 4 inch Model G-4000 Dry Valve Riser Assembly consists of a small profile, light weight, pilot operated valve that is used to separate the water supply from the dry

More information

TECHNICAL DATA Q = C. v P S. 2 Model G-2000 Dry valve. Page 1 of 13

TECHNICAL DATA Q = C. v P S. 2 Model G-2000 Dry valve. Page 1 of 13 Page 1 of 13 1. Description The Viking 2 Model G-2000 Dry Valve Riser Assembly consists of a small profile, light weight, pilot operated valve that is used to separate the water supply from the dry sprinkler

More information

High Precision Gauge Pressure Indicators

High Precision Gauge Pressure Indicators Series 1500 8½" Dial High Precision Gauge Pressure Indicators Series 1500 Gauges have capsule-type pressure elements up to and including the 150-psig range. Ranges of 200 psig and above have Bourdon tubes.

More information

IX1 Spirax-Monnier International Compressed Air Filter Installation and Maintenance Instructions

IX1 Spirax-Monnier International Compressed Air Filter Installation and Maintenance Instructions 0575050/2 IM-P057-04 CH Issue 2 IX1 Spirax-Monnier International Compressed Air Filter Installation and Maintenance Instructions 1 Safety information 2 General product information 3 Installation and commissioning

More information

System Pressure Manager Standard & System Pressure Manager Plus

System Pressure Manager Standard & System Pressure Manager Plus System Pressure Manager Standard & System Pressure Manager Plus Installation, Commissioning & Servicing Instructions Note: THESE INSTRUCTIONS MUST BE READ AND UNDERSTOOD BEFORE INSTALLING, COMMISSIONING,

More information

Premium Series. Economical Price & Reliable Performance. High Quality. Rust-Resistant. Long Life. Tamper-Resistant locking Screw.

Premium Series. Economical Price & Reliable Performance. High Quality. Rust-Resistant. Long Life. Tamper-Resistant locking Screw. Premium Series High Quality Forged Brass Body Rust-Resistant Nickel Plated Finish Long Life Corrosion Resistant Valve Seat Economical Price & Reliable Performance Tamper-Resistant locking Screw Reduces

More information

Unit 24: Applications of Pneumatics and Hydraulics

Unit 24: Applications of Pneumatics and Hydraulics Unit 24: Applications of Pneumatics and Hydraulics Unit code: J/601/1496 QCF level: 4 Credit value: 15 OUTCOME 2 TUTORIAL 9 ACCUMULATORS The material needed for outcome 2 is very extensive so there are

More information

Chlorinator MODEL MK-I

Chlorinator MODEL MK-I Chlorinator MODEL MK-I The New model MK- I is semi automatic gas type. Vacuum operated Solution feed Chlorinator designed to provide a continues & measured quantity of gas while in operation. It is suitable

More information

Medical Gas Control Panel

Medical Gas Control Panel Medical Gas Control Panel INSTALLATION & MAINTENANCE MANUAL 255105 (Rev. 6) 2018-04 2 255105 (Rev. 6) 2018-04 1.0 User Responsibilities The information contained in this installation and operation maintenance

More information

Pressure Equalising Valves Model , for Differential Pressure Gauges

Pressure Equalising Valves Model , for Differential Pressure Gauges Accessories Pressure Equalising Valves Model 910.25, for Differential Pressure Gauges WIKA Data Sheet AC 09.11 Applications T Isolating, pressure equalising as well as purging and air bleeding valves to

More information

Gas Welding Hoses and connections, the weak link

Gas Welding Hoses and connections, the weak link Gas Welding Hoses and connections, the weak link By Leif Andersen, Technical Product Manager Welding, WSS Regulators with flashback arrestors and the shank with its welding or cutting attachment are made

More information

TECHNICAL DATA MAINTENANCE AIR COMPRESSOR MODEL G-1

TECHNICAL DATA MAINTENANCE AIR COMPRESSOR MODEL G-1 Dry 131h 1. DESCRIPTION The Viking Model G-1 Maintenance Air Compressor is an electric motor-driven, aircooled, single-stage, oil-less compressor. The unit is equipped with a check valve and provides a

