The effect of inner tube placement on resistance and work of breathing through tracheostomy tubes: a bench test

Size: px
Start display at page:

Download "The effect of inner tube placement on resistance and work of breathing through tracheostomy tubes: a bench test"

Transcription

1 Anaesthesia 3, 68, 76 8 Original Article doi:./anae.94 The effect of inner tube placement on resistance and work of breathing through tracheostomy tubes: a bench test A. Carter, S. J. Fletcher and R. Tuffin 3 Consultant in Anaesthesia and Intensive Care Medicine, Mid Yorkshire Hospitals, Wakefield, UK Director of Intensive Care, 3 Advanced Trainee in Intensive Care Medicine, Bradford Teaching Hospitals NHS Foundation Trust, Bradford, UK Summary To reduce the risk of tracheostomy tube blockage, a removable inner tube can be used. However, this will reduce the size of the lumen and will increase airflow resistance and work of breathing. The magnitude of this increase in workload is unknown. We undertook a bench test to measure the effect. A lung model was developed to breathe through the tracheostomy tube. We created pressure volume curves from which we calculated work of breathing with and without an inner tube using 6 mm tracheostomy tubes over a range of respiratory rates and tidal volumes. The inner tube increased the resistive work of breathing by an average factor of.. The extra work of breathing imposed easily exceeded the normal total work of breathing. Our results will aid a risk benefit analysis when deciding whether to use inner tubes. Selecting a larger tracheostomy tube is likely to aid weaning from mechanical ventilation.... Correspondence to: S. Fletcher stephen.fletcher@bthft.nhs.uk Accepted: 8 October Failure to wean critically ill patients from mechanical ventilation is largely due to an imbalance between the required work of breathing and weakness of the patient s respiratory pump (due to muscle wasting and critical illness neuromyopathy). Work of breathing is increased due to a combination of poor respiratory system compliance and the imposed resistive work of breathing due to the artificial airway and breathing system. In these patients, a tracheostomy may be formed with the aim of facilitating weaning mainly by reducing the need for sedation but also by reducing airway resistance when compared with trans-laryngeal tracheal tubes [, ]. Blockage of the tracheostomy tube with accumulated secretions may be life-threatening and the use of a replaceable inner tube is suggested as a way of reducing this risk [3]. However, inserting the inner tube decreases the internal diameter of the tracheostomy lumen and therefore will increase the resistance and resistive component of the work of breathing. According to the Hagen Poiseuille law, assuming laminar flow, resistance is inversely proportional to the fourth power of the radius. Thus, a small change in radius may have a significant effect on work of breathing and may adversely affect weaning. If flow is turbulent then the effect of a change in radius is even more marked. We theorised that there was likely to be a mixture of laminar and turbulent flow in the native and artificial airways; furthermore, the proportion of turbulent flow would increase as the peak flow rates increased. For these reasons, the effect of an inner tube on resistance and work of breathing is not amenable to calculation. Therefore, we undertook a 76 Anaesthesia The Association of Anaesthetists of Great Britain and Ireland

2 Carter et al. The work of breathing through tracheostomy inner tubes Anaesthesia 3, 68, 76 8 Table Internal diameters of Portex Blueline Ultra tracheostomy tubes. Tracheostomy tube size External diameter of tracheostomy tube; mm Internal diameter of tracheostomy tube without inner tube; mm Internal diameter of inner tube; mm bench study to assess that effect of the introduction of the inner tube using one of the more commonly used tracheostomy tubes in the UK (Table ). Methods In the first test (steady-state), we looked at resistance to flow through the tracheostomy tube under conditions of constant flow. In the dynamic tests, we measured work of breathing using a lung model. Portex Ò Blueline Ultra Ò tracheostomy tubes of sizes 6 were studied (Smiths Medical International Ltd, Hythe, UK). Measurements were made using a pneumotachograph (VT Plus HF; Fluke Biomedical, Everett, WA, USA). As expiration is normally passive, we only considered inspiratory work of breathing. Five tracheostomy tubes of each size were tested and data were expressed as the mean driving pressure. For the steady-state test, we attached a gas source capable of delivering variable flows to the upstream side of the pneumotachograph via -mm tubing. The tracheostomy tube to be tested was inserted via its 5-mm connector to the downstream side of the pneumotachograph. The driving pressure needed to create flows from to l.min was then recorded at -l.min steps for each size of tracheostomy tube, with and without the inner tube in place. For the dynamic test, we used a model of spontaneous respiration as described previously (Fig. ) [4, 5]. It comprises two mechanically linked bellows: the master bellows are driven by a Siemens Servo 3A ventilator (Siemens Healthcare, Surrey, UK) using a volume control mode, and with an inspiratory to expiratory (I:E) ratio of :. The slave bellows, which represent the lung, were connected via the pneumotachograph to the lower end of an anatomically accurate model trachea (taken from an Figure Diagram of the lung model used for the bench tests. intubation manikin), into which we inserted the test tracheostomy tube [, 6]. The proximal end of the tracheostomy tube was open to atmosphere. The tracheostomy tube was inserted into the model trachea through an incision at a level approximating the space between the second and third tracheal rings. The cuff was inflated and the correct orientation of the tip of the tube parallel to the tracheal wall was confirmed visually. Synchronous pressure and flow data were sampled at 5 Hz throughout the respiratory cycle at different respiratory rates (,, 3, 4 min ) and different tidal volumes (3, 5, 7, 9 ml). The data were then uploaded to Microsoft Excel allowing us to calculate the area of the inspiratory portion of the pressure volume loops to obtain the resistive work of breathing, measured in Joules [7]. To enable comparison, these values were then indexed by tidal volume to Joules per litre (resistive work of breathing (J.l ) = work for one breath (J)/tidal volume (l)). In a bench test not involving biological material, the only sources of measurement variability and error reside in the test subject (in this case, tracheostomy tubes) and in the measuring equipment. The manufacturing tolerances of the Portex Blueline Ultra tracheostomy tubes are.5 mm on the diameter and. mm on the Anaesthesia The Association of Anaesthetists of Great Britain and Ireland 77

