1 ZIRON Oxygen Measuring Technique Oxygen Measuring Device ECO1000 Manual Sales and Service Center Shanghai ZIRON Gasanalyzer Co., Ltd 404/6, No.10, Lanxi Road, Shanghai P.R. China Tel , Fax ,

2 ZIRON Oxygen Measuring Device ECO Table of contents 1 The Company ZIRON... Fehler! Textmarke nicht definiert. 1.1 Business development... Fehler! Textmarke nicht definiert. 1.2 Cooperations... Fehler! Textmarke nicht definiert. 2 General Manual hints Copyright Safety symbols and terms Safety regulations Applications of the ECO General properties Functional description, construction and operation Measuring principle Construction principle of the solid electrolyte measuring cell Electronics Further annotations to the ECO Measuring cell Why calibration? Measuring gas containing organic compounds? Charcoalfilter use! Technical data Installation und initiation Mechanical assembly Power supply Front Rear Operation and calibration General Calibration Zero gas calibration (Offset compensation) Span gas calibration Adjustment span gas concentration Menu navigation Trouble shooting, maintanance and storage Error messages Trouble shooting Appendix Warranty conditions Charcoal filter: Description and application hints...18

3 ZIRON Oxygen Measuring Device ECO Filter construction Filter function and use Charcoal changing Technical data Declaration of conformity...20

4 ZIRON Oxygen Measuring Device ECO General 1.1 Manual hints This operation manual describes design, function and operation of the oxygen measuring device ECO1000. Sales and Service Center Sinom Corp. No.33, Sec.4, Jhongsing Rd., Wugu, New Taipei,Taiwan Tel Fax sales@sinom.com.tw Internet: The manufacturer assures that this operation manual has been written in accordance with the functional and technical features of the delivered ECO1000. This operation manual is not subject to the amendment service. If the manufacturer modifies the ECO1000 with the aim of making technical improvements the user is responsible to insert the additional or updated pages supplied. Secure and proper function of the ECO1000 can only be guaranteed when the contents of this manual are known. Therefore, all chapters of this manual must be read carefully before installation and operation of the ECO1000. The display values shown in this manual are samples or presettings of the manufacturer. The process specific values must be set by user. 1.2 Copyright This manual is copyright protected. It must not be reproduced partially or totally, copied or distributed, without prior written admission from the manufacturer, neither unauthorized use for the purposes of competition or the handing over of the manual to third parties are permitted. All rights reserved.

5 ZIRON Oxygen Measuring Device ECO Safety symbols and terms Symbol for direct imminent hazards This symbol is shown with recommendations for industrial safety if a direct danger exists to the life and health of persons. If these recommendations are not thoroughly observed, fatal injuries may result. Symbol for indirect imminent hazards This symbol informs about situations where indirect hazards occur. Degree and intensity of the damage depend on the triggering actions and the behavior of the persons involved. Disregard of these hints may result in damage to or destruction of the entire ECO1000 or single components, other property, as well as minor injuries. Symbol for proper usage This symbol is found where the observance of instructions, regulations, and proper order of actions/events is important. Not observing these hints may result in damage or destruction of the ECO1000 or its single components. 1.4 Safety regulations The following regulations for industrial safety give basic information about possible hazards during handling and operation of the ECO1000. Therefore, they must be observed and strictly followed by operation and service staff. Secure and proper function of the ECO1000 can only be guaranteed when the contents of this manual are known. Therefore, all chapters of this manual must be read carefully before installation and operation of the ECO1000. The ECO1000 may be used for applications stated as authorized only (see chapter 2). The ECO1000 may only be installed, operated, and maintained by trained staff. In any case, the ECO1000 must be connected to the main supply with a socket with earth connection by supplied cable. Explosive gas mixtures, halogens in high concentration, gas mixtures containing silicon, phosphor and sulfur (e.g. SO 2 ) are not permitted to be measured by ECO1000.

6 ZIRON Oxygen Measuring Device ECO Due to the high operation temperature of the sensor, the ECO1000 produces much heat. It has to take care that the ECO1000 is not covered so strongly that it will be come to overheatings. Vertical installation is not permitted to avoid possible damages to the device by localization of heat. Switch off and disconnect the device from the power supply before opening the housing cover! Special safety recommendations for possible hazards during special work or sequences of work are given in the relevant text. The measuring gas can be reducible for a short time 1 (the electronics show 0). It must not be an explosive mixture the heating of the measuring cell is an ignition source. For measurements in corrosive gases, if condensates or carbon deposition are possible, a consultation by manufacturer is strongly recommended. 1.5 Applications of the ECO1000 In many technological processes the presence of oxygen traces is harmful to the product properties. The rapid and accurate measurement of oxygen traces in the process atmosphere is prerequisite for the identification and prevention of problems. With the ECO1000 the oxygen concentration of a sample is measured and displayed continuously, as well as variations of adjustable setpoints is signalized. With the increasing introduction of quality management systems (eg ISO 9000) the monitoring and documentation of relevant quality parameters are necessary. By ECO1000 the user receives a variety of possibilities for process optimization and quality control. The main areas of the device are monitoring functions in the welding and soldering technology (eg. reflow soldering with inert gas). In this case an charcoal filter must be used. With the help of the filter hydrocarbons, which give a measuring error, are restrained. 1 It has to be distinguished between different conditions of the oxygen in the measuring gas: Free oxygen: There are the oxygen molecules in the gas independently without every relationship relation to the other gas components (inert gases such as N 2 or Ar) Bound oxygen: There aren't any free oxygen molecules, but only in bound form e.g. as steam in the gas. At higher temperatures dissociation is carried out and oxygen molecules then are availably. Since the dissociation degree increases with the temperature, the measurement result is also dependent on the temperature. Circumstances permitting free oxygen can react with perhaps available burning gasses at the hot platinum electrode, a reducing gas then can be also in the result through what.

7 ZIRON Oxygen Measuring Device ECO General properties The compact oxygen measuring device contains a zirconia measuring cell, a measuring gas pump and the electronics. The sensor element is heated by internal heater control. A microprocessor calculates the oxygen concentration according to the NERNST law. The measuring value is shown on the display and given out by analog (4-20 ma) and digital interface (option). Software for data processing is available (option). A span gas calibration can be made by certified test gas. If requested the internal pump can be switched off.

