Instruction Manual MicroChem 2 Chlorine Probes Series CL4000
|
|
- Joel Gaines
- 5 years ago
- Views:
Transcription
1 Instruction Manual MicroChem 2 Chlorine Probes Series CL
2 These instructions describe the installation, operation and maintenance of the subject equipment. Failure to strictly follow these instructions can lead to an equipment rupture that may cause significant property damage, severe personal injury and even death. If you do not understand these instructions, please call De Nora Water Technologies for clarification before commencing any work at and ask for a Field Service Manager. De Nora Water Technologies, Inc. reserves the rights to make engineering refinements that may not be described herein. It is the responsibility of the installer to contact De Nora Water Technologies, Inc. for information that cannot be answered specifically by these instructions. Any customer request to alter or reduce the design safeguards incorporated into De Nora Water Technologies equipment is conditioned on the customer absolving De Nora Water Technologies from any consequences of such a decision. De Nora Water Technologies has developed the recommended installation, operating and maintenance procedures with careful attention to safety. In addition to instruction/operating manuals, all instructions given on labels or attached tags should be followed. Regardless of these efforts, it is not possible to eliminate all hazards from the equipment or foresee every possible hazard that may occur. It is the responsibility of the installer to ensure that the recommended installation instructions are followed. It is the responsibility of the user to ensure that the recommended operating and maintenance instructions are followed. De Nora Water Technologies, Inc. cannot be responsible for deviations from the recommended instructions that may result in a hazardous or unsafe condition. De Nora Water Technologies, Inc. cannot be responsible for the overall system design of which our equipment may be an integral part of or any unauthorized modifications to the equipment made by any party other that De Nora Water Technologies, Inc. De Nora Water Technologies, Inc. takes all reasonable precautions in packaging the equipment to prevent shipping damage. Carefully inspect each item and report damages immediately to the shipping agent involved for equipment shipped F.O.B. Colmar or to De Nora Water Technologies for equipment shipped F.O.B Jobsite. Do not install damaged equipment. De Nora Water Technologies, COLMAR OPERATIONS COLMAR, PENNSYLVANIA, USA ISO 9001: 2008 CERTIFIED
3 TABLE OF CONTENTS 2 1 INTRODUCTION General Description Safety Chlorine Probes, Cables Accessories and Consumables Technical Data / Specifications Principle of Operation INSTALLATION General Mounting Dimensions Chlorine Probes Flow Cell Flow Regulation Components Electrical STARTUP MAINTENANCE Periodic functional Check Electrode Maintenance Electrode Membrane Cleaning the electrode surfaces TROUBLE SHOOTING PROCEDURE...16 FIGURES 1 Chlorine Probe Mounting Dimension Chlorine Probe Components Arrangement, Piping, CL4000 Sensor Probe Cable Connections
4 1 INTRODUCTION 1.1 General Description This manual describes the general description, installation, operation and maintenance of the CL4000 Chlorine probes and accessories. Chlorine is the most common chemical used in disinfection of municipal drinking water, cooling towers and many industrial applications. Once added to water the chlorine will react with chemicals already in the water forming combined chlorine or stay available as free chlorine. A chlorine measuring system incorporating the CL4000 chlorine probes, connected to the MicroChem 2 transmitter, provides automatic measurement/control of chlorine dosage in such systems. The CL4000 chlorine probe measures free chlorine. The transmitter is part of the MicroChem 2 series, which is capable of supporting up to three of the following measurements in all possible combinations: Dissolved Oxygen, Chlorine Dioxide, Residual Chlorine, Bromine, Ozone, ORP, Temperature, ph, Iodine and Conductivity. For a description and instructions regarding the transmitter see the MicroChem 2 transmitter instruction manual. The chlorine probes are 3 electrode amperometric cells covered with a permeable membrane. The probe exterior is manufactured from a durable PVC and stainless steel and has a diameter of 25mm (1 inch). The probes are engineered to fit a specially designed flow cell. De Nora Water Technologies chlorine probes can be used without the need for ph compensation. All the CL4000 probes are designed to run without the need for reagents. The probes have no moving parts and are designed to require minimal maintenance. Figure 1 - Chlorine Probe 1.2 Safety The recommended operating procedures have been designed with careful attention to safety. De Nora Water Technologies has made formal safety reviews of the initial design and any subsequent changes. This procedure is followed for all new products and covers areas in addition to those included in applicable safety standards. Observe the following safety precautions: a. Observe all safety warnings marked on the equipment. These warnings identify areas of immediate hazard, which could result in personal injury, or loss of life. b. Do not use this equipment for any purpose other than described in this manual. c. Disconnect power prior to making any terminal connections within the enclosure. d. Use the recommended connection procedures described in Section 2 of this manual. This equipment should only be installed by suitably qualified personnel
5 1.3 Chlorine Probes, Cables, Accessories and Consumables Chlorine Probe 0-2 ppm, ph tolerable (Note 1,2) Chlorine Probe 0-5 ppm, ph tolerable (Note 1,2) Chlorine Probe 0-10 ppm, ph tolerable (Note 1,2) Foot Cable (Note 3) Foot Cable (Note 3) Cable (Note 3) Sample/Drain Tubing, CLEAR, 3/16 ID X 5/16 OD R-376 Flow Through Cell with integral shut off valve (Note 4) Flow Through Cell and Flow Control Kit (Note 4, 6) Sample Flow Control Kit (Note 7) Flow Control Panel Kit (Note 6) Y Strainer 80 mesh, Max Inlet 75 PSIG Pressure Reducing Valve Kit, 0-25 PSIG, Max Inlet 300 PSIG w/15 PSIG Gauge & Bracket Flow Meter Assembly w/rate Valve, 0-10 GPH Flow Switch (Note 7) Carbon Dechlorinating Filter (Note 9) Annual Maintenance Kit for: and Probes Includes: PN and PN , 23184, 23185, and Probes Includes: PN and PN and Probes Includes: PN and PN M48 Membrane Cap to fit 23183, 23184, 23185, , , and Probes M20 Membrane Cap to fit and Probes G-Holder for 23183, 23184, 23185, and Probes Electrolyte Solution (100ml) for and Probes Electrolyte Solution (100ml) for 23183, 23184, 23185, and Probes Electrolyte Solution (100ml) for and Probes Notes: 1. The MicroChem2 Transmitter/Controller must be fitted with the expanded function board (P/N 1T686B129U12, U13 or U14) in order to provide power to the probe and the software must be Version 2.8 or above. Consult factory on upgrading earlier versions of microchem2 transmitter/controllers to the latest revision , 23184, and 23185, probes with 3-electrode sensor for the measurement of free available chlorine. Temperature compensated (0 to 40 C). ph tolerable within the operating range 4 to Cable , or must be used for probe to function properly. 