More information

MEDICAL GAS SYSTEM & EQUIPMENT

MEDICAL GAS SYSTEM & EQUIPMENT MEDICAL GAS SYSTEM & EQUIPMENT 14 16-17 High Pressure Regulators Vacuum Regulators 19 20 21 22 23 Probes Oxygen Probes Nitrous Oxide Probes Vacuum Probes Medical Air Probes 4 Bar Medical Air Probes 7 Bar

More information

IP2, IP2A and IP2D. Spirax-Monnier International Compressed Air Filter/ Regulators

IP2, IP2A and IP2D. Spirax-Monnier International Compressed Air Filter/ Regulators Page 1 of 6 TI-P510-01 H Issue 5 IP2, IP2 and IP2D ISO 9001 Spirax-Monnier International ompressed ir Filter/ Regulators ert. No. LRQ 0963008 Description The IP range of Spirax-Monnier international compressed

More information

Compact Triple Cabinet Outlet Station Model Installation and Operating Instructions

Compact Triple Cabinet Outlet Station Model Installation and Operating Instructions Compact Triple Cabinet Outlet Station Model 6258-1 Installation and Operating Instructions The Porter Compact Triple Outlet Station (6258-1) provides a quick, safe, and reliable method of connection to

More information

MST21 Stainless Steel Balanced Pressure Thermostatic Steam Trap

MST21 Stainless Steel Balanced Pressure Thermostatic Steam Trap 1250650/6 IM-P125-07 ST Issue 6 MST21 Stainless Steel Balanced Pressure Thermostatic Steam Trap Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation

More information

The Compact, Portable, Efficient and Economical Solution to boost Shop Compressed Air Pressure by 2 or 3 times

The Compact, Portable, Efficient and Economical Solution to boost Shop Compressed Air Pressure by 2 or 3 times Tseries Series AB AB Air Air Boosters Boosters The Compact, Portable, Efficient and Economical Solution to boost Shop Compressed Air Pressure by 2 or 3 times.01/1 MANIFOLD MOUNTED VALVES FOR EASY SERVICING

More information

Installation Instructions for NP100 Series Nitrogen Control Panels

Installation Instructions for NP100 Series Nitrogen Control Panels Installation Instructions for NP100 Series Nitrogen Control Panels Installation of the Nitrogen Control Panel involves installing the rough-in box and making the necessary piping connections in accordance

More information

TECHNICAL DATA. Q = C v P S

TECHNICAL DATA. Q = C v P S Preaction 346a 1. Description The 6 Model G-6000P Electric Release Preaction System Riser Assembly can be used as a Single Interlock Preaction System with Electric Release, or as a Double Interlock Preaction

More information

Pipe threads for tubes and fittings where pressure-tight joints are not made on the threads. (requires PTFE sealing tape or liquid sealant).

Pipe threads for tubes and fittings where pressure-tight joints are not made on the threads. (requires PTFE sealing tape or liquid sealant). Fittings for CO 2 Pipe Threads Pipe thread references quoted in this catalogue conform with the requirements specified in the latest issue and amendments of the following ISO Standards: ISO 7-1 (BS21)

More information

v. Size shall be specified on drawings.

v. Size shall be specified on drawings. SPECIFICATION FOR VACUUM INSULATED PIPING Part 1 General 1. Submittals a. After award of contract and before executing any manufacturing, shop drawings and specifications shall be submitted to the customer

More information

BT6HC Hygienic Sanitary Balanced Pressure Steam Trap for High Capacity and CIP/SIP Applications

BT6HC Hygienic Sanitary Balanced Pressure Steam Trap for High Capacity and CIP/SIP Applications 1800350/6 IM-P180-12 ST Issue 6 BT6HC Hygienic Sanitary Balanced Pressure Steam Trap for High Capacity and CIP/SIP Applications Installation and Maintenance Instructions 1. Safety information 2. General

More information

Gas Pressure Regulator HON 200

Gas Pressure Regulator HON 200 Product information serving the gas industry worldwide Applications, characteristics, technical data Application Gas supply to municipal, industrial and individual consumers Regulator for low-load rails

More information

FLUID POWER FLUID POWER EQUIPMENT TUTORIAL ACCUMULATORS. This work covers part of outcome 2 of the Edexcel standard module:

FLUID POWER FLUID POWER EQUIPMENT TUTORIAL ACCUMULATORS. This work covers part of outcome 2 of the Edexcel standard module: FLUID POWER FLUID POWER EQUIPMENT TUTORIAL ACCUMULATORS This work covers part of outcome 2 of the Edexcel standard module: UNIT 21746P APPLIED PNEUMATICS AND HYDRAULICS The material needed for outcome

More information

TECHNICAL ACCESSORIES

TECHNICAL ACCESSORIES A.R.I. VALVES S-050P AIR PLASTIC RELEASE 16 BAR Automatic Component A.R.I patent, Rolling Seal Mechanism: Dramatically reduces the possibility of obstruction by debris. Discharges high air flow rates upto

More information

TECHNICAL DATA. Trimpac 251a. Spetember 16, 2013

TECHNICAL DATA. Trimpac 251a. Spetember 16, 2013 Spetember 16, 2013 Trimpac 251a 1. DEsCrIpTION DESCRIPTION TRIMPAC Model B-6 and B-6B is a factory assembled trim package for a double interlocked preaction system with an electric/pneu-lectric release

More information

EMPLOYEES STATE INSURANCE CORPORATION SUPER SPECIALITY HOSPITAL SANATHNAGAR, HYDERABAD LIMITED TENDER For ANNUAL MAINTENANCE CONTRACT

EMPLOYEES STATE INSURANCE CORPORATION SUPER SPECIALITY HOSPITAL SANATHNAGAR, HYDERABAD LIMITED TENDER For ANNUAL MAINTENANCE CONTRACT EMPLOYEES STATE INSURANCE CORPORATION SUPER SPECIALITY HOSPITAL SANATHNAGAR, HYDERABAD-500038 LIMITED TENDER For ANNUAL MAINTENANCE CONTRACT FOR THE MEDICAL GAS PIPELINE SYSTEM Not Transferable Tender

More information

South-Tek Systems - Nitrogen Generation Corrosion Inhibiting System. Designed for: Dry or Preaction Fire Protection Systems (FPS)

South-Tek Systems - Nitrogen Generation Corrosion Inhibiting System. Designed for: Dry or Preaction Fire Protection Systems (FPS) South-Tek Systems - Nitrogen Generation Corrosion Inhibiting System Designed for: Dry or Preaction Fire Protection Systems (FPS) 1.0 Description of Work 1.1 The Fire Sprinkler Contractor shall provide

More information

Copper tubes, as we are all aware, have

Copper tubes, as we are all aware, have CRITICAL REQUIREMENTS OF COPPER TUBES & FITTINGS FOR MEDICAL GAS PIPELINE SYSTEMS (MGPS) COVER STORY GURMIT SINGH ARORA Copper tubes, as we are all aware, have various applications including Airconditioning

More information

TECHNICAL DATA. Q = C v P S

TECHNICAL DATA. Q = C v P S January 6, 2012 Preaction 348a 1. Description Viking supervised Surefire Preaction Systems utilize the Viking G-6000P Valve. The small profile, lightweight, pilot operated Viking G-6000P Valve comes complete

More information

TECHNICAL DATA. Q = C v P S

TECHNICAL DATA. Q = C v P S January 6, 2012 Preaction 333a 1. Description Viking supervised Surefire Preaction Systems Utilize the Viking G-3000P Valve. The small profile, lightweight, pilot-operated Viking G-3000P Valve comes complete

More information

Horizontal Bladder Tanks

Horizontal Bladder Tanks DATA SHEET Horizontal Bladder Tanks Features UL Listed and FM Approved for use with various ANSUL proportioners and foam concentrates 175 psi (12.1 bar) maximum allowable working pressure (design pressure)

More information

Tank Blanketing Pressure Regulators RHPS Series

Tank Blanketing Pressure Regulators RHPS Series www.swagelok.com Tank Blanketing Pressure Regulators RHPS Series Types: pressure reducing and vapor recovery 16L stainless steel construction 1/2, 1, and 2 in. end connections Working pressures up to 22

More information

LRS(H)4 USER MANUAL. Read the complete manual before installing and using the regulator.