8 ZIRON Oxygen Measuring Device ECO Functional description, construction and operation 2.1 Measuring principle The NERNST-equation is used as basis for determining the oxygen concentration in gases with the ECO1000 NERNSTequation Dabei ist: U F p O 2,air RT U = ln zf p Cell voltage in mv p O 2, Air O 2, measuringgas R Molar gas constant, R = 8,314 J/(mol K) T Measuring temperature in K p O 2,measuring gas Faraday-constant, F = 9, C/mol Oxygen partial pressure at reference electrode in dry air in Pa Oxygen partial pressure at measuring electrode in Pa 2.2 Construction principle of the solid electrolyte measuring cell (I) The sensor of the ECO1000 is based on the conductivity of oxide ions in a special ceramic substance (zirconia) with stabilising additives. The conductivity of these oxide ions increases exponentially with the temperature and it is sufficiently high for temperatures above 600 C. The gas to be measured passes through the ceramic oxide ion conductor which is a gas-tight pipe. The ceramic pipe is situated axial-symmetrically in a thermic well-insulated heater. The electrodes of the galvanic sensor are made from platinum. The electrode on the outside of the pipe, surrounded by dry air, is used as a reference electrode with a constant, known electrode potential. Measuring cell The ECO1000 is not explosion-proof! 2.3 Electronics The electronics serve the power supply of the measuring cell, calculation of the oxygen concentration from measuring value and output as standard signal (4-20 ma). Additionally a span gas calibration can be made. The electronics realize following functions: Power supply of the measuring cell and heating control Cell and thermo voltage processing to oxygen concentration Oxygen concentration output via standard signal Calibration Measuring gas suction by internal pump

9 ZIRON Oxygen Measuring Device ECO Further annotations to the ECO Measuring cell Due to physical-chemical basics the cell voltage depends on the oxygen concentration. Therefore it does not change with increasing age. Only the inner resistance of the cell increases with time, depending on environmental conditions. This means that the accurate measurement of cell voltage during the aging process is more complicated. The limit of the inner resistance of the cells is approximately kohm and after the experience from Germany this value will be reached after 3-5-years at normal stress (no catalyst poisons or pollutants in the exhaust gas). For devices, which came back for calibration after one year, the inner resistance were 100 to 200 ohms. According to the NERNST law the calculation of oxygen occurs from the voltage of the cell and its temperature. The measuring value is determined thermodynamically if the measuring electrode is not damaged, the cell voltage reflects the concentration of oxygen concentration accurately Why calibration? Cause for the calibration is the thermo couple type K. It can not be integrated in the electrode. Therefore a manufacturing temperature offset between electrode and temperature measurement occurs. This offset can be compensated by calibration. The thermo couple type K ages. A drift of the thermo voltage occurs over the years. The measured temperature (with type K) can more than 10 C deviate from the real electrode temperature. Only by calibration with test gas this error can be corrected and the oxygen calculation readjusted respectively. This type of measuring cell is more cost-saving at manufacturing (low-cost wires, base metal materials) than cells with thermo couple type B or S Measuring gas containing organic compounds? Charcoalfilter use! If the measuring gas contains reducing components (e.g. alcohol or solvent vapour) the concentration of free oxygen cannot be determined correctly because chemical reactions occur at the electrode. In such cases the measuring gas should be filtered by a charcoal filter (see chapter 8). After longer action the filter could be saturated if the oxygen concentration drops suddenly. In this case the whole filter cartridge or the charcoal must be change (see chapter 8).

10 ZIRON Oxygen Measuring Device ECO Technical data Range Vol.-ppm, till 100 Vol.-% on request Accuracy... rel. error < 5% Gas flow l/h (by internal pump) Max. measuring gas pressure mbar over pressure Max. measuring gas temperature C at gas input Dimensions (B x H x T) mm x 123 mm x 235 mm Mass... 3kg Protection degree... IP 20 Gas input... Screw connection for hose 6 x 4 mm Gas output... Hose nipple 4 mm Working conditions C, rel. humidity < 80% at 20 C Storage conditions C, rel. humidity < 95% at 20 C Power supply V AC, Hz (at 110 V AC the pump performance drops strongly) Power consumption VA max. Heating measuring cell V DC, ca. 10 W (controlled internally) Fan V DC, A Keyboard... keypad with 2 keys Clear text display... Double-spaced LCD-Display, 2 x 16 characters Digital interface... RS232 (Option) Analog output... 0/4 20 ma, potential-free, Zero point and endmark free scalable ppm, resistance max. 500 Ohm Option: Voltage output... 0/2 10 V, potential-free, Zero point and endmark free scalable ppm Limit value... < ppm Fuse... 2 x 1 A/T Terminal strip Analog output PIN Designation 1 -I out 2 +I out 3 Alarm 4 Alarm

11 ZIRON Oxygen Measuring Device ECO Installation und initiation 4.1 Mechanical assembly All components are located (main electronics, pump, line filter, measuring cell) in a portable housing. 4.2 Power supply The ECO1000 is connected by rubber connector and delivered cable. It is a portable device. The working position is horizontally. 4.3 Front On the front display and keys are located. Operation and parameterization hints follow in chapter 6. The lateral charcoal filter is installed firmly (Option). Fig. 1: Front view of ECO1000

12 ZIRON Oxygen Measuring Device ECO Rear Gas in- and output, fan, Sub-D-socket for RS232 and rubber connector witch power switch are located on the rear of ECO1000. If a charcoal filter is mounted (Option), the gas input is on the filter! Fig. 2: Rear view of ECO1000 The measuring values and the status messages can be transmitted via RS232 and analog output. Interfaces

13 ZIRON Oxygen Measuring Device ECO Operation and calibration 5.1 General After switching on the warm up starts. Over 400 C the temperature is displayed additionally in the second space. If the measuring cell temperature has reached 750 C the oxygen concentration is also displayed. Warm up: approx. 15 minutes With pressing the cell voltage is displayed in the second space. With help of the booth keys an offset-calibration ( zero point calibration ) in ambient air and a span gas calibration are possible. The whole menu is shown in figure 3, page 16. ***WARM UP*** Temp: 555 C O2: 20.6 Vol.-% Uz: xx.x mv Calibration 5.2 Calibration Especially before measurements with high accuracy request a calibration should be taken (The stability of measuring cell and electronics is so high that a check is only necessary once a year). The device must be in the operation state at least 1 hour before calibration. A calibration in the lower range (< 10 ppm O 2 ) makes sense only without charcoal filter! With filter it can not be measured in this range! 5.3 Zero gas calibration (Offset compensation) This check is very important if the measuring value will be near 20 percent of oxygen. The heater can be changed by small mechanical instabilities or aging. The result is a small temperature deviation at the electrodes and therefore a small cell voltage changing. This missing voltage is compensated by zero gas calibration. For zero gas calibration ambient air must flow through the measuring cell by the expected measuring gas velocity. This status is reached by internal pump. The zero gas calibration is described in chapter 6.6.