4. Flow cell includes a low pressure fitting (installed) and is also provided with a back pressure fitting to prevent the formation of air bubbles in the probe chamber at sample pressures of 6-8 PSIG ( bar). 5. Flow Control Kit (P/N 23318) improves flow and pressure control of CL4000 probes. It includes a PRV Kit (P/N 23425), Flow Meter Assembly (P/N 23426), 10 ft of tubing and connectors. 6. The Flow Control Panel Kit (P/N 23451) (26" H X 28 W X 0.5" thick) is predrilled and tapped for mounting the Microchem2 and 2 acrylic flow cells for chlorine or for use with ph, conductivity, or ORP 12mm probes and is supplied with mounting hardware. 7. Flow Switch: dry contact opens on loss of sample flow. 8. Dechlorinating Filter provides a chlorine free sample stream, used for calibrating the MicroChem
6 1.4 Technical Data / Specifications Quality Management System: The design & manufacturing process of De Nora Water Technologies are certified to ISO9001. Range: Type of measurement: Temperature Compensation: Speed of Response: Ambient/sample Temp: Limit of Detection: Reproducibility: Accuracy: Sample Inlet Pressure: Sample Temperature: Sample Flow: Electrodes: 0-2 mg/l, 0-5 mg/l or 0-10 mg/l (probe dependant) Free chlorine Internal Ninety (90)% step change < 2 minutes. 32 F F (0 C - 40 C) Probe, chlorine, 0-2 ppm : 0.01 ppm Probe, chlorine, 0-5 ppm : 0.03 ppm Probe, chlorine, 0-10 ppm : 0.05 ppm 0.05 mg/l ± 5% of range 0 to 8 psi (0 to.5 bar) 32 F F (0 C - 40 C) 8 gph (30 liters per hour) Working Electrode - gold Counter Electrode - stainless steel Reference Electrode: Silver/Silver-Chloride ph: Chlorine: independent of ph (ph 4 to ph 9.5) Signal loss is 5% per ph unit (ph greater than 7) Dimensions: In flow cell: 14.6 (h) x 2.25 (dia) 370mm x 55 mm Probe alone 1 (dia) x 9 (l), 25mm (dia) x 230mm (l) Weight: probe and flow cell 1lb (450g) Connectors: Cable: Sample Limitations: Ammonia Interference: Permanganate Interference: 2 x compression fitting for flexible tubing ID 3/16, OD 5/16 Various lengths available Samples containing particles 100 microns (0.004 inches) in diameter and larger may require pre-filtration. There should be no surfactants in the water. CL4000 probes in the presence of ammonia will measure free and/or combined chlorine. 1 ppm permanganate = 0.5 ppm chlorine
7 1.5 Principle of Operation The chlorine probes are amperometric cells covered with a permeable membrane. The probes are engineered to fit a specially designed flow cell. A sample of liquid is delivered to the input (bottom) of the flow cell at an approximate rate of 500 ml/min. The sample passes the membrane on the tip of the membrane cap. Chlorine diffuses through the membrane and into the internal filling solution where a small DC current is generated between the stainless steel counter electrode and gold working electrode in direct linear proportion to the amount of residual present in the sample. A third reference electrode establishes a constant potential on the gold working electrode to provide an accurate and stable residual signal. The signal is amplified and transmitted to the MicroChem 2 transmitter/controller via the sensor cable. The probe needs to be loop powered. This power is generated within the MicroChem 2 and supplied to the probe in the same cable which transmits the residual signal. All the CL4000 chlorine probes are designed to run without the need for reagents. The probes have no moving parts and are designed to require minimal maintenance
8 2 INSTALLATION 2.1 General Remove the instrument from its box and carefully inspect each item and report damages immediately. The box will include the chlorine probe (electrode shaft, top cap and membrane cap), electrolyte bottle and a sheet of blue abrasive paper. The chlorine probes are designed for general duty indoor installation. Outdoor installation is possible if the instrument is shielded from dripping water and not mounted in direct sunlight. Ensure that the process and environmental conditions are within specified limits (see specification). Excessive heat and vibration will affect the sensors. 2.2 Mounting Dimensions Sample connectors are compression fitting, requiring flexible tubing 3/16 ID x 5/16 OD Figure 2 - Mounting Dimensions
9 2.3 Chlorine Probes WARNING Do not touch the electrode finger (see Figure 3) Do not touch the bottom of the membrane cap (see Figure 3) To assemble the probes for use: Vent hole Vent Seal When removing the membrane cap, move the vent seal away from the vent hole to allow air into the membrane cap. Not doing so may cause a vacuum to form and damage the membrane. Fill the membrane cap to the top with electrode filling solution (note each electrode has a specially formulated filling solution which is unique to that sensor see parts list for details). Tap the side of the membrane cap to remove any air bubbles that may have formed in the solution. Reposition the vent seal, making sure that the vent hole is completely covered. Fill up the G-holder with filling solution. Be careful that there are no bubbles. Hold the electrode shaft upright and push the electrode finger carefully into the filled G-holder. While holding the electrode shaft vertically with the G-holder in place, screw the filled membrane cap back on to the electrode shaft. Make sure not to press on the vent ring seal or cover the area of the vent hole with a finger when installing the membrane cap. Excess filling solution will escape from top or the vent in the membrane cap. Rinse off any filling solution that may have escaped during the installation with clean water
10 Note: The G-holder is supplied and required for use with the chlorine probes and with the annual maintenance kits containing the M48 membrane cap. See parts list for details. Note: When unscrewing the filled membrane cap from the electrode shaft, first remove the vent seal and ensure that the vent is not covered by a finger (air needs to be allowed through the vent hole to equalize the pressure). Note: Make sure the membrane cap is tightly fastened to the electrode shaft ensuring that no gap remains between the electrode shaft and the membrane cap. Note: It is best to wait 1-4 hours before the first calibration is performed. After one day repeat the calibration. See the MicroChem 2 electronics manual for calibration instructions. Note: Calibration must be made with a continuous flow of liquid at 500 ml/min across the cell membrane
11 Figure 3 - Chlorine Probe Components