3 Anaesthesia 3, 68, 76 8 Carter et al. The work of breathing through tracheostomy inner tubes length (information supplied by Portex). The flow measurement error of the pneumotachograph is %. Therefore, the results derived from one exemplar tracheostomy tube should be valid. However, to confirm this, we ran the steady-state tests with five tracheostomy tubes of each size and calculated the mean and standard deviations for the driving pressure. The variability in the measurements was very small as predicted and the dynamic tests were therefore performed using one representative tube of each size. Pressure (cmho) Results Steady-state test In the steady-state tests, the driving pressures required to produce a given flow increases in a curvilinear fashion confirming turbulent flow in the tracheostomy tubes (Fig. ). Table shows the mean (SD) driving pressures at each flow rate. Dynamic test Figure 3a d show the measured imposed work of breathing in the tracheostomy tubes with and without Flow (l.min ) Figure Pressure-flow characteristics of differing tracheostomy tubes with and without inner tubes: ( ) 6 inner, ( ) 6 no inner, ( ) 7 inner, ( ) 7 no inner, ( ) 8 inner, ( ) 8 no inner, ( ) 9 inner, ( )9no inner, ( ) inner, ( ) no inner. Table Results of the steady-state tests. Figures are mean (SD) driving pressures in cmh O. Tube size inner cannula l.min 4 l.min 6 l.min 8 l.min l.min 6 inner 4.8 (.8) 4.8 (.) 3.78 (.4) 5.94 (.6) (.95) 6 no inner.96 (.6) 7.6 (.9) 5.4 (.3) 5. (.36) (.55) 7 inner 3. (.3).3 (.4). (.5) 36.7 (.9) 54.8 (.8) 7 no inner. (.4) 3.66 (.6) 7.6 (.6).8 (.) 9. (.54) 8 inner.7 (.5) 5.56 (.9).68 (.) 9.84 (.5) 9.84 (.45) 8 no inner.6 (.4).8 (.) 4.64 (.5) 7.94 (.9).9 (.46) 9 inner. (.4) 3.8 (.4) 6.9 (.7).4 (.6) 6.94 (.9) 9 no inner.4 (.4).5 () 3.6 (.5) 5.8 (.4) 7.6 (.8) inner.6 (). (.4) 4.8 (.4) 7. (.7).56 (.6) no inner.3 ().6 (.5).4 (.5) 3.78 (.4) 5.5 (.4) their inner cannula at respiratory rates 4 min. The normal total work of breathing in healthy subjects (thought to be the optimal work of breathing) of.5 J.l is also shown [7]. As the tidal volume and the respiratory rate rises, there is an increase in resistive work of breathing. This increase is more marked in the smaller tracheostomy tubes. Figure 4a d show the additional imposed work of breathing due to insertion of the inner cannula. This was calculated by subtracting the resistive work of breathing of each tracheostomy tube without its inner cannula from the value with its inner cannula for each respiratory rate/tidal volume combination. It can be seen from these graphs that at low respiratory rates the imposed resistive work of breathing added by the inner cannula is modest. However, as the respiratory rates and tidal volumes increase, the added work by placing the inner cannula rapidly exceeds the optimal total work of breathing in the size 6 and 7 tubes. In an attempt to summarise these data, we calculated the average ratio of the work of breathing with and without inner tube for each size of tracheostomy, 78 Anaesthesia The Association of Anaesthetists of Great Britain and Ireland

4 Carter et al. The work of breathing through tracheostomy inner tubes Anaesthesia 3, 68, 76 8 (a) (b) (c) (d) Figure 3 (a d) Imposed resistive work of breathing for differing tracheostomy tubes (with and without inner cannula) at increasing respiratory rates (a, min ; b, min ; c, 3 min ; d, 4 min ). Key for tube sizes inner cannula: ( ) 6 inner, ( ) 6 no inner, ( ) 7 inner, ( ) 7 no inner, ( ) 8 inner, ( ) 8 no inner, ( ) 9 inner, ( ) 9 no inner, ( ) inner, ( ) no inner. ( ) Normal total work of breathing. for each of the test conditions. For the size 6, the work of breathing on average increased by a factor of.; for size 7, a factor of.8; for size 8, a factor of.3; for size 9, a factor of.9; and for size, a factor of.6. Discussion The results from the steady-state test show that flow in the tracheostomy tubes is turbulent, and therefore resistance is inversely proportional to the fifth power of the radius. The driving pressures from the static tests are equivalent to the negative inspiratory pressure a spontaneously breathing patient would have to generate in the trachea to produce an equivalent gas flow. The pressure required to produce greater gas flows increases rapidly in a non-linear fashion, and the effect is amplified with smaller internal diameters. Inspiratory pressures of greater than cmh O are required to produce flows of 6 l.min or more in tubes with an internal diameter of 6 mm or less. Flow rates of more than l.min are often seen in respiratory failure [5]. These data suggest inter alia that it would be almost impossible to sustain these flow rates when a size 6 or 7 tracheostomy of this type is used with its inner cannula. The results from the dynamic tests show that placement of the inner cannula in Portex Blueline Anaesthesia The Association of Anaesthetists of Great Britain and Ireland 79

5 Anaesthesia 3, 68, 76 8 Carter et al. The work of breathing through tracheostomy inner tubes.6 (a) (b) (c) (d) Figure 4 (a d) Additional work of breathing due to inner tube for differing tracheostomy tubes at increasing respiratory rates (a, min ; b, min ; c, 3 min ; d, 4 min ). ( ) Normal total work of breathing. Key for tube sizes: ( ) Size 6, ( ) Size 7, ( ) Size 9, ( ) Size. Ultra tracheostomy tubes significantly increases imposed inspiratory resistive work of breathing. The effect was greatest with the size 7. tube, with an increase in resistive work of breathing by a factor of.8 at clinically relevant respiratory rates and tidal volumes. This is because the inner cannula in the 6. tube only reduces the internal diameter by mm compared with a.5 mm decrease in internal diameter in all other tubes tested (see Table ). The 7. tube therefore has the greatest proportional change in internal diameter when the inner tube is inserted. Total work of breathing in normal subjects is in the range of.3.6 J.l [7]. The optimal work of breathing whilst weaning from a ventilator is difficult to assess; if too low, respiratory muscle wasting can develop; if too high, muscle fatigue and respiratory failure may be induced. The total work of breathing in the normal healthy range would seem to correspond to a reasonable muscle and energy load in both weaning and respiratory failure [7]. Our experiment only considers inspiratory imposed resistive work of breathing. Even so, the results show that in the smaller tubes with the inner cannula in place, the resistive work of breathing starts to exceed normal total work of breathing at fairly low tidal volumes and respiratory rates. Increased imposed work of breathing has been shown to cause failure to wean from mechanical ventilation [8]. In vivo, the imposed work of breathing will be cumulative with the resistive work of breathing of the rest of the bronchial tree and the elastic and viscous work of breathing used to expand the lungs. If the total work is already high due to poor lung compliance or airway narrowing, then even small increases in imposed work are highly likely to cause respiratory decompensation and failure to wean. The increased imposed resistive work of breathing caused by use of the inner cannula does, however, need to be weighed against the important benefits that the inner tube provides. It can be removed readily and cleaned or replaced to remove secretions. Encrusted secretions can be difficult or impossible to remove with suctioning alone. It has been shown that organised secretions in tracheal tubes can reduce the effective internal tube diameter causing a significantly increased resistance and imposed resistive work of breathing [9]. Perhaps the major benefit of the inner cannula is the ability to remove it should the inner lumen become completely blocked. This makes treatment of a potentially life-threatening emergency situation much quicker and easier without the need for an emergency removal and/or replacement of a new tracheostomy tube. 8 Anaesthesia The Association of Anaesthetists of Great Britain and Ireland