14 ZIRON Oxygen Measuring Device ECO Span gas calibration For the span gas calibration a certified test gas with the expected measuring gas concentration flows through the measuring cell (connection via bypass the flow is realized by internal pump). Adjustment of the span gas concentration and the calibration procedure is shown in chapter 6.5 and Adjustment span gas concentration After call SPAN GAS Concentration at calibration menu (see figure 3) the adjustment possibility is activated by. The blinking character can be adjusted by. After changing of all characters this menu is deactivated by and leaved by. Now the calibration can be started.

15 ZIRON Oxygen Measuring Device ECO Menu navigation Fig. 3: Navigation overview

16 ZIRON Oxygen Measuring Device ECO Trouble shooting, maintanance and storage 6.1 Error messages During working process a monitoring of the measuring cell occurs continuously for the identification of typical errors. Error message on display SYSTEM ERROR LOW PROBE TEMP. ERROR THERMOCOUPLE Description During system start the self-test has found a system error. The soll-temperature (750 C) is not reached. Thermo couple or cable are broken. 6.2 Trouble shooting Error message SYSTEM ERROR ERROR THERMOCOUPLE Display does not shine During span gas calibration: CALIBR. FAILED Cause and clearance Please contact manufacturer! Check contacts and connections of the thermo couple, check thermo voltage ATTENTION: Authorized Stuff only! Check power supply Test gas does not correspond to the adjusted span gas concentration. The calibration factor is out of values given by manufacturer. Test gas or adjusted span gas concentration must be checked resp. readjusted. After longer action the filter could be saturated if the oxygen concentration drops suddenly. In this case the whole filter cartridge or the charcoal must be change (see chapter 8). A charcoal filter changing is recommended after 600 operation hours!

17 ZIRON Oxygen Measuring Device ECO Appendix 7.1 Warranty conditions ZIROX warrants that the equipment manufactured and sold by it will be free of defects in workmanship or material upon delivery. Should any failure corresponding to this warranty conditions become apparent during a period of 12 months (sensor) or 24 months (electronic) after the date of delivery, ZIROX will correct such nonconformity by repair or replacement of the defect parts after prompt written notice from the purchaser. The purchaser is not entitled to claim any other legal remedies on the basis of this warranty. ZIROX does not warrant equipment supplied by it against normal deterioration (e.g. reference gas pump). Corrosive gases and solid particles may cause damages and make a repair or a replacement necessary as a consequence of normal wear and tear during the warranty period. The contact of the products with explosive gas mixtures, halogens in high concentrations and sulphur-containing gases (e.g. SO 2 ) is not permissible. The contact of the products with silicon- or phosphor-containing compounds is also not permissible. In the case of combination of ZIROX products with non- ZIROX products, which are not approved by ZIROX, any warranty claim is invalid. Warranty and warranty claims are only accepted if they are in accordance with the "General Sales and Delivery Conditions" of the manufacturer. Warranty and liability claims for injuries to person and/or property damages are void if the claims are due to one or more of the following causes: normal wear unauthorized using disregard of the regulations of the equipment manual improper installation, operation, and maintenance operation of the device without electrical protection (i.e. proper ground connection) unauthorized functional or technical modifications removal of parts or installation of spare parts or additional units which are not delivered or permitted by ZIROX improperly executed repairs or wrong handling Stranger influence and superior force Attention: When installing the equipment, the customer must ensure that all necessary supply lines are connected and the operating temperature of the probe is reached. Experience has shown that products installed but not taken into operation may be damaged by the process or by external influence. ZIROX will not accept any responsibility for such damages.

18 ZIRON Oxygen Measuring Device ECO Charcoal filter: Description and application hints Filter construction The figure above shows the charcoal filter. Two caps with tube connections close the charcoal acryl glass container. Both caps have been passed into the tube (sealed by O-ring). The fittings (tube connection with seal ring) have closing stopper against falling out of the charcoal. For the gas way protection against charcoal particles a prefilter and a fine filter (2 filter paper discs) have been installed into the output of the charcoal filter Filter function and use Organic components of the measuring gas (e.g. alcohols) will be adsorbed by the charcoal. After longer using the cell voltage or the oxygen concentration can be drop unexpectedly. This is a sign for saturation of the filter. The filter or the charcoal filling must be changed Charcoal changing A saturated filter can be recycled by changing of the activated carbon filling only. For this purpose the charcoal must be poured out after removing the tubes (press black part into the connection and pull out the tube) and the tube connection at gas input (wrench size 11). Then new granulated charcoal can be refilled into the container by a small funnel. The powder should be compressed by knocking a piece of wood against the filter housing. Finally, the removed tube connection with seal rings must be screwed onto the filter again. For a connection ECO1000 charcoal filter the tube must be only inserted into the tube connection.

19 ZIRON Oxygen Measuring Device ECO Caution: Do not remove the closing stopper in the Plug connections! Technical data Dimensions (diameter x length) Weight Volume Duration of Use Used charcoal 33 mm x 165 mm approx. 200 g Approx. 100 ml Depending on the components and the concentration of the adsorbed organic components e.g. approx. 1 3 months at raw gas from a fermenting tank (ZKG) granulated or pelleted, 1 3mm

20 ZIRON Oxygen Measuring Device ECO Declaration of conformity

Technical Data Sheet MF010-O-LC

Technical Data Sheet MF010-O-LC Technical Data Sheet MF010-O-LC - 1 - 1. Properties The oxygen measuring system MF010-O-LC determines the oxygen content in gas mixtures up to a temperature of 250 C. It is particularly suitable for the

More information

Oxygen Flue Gas Analyzer BA 2000

Oxygen Flue Gas Analyzer BA 2000 Gas Analysis Oxygen Flue Gas Analyzer BA 2000 Some combustion processes, e.g. process heaters, steam boilers or heating furnaces the air required to achieve optimal system efficiency can easily fluctuate.