12 2.4 Flow Cell The probe is inserted through a fitting which includes a slide ring and o-ring. Loosen or disassemble the fitting to allow the probe to slide through the o-ring. The fitting may include a cap through which the probe cable must be fed. Position the probe so it is about 1 inch (25 mm) from the bottom of the cell. Tighten the fitting into the housing to compress the o-ring around the probe, providing a seal. The inlet valve may be used to provide control the flow to the probe. It should be full open if the sample stream is controlled by other components. The adapter to be used on the outlet from the cell and is dependent on the sample supply. The sample flow as noted in the section 1.4 is required. Excessive air in the sample may adversely affect the readings. The small orifice adapter may be used to provide back pressure to keep the air from outgasing from the sample stream. 2.5 Flow Regulation Components (See Figure 4) To provide a sample to be measured for chlorine content with reduced potential for erratic readings and calibration problems, it is highly recommended that some or all of the components in Figure 4 are incorporated in the sample line. The Y-strainer is for particulate removal. The flow switch provides alarm when flow is not present. The remaining components through the back pressure assembly will ensure that you are providing the proper flow and pressure to allow optimum analysis. The back pressure assembly is used in applications where air bubbles effervesce from the sample
13 Figure 4 Arrangement, Piping, CL4000 Sensor Key Part Number Description Y-strainer - (optional) for particulate removal (See Note 1) Pressure Reducing Valve, 0-25 PSIG (See Note 1) Pressure Gauge, 0-15 PSIG, ¼ NPT SS (See Note 1) Flowmeter with Rate Valve, 0-10 GPH (See Note 2) Flow Switch CL4000 Flow Cell Fitting, ¼ MNPT x 5/16 Tube Fitting, 1/8 MNPT x 5/16 Tube 90 Deg Notes: 1. Included in Flow Control Kit P/N Used to indicate if sample flow is present. For connection to external device or alarm contact
14 2.6 Electrical The chlorine probes are connected to the transmitter analog board at pin 5 and 6 AND to the 24v power supply terminals (see Figure 5). For more details on the analog board see the transmitter manual. WARNING The transmitter power MUST be switched OFF before the probe is attached Terminals 9-12 are +24v Terminals are 0V (G) (6) (+) (5) (-) - + probe S4000 Cable Assembly 4 feet feet feet Figure 5 - Probe Cable Connections WARNING For proper operation of the chlorine probe, one of the above listed cable Assemblies must be used
15 3 STARTUP 1. Check that the entire installation is connected properly. Check the electrode is properly connected via the supplied cable. 2. Configure the transmitter channel that the probe is connected to for CL (Chlorine). Select CL probe (ma) as the probe type. Set the probe range (0-2, 0-5 or 0-10 ppm) as indicated on the chlorine probe label. Refer to the MicroChem 2 Transmitter instruction manual for additional start-up information. 3. A calibration must be performed. Please find the instructions related to the procedure in the MicroChem 2 instruction bulletin. 4. Output from CL4000 probes is never below 3.8 ma. At start-up, the probes may have an output as low as 2 to 2.5 ma, but will rise to about 4 ma after half an hour. If output is less, the electrolyte will need to be changed and the gold probe cleaned. Momentary power losses will also result in low output, but will increase in about 30 minutes. 4 MAINTENANCE 4.1 Periodic Functional Check The MicroChem 2 analyzer, like other analyzers, should be checked once per week to assure the best measurement accuracy. 4.2 Electrode Maintenance There is very little maintenance required on a chlorine electrode. Periodically the electrode should be removed, and rinsed with de-ionised water. CAUTION Do not touch the tip of the electrode Electrode Membrane The membrane cap needs to be replaced periodically (typically annually). The internal filling solution needs to be replaced at the same time (section 2.3) Note: pay particular attention to the notes on removing the membrane cap. If the membrane gets coated it is possible to clean it. Very carefully swirl the membrane in a 1% hydrochloric acid solution Cleaning the electrode surfaces If the electrode has been stored for a long period of time the (gold) working electrode surface may become tarnished and require polishing. Empty out the internal filling solution (section 2.3). Dry the electrode finger carefully using a clean dry cloth.holding the blue abrasive paper flat on a clean dry surface and holding the electrode body vertically move the tip of the electrode across the paper two or three times. Refill and reattach the membrane cap. (see section 2.3)
16 5 TROUBLE SHOOTING PROCEDURE PROBLEM PROBABLE CAUSE CORRECTIVE ACTION No indication No power supply to the transmitter Power the transmitter up or check the power supply fuse. Check power supply jumper setting. Reading on Display is not in agreement with Faulty transmitter Verify transmitter (especially analogue board) the chlorine level The sensor is not continuously in Modify sensor installation present in the sample contact with the sample The connection between the sensor Make correct connections and the transmitter is not correct. Membrane is split or coated; internal Replace membrane and filling solution filling solution contaminated or depleted The working electrode may be Clean the electrode surface tarnished The chlorine cell has been selected rather than the chlorine probe. (RT Fault will be displayed with this error) Reprogram the transmitter software The wrong chlorine range has been Reprogram the transmitter software Reading on Display is fluctuating or erratic set up Not properly calibrated Chlorine level is fluctuating Sample flow rate is too low or variable The sensor is not continuously in contact with the sample Air bubbles in the sample RTD Fault displayed Incorrect chlorine channel definition Wrong chlorine input selected Recalibrate P2 Check the process Resolve flow problems. Add Pressure Regulator and Flow meter for sample control. Modify sensor installation Eliminate the air bubbles or modify the installation Refer to transmitter instruction manual. Select chlorine probe channel definition. Design improvements may be made without notice. Represented By: De Nora Water Technologies 3000 Advance Lane Colmar, PA ph fax web: mail: Registered Trademark All Rights Reserved. De Nora Water Technologies Arley Drive Birch Coppice Business Park Dordon, Tamworth B78 1SA Tel +44 (0) Fax +44 (0) SEP
Instruction Manual Pressure Relief Valve and Rupture Disc Chlorine or Sulfur Dioxide, Series 869
Instruction Manual Pressure Relief Valve and Rupture Disc Chlorine or Sulfur Dioxide, Series 869-1 - 115.6020.15 These instructions describe the installation, operation and maintenance of the subject equipment.
