FTIR Detection of Outgassing Chemicals: Instantaneous and Comprehensive Identification and Quantitation
|
|
- Hope Robertson
- 5 years ago
- Views:
Transcription
1 FTIR Detection of Outgassing Chemicals: Instantaneous and Comprehensive Identification and Quantitation John C. Hulteen, Ph.D. Environmental Specialist 3M St. Paul, MN DISKCON 2000 San Jose, CA September 19, 2000
2 Airborne Molecular Contaminants (AMC) Sources: plastics, tapes, adhesives, pumps, process chemicals, solvents, people, etc. Effects: corrosion, hazing, molecular contamination, adhesion, etc. Major Issue : How do you quantitate and speciate molecular compounds that are outgassed from disk drive materials in a timely manner.
3 An Innovative New Alternative to Existing Technologies. Detection of acids, bases, volatiles and semi-volatiles simultaneously. Quantitation and identification of compounds. Fast collection and analysis time - manufacturing requirement. Low detection limits as required by the microelectronics industry. Rugged method and instrumentation for plant application. Simple equipment maintenance and operation.
4 3M Proprietary FTIR Static Headspace (SHS) temperature controller infrared source and detector heated 4.5 L chamber sample drawer
5 FTIR Static Headspace (SHS): Outgassing Profile FTIR Static HS Gas Cell IR Source Total Cumulative Outgassing vs Time IR Detector Absorbance Units SAMPLE FITR Spectra ng/cm 2 Time
6 FTIR Static Headspace (SHS): Outgassing Profile FTIR Static HS Gas Cell Total Cumulative Outgassing vs Time Absorbance Units SAMPLE FITR Spectra ng/cm 2 Time
7 FTIR Static Headspace (SHS): Outgassing Profile FTIR Static HS Gas Cell Total Cumulative Outgassing vs Time Absorbance Units SAMPLE FITR Spectra ng/cm 2 Time
8 FTIR Static Headspace (SHS): Outgassing Profile FTIR Static HS Gas Cell Total Cumulative Outgassing vs Time Absorbance Units SAMPLE FITR Spectra ng/cm 2 Time
9 FTIR Static Headspace (SHS): Outgassing Profile FTIR Static HS Gas Cell Total Cumulative Outgassing vs Time Absorbance Units SAMPLE FITR Spectra ng/cm 2 Time
10 Determination of Extent of Outgassing in SHS FTIR 9 8 ppm v of acrylate in cell TOTAL OUTGASSING OF THE SATURATED ACRYLATE HEADSPACE WAS REACHED AT THIS TEMPERATURE OR Purge with N 2 TOTAL OUTGASSING Time (min) Stop N 2 purge and reseal cell Label Tape at 85 C
11 The Basic Principle of Infrared Spectroscopy Formic Acid O C 1775 cm -1 H O H O 638 cm -1 C H O H 1105 cm-1 C H O H O Absorbance 2945 cm -1 H O C O H Energy (wavenumbers cm -1 ) Fingerprint Region
12 FTIR Detection of Acids Collection Time < 2 minutes hydrofluoric acid (HF) 3 ppm v Absorbance acrylic acid 16 ppm v hydrochloric acid (HCl) 29 ppm v acetic acid 11 ppm v Energy (wavenumbers cm -1 )
13 Example of Outgassing Acids in Tape Products Tape Sample A Formic acid detected in outgas after 2 minutes. Formic Acid Standard Fully outgassed at 570 ng/cm 2 or 41 ppm w after 120 minutes. FTIR outgassing spectra of a tape sample and standard at 85 o C Cummulative Outgassing Concentration (ng/cm2) :00 0:28 0:57 1:26 1:55 2:24 Time (hr:min)
14 Example of Outgassing Acids in Tape Products Tape Sample B Acetic acid detected in outgas after 2 minutes. Acetic Acid Standard Fully outgassed at 740 ng/cm 2 or 33 ppm w after 14 hours. FTIR outgassing spectra of a tape sample and standard at 85 o C Cummulative Outgassing Concentration (ng/cm2) :00 2:24 4:48 7:12 9:36 12:00 14:24 16:48 Time (hr:min)
15 FTIR Detection of Bases Collection Time < 2 minutes ammonia 8.5 ppm v Absorbance diethyl amine 25 ppm v methyl amine 40 ppm v Energy (wavenumbers cm -1 )
16 Example of Outgassing Bases in Tape Products Tape Sample C Ammonia Standard Ammonia detected in outgas after 2 minutes. Fully outgassed at 245 ng/cm 2 or 32 ppm w after 30 minutes. FTIR outgassing spectra of a tape sample and standard at 85 o C Cummulative Outgassing Concentration (ng/cm2) :00 0:14 0:28 0:43 0:57 1:12 Time (hr:min)
17 FTIR Detection of Semi-Volatiles Collection Time < 2 minutes Octamethylcyclotetrasiloxane 10 ppm v Absorbance ethylhexyl acrylate 21 ppm v phenol 10 ppm v Energy (wavenumbers cm -1 )