More information

Columbus Instruments

Columbus Instruments 0215-003M Portable O 2 /CO 2 /CH 4 Meter User s Manual Columbus Instruments 950 NORTH HAGUE AVENUE TEL:(614) 276-0861 COLUMBUS, OHIO 43204, USA FAX:(614) 276-0529 1 www.colinst.com TOLL FREE 1-800-669-5011

More information

Gas Sampling Fitting with MK2.1 and L-Probe

Gas Sampling Fitting with MK2.1 and L-Probe measuring controlling automation measuring controlling automation measuring controlling automation Page i von 8 Operation Manual Gas Sampling Fitting with MK2.1 and L-Probe Page ii von 8 Manufacturer MESA

More information

FTC130 Transmitter. Operating Manual

FTC130 Transmitter. Operating Manual Seite 1 von 11 FTC130 Transmitter Fast Thermal Conductivity Analyzer Operating Manual 1.09KD180724CWI1V05 Messkonzept GmbH Seite 2 von 11 1. Intended Use... 3 2. Description... 4 3. Measuring gases and

More information

ZIRCONIA OXYGEN DETECTOR SERVICE MANUAL

ZIRCONIA OXYGEN DETECTOR SERVICE MANUAL Service Manual ZIRCONIA OXYGEN DETECTOR SERVICE MANUAL TYPE: ZFK 3, 4, 7 TN5A08a-E CONTENTS. PREFACE.... MEASUREMENT PRINCIPLE... 3. ADJUSTMENT...3 3. Zero (Air) Adjustment... 3 3. Span Adjustment... 3

More information

OXYGEN MONITORING SYSTEM OMS

OXYGEN MONITORING SYSTEM OMS User Manual OXYGEN MONITORING SYSTEM OMS 420 Read and observe before commissioning! Ex Situ oxygen analyzer based on ZrO 2 sensor ENGLISH Caution! Please inspect in the presence of the transporting agent

More information

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet: OXY Integral INTERCON ENTERPRISES INC Tel: 800 665 6655 Fax: 604 946 5340 E-Mail: sales@intercononline.com Internet: www.intercononline.com Manual Integral 2006 1 INDEX 2-3 PREFACE 4 INTRODUCTION 5 Principle

More information

GasSense NDIR User Manual

GasSense NDIR User Manual INDEX INDEX... 1 1. OVERVIEW... 2 2. TECHNICAL DATA... 3 3. SPECIFICATIONS... 4 4. PRODUCT DESCRIPTION... 5 4.1 Mechanical details... 5 4.2 Piping... 7 4.3 Connections... 7 5. INSTALLATION... 10 6. CALIBRATION

More information

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante810 Benchtop Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante810 benchtop dissolved oxygen

More information

EuroFID Total Hydrocarbon Analyzer. Precise Determination of Total Hydrocarbons in Air for Corrosive as well as Condensing Gases

EuroFID Total Hydrocarbon Analyzer. Precise Determination of Total Hydrocarbons in Air for Corrosive as well as Condensing Gases P r o d u c t i n f o r m at i o n EuroFID Total Hydrocarbon Analyzer Precise Determination of Total Hydrocarbons in Air for Corrosive as well as Condensing Gases Proven Analyzer Technology Measuring total

More information

Transmitter CS 21 Operation Manual

Transmitter CS 21 Operation Manual Transmitter CS 21 Operation Manual Content Page For your Safety 3 General Description 3 Detection Principle 4 Operational Notes 4 Design 4 Mounting Position of CS21 5 Mounting 6 Installation of Electrical

More information

User s Guide Temperature Sensor Converter TSC-599

User s Guide Temperature Sensor Converter TSC-599 User s Guide Temperature Sensor Converter TSC-599 ILX Lightwave Corporation 31950 Frontage Road Bozeman, MT, U.S.A. 59715 U.S. & Canada: 1-800-459-9459 International Inquiries: 406-556-2481 Fax 406-586-9405

More information

Instruction Manual BA90 Precision Barometer

Instruction Manual BA90 Precision Barometer Instruction Manual BA90 Precision Barometer halstrup-walcher GmbH Stegener Straße 10 D-79199 Kirchzarten Germany Phone: +49 (0) 76 61/39 63 0 Fax: +49 (0) 76 61/39 63 99 E-Mail: info@halstrup-walcher.com

More information

BI-680 Online Dissolved Oxygen Controller Instruction Manual

BI-680 Online Dissolved Oxygen Controller Instruction Manual BI-680 Online Dissolved Oxygen Controller Instruction Manual BANTE INSTRUMENTS CO., LTD BI-680 Online Dissolved Oxygen Controller 1 Introduction Thank you for selecting the BI-680 online dissolved oxygen

More information

Installation and Operation Manual

Installation and Operation Manual Manual Static pressure transducer with controller Differential static pressure transducer with analog output and optional PI control mode Large diaphragm element with differential transformer Transducer

More information

MEGAS 2.0. Gas analysis with direct C-Level Calculation and Bus-Connection. Data sheet

MEGAS 2.0. Gas analysis with direct C-Level Calculation and Bus-Connection. Data sheet Data sheet MEGAS 2.0 Gas analysis with direct C-Level Calculation and Bus-Connection Mesa Industrie-Elektronik GmbH Neckarstraße 19, D-45768 Marl info@mesa-gmbh.de +49 (0) 2365-97 45 1-0 +49 (0) 2365-97

More information

Gas density monitor With integrated transmitter Model GDM-100-TI

Gas density monitor With integrated transmitter Model GDM-100-TI SF 6 gas solutions Gas density monitor With integrated transmitter Model GDM-100-TI grid Products WIKA data sheet SP 60.05 for further approvals see page 5 Applications Gas density monitoring of closed

More information

E8AA. Pressure Sensor of Stainless Steel Construction Is Ideal for a Wide Range of Applications. Pressure Sensor (Stainless Steel Diaphragm)

E8AA. Pressure Sensor of Stainless Steel Construction Is Ideal for a Wide Range of Applications. Pressure Sensor (Stainless Steel Diaphragm) Pressure Sensor (Stainless Steel Diaphragm) CSM DS_E_3_1 Pressure Sensor of Stainless Steel Construction Is Ideal for a Wide Range of Applications Incorporates double diaphragms consisting of SUS316L stainless

More information

Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter

Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter Overview C % ppm 4~20mA Analog Output ppb Power Supply 100~240 VAC mg/l Dimensions 96 x 96 x 132mm RS-485 Digital Output (for DC-5310-RS only)

More information

Instruction Manual EMA 84 Digital Pressure Gauge

Instruction Manual EMA 84 Digital Pressure Gauge Instruction Manual EMA 84 Digital Pressure Gauge halstrup-walcher GmbH Stegener Straße 10 D-79199 Kirchzarten Phone: +49 (0) 76 61/39 63 0 Fax: +49 (0) 76 61/39 63 99 E-Mail: info@halstrup-walcher.com

More information

Zirconia Oxygen Sensors. Model 1230 Series

Zirconia Oxygen Sensors. Model 1230 Series Zirconia Oxygen Sensors Model 1230 Series March 2003 USING THIS MANUAL CAUTION 1 The probe or sensor heater is supplied with mains voltage. This supply has electrical shock danger to maintenance personnel.