More informationInstruction Manual - Diaframless TM Ejector Chlorine, Sulfur Dioxide and Ammonia
Instruction Manual - Diaframless TM Ejector Chlorine, Sulfur Dioxide and Ammonia - 1-122.6010.6 These instructions describe the installation, operation and maintenance of the subject equipment. Failure
More informationInstruction Manual O-Ring Ejector Chlorine & Sulfur Dioxide to 500 PPD (10 kg/h)
Instruction Manual O-Ring Ejector Chlorine & Sulfur Dioxide to 500 PPD (10 kg/h) 100 PPD (2 kg/h) Maximum 250 PPD (5 kg/h) Maximum 500 PPD (10 kg/h) Maximum - 1-122.6006.9 These instructions describe the
More informationCAPITAL CONTROLS SERIES NXT3000
CAPITAL CONTROLS SERIES NXT3000 Modular design gas feed system with self contained automatic switchover capability. The Series NXT3000 Gas Feed System is a family of vacuumoperated, solution-feed gas dispensing
More informationCCS240 and CCS241. Technical Information. Sensors for chlorine dioxide Amperometric, membrane-covered sensors for installation in the CCA250 assembly
Technical Information CCS240 and CCS241 Sensors for chlorine dioxide Amperometric, membrane-covered sensors for installation in the CCA250 assembly Application Chlorine dioxide is used for disinfection
More informationSeries 4000 Transmitter Analyzer Controller. Series CL4000 Chlorine Probes used with MicroChem 2
www.severntrentservices.com MicroChem 2 Series 4000 Series 4000 Transmitter Analyzer Controller Series CL4000 Chlorine Probes used with MicroChem 2 Series 4000 Transmitter and CL4000 Bufferless Probe Series
More informationCapital Controls Series NXT3000 Modular design gas feed system with self-contained automatic switchover capability.
Capital Controls Series NXT3000 Modular design gas feed system with self-contained automatic switchover capability. The Series NXT3000 Gas Feed System is a family of vacuum-operated, solution-feed gas
More informationHighest Quality. Gas Chlorination Systems
Highest Quality Gas Chlorination Systems Application: Gas chlorination systems are used for water disinfection and other purposes in a variety of applications. Compared to other disinfection methods, chlorine
More informationACV-10 Automatic Control Valve
ACV-10 Automatic Control Valve Installation, Operation & Maintenance General: The Archer Instruments ACV-10 is a precision automatic feed rate control valve for use in vacuum systems feeding Chlorine,
More informationNon-Membrane Disinfection Sensors Instruction Manual
W A L C H E M IWAKI America Inc. Disinfection Sensors Non-Membrane Disinfection Sensors Instruction Manual Five Boynton Road Hopping Brook Park Holliston, MA 01746 USA TEL: 508-429-1110 WEB: www.walchem.com
More informationSTERILIZABLE DISSOLVED OXYGEN SENSORS
STERILIZABLE DISSOLVED OXYGEN SENSORS USER MANUAL Table of Contents 1. Introduction 2. Operation 3. Calibration 4. Electrode Verification 5. Mounting 6. Maintenance 7. Storage 8. Instructions for Exchange
More informationT EK-SUB 4800C 19 mm Submersible Level Transmitter
Technology Solutions T EK-SUB 4800C 19 mm Submersible Level Transmitter Instruction Manual Document Number: IM-4800C www.tek-trol.com Table of Contents 1 Safety Instructions... 2 1.1 Intended Use... 2
More informationModel PDT Dewpoint Transmitter
Model PDT Dewpoint Transmitter Instruction Manual Alpha Moisture Systems Alpha House 96 City Road Bradford BD8 8ES England Tel: +44 1274 733100 Fax: +44 1274 733200 email: mail@amsytems.co.uk web: www.amsystems.co.uk
More informationBI-680 Online Dissolved Oxygen Controller Instruction Manual
BI-680 Online Dissolved Oxygen Controller Instruction Manual BANTE INSTRUMENTS CO., LTD BI-680 Online Dissolved Oxygen Controller 1 Introduction Thank you for selecting the BI-680 online dissolved oxygen
More informationDissolved Oxygen Sensor
Instruction Manual 499ADO and 499ADO-VP LIQ_MAN_ABR_499ADO May 2014 Dissolved Oxygen Sensor CAUTION For additional information, please visit our website at www.rosemountanalytical.com SENSOR/PROCESS APPLICATION
More informationFree Chlorine Sensor For additional information, please visit our website at
Instruction Manual 499ACL01 LIQ_MAN_ABR_499ACL01 April 2014 Free Chlorine Sensor For additional information, please visit our website at www.rosemountanalytical.com. CAUTION Sensor/Process Application
More informationQuick Start Guide LIQ-QSG-499AOZ, Rev M July Rosemount 499AOZ. Dissolved Ozone Sensor
Quick Start Guide LIQ-QSG-499AOZ, Rev M July 2017 Rosemount 499AOZ Dissolved Ozone Sensor Safety information CAUTION! SENSOR/PROCESS APPLICATION COMPATIBILITY The wetted sensor material may not be compatible
More informationWater Weir Flow Controller. Introduction. Safety Precautions. Mounting the Hardware
57007-88 Introduction Safety Precautions This instruction sheet describes how to set up and use the Hach (Figure 1). A water weir is a device that raises or diverts water to regulate the flow. Hach s water
More informationTechnical Information
Technical Information Installation, Operation, Installation, Operation and Maintenance and Maintenance Manual Manual Balston Model 75700-K728 Nitrogen Generator Figure 1-75700-K728 Overall Dimensions These
More informationPART 1 GENERAL..2 Components of the gas chlorination system include, but are not limited to the following:
PART 1 GENERAL 1.1 DESCRIPTION.1 This section includes the supply, delivery and installation, testing and placement into operation of a gas chlorination system, and appurtenances as specified herein and
More informationWelker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual
Installation, Operation, and Maintenance Manual Welker Sampler Model GSS-1 The information in this manual has been carefully checked for accuracy and is intended to be used as a guide to operations. Correct
More informationINSTRUCTION MANUAL. 840 Transmitter. Loop Powered Dissolved Oxygen Transmitter
INSTRUCTION MANUAL Loop Powered Dissolved Oxygen Transmitter Contents Contents...2 Introduction...3 Range and Display Settings...4 Installation...5 Electrical Installation...6 Calibration...7 EasyCal...8
More informationBante810 Benchtop Dissolved Oxygen Meter Instruction Manual
Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante810 Benchtop Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante810 benchtop dissolved oxygen
More informationInstallation, Operation, and Maintenance Manual
Installation, Operation, and Maintenance Manual Welker The information in this manual has been carefully checked for accuracy and is intended to be used as a guide for the installation, operation, and
More information256 Pneumatic Pressure Indicator
256 Pneumatic Pressure Indicator 51425699 Copyright 2002 Slope Indicator Company. All Rights Reserved. This equipment should be installed, maintained, and operated by technically qualified personnel. Any
More informationCSA Sample Draw Aspirator Adapter Operator s Manual
30-0951-CSA Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0367 Revision: 0 Released: 4/30/15 www.rkiinstruments.com WARNING Read and understand this instruction manual before operating
More informationElectropneumatic Transducer (I/P) Installation, Operation and
Nitra NCP1 Series Electropneumatic Transducer (I/P) Installation, Operation and Maintenance Instructions Ordering Information Contents Part Number I/P Transducers Output Range Input psig bar Section Description
More informationInstruction Bulletin Addendum For Pressure-Balanced In Situ Oxygen Probes
Instruction Bulletin Addendum For Pressure-Balanced In Situ Oxygen Probes Instruction Bulletin Addendum IB PB1000 Part no. Serial no. Order no. Oxymitter 4000 Applicability: Used with pressure-balancing
More informationModel VR6 System. Installation, Operation & Maintenance
Model VR6 System Installation, Operation & Maintenance General: All Archer Instruments chlorination systems are carefully designed and tested for years of safe, accurate field service. All Archer Instruments
More informationCS Controlled Dissolved Gas assembly with optional CS Air Aspirator assembly. Installation and Operation information
CS-425-01 Controlled Dissolved Gas assembly with optional CS-603-01 Air Aspirator assembly Installation and Operation information Serial # CS-XXX Date of shipment:, 2005 Shipped to: Purchase order #: xxxxxxx
More informationUsER manual for Watersens ph -REDOX
UsER manual for Watersens -REDOX Cl 8 1 2 6 3 3 7 7 4 4 4 4 Parts List 1 Redox Probe 1 x 2 PH Probe 1 x 5 Tube Weight 2 x 6 Connection Valve 1 x chlorine 3 Chlorine and Pumps 2 x 7 Dosing Valve 2 x 5 5
More informationBante820 Portable Dissolved Oxygen Meter Instruction Manual
Bante820 Portable Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante820 Portable Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante820 portable dissolved oxygen
More informationChlorine Supply Containers And Piping
Chlorine Supply Containers And Piping Containers Chlorine is packaged in containers as a liquefied gas under pressure, resulting in the container having both a liquid phase and a gas phase. When removing
More informationGILMONT ACCUCAL FLOWMETERS
OPERATING MANUAL GILMONT ACCUCAL FLOWMETERS 28W092 Commercial Ave. Barrington, IL U.S.A. 60010-2392 (847) 381-4888 (847) 381-7053 (Fax) 800-962-7142 www.barnant.com e-mail: barnant@barnant.com A-1299-0766
More informationOxygen Measurement Systems High Reliability and Wide Application Coverage
Introduction High Process Efficiency Under Any Condition Oxygen Measurement Systems High Reliability and Wide Application Coverage METTLER TOLEDO provides sensors to measure dissolved oxygen (DO) in demanding
More informationDextens Thermo Conductivity Sensors. Type 52101, 54101, 55101
Dextens Thermo Conductivity Sensors Type 52101, 54101, 55101 Operating Manual May 2012 Table of Content 1 GENERAL INFORMATION...4 1.1 DISCLAIMER...4 1.2 SYMBOLS AND CONVENTIONS...4 1.3 SERVICES AND REPAIRS...4
More informationSTEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE
STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE GENERAL INFORMATION This family of dissolved oxygen electrodes is of the steam sterilizable polarographic type, designed to be interchangeable with NBS
More informationQUICK START GUIDE TO DRINKING WATER MONITORING
APPENDIX F QUICK START GUIDE TO DRINKING WATER MONITORING Introduction: This Quick Start guide is designed to allow users who are already familiar with the basic use of YSI 6-series sondes for surface
More informationAir Check Lite Toxic Gas Monitor
Air Check Lite Toxic Gas Monitor Instruction Manual PureAire Monitoring Systems, Inc. 557 Capital Drive Lake Zurich, Illinois 60047 Phone: 847-726-6000 Fax: 847-726-6051 Toll-Free: 888-788-8050 www.pureairemonitoring.com
More informationSanitizers. Sanitizers. Sanitizers for Pools & Spas
for Pools & Spas 207 IntelliChlor Salt Chlorinator Featured Highlights IntelliChlor Salt Chlorinator IntelliChlor Salt Chlorinator uses common table salt to produce all the chlorine a pool needs, safely,
More informationDTG - LCD Digital Temperature Gauge USER MANUAL
DTG - LCD USER MANUAL Page 1 of 7 USERS GUIDE page 1. Description 3 2. Function 3 3. Safety Instruction 3 3.1. Safety Conventions 3 3.2. Proper Use 3 4. Installation 4 4.1. Unpacking 4 4.2. Storage 4 4.3.