18 Example of Outgassing Semi- Volatiles in Tape Products Tape Sample D Siloxane (OMCTS) Standard Siloxanes detected in outgas after 2 minutes. Fully outgassed at 20 ng/cm 2 or 2 ppm w after 30 minutes. 25 FTIR outgassing spectra of a tape sample and standard at 85 o C Note: OMCTS is a surrogate standard for all siloxanes through the 800 cm -1 band. Cummulative Outgassing Concentration (ng/cm2) :00 0:07 0:14 0:21 0:28 0:36 0:43 0:50 0:57 1:04 Time (hr:min)
19 Simultaneous Detection of Multiple Compounds in Tape Samples 3 Cummulative Outgas Concentration (ppmv) acetic acid (MA) ammonia (MB) methacrylic acid (MA) siloxanes (MC) 0 0:00 0:07 0:14 0:21 0:28 0:36 0:43 0:50 0:57 1:04 Time (hr:min)
20 FTIR Static Headspace Can quantitate and speciate, in a timely manner, molecular compounds that are outgassed from materials destined for the hard disk drive. Useful on manufacturing floor for routine evaluation of semi-finished or finished goods for unwanted contaminants and control charting. Useful in product development and product certification activities.
Gas monitoring using GASMET FTIR
Gas monitoring using GASMET FTIR Dominic Duggan Quantitech dd@quantitech.co.uk Andrew Hobson Quantitech Antti Heikkila Gasmet Outline of the workshop: FTIR for emissions measurement Applications in ambient
More informationHoneywell Analytics. December 2, Overview of Gas Detection Capabilities. IH Considerations for a TGMS
Honeywell Analytics December 2, 2008 Overview of Gas Detection Capabilities IH Considerations for a TGMS Agenda Semiconductor Industry Issues Gas Monitoring Systems - Code & Corporate Requirements - Industrial
More informationMB3000-CH90 and MBGAS-3000CH FT-IR solutions for laboratory and process gas phase measurements R&D, QA/QC, process optimization and air monitoring
Measurement & Analytics Measurement made easy MB3000-CH90 and MBGAS-3000CH FT-IR solutions for laboratory and process gas phase measurements R&D, QA/QC, process optimization and air monitoring FT-IR technology
More informationMATRIX-MG Series. Innovation with Integrity. Automated High-Performance Gas Analyzers FT-IR
MATRIX-MG Series Automated High-Performance Gas Analyzers Innovation with Integrity FT-IR The MATRIX-MG Series comprises three high-performance FT-IR gas analyzers in a compact and rugged housing. They
More informationFT-IR solutions for laboratory and process gas phase measurements R&D, QA/QC, process optimization and air monitoring
A B B M E A S U R E M E N T & A N A LY T I C S FT-IR solutions for laboratory and process gas phase measurements R&D, QA/QC, process optimization and air monitoring 2 3 F T- I R S O L U T I O N S F O R
More informationPurpose and Scope: Responsibilities: Warnings and precautions: Background Beers law
Purpose and Scope: This document describes the procedures and policies for using the MSE Bruker FTIR Spectrometer. The scope of this document is to establish user procedures. Instrument maintenance and
More informationFOURIER TRANSFORM INFRARED SPECTROSCOPY
Purpose and Scope: This document describes the procedures and policies for using the MSE Bruker FTIR Spectrometer. The scope of this document is to establish user procedures. Instrument maintenance and
More informationFTIR Users Group Meeting FTIR development
FTIR Users Group Meeting FTIR development A bit of history From Oofty. To Spectronus A bit of history Oofty, Univ. Of Wollongong, 1995 They developed a method of trace gas analysis based on 1-cm -1 resolution
More informationInnovative gas and outgassing analysis and monitoring
Innovative gas and outgassing analysis and monitoring Guillaume Genoud - guillaume.genoud@vtt.fi VTT Technical Research Centre of Finland Ltd Environmental Metrology This project has received funding from
More informationHCL MONITORING. ICAC MARAMA WORKSHOP May 18-19th, Presented by: Ty Smith. CEMTEK
HCL MONITORING ICAC MARAMA WORKSHOP May 18-19th, 2011 Presented by: Ty Smith. CEMTEK 1 HCL MONITORING Applications Existing Measurement Technologies New Technologies Calibration and issues 2 APPLICATION
More informationFOUP material influence on HF contamination during queue-time