More information

Electro-Pneumatic Converter YT-940 SERIES

Electro-Pneumatic Converter YT-940 SERIES Electro-Pneumatic Converter YT-940 SERIES PRODUCT MANUAL VERSION 1.00 Contents 1. Introduction 3 1.1 General information for the users. 3 1.2 Manufacturer Warranty 3 1.3 Explosion Proof Warning. 4 2. Product

More information

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc. Advanced Calibration Designs, Inc. Calibration Gas Instrument INSTRUCTION MANUAL Release I www.goacd.com Instruction Manual Gas Generator Release I TABLE OF CONTENTS I. General Description Page 2 II. Start-Up

More information

Instruction Manual Differential Pressure Transmitter

Instruction Manual Differential Pressure Transmitter DE61 Instruction Manual Differential Pressure Transmitter Table of Contents 1. Safety Instructions 2. Intended Applications 3. Product Description and Functions 4. Installation 5. Commissioning 6. Maintenance

More information

Two-Wire In Situ Oxygen Analyzer (550 to 1400 C)

Two-Wire In Situ Oxygen Analyzer (550 to 1400 C) Product Data Sheet PDS 106-581.A01 March, 2004 Model 5081FG Two-Wire In Situ Oxygen Analyzer (550 to 1400 C) Intrinsically safe: - CENELEC EEx ia IIC - Class I Div. I Gr. B, C, D (pending) Operates at

More information

Sensoric 4-20 ma Transmitter Board Operation Manual

Sensoric 4-20 ma Transmitter Board Operation Manual Sensoric 4-20 ma Transmitter Board Operation Manual 1 Content Features Operation Manual Technical Data Mechanical Dimensions Remarks & Contact information 2 Features Soldered sensor cell (non replaceable)

More information

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

Bante820 Portable Dissolved Oxygen Meter Instruction Manual Bante820 Portable Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante820 Portable Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante820 portable dissolved oxygen

More information

SPECIFICATIONS APCEPH1

SPECIFICATIONS APCEPH1 APCEPH1 ph CONTROLLER SPECIFICATIONS APCEPH1 Input voltage 120 Volts AC Maximum amperage 14.5 amps @ 120 VAC ph Accuracy +/- 0.2 ph ph Control range Adjustable 4.5 8.5 ph Weight < 1 lbs Dimensions 3" x

More information

Specifications of Oxygen Sensor Module V-8344A

Specifications of Oxygen Sensor Module V-8344A Specifications of Oxygen Sensor Module V-8344A 1. General: This document describes about the specification of Ceramic Oxygen Sensor module, FUJIKURA's FCX-MWL-F-(AC). Table shown below is revision records

More information

RAM 4021 Operation Manual

RAM 4021 Operation Manual RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For your safety...3 Description...3 Set-up mode...4 Annunciator lights/alarms...4 Operation...5 Calibration...6

More information

Portable Gas Monitor GX User Maintenance Manual (H4-0050)

Portable Gas Monitor GX User Maintenance Manual (H4-0050) H4E-0050 Portable Gas Monitor GX-8000 User Maintenance Manual (H4-0050) Need of Maintenance and Servicing This gas monitor must be maintained in a normal state at all times to prevent accidents due to

More information

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars.

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars. Simply a question of better measurement SCHMIDT Flow Sensor SS 20.261 The cost-effective alternative in pressurised systems up to 10 bars. Compressed air technology Industrial processes A cost analysis

More information

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual BANTE INSTRUMENTS CO., LTD DOscan10 Pocket Dissolved Oxygen Tester 1 Thank you for selecting the DOscan10 pocket dissolved oxygen tester. This

More information

METROTEC GSM. Type GSM. Oxygen measurement and control unit. *** Version 1.6 ***

METROTEC GSM. Type GSM. Oxygen measurement and control unit. *** Version 1.6 *** Systems for oxygen measurement GSM Oxygen measurement and control unit Type GSM *** Version 1.6 *** Declaration of Conformity for Oxygen measurement and control unit GSM This device has been designed for

More information

O3 3E 1 F Gas Sensor Module

O3 3E 1 F Gas Sensor Module Product Information Pack O3 3E 1 F Gas Sensor Module (Ozone) CONTENTS Product Data Sheet Product Specification 2 Poisoning and Cross Sensitivities 3 Operating Instructions Introduction 4 Electrostatic

More information

O 2 SENSORS Zirconium Dioxide (ZrO 2 ) Oxygen Sensor Operating Principle Guide

O 2 SENSORS Zirconium Dioxide (ZrO 2 ) Oxygen Sensor Operating Principle Guide O 2 SENSORS Zirconium Dioxide (ZrO 2 ) Oxygen Sensor Operating Principle Guide This document describes the physics and concepts behind CO2Meter s range of dynamic and highly accurate oxygen sensors. When

More information

User Manual. Quantos Automated Dosing Liquid Module

User Manual. Quantos Automated Dosing Liquid Module User Manual Liquid Module 1 Safety Information 1.1 Definition of warnings and symbols Signal Words WARNING for a hazardous situation with medium risk, possibly resulting in severe injuries or death if

More information

Nitrogen Supply System

Nitrogen Supply System ATMOSCOPE Nitrogen Supply System NS500 Users Guide Copyright 2001 EDSYN, INC. ATMOSCOPE Nitrogen Supply Systems TABLE OF CONTENTS Introduction............................. 3 General Information.......................

More information

Instruction Manual Contact Pressure Vacuum Gauge

Instruction Manual Contact Pressure Vacuum Gauge MS10 Instruction Manual Contact Pressure Vacuum Gauge Table of Contents 1. Safety Instructions 2. Intended Applications 3. Product Description and Functions 4. Installation 5. Commissioning 6. Maintenance

More information

MASS-STREAM TM Instruction Manual. D-6200 Analog Mass Flow Meters / Controllers

MASS-STREAM TM Instruction Manual. D-6200 Analog Mass Flow Meters / Controllers MASS-STREAM TM Instruction Manual D-6200 Analog Mass Flow Meters / Controllers Doc. No.: 9.17.112A Date: 17-03-2017 ATTENTION Before installing and operating the instrument it is strongly recommended that

More information

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr A DIVISION OF THE FREDERICKS COMPANY 2400 PHILMONT AVE. HUNTINGDONVALLEY, PA 19006 PARTS LIST 1 3 4 2 # QTY ITEM DESCRIPTION

More information

Safety. Operating instructions Solenoid valve for gas VG 6 VG 15/10 DANGER. Contents WARNING CAUTION. Changes to edition 09.14

Safety. Operating instructions Solenoid valve for gas VG 6 VG 15/10 DANGER. Contents WARNING CAUTION. Changes to edition 09.14 15 Elster GmbH Edition 7.15 Translation from the German 519 D F NL I E DK S N P GR TR CZ PL RUS H www.docuthek.com Operating instructions Solenoid valve for gas VG VG 15/1 Contents Solenoid valve for gas

More information

Membrane covered optical CO 2 sensor

Membrane covered optical CO 2 sensor Membrane covered optical CO 2 sensor - Manual - 30 mg/l CO 2, 4-20 ma output Exclusively Distributed & Supported By: www.edaphic.com.au info@edaphic.com.au Ph: 1300 430 928 Warning! Do never try to unscrew