More informationSPECIFICATIONS ATTENTION
VPS 504 S06 Installation Manual - P/N 80122 - Ed. 01/09 VPS 504 S06 and S05 Valve Proving System Installation Instructions VPS 1 6 Gases Natural gas, air and other inert gases. NOT suitable for butane
More informationLogIT Dissolved Oxygen Probe
LogIT Dissolved Oxygen Probe INTRODUCTION The LogIT Dissolved Oxygen Probe (DO2) has been designed especially for LogIT and is easy to set up and use. It features simple single control calibration and
More informationInstalling a Probe. Typical Installation
for hazardous locations. Prior to any installation in a Classified Hazardous Location, verify installation methods by the Control Drawing referenced on the product s name tag 2. The Feed Through Assemblies
More informationVariable Area Flow Meters
Variable Area Flow Meters Over sixty years of variable area meter expertise Brooks started designing and manufacturing variable area (VA) meters, also referred to as rotameters, in 1946. Now, with more
More informationTECHNICAL INFORMATION
TECHNICAL INFORMATION Installation, Operation, and Maintenance Balston Model 75-77 Nitrogen Generator Figure 1-75-77 Overall Dimensions These instructions must be thoroughly read and understood before
More informationInstallation, Operation, and Maintenance Manual
Installation, Operation, and Maintenance Manual Welker Probe Instrument Regulator Model The information in this manual has been carefully checked for accuracy and is intended to be used as a guide for
More informationTEK-SUB 4800B. Submersible Level Transmitter. Flow Level Temperature Pressure Valves Analyzers Accessories TekValSys LEVEL
Technology Solutions TEK-SUB 4800B Submersible Level Transmitter LEVEL Flow Level Temperature Pressure Valves Analyzers Accessories TekValSys Introduction The Tek-Sub 4800B Submersible Level Transmitter
More informationChemical Feed Equipment Floor-Mounted Gas Dispenser - Series CH4200
Specification Bulletin (rev.01) Series CH4200 Chemical Feed Equipment Floor-Mounted Gas Dispenser - Series CH4200 Easy, fast& accurate digital with 11 points calibration Laser sensor offering actual feed
More informationEASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:
2 Table Of Contents 1. Introduction 3 2. About this Manual 3 3. Contacting YZ Systems 3 4. Vessel Components 4 5. Specifications 5 6. Application 6 7. Theory of Operation 7 8. DuraSite Installation & Use
More informationSanitizers. Sanitizers. Sanitizers for Pools & Spas
Sanitizers Sanitizers for Pools & Spas 213 213 IntelliChlor Salt Chlorinator Featured Highlights IntelliChlor Salt Chlorinator IntelliChlor Salt Chlorinator uses common table salt to produce all the chlorine
More informationAvenger Model 33M Installation & Operation Instructions
Avenger Model 33M Installation & Operation Instructions Manufacturing Contact Information A+ Corporation, LLC Call for expert product application assistance: 41041 Black Bayou Rd. Gonzales, LA 70737 Phone:
More informationDOscan10 Pocket Dissolved Oxygen Tester Instruction Manual
DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual BANTE INSTRUMENTS CO., LTD DOscan10 Pocket Dissolved Oxygen Tester 1 Thank you for selecting the DOscan10 pocket dissolved oxygen tester. This
More information36H SERIES Combination Gas Valve
FAILURE TO READ AND FOLLOW ALL INSTRUCTIONS CAREFULLY BEFORE INSTALLING OR OPERATING THIS CONTROL COULD CAUSE PERSONAL INJURY AND/OR PROPERTY DAMAGE. DESCRIPTION The 36H series combination gas valve is
More informationChemical Injection Technologies Installation/Service Bulletin
Bulletin 4002 Chemical Injection Technologies Installation/Service Bulletin SUPERIOR Series CL-16/26/56 Automatic Switchover Gas Chlorinator - Installation & Operation IMPORTANT!! READ THESE PRECAUTIONS
More informationInstruction Manual ADVANCE TM Wall Panel Gas Feeders Series WP70CV3
LIQUID FIL DO NOT TAM Instruction Manual ADVANCE TM Wall Panel Gas Feeders Series WP70CV3 LIQUID FIL DO NOT TAM 500 PPD (10 kg/h Chlorine & Sulfur Dioxide 3000 PPD (60 kg/h Chlorine & Sulfur Dioxide Auto
More informationCalcimeter Instruction Manual
Hohner (UK - Canada - Texas) Calcimeter Instruction Manual The Hohner Calcimeter is based on industry standard versions, and is used to measure the calcium carbonate and magnesium carbonate in samples.
More informationModel 130M Pneumatic Controller
Instruction MI 017-450 May 1978 Model 130M Pneumatic Controller Installation and Operation Manual Control Unit Controller Model 130M Controller is a pneumatic, shelf-mounted instrument with a separate
More informationRosemount 140/141/142
Rosemount 140/141/142 Contacting Conductivity Sensors Instruction Manual LIQ-MAN-140 Rev. F May 2017 hasgkas Essential Instructions Read this page before proceeding! Emerson designs, manufactures and tests
More informationInstallation, Operation, and Maintenance Manual Parker Balston Model N2-04 Nitrogen Generator
Installation, Operation, and Maintenance Manual Parker Balston Figure 1 - N2-04 Nitrogen Generator These instructions must be thoroughly read and understood before installing and operating this product.