6 Carter et al. The work of breathing through tracheostomy inner tubes Anaesthesia 3, 68, 76 8 In 8, the UK Intensive Care Society published Standards for the care of adult patients with a temporary tracheostomy in which they state Tracheostomy tubes with an inner cannula are inherently safer and are normally preferred [3]. The document also states: As a general rule, most adult females can accommodate a tube with an outside diameter of mm, whilst a tube with an outside diameter of mm is suitable for most adult males. Using Portex Blueline Ultra tracheostomy tubes, this gives a maximum size of 7 in males and size 6 in females. With the inner cannula in situ, the internal diameter is 5.5 and 5 mm, respectively. Our results show that these tube sizes produce a marked restriction to flow and increased work of breathing and thus the guidelines may not be appropriate. Whilst a larger diameter lumen would appear beneficial, we are limited by the resulting outside diameter that the trachea is able to accommodate. Reducing the difference between the internal and external diameter would seem to be advantageous, but would require a thinner tube wall, whilst retaining wall strength. Portex Blueline Ultra tubes are made from thermosensitive PVC, which is designed to be rigid enough for percutaneous insertion and then soften once in situ to accommodate individual anatomy. Whilst thinner tubes are available, they tend to be rigid, for example, the Shiley and Trachoe. A more rigid tube will transmit more of the weight of the breathing system onto the tracheal wall, arguably increasing the risk of mucosal ischaemia and resultant fibrosis. Imposed work of breathing during spontaneous respiration can be offset to a certain degree by using inspiratory assist. Our results suggest, however, that conventional pressure support ventilation may not fully overcome the resistive work imposed by smaller diameter tracheostomy tubes. Automated tube compensation employs a mathematical model to calculate the tube resistance for specific tube sizes at differing flow rates. The delivered pressure support can then be adjusted to offset the resistance. Automated tube compensation has been shown to be more effective at reducing imposed work of breathing compared with simple pressure support ventilation alone, whilst improving patient comfort, and maintaining cardiovascular stability [ ]. The use of such ventilatory modes may improve weaning success, particularly when smaller tracheostomy tubes are employed. However, it is important to remember to set the actual internal diameter of the tube when an inner cannula is used, rather than the size written on the box or cuff e.g. a size 8. Portex Blueline Ultra with an inner cannula in situ has an inner diameter of 6.5 mm not 8 mm. It is likely that the values for work of breathing for any given lumen size that we measured will be similar for the same sizes but different makes of tracheostomy tube. Differences may occur due to differing design (for example different radii, different lengths). Our study has some limitations. At higher inspiratory pressures (> 3 cmh O), the slave bellows tended to collapse; therefore, we could not generate data for the smaller tubes at the extremes of the test protocol. However, it is equally likely that a human would be unable to generate such negative pressures. Thus, loss of these data does not alter our conclusions. Another potential criticism is that we did not look at the expiratory work of breathing. Whilst expiration is normally passive, at expiratory pressures greater than cmh O, expiration becomes active [3]. Expiratory pressure will correlate with inspiratory pressure and thus we are likely to have underestimated total resistive work of breathing with smaller tubes and higher respiratory settings. However, this makes our findings more, not less, significant. Our results will not guide decisions whether to use fenestrated tubes or deflate the tracheostomy cuff as part of the weaning process. However, if the tracheostomy tube is capped off, forcing the patient to breathe either through the fenestration and/or around the tube, then thought needs to be given to the size of the fenestration and the cross-sectional area of trachea between the tube and the tracheal wall. As we have demonstrated, small changes in airway dimensions can impose significant increases in respiratory workload. In conclusion, although the decision of whether to use an inner cannula is complex in patients with weaning difficulty, our data provide an estimate of the increase in imposed work of breathing when they are used and thus should inform this decision making. We would suggest the following: Select the largest possible tracheostomy tube. Anaesthesia The Association of Anaesthetists of Great Britain and Ireland 8

7 Anaesthesia 3, 68, 76 8 Carter et al. The work of breathing through tracheostomy inner tubes When a patient is failing to wean, consideration should be given to the possibility that the artificial airway is imposing excessive respiratory work. 3 Do not routinely use inner tubes where there is adequate nursing and suitable medical supervision, but check regularly for accumulation of secretions. 4 When a patient is to be discharged to a non-critical care environment with a tracheostomy in situ, then an inner tube should be inserted. A 4-h trial to demonstrate that the patient can cope with the extra work of breathing due to the inner tube should be performed before discharge. 5 During weaning, automated tube compensation may help to offset the imposed work of breathing due to presence of an inner tube. Competing interests No external funding and no competing interests declared. References. Yung MW, Snowdon SL. Respiratory resistance of tracheostomy tubes. Archives of Otolaryngology 984; : Epstein SK. Anatomy and physiology of tracheostomy. Respiratory Care 5; 5: Intensive Care Society (UK). Standards for the care of adult patients with a temporary tracheostomy, London: Intensive Care Society, Lyle DJR, Nunn JF, Hawes DWC, Dickins J, Tate M, Baker JA. A model lung with the capacity for simulated spontaneous breathing. British Journal of Anaesthesia 984; 56: Glover GW, Fletcher SJ. Assessing the performance of the Whisperflow continuous positive airway pressure generator: a bench study. British Journal of Anaesthesia 9; : Mehta S, Myat HM. The cross-sectional shape and circumference of the human trachea. Annals of the Royal College of Surgeons of England 984; 66: van Heerde M, Roubik K, Kopelent V, Pl otz FB, Markhorst DG. Unloading work of breathing during high-frequency oscillatory ventilation: a bench study. Critical Care 6; : R3. 8. Kirton OC, DeHaven CB, Morgan JP, Windsor J, Civetta JM. Elevated imposed work of breathing masquerading as ventilator weaning intolerance. Chest 995; 8: Wilson AM, Gray DM, Thomas JG. Increases in endotracheal tube resistance are unpredictable relative to duration of intubation. Chest 9; 36: Haberth ur C, Fabry B, Stocker R, Ritz R, Guttmann J. Additional inspiratory work of breathing imposed by tracheostomy tubes and non-ideal ventilator properties in critically ill patients. Intensive Care Medicine 999; 5: Guttmann J, Bernhard H, Mols G, et al. Respiratory comfort of automatic tube compensation and inspiratory pressure support in conscious humans. Intensive Care Medicine 997; 3: Guttmann J, Haberth ur C, Mols G, Lichtwarck-Aschoff M. Automatic tube compensation (ATC). Minerva Anestesiologica ; 68: Nunn J. Nunns Applied Respiratory Physiology. 4th edn. Cambridge: Butterworth Heinemann, Anaesthesia The Association of Anaesthetists of Great Britain and Ireland

RESPIRATORY PHYSIOLOGY, PHYSICS AND PATHOLOGY IN RELATION TO ANAESTHESIA AND INTENSIVE CARE

RESPIRATORY PHYSIOLOGY, PHYSICS AND PATHOLOGY IN RELATION TO ANAESTHESIA AND INTENSIVE CARE Course n : Course 3 Title: RESPIRATORY PHYSIOLOGY, PHYSICS AND PATHOLOGY IN RELATION TO ANAESTHESIA AND INTENSIVE CARE Sub-category: Intensive Care for Respiratory Distress Topic: Pulmonary Function and

More information

Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 6th Edition by Cairo

Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 6th Edition by Cairo Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 6th Edition by Cairo Link full download: http://testbankair.com/download/test-bank-for-pilbeams-mechanicalventilation-physiological-and-clinical-applications-6th-edition-by-cairo/

More information

RSPT 1060 OBJECTIVES OBJECTIVES OBJECTIVES EQUATION OF MOTION. MODULE C Applied Physics Lesson #1 - Mechanics. Ventilation vs.