LRS(H)4 USER MANUAL. Read the complete manual before installing and using the regulator. LRS(H)4 USER MANUAL Read the complete manual before installing and using the regulator. WARNING INCORRECT OR IMPROPER USE OF THIS PRODUCT CAN CAUSE SERIOUS PERSONAL INJURY AND PROPERTY DAMAGE. Due to the

More information

PART 1 GENERAL..2 Components of the gas chlorination system include, but are not limited to the following:

PART 1 GENERAL..2 Components of the gas chlorination system include, but are not limited to the following: PART 1 GENERAL 1.1 DESCRIPTION.1 This section includes the supply, delivery and installation, testing and placement into operation of a gas chlorination system, and appurtenances as specified herein and

More information

CENTRAL GAS SUPPLY SYSTEMS : CYLINDER BATTERIES AND HIGH PRESSURE HOSES

CENTRAL GAS SUPPLY SYSTEMS : CYLINDER BATTERIES AND HIGH PRESSURE HOSES P0170E-1.DOCX/08.16/MG CENTRAL GAS SUPPLY SYSTEMS : CYLINDER BATTERIES AND HIGH PRESSURE HOSES With a Central Gas Supply System, individual work places or gas take-off stations are supplied with gas through

More information

Radiator thermostats type RA2000, valve bodies type RA-FN (series D) and RA-G

Radiator thermostats type RA2000, valve bodies type RA-FN (series D) and RA-G Data sheet Radiator thermostats type RA2000, valve bodies type RA-FN (series D) and RA-G Application RA2000 013G2910 Built-in sensor RA2000 013G2920 tamperproof RA2000 013G5062 remote temperature adjuster

More information

Compact Triple Cabinet Outlet Station Model B Installation and Operating Instructions

Compact Triple Cabinet Outlet Station Model B Installation and Operating Instructions PORTER Parker Hannifin Corporation Porter Instrument Division 245 Township Line Rd. P.O. Box 907 Hatfield, PA 19440-0907 USA (215) 723-4000 / fax (215) 723-5106 Compact Triple Cabinet Outlet Station Model

More information

TECHNICAL DATA. TRIMPAC Model B-5 & B-5B

TECHNICAL DATA. TRIMPAC Model B-5 & B-5B September 16, 2013 Trimpac 250a 1. DESCRIPTION DEsCRIPTIoN is a factory assembled trim package for a ed an electric/pneumatic release module in a metal enclosure. The standard trim normally required on

More information

OPERATING INSTRUCTIONS MANUAL FOR QACV PNEUMATIC DOSING PUMP

OPERATING INSTRUCTIONS MANUAL FOR QACV PNEUMATIC DOSING PUMP This operating instructions contains safety information that if ignored can endanger life or result in serious injury. They are indicated by this icon. Use of this pump with radioactive chemicals is forbidden!

More information

PRESSURE CONTROL REGULATORS

PRESSURE CONTROL REGULATORS REGULATOR HARRIS STYLE SINGLE STAGE 201 Series Medium Duty Single stage construction Forged brass body for maximum strength Chrome bonnet Neoprene diaphragms Sintered metal inlet filter to trap impurities

More information

Vertical Bladder Tanks

Vertical Bladder Tanks DATA SHEET Vertical Bladder Tanks Features UL Listed and FM Approved for use with various ANSUL proportioners and foam concentrates 175 psi (12.1 bar) maximum allowable working pressure (design pressure)

More information

INSTRUCTION MANUAL TIPPING BUCKET RAIN GAUGE. MODEL TB3/0.1mm

INSTRUCTION MANUAL TIPPING BUCKET RAIN GAUGE. MODEL TB3/0.1mm INSTRUCTION MANUAL TIPPING BUCKET RAIN GAUGE MODEL TB3/0.1mm QUALITY SYSTEM ISO 9001 2008 CERTIFIED HYQUEST SOLUTIONS PTY LTD PO BOX 332, LIVERPOOL B.C NSW 1871, AUSTRALIA Phone:(Int.) 612 9601 2022 Fax:

More information

Chemical Feed Equipment Floor-Mounted Gas Dispenser - Series CH4200

Chemical Feed Equipment Floor-Mounted Gas Dispenser - Series CH4200 Specification Bulletin (rev.01) Series CH4200 Chemical Feed Equipment Floor-Mounted Gas Dispenser - Series CH4200 Easy, fast& accurate digital with 11 points calibration Laser sensor offering actual feed