More informationAir Operated Hydraulic Pumping Systems to 50,000 psi
High Pressure Equipment Air Operated Hydraulic Pumping Systems to 50,000 psi PS-10: 10,000 psi PS-20: 20,000 psi PS-30: 30,000 psi PS-40: 40,000 psi PS-50: 50,000 psi PS-90: 90,000 psi High Pressure air
More informationRK-IR Sample Draw Aspirator Adapter Operator s Manual
30-0951RK-IR Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0018RK Revision: A Released: 6/2/10 www.rkiinstruments.com Product Warranty RKI Instruments, Inc. warrants gas alarm equipment
More informationOxymax W COS41. Technical Information. Potentiostatic amperometric two-electrode sensor Long-term stable sensor for water and wastewater applications
Technical Information Oxymax W COS41 Potentiostatic amperometric two-electrode sensor Long-term stable sensor for water and wastewater applications Application The continuous measurement of the dissolved
More informationColumbus Instruments
0215-003M Portable O 2 /CO 2 /CH 4 Meter User s Manual Columbus Instruments 950 NORTH HAGUE AVENUE TEL:(614) 276-0861 COLUMBUS, OHIO 43204, USA FAX:(614) 276-0529 1 www.colinst.com TOLL FREE 1-800-669-5011
More informationOPERATION MANUAL NTF-15
OPERATION MANUAL NTF-15 Nitrogen Tire Filling Valve Stem Caps (Qty=200) Order P/N 436075 RTI Technologies, Inc 10 Innovation Drive York, PA 17402 800-468-2321 www.rtitech.com 035-81235-00 (Rev B) TABLE
More informationModel TX-FM/FN Sensor Head
Model TX-FM/FN Sensor Head Instruction Manual PureAire Monitoring Systems, Inc. 557 Capital Drive Lake Zurich, Illinois 60047 Phone: 847-726-6000 Fax: 847-726-6051 Toll-Free: 888-788-8050 Pureairemonitoring.com
More informationINSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y
INSTALLATION and INSTRUCTION MANUAL for QUALITY AIR BREATHING SYSTEMS Model 50 Systems Outfitted with ABM-725 Monitor M A R T E C H S E R V I C E S C O M P A N Y OFFICE: (507) 843-4700 P.O. BOX 7079 Toll
More informationRejuvenation Instructions
Rejuvenation Instructions #401 Air Systems UPR This NRI covers the following: Understanding the applications and operation of flow meters. Understand the application and operation of test pressure gauges.
More informationRAM 4021 Operation Manual
RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For your safety...3 Description...3 Set-up mode...4 Annunciator lights/alarms...4 Operation...5 Calibration...6
More informationOEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE
OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE 1 These instructions generally describe the installation, operation, and maintenance of subject equipment. The manufacturer reserves the right
More informationABRASIVE SYSTEM ABRASIVE REGULATOR II 800# BULK ABRASIVE DELIVERY POT Abrasive Regulator II and 800# Pot Manual
ABRASIVE SYSTEM ABRASIVE REGULATOR II 800# BULK ABRASIVE DELIVERY POT TABLE OF CONTENTS Contact Information and Customer & Technical Service... 1 1 Introduction... 1 2 Abrasive Regulator... 1 Description:...
More informationTANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel
TANK MANAGER FOR TWO TANKS OPERATING MANUAL 10/31/11 C-More T6C L color touch panel 1 TABLE OF CONTENTS GENERAL...3 INSTALLATION...4 STONE TEST PROCEDURE...7 OPERATIONAL SUMMARY...7 AUTO CARBONATION...10
More informationPROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road
PROPORTIONING VALVE Model 150 INSTRUCTION MANUAL March 2017 IMS Company 10373 Stafford Road Telephone: (440) 543-1615 Fax: (440) 543-1069 Email: sales@imscompany.com 1 Introduction IMS Company reserves
More informationQBS Electronic Pressure Regulator
INSTALLATION AND MAINTENANCE INSTRUCTIONS DESCRIPTION / IDENTIFICATION The QBS stainless steel series is an electronically controlled pressure regulator. The QBS controls the pressure on its output port
More informationRCD - Ultra-Rugged Differential Pressure Flowmeter
0.15-1 to 85-600 GPM Liquids 2.5-25 to 300-1700 SCFM Gases 1/2 to 3 Line Sizes Bronze or 316-Ti Stainless Steel Mechanical Indicator, Digital Indicator, Transmitters and Switches Available Custom Calibrations
More information1020 Industrial Drive, Orlinda, TN fax
Operation Manual Ultrafiltration for High Purity Distribution K-A-HPTUF Series 615-654-4441 sales@specialtyh2o.com 615-654-4449 fax TABLE OF CONTENTS Section 1 GENERAL 1.1 Warnings and Cautions... 1 1.2
More informationWalchem Disinfection Sensors WCDB, WFCB, WOZB and WPAB Instruction Manual
W A L C H E M An Iwaki America Company Disinfection Sensors Walchem Disinfection Sensors WCDB, WFCB, WOZB and WPAB Instruction Manual Five Boynton Road Hopping Brook Park Holliston, MA 01746 USA TEL: 508-429-1110
More informationNB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE
NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE INSTALLATION, OPERATION & MAINTENANCE MANUAL INTERFACE DEVICES, INC. 230 Depot Road, Milford, CT 06460 Ph: (203) 878-4648, Fx: (203) 882-0885, E-mail: info@interfacedevices.com
More informationTECHNICAL INFORMATION
TECHNICAL INFORMATION Installation, Operation, and Maintenance Balston 75-96 and 75-98 Nitrogen Generation System Tower Pressure Gauge 5'-1" 3'-6" 7'-4" 6'-10 1/2" PLAN VIEW Figure 1 - Balston 75-96 and
More informationGas Lines. Technical Manual. Sumitomo (SHI) Cryogenics of America, Inc Vultee Street Allentown, PA U.S.A.
Gas Lines Technical Manual Sumitomo (SHI) Cryogenics of America, Inc. 1833 Vultee Street Allentown, PA 18103-4783 U.S.A. Revision F: April 2008 261320A TABLE OF CONTENTS Page DESCRIPTION...1 SPECIFICATIONS...2
More informationOC Panel High Limit Aquastat Kit, Manual Reset p/n
OC Panel High Limit Aquastat Kit, Manual Reset p/n 233202 Instruction Sheet APPLICATION The OC (Option Control) Panel High Limit Aquastat Kit provides electronic temperature sensing in a UL limit-rated
More informationP5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.
Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 TestEquipmentDepot.com P5513 Pneumatic Comparison Test Pump Users Manual PN 3963372 November 2010 2010 Fluke Corporation. All
More informationGenie Supreme TM Model 123HP Installation & Operation Instructions
Genie Supreme TM Model 123HP Installation & Operation Instructions Manufacturing Contact Information A+ Corporation, LLC Call for expert product application assistance: 41041 Black Bayou Rd. Gonzales,
More information299H Series. Introduction. P.E.D. Categories. Specifications. Installation. Warning. Installation Guide English September 2012
Installation Guide English September 2012 299H Series Introduction This Installation Guide provides instructions for installation, startup, and adjustment of 299H Series regulators. To receive a copy of
More informationSS32L Dissolved O 2 Probe
BIOPAC WWW.biopac.com Systems, Inc. Application Note PH-185 SS32L Dissolved O 2 Probe SS32L Dissolved O 2 Probe BIOPAC Software SS32L Specifications BSL PRO v. 3.6.6 BIOPAC Hardware SS32L Dissolved O 2
More informationSERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL
www.modsafe.com SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL WARNING: Do not attempt to operate this equipment without first reading and understanding the service manual enclosed with this device.
More informationOPERATION MANUAL NTF-60 Plus
OPERATION MANUAL NTF-60 Plus Nitrogen Tire Filling Valve Stem Caps (Qty=200) Order P/N 436075 RTI Technologies, Inc 10 Innovation Drive York, PA 17402 800-468-2321 www.rtitech.com 035-81264-00 (Rev A)
More informationWATER MADE EASY MARINE ENERGY MUNICIPAL INDUSTRIAL
Capital Controls Series 71V3000 The Series 71V3000 electrically-heated vaporizer automatically vaporizes and superheats liquid chlorine, sulfur dioxide or ammonia at a rate controlled by the gas feed system.
More informationAW400 cell Chlorine/Chlorine dioxide/ozone
ABB MEASUREMENT & ANALYTICS USER GUIDE IM/AW4SR REV. F AW400 cell Chlorine/Chlorine dioxide/ozone Measurement made easy AW400 cell For more information Further publications are available for free download
More informationFloat Operated Level Controllers
CONTENTS Float Operated Level Controllers IM0015 Nov. 2014 PAGE Introduction 1 Scope 1 Description 1 Specification 1 Control Installation 2 INTRODUCTION Side Mount Back Mount Prior to installing, the instructions
More informationLDO II AQ 2/2014, Edition 1.5
9623800 LDO II AQ 2/2014, Edition 1.5 USER MANUAL Table of Contents Specifications on page 2 General information on page 3 Product overview on page 4 Installation on page 5 Operation on page 7 Maintenance
More informationInstallation and Use Manual
Installation and Use Manual EMASMB & LMASMB Surface Mount Bottle Filling Stations Model EMASMB IMPORTANT THIS IS AN INDOOR APPLICATION ONLY! ALL SERVICE TO BE PERFORMED BY AN AUTHORIZED SERVICE PERSONNEL.
More informationDiscontinued. Powers Controls. Technical Instructions Document No P25 RV Rev. 1, May, RV 201 Pressure Reducing Valves.
Powers Controls RV 201 Pressure Reducing Valves Description Features Product Numbers Dual Pressure PRV Technical Instructions Document No. 155-049P25 RV 201-1 Single Pressure PRV The RV 201 Pressure Reducing
More informationLRS(H)4 USER MANUAL. Read the complete manual before installing and using the regulator.
LRS(H)4 USER MANUAL Read the complete manual before installing and using the regulator. WARNING INCORRECT OR IMPROPER USE OF THIS PRODUCT CAN CAUSE SERIOUS PERSONAL INJURY AND PROPERTY DAMAGE. Due to the
More informationTECHNICAL DATA. TRIMPAC Model B-5 & B-5B
September 16, 2013 Trimpac 250a 1. DESCRIPTION DEsCRIPTIoN is a factory assembled trim package for a ed an electric/pneumatic release module in a metal enclosure. The standard trim normally required on
More informationINSTALLATION & MAINTENANCE INSTRUCTIONS
DESCRIPTION / IDENTIFICATION The QBX series valve uses Proportion-Air closed loop technology for Pressure control. It gives an output pressure proportional to an electrical command signal input. The QB1X
More informationOperation Manual - PN A MENSOR MODEL 73 SHOP AIR BOOSTER
Operation Manual - PN 0017946001 A MENSOR MODEL 73 SHOP AIR BOOSTER Mensor Model 73 Shop Air Booster System (750 psi Version) April 23, 2012 Trademarks / Copyright Mensor is a registered trademark of Mensor
More informationQuick Start Guide LIQ-QSG-499ACL-03, Rev I July Rosemount 499ACL-03. Monochloramine Sensor
Quick Start Guide LIQ-QSG-499ACL-03, Rev I July 2017 Rosemount 499ACL-03 Monochloramine Sensor Safety information CAUTION! SENSOR/PROCESS APPLICATION COMPATIBILITY The wetted sensor material may not be
More informationUn-Pressurized Orefice Fittings FIO EZ. Parts List and Operation Instructions TECHNICAL MANUAL. Dn 2-6 Class Lbs
Un-Pressurized Orefice Fittings FIO EZ Parts List and Operation Instructions TECHNICAL MANUAL Dn 2-6 Class 150-600 Lbs US US 2 FIO EZ - MT 108-US - 05-2016 FIO EZ Important Instructions US Pietro Fiorentini
More informationPERFORM Operating Document
PERFORM Operating Document Use and Maintenance of CO 2 Incubator PC-POD-CA-007-v03 Revision History Version Reason for Revision Date 01 New POD 30-Sep-13 02 Minor revisions for section 2.3, 3.1, 4.3. 14-April-16
More information