Solid State Phenomena Online: 2014-09-26 ISSN: 1662-9779, Vol. 219, pp 251-255 doi:10.4028/www.scientific.net/ssp.219.251 2015 Trans Tech Publications, Switzerland FOUP material influence on HF contamination
More informationOnline DGA-monitoring of power transformers
Online DGA-monitoring of power transformers Senior Industry Expert, Vaisala Finland Member of CIGRE A2/D1.47 and D1.52 WGs April 2018 Vaisala Webinar content 1. Dissolved gases why and which gases 2. Fault
More informationS on equipment and pipelines along with the constant threat to personnel safety justify major investment in H 2
Application Summary Analytes: S (hydrogen sulfide) Detector: OMA-300 S Analyzer Process Stream: natural gas Introduction Raw natural gas contains hydrogen suflide, the level of which varies by field. Due
More informationDr. Tetsuzo KITAGAWA, Tokihiko YONEDA, Shigefumi NAKAJIMA. 1. Preface
PORTABLE BREATH ALCOHOL ANALYZER BY NDIR METHOD Dr. Tetsuzo KITAGAWA, Tokihiko YONEDA, Shigefumi NAKAJIMA KOMYO RIKAGAKU KOGYO K. K. 2-14,2-chome, Nakane, Meguro-ku, Tokyo 152, Japan 1. Preface For the
More informationBackground Statement for SEMI Draft document 4657B NEW STANDARD: SPECIFICATIONS FOR TUNGSTEN HEXAFLUORIDE (WF 6 )
Background Statement for SEMI Draft document 4657B NEW STANDARD: SPECIFICATIONS FOR TUNGSTEN HEXAFLUORIDE (WF 6 ) Note: This background statement is not part of the balloted item. It is provided solely
More informationTransfeu WP2 Subtask Development of small-scale test method for fire effluents
Pagina 83 di 89 Transfeu WP2 Subtask 2.1.2 Development of small-scale test method for fire effluents Step 2: Influence of 2 nd filter on FTIR sampling line to prevent HCl and other halogen gases trapping
More informationMid-IR Lasers Market Review and Forecast 2010
Mid-IR Market Review and Forecast 2010 Report OM-56 September 2010 Strategies Unlimited 201 San Antonio Circle, Suite 225 Mountain View, California 94040 Phone: (650) 941-3438 Fax: (650) 941-5120 e-mail:
More informationGravimetric preparation of NO 2 primary reference gas mixtures at the NMISA. Tshepiso Mphamo 03 September 2012
Gravimetric preparation of NO 2 primary reference gas mixtures at the NMISA Tshepiso Mphamo 03 September 2012 NMISA 2012 Contents Introduction Experimental Verification Uncertainty CCQM- Pilot study results
More informationacrolein, acetaldehyde and acetone( cm -1 ); methanol (1306 cm -1 ); ethylene (949 cm -1 ); and isoprene (893 cm -1 ).
acrolein, acetaldehyde and acetone(1550 1800 cm -1 ); methanol (1306 cm -1 ); ethylene (949 cm -1 ); and isoprene (893 cm -1 ). 5 Figure 4a 6 Figure 4b Figure 4c 7 Figure 5 Questions in Student Handout
More informationSAFETY DATA SHEET. Permabond TA4246
SAFETY DATA SHEET 1. IDENTIFICATION OF THE SUBSTANCE / PREPARATION AND THE COMPANY Product name Company Permabond Engineering Adhesives Ltd Wessex Business Park Wessex Way Colden Common Hampshire SO21
More informationCHEMTRACE Fremont Blvd. Fremont, CA 94538, Tel: (510) Fax: (510)
Report Date: 2/12/04 Page 1 of 5 Analysis Requested: Organic Outgassing Analysis by GC-MS at 100 C for 30 minutes. Sample Information: Sample ID: Wafer Carrier Material Samples were received wrapped in
More informationGCMSD-Headspace Analysis SOP
Before you start GCMSD-Headspace Analysis SOP Method and Sequences names are restricted to 50 characters (MSD program will crash otherwise) The GC oven and Injection Ports need to be cooled to 50 oc to
More informationTagging - Teltag System
First Used Initials I - Inspection R - Repair 56 Tagging - System Two Part Label The two part set consists of firstly a rigid polycarbonate holder, back printed, so that the details of the equipment can
More informationIt is widely accepted that what you
Getting a handle on HAPs The Clean Air Act depends on our ability to measure hazardous air pollutants but this is no simple task, as Jim McKinley explains. It is widely accepted that what you cannot measure,
More informationElimination of Percent Level H 2 S Calibration Gas from Flare Gas Monitoring Systems Measuring Total Sulfur, H 2 S and BTU.