More information

Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL -

Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL - Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL - Table of Contents 1. Introduction 3 2. Instrument Setup 4 2.1 Instrument Checklist 4 2.2 Optional Accessories 4 2.3 Gas Concentration Ranges Available

More information

T EK-SUB 4800C 19 mm Submersible Level Transmitter

T EK-SUB 4800C 19 mm Submersible Level Transmitter Technology Solutions T EK-SUB 4800C 19 mm Submersible Level Transmitter Instruction Manual Document Number: IM-4800C www.tek-trol.com Table of Contents 1 Safety Instructions... 2 1.1 Intended Use... 2

More information

Operating Instructions Temperature Measuring System MF420-5T-100

Operating Instructions Temperature Measuring System MF420-5T-100 Operating Instructions Temperature Measuring System Dnhqdhnjah MF420-5T-100 Dmsklakla Read before use! Observe all safety instructions! Keep for future reference! Fon +49/7221/64103 Fax +49/7221/17103

More information

BGA244 Binary Gas Analyzer

BGA244 Binary Gas Analyzer Quick Start Guide Revision 1.0 Certification Warranty Service certifies that this product met its published specification at the time of shipment. This product is warranted against defects in materials

More information

Please read these instructions carefully before use! Please consider the safety instructions! Please keep for future reference!

Please read these instructions carefully before use! Please consider the safety instructions! Please keep for future reference! Z11.0.1X.6C-18 Mounting and Operating Manual Atmospheric Oxygen Sensor Element GOEL 370, GOEL 381 Please read these instructions carefully before use! Please consider the safety instructions! Please keep

More information

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer July 2011 OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer WARNING! Before operating this product, read and understand this Operator s Manual. Become familiar with the potential hazards of this unit. Contact

More information

Instruction and Maintenance Manual

Instruction and Maintenance Manual Instruction and Maintenance Manual GRYF OXY ZM 02/100/2 E Contact GRYF HB, spol. s r.o. Cechova 314 Havlickuv Brod 580 01 tel.: +420 569 426 627 fax: +420 569 426 627 www.gryf.eu ver.: 11.2.2015 OEM module

More information

E2K-L. Liquid Level Sensor That Is Unaffected by the Color of the Pipe or Liquid. Liquid Level Sensor. Ordering Information

E2K-L. Liquid Level Sensor That Is Unaffected by the Color of the Pipe or Liquid. Liquid Level Sensor. Ordering Information Liquid Level EK-L CSM_EK-L_DS_E 3 Liquid Level That Is Unaffected by the Color of the or Liquid Mount to bypass pipes. Fit a wide range of pipe diameters: 8 to mm or to mm Built-in Amplifiers to save space.

More information

Safety instructions and operating manual

Safety instructions and operating manual 5-Master Safety instructions and operating manual TBF-PyroTec GmbH Lichterfelder Str. 5 A 21502 Geesthacht Tel.: + 49 (0)4152 157 9950 Fax: + 49 (0)4152 157 9951 Flame projector 5-Master Safety instructions

More information

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model User's Manual Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved

More information

Transmitter CS 21 Operation Manual

Transmitter CS 21 Operation Manual Transmitter CS 21 Operation Manual 1194 Oak Valley Drive, Suite 20, Ann Arbor, MI 48108 800-959-0573 734-769-1888 Content Page For your Safety 3 General Description 3 Detection Principle 4 Operation 4

More information

Oxymax H COS21. Technical Information. Sensor with long-term stability for frequent sterilization and autoclaving

Oxymax H COS21. Technical Information. Sensor with long-term stability for frequent sterilization and autoclaving Technical Information Oxymax H COS21 Sensor with long-term stability for frequent sterilization and autoclaving Application Process control in enzyme production Control of culture growth Biotechnological

More information

MANUAL DIRECT PURGE OPTION. UNION Instruments GmbH CWD2005 PLUS. General information, safety standards and regulations for direct purge option

MANUAL DIRECT PURGE OPTION. UNION Instruments GmbH CWD2005 PLUS. General information, safety standards and regulations for direct purge option MANUAL DIRECT PURGE OPTION UNION Instruments GmbH CWD2005 PLUS General information, safety standards and regulations for direct purge option Version: V1.00 Stand: 24.03.2010 Union Instruments GmbH Zeppelinstr.

More information

Pneumatic high-pressure controller Model CPC7000

Pneumatic high-pressure controller Model CPC7000 Calibration technology Pneumatic high-pressure controller Model CPC7000 WIKA data sheet CT 27.63 Applications Automotive and avionics industry Industry (laboratory, workshop and production) Transmitter

More information

LK-SX VOC. Application. Security Advice Caution. Notes on Disposal. Combined sensor mixed gas. Data sheet

LK-SX VOC. Application. Security Advice Caution. Notes on Disposal. Combined sensor mixed gas. Data sheet LK-SX VOC Combined sensor mixed gas Data sheet Subject to technical alteration Issue date: 24.08.2015 Application The sensor detects air quality in ventilation ducts. A stronger output signal of the sensor

More information

AQUARIUS. Operating Instructions for the. Type 70 Electrolytic Gas Soldering / Welding Unit. Table of Contents

AQUARIUS. Operating Instructions for the. Type 70 Electrolytic Gas Soldering / Welding Unit. Table of Contents AQUARIUS Operating Instructions for the Type 70 Electrolytic Gas Soldering / Welding Unit Table of Contents 1. Important General Remarks 2 2. Safety Instructions 2 3. Description of the Soldering Unit

More information

RAM Operation Manual

RAM Operation Manual RAM 4021-1 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021-1 For Your Safety... 2 Description... 2 Setup Mode... 3 Lights/Alarms... 3 Operation... 4 Calibration...

More information

Instruction and Maintenance Manual

Instruction and Maintenance Manual Instruction and Maintenance Manual GRYF OXY Z 02/100/2 E Contact GRYF HB, spol. s r.o. Cechova 314 Havlickuv Brod 580 01 tel.: +420 569 426 627 fax: +420 569 426 627 Czech Republic www.gryf.eu Technical

More information

Measurement accessories METPOINT OCV for the measurement in systems up to 40 bar

Measurement accessories METPOINT OCV for the measurement in systems up to 40 bar EN - english Instructions for installation and operation Measurement accessories METPOINT OCV for the measurement in systems up to 40 bar Dear customer, Thank you for deciding in favour of the METPOINT

More information

Model FS5001B VA.1. MEMS Mass Flow Sensors ISO 9001 ISO OHSAS 18001

Model FS5001B VA.1. MEMS Mass Flow Sensors ISO 9001 ISO OHSAS 18001 Model FS5001B MEMS Mass Flow Sensors VA.1 ISO 9001 ISO 14001 OHSAS 18001 Model FS5001B Features Low mass flow range from 0 ~ 00 sccm up to 0 ~ 1000 sccm Outstanding accuracy of ±.0 % Response time < 10

More information

LK-SX CO 2 +VOC. Application. Security Advice Caution. Notes on Disposal

LK-SX CO 2 +VOC. Application. Security Advice Caution. Notes on Disposal LK-SX CO 2 +VOC Sensor for detection of carbon dioxide (CO 2) and mixed gas content in air ducts Data sheet Subject to technical alteration Issue date: 24.08.2015 Application For detection of CO2 and VOC.