RSPT 1060 OBJECTIVES OBJECTIVES OBJECTIVES EQUATION OF MOTION. MODULE C Applied Physics Lesson #1 - Mechanics. Ventilation vs. RSPT 1060 MODULE C Applied Physics Lesson #1 - Mechanics OBJECTIVES At the end of this module, the student should be able to define the terms and abbreviations used in the module. draw & explain the equation

More information

Completed downloadable Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 5th Edition by Cairo

Completed downloadable Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 5th Edition by Cairo Completed downloadable Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 5th Edition by Cairo Link full download: http://testbankcollection.com/download/pilbeams-mechanicalventilation-physiological-and-clinical-applications-5th-edition-test-bank-cairo

More information

Instructions/Procedure for Use of the Speaking Valve

Instructions/Procedure for Use of the Speaking Valve Instructions/Procedure for Use of the Speaking Valve Notes on the use of the speaking valve for ventilator-dependent and non-ventilator dependent patients The speaking valve is a one-way valve that allows

More information

Mechanical Ventilation. Which of the following is true regarding ventilation? Basics of Ventilation

Mechanical Ventilation. Which of the following is true regarding ventilation? Basics of Ventilation Mechanical Ventilation Jeffrey L. Wilt, MD, FACP, FCCP Associate Professor of Medicine Michigan State University Associate Program Director MSU-Grand Rapids Internal Medicine Residency Which of the following

More information

Chapter 4: Ventilation Test Bank MULTIPLE CHOICE

Chapter 4: Ventilation Test Bank MULTIPLE CHOICE Instant download and all chapters Test Bank Respiratory Care Anatomy and Physiology Foundations for Clinical Practice 3rd Edition Will Beachey https://testbanklab.com/download/test-bank-respiratory-care-anatomy-physiologyfoundations-clinical-practice-3rd-edition-will-beachey/

More information

Sign up to receive ATOTW weekly -

Sign up to receive ATOTW weekly - THE PHYSICS OF FLOW ANAESTHESIA TUTORIAL OF THE WEEK 84 9TH APRIL 2008 Paul Clements, SpR in Anaesthetics, Hope Hospital, Salford, UK. Carl Gwinnutt, Consultant Anaesthetist, Hope Hospital, Salford, UK.

More information

Mechanical Ventilation

Mechanical Ventilation Mechanical Ventilation Understanding Modes Rob Chatburn, RRT-NPS, FAARC Research Manager Respiratory Therapy Cleveland Clinic Associate Professor Case Western Reserve University 1 Overview Characteristics

More information

Basics of Mechanical Ventilation. Dr Shrikanth Srinivasan MD,DNB,FNB,EDIC Consultant, Critical Care Medicine Medanta, The Medicity

Basics of Mechanical Ventilation. Dr Shrikanth Srinivasan MD,DNB,FNB,EDIC Consultant, Critical Care Medicine Medanta, The Medicity Basics of Mechanical Ventilation Dr Shrikanth Srinivasan MD,DNB,FNB,EDIC Consultant, Critical Care Medicine Medanta, The Medicity Overview of topics 1. Goals 2. Settings 3. Modes 4. Advantages and disadvantages

More information

Selecting the Ventilator and the Mode. Chapter 6

Selecting the Ventilator and the Mode. Chapter 6 Selecting the Ventilator and the Mode Chapter 6 Criteria for Ventilator Selection Why does the patient need ventilatory support? Does the ventilation problem require a special mode? What therapeutic goals

More information

COMPARISON OF PORTABLE EMERGENCY VENTILATORS USING A LUNG MODEL

COMPARISON OF PORTABLE EMERGENCY VENTILATORS USING A LUNG MODEL British Journal of Anaesthesia 1993; 70: 2-7 APPARATUS COMPARISON OF PORTABLE EMERGENCY VENTILATORS USING A LUNG MODEL L. ATTEBO, M. BENGTSSON AND A. JOHNSON SUMMARY A lung model was used to test the performance

More information

CARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS

CARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS Brit. J. Anaesth. (1956), 28, 196 CARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS BY RUSSELL M. DAVIES, I. R. VERNER Queen Victoria Hospital, East Grinstead AND A. BRACKEN Research and Development Centre,

More information

Automatic Transport Ventilators. ICU Quality Ventilation on the Street.

Automatic Transport Ventilators. ICU Quality Ventilation on the Street. Automatic Transport Ventilators ICU Quality Ventilation on the Street. Kevin Bowden, March 20 th 2014 Ventilator Definition A ventilator is an automatic mechanical device designed to provide all or part

More information

Health Professional Info

Health Professional Info Health Professional Info Mouthpiece Ventilation (MPV) What is MPV? MPV is a less intrusive form of noninvasive ventilation that uses a portable home mechanical ventilator (HMV) with a single-limb open-circuit

More information

MEDICAL EQUIPMENT IV MECHANICAL VENTILATORS. Prof. Yasser Mostafa Kadah

MEDICAL EQUIPMENT IV MECHANICAL VENTILATORS. Prof. Yasser Mostafa Kadah MEDICAL EQUIPMENT IV - 2013 MECHANICAL VENTILATORS Prof. Yasser Mostafa Kadah Mechanical Ventilator A ventilator is a machine, a system of related elements designed to alter, transmit, and direct energy

More information

4/18/12 MECHANISM OF RESPIRATION. Every Breath You Take. Fun Facts

4/18/12 MECHANISM OF RESPIRATION. Every Breath You Take. Fun Facts Objectives MECHANISM OF RESPIRATION Dr Badri Paudel Explain how the intrapulmonary and intrapleural pressures vary during ventilation and relate these pressure changes to Boyle s law. Define the terms

More information

Advanced Ventilator Modes. Shekhar T. Venkataraman M.D. Professor Critical Care Medicine and Pediatrics University of Pittsburgh School of Medicine

Advanced Ventilator Modes. Shekhar T. Venkataraman M.D. Professor Critical Care Medicine and Pediatrics University of Pittsburgh School of Medicine Advanced Ventilator Modes Shekhar T. Venkataraman M.D. Shekhar T. Venkataraman M.D. Professor Critical Care Medicine and Pediatrics University of Pittsburgh School of Medicine Advanced modes Pressure-Regulated

More information

Ventilators. Dr Simon Walton Consultant Anaesthetist Eastbourne DGH KSS Basic Science Course

Ventilators. Dr Simon Walton Consultant Anaesthetist Eastbourne DGH KSS Basic Science Course Ventilators Dr Simon Walton Consultant Anaesthetist Eastbourne DGH KSS Basic Science Course Objectives Discuss Classification/ terminology Look at Modes of ventilation How some specific ventilators work

More information

Lab 3. The Respiratory System (designed by Heather E. M. Liwanag with T.M. Williams)

Lab 3. The Respiratory System (designed by Heather E. M. Liwanag with T.M. Williams) Name Lab Partners Lab 3. The Respiratory System (designed by Heather E. M. Liwanag with T.M. Williams) Part 1. Lung Volumes and Capacities Objectives 1. Obtain graphical representation of lung capacities

More information

Organis TestChest. Flight Simulator for Intensive Care Clinicians

Organis TestChest. Flight Simulator for Intensive Care Clinicians Organis TestChest Flight Simulator for Intensive Care Clinicians Organis TestChest Critical Care challenges Training on ventilation modes with simulation is crucial for patient safety The equipment and

More information

PART EIGHT HIGH FREQUENCY PERCUSSIVE OSCILLATION (HFPOV )

PART EIGHT HIGH FREQUENCY PERCUSSIVE OSCILLATION (HFPOV ) PART EIGHT HIGH FREQUENCY PERCUSSIVE OSCILLATION (HFPOV ) Note: For maximal comparative understanding, FIRST read PART SEVEN which defines the concept of High Frequency Oscillatory Ventilation (HFOV).