Elimination of Percent Level H 2 S Calibration Gas from Flare Gas Monitoring Systems Measuring Total Sulfur, H 2 S and BTU. Control Analytics, Inc 6017 Enterprise Dr Export, PA 15632 (724) 387 CEMS (2367)
More informationSorbent Pen Heavy & Labile SVOC s Taints, Off-Flavors, Odors
Revolutionary extraction technology quantitative by design. Say Goodbye to SPME... and hello to Sorbent Pens! A revolutionary new extraction technology. Sorbent Pen Heavy & Labile SVOC s Taints, Off-Flavors,
More informationMMS Mercury Monitoring System for Natural Gas
MMS Mercury Monitoring System for Natural Gas Natural gas often contains mercury at concentrations that vary from below 1 to above 10000 µg/m 3. Mercury is both toxic and potentially damaging: it can corrode
More informationDegassing ERC KK Saitama, Japan
Degassing Degassing Solutions IDEX Health & Science degassers improve fluidic instrument precision and reliability by removing dissolved gases from fluids before they outgas and form problem-causing bubbles
More informationAdvanced Technology Center. Vacuum Optics. Mark Sullivan December 2, Vacuum Optics 1
Advanced Technology Center Vacuum Optics Mark Sullivan December 2, 2006 Vacuum Optics 1 Outline System design approach Specific steps to build successful vacuum optics ASTM E 1559 outgassing test Resources
More information3M DICHROIC Glass Finishes
3M DICHROIC Glass Finishes DF-PA DF-PA Chill Chill & & DF-PA DF-PA Blaze Blaze Doc TypeInstallation Product #Guide (opt.) Revision D, April 2018 Product Description 3M DICHROIC Glass Finish DF-PA (the
More informationInprocess Operator Training Programme
2016 Inprocess Operator Training Programme ABSORPTION COLUMN These exercises are intended to provide an understanding of absorption columns and the fundamental principles they use to eliminate pollutants
More informationCalibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.
Advanced Calibration Designs, Inc. Calibration Gas Instrument INSTRUCTION MANUAL Release I www.goacd.com Instruction Manual Gas Generator Release I TABLE OF CONTENTS I. General Description Page 2 II. Start-Up
More informationSchool of Chemistry SOP For Operation Of Glove Boxes
School SOP for Operation of Glove Boxes: The following SOP provides guidelines on how to adequately and safely operating a standard laboratory glove box and its associated devices. General Information:
More informationDetector Tubes and Badges
Detector Tubes and Badges 1 Detector Tubes Elements of detector tube system Types of tubes Active (short- and long-term) multi-level Passive (diffusion controlled) References: pp. 692-708 in Perkins Overview
More informationINTRODUCTION TO THE SPECTROPHOTOMETER AND PIPETTING SKILLS
INTRODUCTION TO THE SPECTROPHOTOMETER AND PIPETTING SKILLS Section A: Intro to the spectrophotometer A commonly used instrument in the analysis of cellular extracts is the Spectrophotometer. Today you
More informationThermo K-Alpha XPS Standard Operating Procedure
Thermo K-Alpha XPS Standard Operating Procedure Quick Guide Draft v.0.1 Procedure overview 1. Vent the loadlock 2. Secure your sample to the stage using clips, check the height of the final assembly. 3.
More informationGenerating Calibration Gas Standards
Technical Note 1001 Metronics Inc. Generating Calibration Gas Standards with Dynacal Permeation Devices Permeation devices provide an excellent method of producing known gas concentrations in the PPM and
More information(2 pts) Draw the line of best fit through the data and estimate the concentration of Fe in your sample solution.
Name: 1. You are asked to measure the percent iron in a steel sample you believe to be about 50% Fe by mass. You dissolve four replicate samples in nitric acid and dilute them by mass to a concentration
More informationCambridge Refrigeration Technology
Cambridge Refrigeration Technology www.crtech.co.uk. Guide to Sample Taking and Flame Testing Refrigeration Units and Bottles for Chlorine Contamination 8 th May 2013 Cambridge Refrigeration Technology
More informationSelection of Materials and Processes for Space Use. Ruben Edeson, Mechanical Engineer, SSTD.