More information

EC214 EC215 - EC215R Bench Conductivity Meters

EC214 EC215 - EC215R Bench Conductivity Meters Instruction Manual EC214 EC215 - EC215R Bench Conductivity Meters http://www.hannainst.com These Instruments are in Compliance with the CE Directives Dear Customer, Thank you for choosing a Hanna Instruments

More information

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021-PR Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021-PR For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4

More information

Instruction Manual AD 1000 / BA 1000 Absolute Pressure Sensor

Instruction Manual AD 1000 / BA 1000 Absolute Pressure Sensor Instruction Manual AD 1000 / BA 1000 Absolute Pressure Sensor halstrup-walcher GmbH Stegener Straße 10 D-79199 Kirchzarten Phone: +49 (0) 76 61/39 63 0 Fax: +49 (0) 76 61/39 63 99 E-Mail: info@halstrup-walcher.com

More information

PH01 Perfusion Cannula Manual

PH01 Perfusion Cannula Manual PH01 Perfusion Cannula Manual Information in this document is subject to change without notice. No part of this document may be reproduced or transmitted without the express written permission of Multi

More information

Mass Flow Controller (MFC) for Gases

Mass Flow Controller (MFC) for Gases Mass Flow Controller (MFC) for Gases Bypass MFC with capillary technology for nominal flow rates from 5 ml N /min to 15 l N /min Applicable for aggressive gases Compact design and digital communication

More information

Operation manual Oxygen sensor

Operation manual Oxygen sensor Using oxygen sensors made by SENSORE, oxygen can be measured in gaseous media within a concentration range from 10 ppm O 2 up to 96 % O 2 depending on the sensor type. The sensor is thus characterized

More information

OXY 3690 MP. Measuring transducer for air oxygen

OXY 3690 MP. Measuring transducer for air oxygen T02.0.4X.6C-11 Measuring transducer for air oxygen OXY 3690 MP as of version V1.3 operating manual OXY 3690 MP - LO WEEE-Reg.-Nr. DE 93889386 GHM Messtechnik GmbH Standort Greisinger Hans-Sachs-Str. 26

More information

AMS 2710 PCB pressure sensor module with V output

AMS 2710 PCB pressure sensor module with V output FEATURES Universal pressure sensor module with 0.. 10 V voltage output Fully calibrated and temperature compensated sensor module Variants for (bidirectional) differential, gage, absolute and barometric

More information

Mass Flow Controller (MFC) for Gases

Mass Flow Controller (MFC) for Gases Mass Flow Controller (MFC) for Gases Type 8713 can be combined with... Direct flow measurement by MEMS- Technology for nominal flow rates from 1 ml N /min to 8 l N /min (N 2 ) High accuracy and repeatability

More information

KERN EG/EW Version /02

KERN EG/EW Version /02 E KERN EG/EW Version 1.5 07/02 Operating Instructions Electronic Precision Balances Contents 1 TECHNICAL DATA... 20 2 UNPACKING AND STANDARD ACCESSORIES... 23 3 SETTING UP THE BALANCE... 23 4 EXTERNAL

More information

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4 Calibration....

More information

Quick Installation Guide Doc nr.: C Date:

Quick Installation Guide Doc nr.: C Date: Quick Installation Guide Doc nr.: 9.17.093C Date: 13-05-2016 SCOPE OF THIS GUIDE mini CORI-FLOW instruments are highly accurate instruments for measuring and controlling the mass flow rate of liquids and/or

More information

RAM 4021-DPX Operation Manual

RAM 4021-DPX Operation Manual RAM 4021-DPX Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS ABL 4021-DPX / RAM 4021-DPX For Your Safety... 3 Description... 3 Setup Mode... 4 Lights/Alarms... 4 Operation...

More information

HUMOR 20. High-precision Humidity Calibrator HUMOR 20. Operation

HUMOR 20. High-precision Humidity Calibrator HUMOR 20. Operation High-precision Humidity Calibrator The role of humidity calibrations that are accurate, reproducible, and documentable is becoming more and more important. ISO quality guidelines and regulations according

More information

Two-Wire In Situ Oxygen Analyzer (550 to 1400 C)

Two-Wire In Situ Oxygen Analyzer (550 to 1400 C) Product Data Sheet PDS 106-581.A01 September, 2007 Model 5081FG Two-Wire In Situ Oxygen Analyzer (550 to 1400 C) Intrinsically safe: ATEX, Zone 1, EEx ia IIC T4 Class I, Div. I, Gr. A, B, C, D T4 Operates

More information

MEMS VOC Gas Sensor. (Model No.:GM-503A) Manual. Version: 1.0. Valid from: Zhengzhou Winsen Electronics Technology Co.

MEMS VOC Gas Sensor. (Model No.:GM-503A) Manual. Version: 1.0. Valid from: Zhengzhou Winsen Electronics Technology Co. MEMS VOC Gas Sensor (Model No.:GM-503A) Manual Version: 1.0 Valid from: 2015.12.1 Zhengzhou Winsen Electronics Technology Co., Ltd Statement This manual copyright belongs to Zhengzhou Winsen Electronics

More information

Precision pressure sensor Basic version Model CPT6020

Precision pressure sensor Basic version Model CPT6020 Calibration Precision pressure sensor Basic version Model CPT6020 WIKA data sheet CT 25.13 Applications Calibration technology High-accuracy pressure monitoring Pressure sensing in critical applications

More information

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4 Calibration....