More information

Introduction. Respiration. Chapter 10. Objectives. Objectives. The Respiratory System

Introduction. Respiration. Chapter 10. Objectives. Objectives. The Respiratory System Introduction Respiration Chapter 10 The Respiratory System Provides a means of gas exchange between the environment and the body Plays a role in the regulation of acidbase balance during exercise Objectives

More information

Spontaneously Breathing Lung Model Comparison of Work of Breathing Between Automatic Tube Compensation and Pressure Support

Spontaneously Breathing Lung Model Comparison of Work of Breathing Between Automatic Tube Compensation and Pressure Support Spontaneously Breathing Lung Model Comparison of Work of Breathing Between Automatic Tube Compensation and Pressure Support Yuji Fujino MD, Akinori Uchiyama MD, Takashi Mashimo MD, and Masaji Nishimura

More information

EMS INTER-FACILITY TRANSPORT WITH MECHANICAL VENTILATOR COURSE OBJECTIVES

EMS INTER-FACILITY TRANSPORT WITH MECHANICAL VENTILATOR COURSE OBJECTIVES GENERAL PROVISIONS: EMS INTER-FACILITY TRANSPORT WITH MECHANICAL VENTILATOR COURSE OBJECTIVES Individuals providing Inter-facility transport with Mechanical Ventilator must have successfully completed

More information

PROBLEM SET 8. SOLUTIONS April 15, 2004

PROBLEM SET 8. SOLUTIONS April 15, 2004 Harvard-MIT Division of Health Sciences and Technology HST.542J: Quantitative Physiology: Organ Transport Systems Instructors: Roger Mark and Jose Venegas MASSACHUSETTS INSTITUTE OF TECHNOLOGY Departments

More information

PART SEVEN THE HISTORY AND APPLICATION OF HIGH FREQUENCY OSCILLATORY VENTILATION (HFOV)

PART SEVEN THE HISTORY AND APPLICATION OF HIGH FREQUENCY OSCILLATORY VENTILATION (HFOV) PART SEVEN THE HISTORY AND APPLICATION OF HIGH FREQUENCY OSCILLATORY VENTILATION (HFOV) Reciprocating pistons with an eccentric travel speed, moving to and fro within a cylinder (with a common inlet/outlet),

More information

Keeping anaesthetic pollution to minimum in theatre environment

Keeping anaesthetic pollution to minimum in theatre environment Vet Times The website for the veterinary profession https://www.vettimes.co.uk Keeping anaesthetic pollution to minimum in theatre environment Author : Gianluca Bini Categories : General, Vets Date : September

More information

Respiratory Physiology Gaseous Exchange

Respiratory Physiology Gaseous Exchange Respiratory Physiology Gaseous Exchange Session Objectives. What you will cover Basic anatomy of the lung including airways Breathing movements Lung volumes and capacities Compliance and Resistance in

More information

Mechanical Ventilation

Mechanical Ventilation Mechanical Ventilation Chapter 4 Mechanical Ventilation Equipment When providing mechanical ventilation for pediatric casualties, it is important to select the appropriately sized bag-valve mask or endotracheal

More information

The physiological functions of respiration and circulation. Mechanics. exercise 7. Respiratory Volumes. Objectives

The physiological functions of respiration and circulation. Mechanics. exercise 7. Respiratory Volumes. Objectives exercise 7 Respiratory System Mechanics Objectives 1. To explain how the respiratory and circulatory systems work together to enable gas exchange among the lungs, blood, and body tissues 2. To define respiration,

More information

Solutions. Digital pressure management solutions. Copyright 2013 GaleMed Corporation

Solutions. Digital pressure management solutions. Copyright 2013 GaleMed Corporation Solutions Digital pressure management solutions Content 1. About GiO Solutions 2. What is Gio 3. Clinical Applications 4. Product Range 5. Features 6. Specifications 7. Accessories 8. Order Information

More information

Day-to-day management of Tracheostomies & Laryngectomies

Day-to-day management of Tracheostomies & Laryngectomies Suctioning Suctioning the airway is an essential part of routine care of the tracheostomy and laryngectomy patient. Sputum is continually produced in health and our native airways deal with this load without

More information

http://www.priory.com/cmol/hfov.htm INTRODUCTION The vast majority of patients who are admitted to an Intensive Care Unit (ICU) will need artificial ventilation (Jones et al 1998). The usual means through

More information

Technical Data and Specifications

Technical Data and Specifications Technical Data and Specifications INTENDED USE Ventilator designed to provide Invasive and Non-invasive ventilation for the critical care management of adult, pediatric and neonate-infant (including premature)

More information

RESISTANCE TO AIRFLOW IN ANAESTHETIC BREATHING SYSTEMS

RESISTANCE TO AIRFLOW IN ANAESTHETIC BREATHING SYSTEMS Br. J. Anaesth. (1989), 62, 456-461 RESISTANCE TO AIRFLOW IN ANAESTHETIC BREATHING SYSTEMS D. G. MARTIN, K. L. KONG AND G. T. R. LEWIS The resistance to airflow of the components of breathing systems have

More information

TOPO. operating Manual

TOPO. operating Manual TOPO Dual Mode Ventilator operating Manual Rev. B1a September, 2009 Table of Contents General Information... 3 Contact Information... 3 Disclaimer... 3 Warranty... 3 Repairs... 3 Customer Satisfaction...