Selection of Materials and Processes for Space Use Ruben Edeson, Mechanical Engineer, SSTD. Materials Selection for Space Mechanical properties, ie specific stiffness Thermal and optical properties Cost
More informationUSER'S MANUAL HOT-WIRE TYPE GAS SENSOR NAP-100A SERIES ( NET10A(R) ) (For All Combustible Gases)
(June 27, 2002) USER'S MANUAL HOT-WIRE TYPE GAS SENSOR NAP-100A SERIES ( NET10A(R) ) (For All Combustible Gases) 1. General Being designed for light industrial applications, NAP-100A series also available
More informationGAF Safety Data Sheet SDS # 2104 SDS Date: December 2014
GAF Safety Data Sheet SDS # 2104 SDS Date: December 2014 SECTION 1: PRODUCT AND COMPANY INFORMATION PRODUCT NAME: TRADE NAME: CHEMICAL NAME / SYNONYM: CHEMICAL FAMILY: MANUFACTURER: EverGuard H2O Bonding
More information3. Coupled Techniques in Thermal Analysis
3. Coupled Techniques in Thermal Analysis Coupled Techniques in Thermal Analysis as Evolved Gas Analysis (EGA) TG/FT-IR Technique for analyzing gas evolved from TG with FT-IR TG/MS Technique for analyzing
More informationReal-time Analysis of Industrial Gases with Online Near- Infrared Spectroscopy
Real-time Analysis of Industrial Gases with Online Near- Infrared Branch 5 Petrochemical, Biofuels Keywords Process, NIR, Gases, Gas Cell, Ethylene, Carbon Dioxide, Propane, Butane, Acetylene Summary Near-infrared
More informationChemiPPR. Response of hazardous materials LAMK. Sakari Halmemies 26 Sep2016
ChemiPPR Response of hazardous materials LAMK Sakari Halmemies 26 Sep2016 Course Objectives General (1/2) - to learn to know principles of tactics and methods when responding to a chemical accident - to
More informationSetting a New Standard in Flash Chromatography Performance
Setting a New Standard in Flash Chromatography Performance What s New Faster Flow rates, up to 300 ml/min Higher pressures, up to 300 psi (20 bar) Bigger touchscreen, 12 or 15 inches Wider Dynamic Range
More informationColumbus Instruments
Micro-Oxymax Layman s Guide to Measurement Principles Introduction: The Micro-Oxymax Respirometer (Figure 1) measures the production and/or consumption of many gases simultaneously. Data is typically reported
More informationContact Person(s) : Isabel M. Fisenne APPLICATION
Ra-02-RC RADIUM-226 - EMANATION PROCEDURE Contact Person(s) : Isabel M. Fisenne APPLICATION This procedure is specific and may be applied to almost any matrix which can be converted to a homogeneous solution.
More informationChemical Spill Clean-up Protocol
Protocol Department of www.rms.ubc.ca UBC-RMS-OHS-PRO 17-002 Date created: November 3, 2015 Date revised: April 28 th, 2017 Chemical Spill Clean-up Protocol Scope Laboratory-specific safe work procedures
More informationStandard Operating Procedure (SOP)
INTRODUCTION This SOP applies to. FORMALDEHYDE exposure in Anatomy Lab will be monitored at the beginning of each semester dissections are performed. Environmental Health and Safety will perform monitoring
More informationUsing Vaporized Hydrogen Peroxide
for Bio-Decontamination D-111046-2013 Using Vaporized Hydrogen Peroxide In 1994, Dräger introduced the first electrochemical hydrogen peroxide (H2O2) sensor for monitoring low concentrations of vaporized
More informationABB MEASUREMENT & ANALYTICS. Continuous gas analyzers EL3060 The specialists for hazardous areas
ABB MEASUREMENT & ANALYTICS Continuous gas analyzers EL3060 The specialists for hazardous areas The specialists for hazardous areas ABB Analytical possesses decades of experience in the area of explosion
More informationSection 1 Chemical Property and Company Identification. Section 2 Hazards Identification. Emergency Overview
Safety Data Sheet FreeDerm Pump Spray and Ampoule Section 1 Chemical Property and Company Identification SDS Name: FreeDerm Pump Spray Company Identification: BioDerm, Inc. 12320 73 rd Court N Largo, FL
More informationRapid and Reliable Detection of Dissolved Gases in Water
Rapid and Reliable Detection of Dissolved Gases in Water Andrea Caruso and Massimo Santoro Thermo Fisher Scientific, Milan, Italy Application Note 005 Key Words Chromeleon CDS, Environmental, Fracking,
More informationUT 3000 Mercury Ultratracer
MERCURY INSTRUMENTS ANALYTICAL TECHNOLOGIES UT 3000 Mercury Ultratracer Measuring of Total Gaseous Mercury (TGM) in Air and other Gases at Ultra Trace Levels. Applications: Air quality studies Monitoring
More informationDate: Review Date: Assessment Reference: Experiment or Procedure (include a brief description, including equipment used and frequency of exposure):