More information

Transferring and Storage

Transferring and Storage Organometallics Transferring and Storage A dip tube is available for Container 2 which can be screwed into the neck to permit safe transfer of product. Small quantities of metal alkyl can be withdrawn

More information

Forged Brass Bellows Valves

Forged Brass Bellows Valves Forged Brass Bellows Valves Designed Specifically For High Purity Gas Service Size Range - 0.625" to 4.125" Outer Diameter Size Product Features: Forged Brass Bodies - No Sand Castings Clean Area Assembled

More information

TECHNICAL DATA. Q = C v P S

TECHNICAL DATA. Q = C v P S January 6, 2012 Preaction 347a 1. Description Viking supervised Double-Interlocked Electric/Pneumatic Release Preaction Systems utilize the Viking G-6000P Valve. The small profile, lightweight, pilot operated

More information

Radiator thermostats type RA2000, valve bodies type RA-FN (series D) and RA-G

Radiator thermostats type RA2000, valve bodies type RA-FN (series D) and RA-G Data sheet Radiator thermostats type RA2000, valve bodies type RA-FN (series D) and RA-G EN 215 Application RA2000 013G2910 Built-in sensor RA2000 013G2920 tamperproof RA2000 013G5062 remote temperature

More information

INSTALLATION OPERATION MAINTENANCE

INSTALLATION OPERATION MAINTENANCE Bermad Electrically Controlled On-Off Deluge Valve Model: 400E-3D INSTALLATION OPERATION MAINTENANCE Application Engineering BERMAD 1. Safety First BERMAD believes that the safety of personnel working

More information

Installation, Operating and Maintenance Instructions. V914 Cast Iron Flanged Swing Check Valve V914

Installation, Operating and Maintenance Instructions. V914 Cast Iron Flanged Swing Check Valve V914 Installation, Operating and Maintenance Instructions V914 Cast Iron Flanged Swing Check Valve V914 THE PRESSURE EQUIPMENT DIRECTIVE 97/23/EC and CE MARKING The Pressure Equipment Regulations 1999 (SI 1999/2001)

More information

Safety Policy Manual Policy No. 133

Safety Policy Manual Policy No. 133 Policy: Installation, Testing, and Maintenance of Medical Gas Page 1 of 2 APPLICATION NYU Langone Medical Center (NYULMC) PURPOSE To provide guidelines for safe and proper installation, testing and maintenance

More information

CLEAN AGENT FIRE EXTINGUISHING SYSTEM MX 1230 Extinguishing Agent Cylinders

CLEAN AGENT FIRE EXTINGUISHING SYSTEM MX 1230 Extinguishing Agent Cylinders The technical data described herein is for components of the Minimax Fire Products MX 30 Clean Agent System. For design, installation, and operation instructions, refer to the latest edition of the original

More information

Gas plug-in hose, straight-through form. Instructions for Use. suitable for gas socket model G2019T, G2016T. G2023 from 01/1990.

Gas plug-in hose, straight-through form. Instructions for Use. suitable for gas socket model G2019T, G2016T. G2023 from 01/1990. Gas plug-in hose, straight-through form Instructions for Use suitable for gas socket model G2019T, G2016T Model Year built: G2023 from 01/1990 en_int Gas plug-in hose, straight-through form 2 from 12 Table

More information

Serie ECO 3T. Threaded end back flow preventer with controllable reduced pressure zone. made in. Application fields. Protection E U R O P E

Serie ECO 3T. Threaded end back flow preventer with controllable reduced pressure zone. made in. Application fields. Protection E U R O P E Serie ECO T Threaded end back flow preventer with controllable reduced pressure zone BRANDONI made in E U R O P E Application fields WATER FIRE FIGHTING DRINKING WATER 74 www.brandoni.it Serie ECO T The

More information

SUPPLY OF ISOLATION AND APPLIANCE VALVES PTS - APPLIANCE BALL VALVE

SUPPLY OF ISOLATION AND APPLIANCE VALVES PTS - APPLIANCE BALL VALVE PTS - APPLIANCE BALL VALVE 2 30.06.17 Issued for Tender AS ADE GS 1 12.04.17 Issued for Tender AS AD GS 0 06.03.17 Issued for Client s Comments AS AD GS Rev. Date Subject of revision Author Checked Approved

More information