More information

TESTCHEST RESPIRATORY FLIGHT SIMULATOR SIMULATION CENTER MAINZ

TESTCHEST RESPIRATORY FLIGHT SIMULATOR SIMULATION CENTER MAINZ TESTCHEST RESPIRATORY FLIGHT SIMULATOR SIMULATION CENTER MAINZ RESPIRATORY FLIGHT SIMULATOR TestChest the innovation of lung simulation provides a breakthrough in respiratory training. 2 Organis is the

More information

INTELLiVENT -ASV. The world s first Ventilation Autopilot

INTELLiVENT -ASV. The world s first Ventilation Autopilot INTELLiVENT -ASV The world s first Ventilation Autopilot Intelligent Ventilation since 1983 We live for ventilation technology We live for ventilation technology that helps caregivers improve the lives

More information

Medical Ventilators. Ryan Sanford Daniel Tajik

Medical Ventilators. Ryan Sanford Daniel Tajik Medical Ventilators Ryan Sanford Daniel Tajik Medical Ventilators Assist patients that have trouble breathing on their own. 2 Oxygen Intake The Respiratory System has the responsibility of providing the

More information

Principles of mechanical ventilation. Anton van Kaam, MD, PhD Emma Children s Hospital AMC Amsterdam, The Netherlands

Principles of mechanical ventilation. Anton van Kaam, MD, PhD Emma Children s Hospital AMC Amsterdam, The Netherlands Principles of mechanical ventilation Anton van Kaam, MD, PhD Emma Children s Hospital AMC Amsterdam, The Netherlands Disclosure Research grant Chiesi Pharmaceuticals Research grant CareFusion GA: 27 weeks,

More information

The Manley ventilator

The Manley ventilator Anaesthesia, 1995, Volume 50, pages 6471 CLASSIC PAPER The Manley ventilator Roger Manley was senior house officer (SHO) in the anaesthetic department of Westminster Hospital when he made and tested his

More information

Measuring Lung Capacity

Measuring Lung Capacity Name Class Date Chapter 37 Circulatory and Respiratory Systems Measuring Lung Capacity Introduction The amount of air that you move in and out of your lungs depends on how quickly you are breathing. The

More information

Endotracheal Tubes - VentiSeal (High Volume Low Pressure)

Endotracheal Tubes - VentiSeal (High Volume Low Pressure) Endotracheal Tubes - VentiSeal (High Volume Low Pressure) VentiSeal reduces the risk of vocal cord injury and tracheal necrosis through an optimal volume cuff with ideal pressure distribution and effective

More information

Medical Instruments in the Developing World

Medical Instruments in the Developing World 2.2 Ventilators 2.2.1 Clinical Use and Principles of Operation Many patients in an intensive care and the operating room require the mechanical ventilation of their lungs. All thoracic surgery patients,

More information

The comparison of Work Of Breathing methodologies on a patient model

The comparison of Work Of Breathing methodologies on a patient model The comparison of Work Of Breathing methodologies on a patient model E. Koumoundouros 1, J. Santamaria 2 and J. Patterson 3. 1 Department of Electrical and Electronic Engineering, The University of Melbourne,

More information

Using Common Ventilator Graphics to Provide Optimal Ventilation

Using Common Ventilator Graphics to Provide Optimal Ventilation Using Common Ventilator Graphics to Provide Optimal Ventilation David Vines, MHS, RRT, FAARC Associate Professor Chair / Program Director Department of Respiratory Care RUSH UNIVERSITY MEDICAL CENTER Disclosure

More information

Respiratory Physiology 2

Respiratory Physiology 2 Respiratory Physiology 2 Session Objectives. What you will cover Gaseous Exchange Control of Breathing Rate Your objectives are State the function of support structures and epithelia of the bronchial tree

More information

Indications for Mechanical Ventilation. Mechanical Ventilation. Indications for Mechanical Ventilation. Modes. Modes: Volume cycled

Indications for Mechanical Ventilation. Mechanical Ventilation. Indications for Mechanical Ventilation. Modes. Modes: Volume cycled Mechanical Ventilation Eric A. Libré, MD VCU School of Medicine Inova Fairfax Hospital and VHC Indications for Mechanical Ventilation Inadequate ventilatory effort Rising pco2 with resp acidosis (7.25)

More information

ONLINE DATA SUPPLEMENT. First 24 hours: All patients with ARDS criteria were ventilated during 24 hours with low V T (6-8 ml/kg

ONLINE DATA SUPPLEMENT. First 24 hours: All patients with ARDS criteria were ventilated during 24 hours with low V T (6-8 ml/kg APPENDIX 1 Appendix 1. Complete respiratory protocol. First 24 hours: All patients with ARDS criteria were ventilated during 24 hours with low V T (6-8 ml/kg predicted body weight (PBW)) (NEJM 2000; 342

More information

Collin County Community College. Lung Physiology

Collin County Community College. Lung Physiology Collin County Community College BIOL. 2402 Anatomy & Physiology WEEK 9 Respiratory System 1 Lung Physiology Factors affecting Ventillation 1. Airway resistance Flow = Δ P / R Most resistance is encountered

More information

Mechanics of proportional-assist ventilation

Mechanics of proportional-assist ventilation CHAPTER 10 Mechanics of proportional-assist ventilation A.C. Lua & K.C. Shi School of Mechanical and Aerospace Engineering, Nanyang Technological University, 50, Nanyang Avenue, Singapore 639798. Abstract

More information

Inspire rpap REVOLUTION FROM THE FIRST BREATH

Inspire rpap REVOLUTION FROM THE FIRST BREATH Inspire rpap TM REVOLUTION FROM THE FIRST BREATH The Inspire rpap The Inspire rpap is a revolutionary, non-invasive system for the initial stabilisation and resuscitation of infants. TM Its innovative,

More information

EXCESSIVE WORK OF BREATHING DURING INTERMITTENT MANDATORY VENTILATION

EXCESSIVE WORK OF BREATHING DURING INTERMITTENT MANDATORY VENTILATION Br. J. Anaesth. (1986), 58, 1048-1054 EXCESSIVE WORK OF BREATHING DURING INTERMITTENT MANDATORY VENTILATION J. S. MECKLENBURGH, I. P. LATTO, T. A. A. AL-OBAIDI, E. A. SWAI AND W. W. MAPLESON Ventilators

More information

Advanced nasal CPAP system www.hamilton-medical.com/arabella Noninvasive, low-cost, effective and safe Neonatal nasal CPAP therapy Nasal CPAP therapy aims to support neonates, especially pre-term and

More information

Lung recruitment maneuvers

Lung recruitment maneuvers White Paper Lung recruitment maneuvers Assessment of lung recruitability and performance of recruitment maneuvers using the P/V Tool Pro Munir A Karjaghli RRT, Clinical Application Specialist, Hamilton

More information

AUTOVENT 4000 VENTILATOR

AUTOVENT 4000 VENTILATOR OVERVIEW AUTOVENT 4000 Only properly trained and approved Escambia County Bureau of Public Safety Paramedics are to use the AutoVent 4000 ventilator manufactured by LSP to transport patients already on

More information

INTRODUCTION TO BI-VENT (APRV) INTRODUCTION TO BI-VENT (APRV) PROGRAM OBJECTIVES

INTRODUCTION TO BI-VENT (APRV) INTRODUCTION TO BI-VENT (APRV) PROGRAM OBJECTIVES INTRODUCTION TO BI-VENT (APRV) INTRODUCTION TO BI-VENT (APRV) PROGRAM OBJECTIVES PROVIDE THE DEFINITION FOR BI-VENT EXPLAIN THE BENEFITS OF BI-VENT EXPLAIN SET PARAMETERS IDENTIFY RECRUITMENT IN APRV USING

More information

ASV. Adaptive Support Ventilation

ASV. Adaptive Support Ventilation ASV Adaptive Support Ventilation Intelligent Ventilation since 1983 We live for ventilation technology We live for ventilation technology that helps caregivers improve the lives of their critically ill

More information

SLE4000. Infant Ventilator with touch-screen operation. When the smallest thing matters