Date: Review Date: Assessment Reference: Workers involved: Experiment or Procedure (include a brief description, including equipment used and frequency of exposure): Phytolith extraction from soil, separation
More informationCommercial Practice Test Method Internal Vapor Analysis of Hermetic Devices
Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices Oneida Research Services, Inc. 8282 Halsey Road Whitesboro, NY 13492 Phone: (315) 736-5480 FAX: (315) 736-9321 1.0 Purpose The
More informationABB MEASUREMENT & ANALYTICS. Continuous gas analyzers EL3060 The specialists for hazardous areas
ABB MEASUREMENT & ANALYTICS Continuous gas analyzers EL3060 The specialists for hazardous areas The specialists for hazardous areas ABB Analytics possesses decades of experience in the area of explosion
More informationWhat I have learned about SF 6 gas testing.a Practical explanation
What I have learned about SF 6 gas testing.a Practical explanation Over the past few years, testing of SF6 gas quality has become increasingly important and with the EPA in USA, setting guidelines for
More informationGas Release from Solids, Materials for Vacuum and Leak Detector
VACUUM AND PLASMA SCIENCE AND TECHNOLOGY W9 Gas Release from Solids, Materials for Vacuum and Leak Detector Ming-Show Wong Nov. 2015 Gas Release from Solids Vaporization Thermal Desorption First Order
More informationHIGH ACCURACY MULTI-GAS MONITORING USING AUTOMATED SELF- CALIBRATION
HIGH ACCURACY MULTI-GAS MONITORING USING AUTOMATED SELF- CALIBRATION SCENTROID is the global leader in sensory technology providing intelligent environmental solutions across a wide range of industries
More informationLaser Spectrometers for Online Moisture Measurement in Natural Gas. Ken Soleyn GE M&C
Laser Spectrometers for Online Moisture Measurement in Natural Gas Ken Soleyn GE M&C ken.soleyn@ge.com Introduction TDLAS (Tunable Diode Lase Absorption Spectroscopy) Moisture Analyzers have become the
More informationSUBSTATION MAINTENANCE ELECTRICAL OPERATING PROCEDURE
Page 1 of 6 A. PURPOSE AND SCOPE This procedure establishes safe working practices for handling sulfur hexafluoride (SF 6 ) gas and release minimization practices. This document also specifies methods
More informationPlasma-Therm PECVD. Operating Characteristics. Operating Instructions. Typical Processes. I. Loading. II. Operating
Plasma-Therm PECVD A PECVD (plasma enhanced chemical vapor deposition) reacts gases in a RF (radio frequency) induced plasma to deposit materials such as silicon dioxide and silicon nitride. This PECVD
More informationASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist
ASTM WK 43267 Standard Test Method for Dissolved Gases Anne Jurek Applications Chemist Rationale Hydraulic Fracturing is becoming more and more common. Nearby water well can be affected during the fracking
More informationSupporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial
Supporting Information for Micro-Collection of Gases in a Capillary Tube: Preservation of Spatial and Temporal Resolution AUTHOR INFORMATION Kristin A. Herrmann Favela, 1 * Pieter Tans, 2 Thomas Jaeckle,
More informationFINAL REVISED TECHNICAL SPECIFICATION OF INDUCTIVELY COUPLED PLASMA MASS SPECTROMETER (ICP-MS) WILL BE AS FOLLOWS:- GENERAL SPECIFICATIONS
Ref No. INST/12(121)/2018-Pur Date: 31/12/2018 CORRIGENDUM Reference to NIT no. INST/12(121)/2018-Pur published in national newspapers on 07/09/2018 for purchase of equipment: ICP-MS WITH ACCESSORIES.
More informationIonic outgassing from photoresist compositions upon irradiation at 13.5 nm
Ionic outgassing from photoresist compositions upon irradiation at 13.5 nm Grace H. Ho, Chih- Jen Liu, Chih- Han Yen, Ming- Hsuan Ho, Shih-Yu Wu, and Yu-Hsian Shih Department of Applied Chemistry National
More informationEU Safety Data Sheet. habasit. Vulcol (Version 99)
Company Page 1 of 8 Habasit AG, Römerstrasse 1, CH-4153 Reinach Tel. ++41 (0)61/ 715 15 15, Telefax ++41 (0)61/ 715 15 55 1 1.1 Commercial product name and supplier Commercial product name / Designation
More informationMATERIAL SAFETY DATA SHEET
Effective and Print Date: 12/01/2014 MSDS #1222014 Page #1 SECTION I - PRODUCT AND COMPANY IDENTIFICATION Product Code: 122 Part A (Resin) Product Name: NSP 122 Industrial Floor Coating Product Class:
More informationOff-gassing from III-V Wafer Processing -- A Collaboration of SEMATECH and IMEC
Off-gassing from III-V Wafer Processing -- A Collaboration of SEMATECH and IMEC CMPUG Spring Meeting Albany, NY April 16, 2015 Hsi-An Kwong, P.E. Alain Pardon Steve Trammell, P.E. SEMATECH IMEC EORM hsi-an.kwong@sematech.org
More information2B Technologies, Inc.