More information

FTS SUBMERSIBLE PRESSURE TRANSMITTER USER S MANUAL

FTS SUBMERSIBLE PRESSURE TRANSMITTER USER S MANUAL FTS SUBMERSIBLE PRESSURE TRANSMITTER USER S MANUAL TABLE OF CONTENTS PRODUCT OVERVIEW 2 I - USE AND CARE: 3 II - INSTALLATION: 5 III - GENERAL MAINTENANCE TIPS: 5 IV - APPENDIX: A-1 2-WIRE CURRENT LOOP

More information

Operating instructions Safety Rope Emergency Stop Switches ZB0052 / ZB0053 ZB0072 / ZB0073

Operating instructions Safety Rope Emergency Stop Switches ZB0052 / ZB0053 ZB0072 / ZB0073 Operating instructions Safety Rope Emergency Stop Switches UK ZB0052 / ZB0053 ZB0072 / ZB0073 7390878 / 02 03 / 2011 Contents 1 Safety instructions...3 2 Installation / set-up...4 2.1 Applications...4

More information

Pneumatic high-pressure controller Model CPC7000

Pneumatic high-pressure controller Model CPC7000 Calibration technology Pneumatic high-pressure controller Model CPC7000 WIKA data sheet CT 27.63 Applications Healthcare and avionics industry Industry (laboratory, workshop and production) Transmitter

More information

TC65M/15M TEMPERATURE CALIBRATOR USER S MANUAL

TC65M/15M TEMPERATURE CALIBRATOR USER S MANUAL TC65M/15M TEMPERATURE CALIBRATOR USER S MANUAL Thank you for purchasing a Scan-Sense temperature calibrator. The products are manufactured by Scan-Sense AS in accordance with our high quality standards

More information

HI 2314 HI 2315 HI 23151

HI 2314 HI 2315 HI 23151 Instruction Manual HI 2314 HI 2315 HI 23151 Multi-Range Conductivity Meters for Laboratories www.hannainst.com Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction

More information

In Vivo Scientific, LLC INSTRUCTION MANUAL

In Vivo Scientific, LLC INSTRUCTION MANUAL CO 2 Controller In Vivo Scientific, LLC INSTRUCTION MANUAL CONTENTS CONTENTS...1 ABOUT THIS MANUAL...2 INTRODUCTION...2 Cautions and Warnings...2 Parts List...2 Unpacking...2 INSTRUMENT DESCRIPTION...3

More information

High-performance submersible pressure transmitter For level measurement Model LH-10

High-performance submersible pressure transmitter For level measurement Model LH-10 Electronic pressure measurement High-performance submersible pressure transmitter For level measurement Model LH-10 WIKA data sheet PE 81.09 Applications Level measurement in rivers and lakes Deep well

More information

...makes sense! Combined CO 2 /O 2 analyzers. Monitor your process in real-time: OUR CER. BlueSens.com

...makes sense! Combined CO 2 /O 2 analyzers. Monitor your process in real-time: OUR CER. BlueSens.com ...makes sense! Combined CO 2 /O 2 analyzers Monitor your process in real-time: BlueSens.com OUR CER RQ ...makes sense! The BlueInOne sensor BlueSens is providing modern and inexpensive measuring solutions

More information

OxiQuant S. Operating Instructions

OxiQuant S. Operating Instructions ENVITEC -WISMAR GMBH Environmental & Medical Sensor Technology OxiQuant S Operating Instructions P/N: 46-00-0021 EnviteC-Wismar GmbH Alter Holzhafen 18, D-23966 Wismar, Germany Tel.: +49-(0)-3841-360-1

More information

SVEA II GSM Connector User Manual

SVEA II GSM Connector User Manual www.possio.com SVEA II GSM Connector User Manual 1. Introduction Before using SVEA II please read this manual carefully and keep it for future reference. SVEA II GSM Connector is a device for realization

More information

AC1810 / AC1810-A TECHNICAL SPECIFICATIONS. Operating Pressure psi ( kgs/cm²) [AC1810] Displacement. Net Weight

AC1810 / AC1810-A TECHNICAL SPECIFICATIONS. Operating Pressure psi ( kgs/cm²) [AC1810] Displacement. Net Weight Technical Specifications Operating Instructions Maintenance Information Troubleshooting Guide Parts Diagrams AC1810 / AC1810-A THE EVOLUTION OF PERFECTION CAUTION: Before attempting to use or service this

More information

Instruction Manual Dräger MSI P7 and MSI P7 plus

Instruction Manual Dräger MSI P7 and MSI P7 plus Dräger MSI GmbH Rohrstraße 32 58093 Hagen Tel.: +49-2331 / 9584-0 Fax: +49-2331 / 9584-29 e-mail: info@draeger-msi.de D 923; Edition 2011-01-01 Content 1. General Hints Page 4 2. The Instrument 2.1 Front

More information

AIR HAMMERS MODEL NO: CAT138/CAT139 OPERATING & MAINTENANCE INSTRUCTIONS PART NO: / GC064

AIR HAMMERS MODEL NO: CAT138/CAT139 OPERATING & MAINTENANCE INSTRUCTIONS PART NO: / GC064 AIR HAMMERS MODEL NO: CAT138/CAT139 PART NO: 3120152 /3120153 OPERATING & MAINTENANCE INSTRUCTIONS GC064 INTRODUCTION Thank you for purchasing this CLARKE Air Hammer. Before attempting to use this product,

More information

Model GFC 7000T CO 2 Analyzer (Addendum to GFC 7000TA Manual, PN 07272)

Model GFC 7000T CO 2 Analyzer (Addendum to GFC 7000TA Manual, PN 07272) Manual Addendum Model GFC 7000T CO 2 Analyzer (Addendum to GFC 7000TA Manual, PN 07272) P/N M07273 DATE 06/11/13 TELEDYNE ELECTRONIC TECHNOLOGIES Analytical Instruments 16830 Chestnut Street City of Industry,

More information

GASGUARD VENT LINE3 Ammonia Sensor OPERATING & INSTALLATION MANUAL

GASGUARD VENT LINE3 Ammonia Sensor OPERATING & INSTALLATION MANUAL GASGUARD VENT LINE3 Ammonia Sensor OPERATING & INSTALLATION MANUAL Operating and Installation Manual Warning Use this product only in the manner described in this manual. If the equipment is used in a

More information

Supplementary Operator s Manual 42/24-14 EN Rev. 3

Supplementary Operator s Manual 42/24-14 EN Rev. 3 Advance Optima Continuous Gas Analyzers AO2000 Series with Uras26, Magnos206, Caldos25, Caldos27 Designed per Category 3G for Measurement of Flammable Gases ( Safety Concept ) and Non-flammable Gases Supplementary

More information

* * Data Sheet and Operating Manual. Digital manometer ME01 ## # 87 # HL S#### Dust explosion protection Zone 22.

* * Data Sheet and Operating Manual. Digital manometer ME01 ## # 87 # HL S#### Dust explosion protection Zone 22. *09005087* DB_BA_EN_ME01_S Rev.A 12/16 *09005087* d e v e l o p i n g s o l u t i o n s Data Sheet and Operating Manual ME01 Digital manometer Table of Contents ME01 ## # 87 # HL S#### Dust explosion protection

More information