SLE4000. Infant Ventilator with touch-screen operation. When the smallest thing matters SLE4000 Infant Ventilator with touch-screen operation When the smallest thing matters SLE4000 - The Total Solution for Conventional Infant Ventilation SLE is a world leader in the design and manufacture

More information

VENTILATION STRATEGIES FOR THE CRITICALLY UNWELL

VENTILATION STRATEGIES FOR THE CRITICALLY UNWELL VENTILATION STRATEGIES FOR THE CRITICALLY UNWELL Dr Nick Taylor Visiting Emergency Specialist Teaching Hospital Karapitiya Senior Specialist and Director ED Training Clinical Lecturer, Australian National

More information

Peter Kremeier, Christian Woll. 2nd. Understanding and comparing modes of ventilation. The Kronberg List of Ventilation Modes

Peter Kremeier, Christian Woll. 2nd. Understanding and comparing modes of ventilation. The Kronberg List of Ventilation Modes Extended Edition Revised and 2nd Peter Kremeier, Christian Woll Understanding and comparing modes of ventilation The Kronberg List of Ventilation Modes Contents Preface... 3 Comparison Table - Parameters...6

More information

Clinical Skills. Administering Oxygen

Clinical Skills. Administering Oxygen Clinical Skills Administering Oxygen Updated July 2017 Clare Cann Original 2012 Carole Loveridge, Lecturer in Women`s Health Aims and Objectives Aims and Objectives The aim of this module is to facilitate

More information

THE MECHANICS of RESPIRATION. Introduction

THE MECHANICS of RESPIRATION. Introduction THE MECHANICS of RESPIRATION Dr. James Duffin Departments of Physiology and Anaesthesia General Learning Objectives: 1. How is air moved into and out of the lungs? 2. What mechanical factors affect the

More information

April KHALED MOUSA BACHA. Physiology #2. Dr. Nayef AL-Gharaibeh. Pulmonary volumes & capacities

April KHALED MOUSA BACHA. Physiology #2. Dr. Nayef AL-Gharaibeh. Pulmonary volumes & capacities 25 th April Physiology #2 Pulmonary volumes & capacities Dr. Nayef AL-Gharaibeh KHALED MOUSA BACHA We will start this lecture by explaining an important concept from the previous one: Intrapleural pressure

More information

Operation and Maintenance of the EPV200 Portable Ventilator

Operation and Maintenance of the EPV200 Portable Ventilator Operation and Maintenance of the EPV200 Portable Ventilator 1 Applications of the EPV200 The EPV200 Portable Ventilator is a gas powered electronically controlled mechanical ventilator, designed to provide

More information

Initiation and Management of Airway Pressure Release Ventilation (APRV)

Initiation and Management of Airway Pressure Release Ventilation (APRV) Initiation and Management of Airway Pressure Release Ventilation (APRV) Eric Kriner RRT Pulmonary Critical Care Clinical Specialist Pulmonary Services Department Medstar Washington Hospital Center Disclosures

More information

Unit 15 Manual Resuscitators

Unit 15 Manual Resuscitators 15-1 Unit 15 Manual Resuscitators GOAL On completion of this unit, the student should comprehend the proper operation of self-inflating resuscitation bags, flow-inflating resuscitation bags and gas-powered

More information

APRV: Moving beyond ARDSnet

APRV: Moving beyond ARDSnet APRV: Moving beyond ARDSnet Matthew Lissauer, MD Associate Professor of Surgery Medical Director, Surgical Critical Care Rutgers, The State University of New Jersey What is APRV? APRV is different from

More information

Section 2.9 Decannulation

Section 2.9 Decannulation Bite- sized training from the GTC Section 2.9 Decannulation This is one of a series of bite- sized chunks of educational material developed by the Global Tracheostomy Collaborative. The GTC has developed

More information

Automatic Transport Ventilator

Automatic Transport Ventilator Automatic Transport Ventilator David M. Landsberg, MD, FACP, FCCP, EMT-P Luke J. Gasowski, RRT, NPS, ACCS, CCP-C, FP-C Christopher J. Fullagar, MD, FACEP, EMT-P Stan Goettel, MS, EMT-P Author credits /

More information

HAMILTON-C2 HAMILTON-C2. The universal ventilation solution

HAMILTON-C2 HAMILTON-C2. The universal ventilation solution HAMILTON-C2 HAMILTON-C2 The universal ventilation solution The universal ventilation solution HAMILTON-C2 - The compact ventilation solution The HAMILTON-C2 mechanical ventilator is a universal ventilation

More information

Update to RS-232 commands. Changing patient from NIV to INVASIVE Vent Type. SNDF command

Update to RS-232 commands. Changing patient from NIV to INVASIVE Vent Type. SNDF command Changing patient from NIV to INVASIVE Vent Type Table 9 shows automatic settings changes that occur when changing the same patient from NIV to INVASIVE Vent Type. Table 9: Automatic settings changes NIV

More information

Principles of Mechanical Ventilation: A Graphics-Based Approach

Principles of Mechanical Ventilation: A Graphics-Based Approach Principles of Mechanical Ventilation: A Graphics-Based Approach Steven M. Donn, MD, FAAP Professor of Pediatrics Neonatal-Perinatal Medicine C.S. Mott Children s Hospital University of Michigan Health

More information

Quick Response Pneupac VR1 Emergency Ventilator AIRWAYMANAGEMENT

Quick Response Pneupac VR1 Emergency Ventilator AIRWAYMANAGEMENT Quick Response Pneupac VR1 Emergency Ventilator AIRWAYMANAGEMENT Responding to your needs One of the best selling resuscitators... Smiths Medical is bringing you the next generation of emergency ventilators.

More information

VT PLUS HF performance verification of Bunnell Life-Pulse HFJV (High Frequency Jet Ventilator)

VT PLUS HF performance verification of Bunnell Life-Pulse HFJV (High Frequency Jet Ventilator) VT PLUS HF performance verification of Bunnell Life-Pulse HFJV (High Frequency Jet Ventilator) VT PLUS HF provides a special mode for evaluating the performance of high frequency ventilators while connected

More information

Panther 5 Acute Care Ventilator

Panther 5 Acute Care Ventilator 1 HIGHLIGHTS High performance and advanced features target the ICU environment Utilizes an internal blower with a specially designed flow control valve removing the need for using compressed air without

More information

3100A Competency Exam

3100A Competency Exam NAME DATE (Circle the appropriate answer) 3100A Competency Exam 1. Of the following, which best describes the mechanics of ventilation used by the 3100A? a. Active inspiration with passive exhalation b.