2B Technologies, Inc. An InDevR Company Technical Note No. 018 Cleaning Procedure for Model 202 Ozone Monitor Date: 22 September 2008; Updated 8 July 2016 Author: John Birks Summary: It is recommended
More informationFocus on VOC Emissions Reduction Using an Oxygen Based Inerting Control System For Inert Gas Blanketing of Chemical Process Vessels
Inerting Control Systems by NTRON PROCESS ANALYZER DIVISION of NEUTRONICS INC. EXTON. PA. 19341 RJN Rev. A June 2001 Focus on VOC Emissions Reduction Using an Oxygen Based Inerting Control System For Inert
More informationHot extractive multi-channel gas analyser for continuous process and emissions monitoring.
Product Data Sheet CODEL Continuous Emission Monitoring GCEM4100 Extractive Gas Analyser CO, CO2, NO, NO2, SO2, HCL, CH4 & H2O Hot extractive multi-channel gas analyser for continuous process and emissions
More informationLeak Checking Large Vacuum Chambers
Leak Checking Large Vacuum Chambers Technical Overview Vacuum Technologies Introduction Understanding the pump-down characteristics of a large vacuum vessel is critical for determining whether the vacuum
More informationSection 1. Registration Information
Section 1. Registration Information Source Identification Facility Name: Parent Company #1 Name: Parent Company #2 Name: Submission and Acceptance Submission Type: Subsequent RMP Submission Reason: Description:
More informationQuantifying Cryocooler Contamination
C19_056 1 Quantifying Cryocooler Contamination P. S. Spoor Chart Industries Troy, NY 12180 ABSTRACT Contamination of the cryocooler working gas (usually helium) with other species results in performance
More informationMETHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the
1151 METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS 1.0 Scope and Application. 1.1 Analytes. Analyte Volatile Organic Compounds (VOC) CAS No. No CAS number assigned 1.2 Scope. This method
More informationSPILL KITS UNIVERSAL / CHEMICAL. Suitable for water, oils, fuels, solvents, acids and bases OIL-ONLY.
SPILL KITS - Contain and clean up spills quickly - Reduce clean up and disposal costs - Kits are stocked to handle spills up to 310 liters - Choose either Universal or Oil-Only for your facility UNIVERSAL
More informationCHLORINE DIOXIDE GENERATOR
CHLORINE DIOXIDE GENERATOR M I N I Pressurized system capacities up to 20gr/h A I R Pressure-less system for multi-point injection up to 8gr/h M A X I Pressurized system capacities up to 1000gr/h E A S
More informationMcIlvaine Company Hot Topic Hour. April 9, Dry Sorbent Injection Options and Issues
McIlvaine Company Hot Topic Hour April 9, 2015 Dry Sorbent Injection Options and Issues Marty Dillon Flue Gas Treatment Specialist Lhiost North America Presentation Overview Lhoist Overview Dry Sorbent
More informationHSE PRACTICES FOR OFFSHORE AND PETROLEUM INDUSTRIES SAFETY PRACTICES NPTEL TUTORIAL: 12
HSE PRACTICES FOR OFFSHORE AND PETROLEUM INDUSTRIES SAFETY PRACTICES TUTORIAL: 12 COURSE CORDINATOR Prof. SRINIVASAN CHANDARSEKARAN B.E (civil engg)., M.Tech (Structural engg- IIT Delhi); Ph.D (IIT Delhi).,
More informationppm mg/m3 ppm mg/m3 ppm mg/m3 Bisphenol A/Epichlorohydrin Based Epoxy Resin < 5
1. CHEMICAL PRODUCT AND COMPANY IDENTIFICATION Product Name: www.tra-con.com TRA-BOND F113SC Hardener TRA-CON, INC. 45 Wiggins Avenue Bedford, MA 01730 781-275-6363 1-800-872-2661 Emergency Telephone:
More informationCalibration and Bump Test Requirements for Direct Reading Portable Gas Monitors
: Calibration and Bump Test Requirements for Direct Reading Portable Gas Monitors Manufacturers and regulatory agencies agree the safest and most conservative approach is to perform a functional test by
More informationPlasma Sources and Feedback Control in Pretreatment Web Coating Applications
Plasma Sources and Feedback Control in Pretreatment Web Coating Applications Joseph Brindley, Benoit Daniel, Victor Bellido-Gonzalez, Dermot Monaghan Gencoa Ltd., Physics Rd, L24 9HP Liverpool, UK (+44)
More informationConfined Space Entry Appendix - Sewerage Transfer. 1. Purpose Sewerage Transfer Assets. 2. Scope. 3. Procedure