More information

Respiratory Pulmonary Ventilation

Respiratory Pulmonary Ventilation Respiratory Pulmonary Ventilation Pulmonary Ventilation Pulmonary ventilation is the act of breathing and the first step in the respiratory process. Pulmonary ventilation brings in air with a new supply

More information

VENTILATORS PURPOSE OBJECTIVES

VENTILATORS PURPOSE OBJECTIVES VENTILATORS PURPOSE To familiarize and acquaint the transfer Paramedic with the skills and knowledge necessary to adequately maintain a ventilator in the interfacility transfer environment. COGNITIVE OBJECTIVES

More information

Ventrain starts a new era in ventilation

Ventrain starts a new era in ventilation Ventrain starts a new era in ventilation What is new? full ventilation through a 2 mm catheter active expiration up to 7 litres minute volume connection for capnometry Ventrain is the first ventilation

More information

Accumulation of EEV Barotrauma Affect hemodynamic Hypoxemia Hypercapnia Increase WOB Unable to trigger MV

Accumulation of EEV Barotrauma Affect hemodynamic Hypoxemia Hypercapnia Increase WOB Unable to trigger MV Complicated cases during mechanical ventilation Pongdhep Theerawit M.D. Pulmonary and Critical Care Division Ramathibodi Hospital Case I Presentation Male COPD 50 YO, respiratory failure, on mechanical

More information

Mechanical Ventilation. Mechanical Ventilation is a Drug!!! is a drug. MV: Indications for use. MV as a Drug: Outline. MV: Indications for use

Mechanical Ventilation. Mechanical Ventilation is a Drug!!! is a drug. MV: Indications for use. MV as a Drug: Outline. MV: Indications for use Mechanical Ventilation is a Drug!!! Mechanical Ventilation is a drug I am an employee of Philips Healthcare Hospital Respiratory Care Group and they help me pay for my kids education Jim Laging, RRT, RCP

More information

Selecting and Connecting Breathing Systems

Selecting and Connecting Breathing Systems Selecting and Connecting Breathing Year Group: BVSc3 + Document number: CSL_A03 Equipment for this station: Equipment list: Pen Paper Calculator T-piece (in CSL a strip of white tape is around this system)

More information

Hospital and Transport for Controlled Breathing

Hospital and Transport for Controlled Breathing Hospital and Transport for led Breathing When transporting a critically ill patient you need a ventilator that can go anywhere in any situation. Smiths Medical Pneupac small portable gas powered ventilators

More information

SLE5000 Infant Ventilator with HFO

SLE5000 Infant Ventilator with HFO SLE5000 Infant Ventilator with HFO When the smallest thing matters SLE5000 - The Total Solution for Infant Ventilation Shown on optional roll stand. Ventilator may be mounted either way round on stand.

More information

Inspiration 7i Ventilator

Inspiration 7i Ventilator Inspiration 7i Ventilator Clinician Focused Products is not just a tag line, rather a belief system. Low cost of ownership High performance PSOL and active exhalation Comprehensive graphics and trending

More information

Endotracheal Suctioning: In Line ETT

Endotracheal Suctioning: In Line ETT Approved by: Endotracheal Suctioning: In Line ETT Gail Cameron Senior Director Operations, Maternal, Neonatal & Child Health Programs Dr. Paul Byrne Medical Director, Neonatology Neonatal Policy & Procedures

More information

Chapter 9 Airway Respirations Metabolism Oxygen Requirements Respiratory Anatomy Respiratory Anatomy Respiratory Anatomy Diaphragm

Chapter 9 Airway Respirations Metabolism Oxygen Requirements Respiratory Anatomy Respiratory Anatomy Respiratory Anatomy Diaphragm 1 Chapter 9 Airway 2 Respirations Every cell of the body requires to survive Oxygen must come in and carbon must go out 3 Metabolism Metabolism--Process where the body s cells convert food to Adequate

More information

Stephanie. Pediatrics. The ventilation system for neonatology

Stephanie. Pediatrics. The ventilation system for neonatology Pediatrics Stephanie The ventilation system for neonatology + Conventional ventilator + High-frequency oscillator + Proportional Assist Ventilation (PAV) + Integrated patient gas humidifier + Minimum volume

More information

mechanical ventilation Arjun Srinivasan

mechanical ventilation Arjun Srinivasan Respiratory mechanics in mechanical ventilation Arjun Srinivasan Introduction Mechanics during ventilation PV curves Application in health & disease Difficulties & pitfalls The future. Monitoring Mechanics

More information

Intelligent Ventilation solution from ICU to MRI

Intelligent Ventilation solution from ICU to MRI Intelligent Ventilation solution from ICU to MRI www.hamilton-medical.com/mr1 Supporting you in saving lives! HAMILTON-MR1 ventilators feature a compact, powerful design that increase the availability

More information

Presentation Overview. Monitoring Strategies for the Mechanically Ventilated Patient. Early Monitoring Strategies. Early Attempts To Monitor WOB

Presentation Overview. Monitoring Strategies for the Mechanically Ventilated Patient. Early Monitoring Strategies. Early Attempts To Monitor WOB Monitoring Strategies for the Mechanically entilated Patient Presentation Overview A look back into the future What works and what may work What s all the hype about the WOB? Are ventilator graphics really

More information

Your patients, our primary focus. Siesta i Whispa anesthesia machine

Your patients, our primary focus. Siesta i Whispa anesthesia machine Your patients, our primary focus Siesta i Whispa anesthesia machine Functional, and flexible At Dameca we are not just committed to meeting the healthcare needs of today. We also think about tomorrow.

More information

Blease900 Ventilator PATIENT CENTERED VENTILATION

Blease900 Ventilator PATIENT CENTERED VENTILATION Blease900 Ventilator PATIENT CENTERED VENTILATIN PATIENT CENTERED VENTILATIN From the first Blease/Manley ventilator sixty years ago, Spacelabs Healthcare has built a heritage of innovative and intuitive

More information

Please, after a carefully reading of given references, answer the following questions:

Please, after a carefully reading of given references, answer the following questions: Report Session 1: Introduction of Respiratory System Student 1 name: Student 2 name: Questionnaire Please, after a carefully reading of given references, answer the following questions: (1) How gas exchange

More information

Initiating In-Line Ventilator Speaking Valve Protocols: Why & How? Kristin C. Dolan; CCC-SLP Kindred Hospital Melbourne, FL

Initiating In-Line Ventilator Speaking Valve Protocols: Why & How? Kristin C. Dolan; CCC-SLP Kindred Hospital Melbourne, FL Initiating In-Line Ventilator Speaking Valve Protocols: Why & How? Kristin C. Dolan; CCC-SLP Kindred Hospital Melbourne, FL Disclosure of Relevant Financial Relationships Passy-Muir, Inc consultant Expected

More information

PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: INSTRUCTOR:

PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: INSTRUCTOR: PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: 1. **Identify and name the filters on the 7200ae. 2. **Explain how each filter is sterilized. 3. **Trace the gas flow through the ventilator

More information

HAMILTON-C3 HAMILTON-C3. The compact high-end ventilator

HAMILTON-C3 HAMILTON-C3. The compact high-end ventilator HAMILTON-C3 HAMILTON-C3 The compact high-end ventilator The compact high-end ventilator HAMILTON-C3 - The all-rounder for ICUs The HAMILTON-C3 ventilator is a modular high-end ventilation solution for

More information

NOTE: If not used, provider must document reason(s) for deferring mechanical ventilation in a patient with an advanced airway

NOTE: If not used, provider must document reason(s) for deferring mechanical ventilation in a patient with an advanced airway APPENDIX: TITLE: Mechanical Ventilator Use REVISED: November 1, 2017 I. Introduction: Mechanical Ventilation is the use of an automated device to deliver positive pressure ventilation to a patient. Proper

More information