1. Purpose Sewerage Transfer Assets This procedure defines specific requirements in addition to CORP H&S 053 Confined Spac e Entry and is designed to be read in conjunction with that document for work
More informationGas Detection. Portable Gas Detection for Safety in Confined Spaces
Gas Detection Portable Gas Detection for Safety in Confined Spaces Portable Gas Detection for Safety in Confined Spaces An Overview In today s working environment confined spaces and confined space entry
More informationRemoval of Lead-Based Paint
Safety Management System SP24-61 Process Authority: HS Manager Removal of Lead-Based Paint Approval: Chief Operating Officer Version Date:30/03/2015 Revision: B 1 Purpose To define the requirements for
More informationDocumentation Beats Confrontation in Gas Detection
Session No. 770 Documentation Beats Confrontation in Gas Detection Andrew Saunders, CQT Applications and Training Specialist BW Technologies by Honeywell Introduction Deaths of workers in confined spaces
More informationDOW CORNING CORPORATION Material Safety Data Sheet
Page: 1 of 9 1. IDENTIFICATION OF THE SUBSTANCE AND OF THE COMPANY Dow Corning Corporation South Saginaw Road Midland, Michigan 48686 24 Hour Emergency Telephone: Customer Service: Product Disposal Information:
More informationSINEO Microwave Digestion Core Technology (2012)
SINEO Microwave Digestion Core Technology (2012) Ⅰ. Technology of Vessel: 1. Composite fiber outer vessel In recent years, the application of composite fibers is becoming increasingly popular, and it is
More informationRevision No:06 Compiler: Mr F M Jachim. Liquid - 1,5, & 20 litre cans.
SAFETY DATA SHEET - REGULATIONS 1994 (CHIP 2) WE urge the customer receiving this Material Safety Data Sheet to study it carefully to become aware of hazards, if any, of the product involved. In the interests
More informationBasic & Controlled Atmosphere Glove Boxes
I N T R O D U C T I O N Precise Basic and Controlled Atmosphere Glove Boxes provide a leak-tight environment for work with contamination-sensitive materials. Precise Basic Glove Boxes are simple, economical
More informationPage: 1 of 8 Version: 1.9 Revision Date: 06/24/2014 SILA-SEAL CLEAR INDUSTRIAL GRADE SILICONE SEALANT. MSDS No.: Revision Date: 06/24/2014
I Page: 1 of 8 1. PRODUCT AND COMPANY IDENTIFICATION SILA-SEAL 5167 East 65th Street Indianapolis, Indiana 46220 24 Hour Emergency Telephone: Customer Service: Product Disposal Information: CHEMTREC: (989)
More informationA NOVEL SENSOR USING REMOTE PLASMA EMISSION SPECTROSCOPY FOR MONITORING AND CONTROL OF VACUUM WEB COATING PROCESSES
A NOVEL SENSOR USING REMOTE PLASMA EMISSION SPECTROSCOPY FOR MONITORING AND CONTROL OF VACUUM WEB COATING PROCESSES F. Papa 1, J. Brindley 2, T. Williams 2, B. Daniel 2, V. Bellido-Gonzalez 2, Dermot Monaghan
More informationEXPERIMENT XI. Careful!! Improper handling of the vacuum line may result in the release of SO 2 which is an irritating and suffocating gas.
Chem 366-3 Page XI - 1 EXPERIMENT XI INFRARED SPECTRUM OF SO2 (S&G, 5th ed. Expt 36, 6th ed. Expt. 35) 1. Pre-Lab preparation. The description of this experiment has disappeared from the more recent editions
More information973-SF 6 Analyzer. Precise and Stable SF 6 Gas Analyzer REFLECTING YOUR STANDARDS
Precise and Stable SF 6 Gas Analyzer Simultaneous measurement of humidity, SF 6 purity and SO 2 concentration Integrated gas recovery system with automatic pump back Fully automated SF 6 gas testing Fundamental
More informationvermiculite, dry sand, earth), and place in a chemical waste container. 5-2
MSDS for MONOETHANOLAMINE 1. Product Identification Synonyms: Monoethanolamine; Ethanolamine; beta-aminoethanol, ethylolamine; glycinol CAS No.: 141-43-5 Molecular Weight: 61.08 Chemical Formula: NH2CH2CH2OH
More informationAgilent 5500 Series FTIR Operation Manual
Agilent 5500 Series FTIR Operation Manual Agilent 5500 Series FTIR Operation Manual Notices Agilent Technologies, Inc. 2011 No part of this manual may be reproduced in any form or by any means (including
More informationChapter 6 Leak detection
Chapter 6 Leak detection 1 2 1. Leak rate q L = V p t mbar l s -1 A leak rate of 1 mbar l s -1 exists in a closed vessel having a volume of 1 liter when the pressure increases by 1 mbar within one second,
More information