ZX2 Operating Manual Issue L D1252 June 2015 THE RECOMMENDED SAFETY GOGGLES MUST BE WORN BY BOTH THE PATIENT & OPERATOR AT ALL TIMES DURING TREATMENT

Size: px
Start display at page:

Download "ZX2 Operating Manual Issue L D1252 June 2015 THE RECOMMENDED SAFETY GOGGLES MUST BE WORN BY BOTH THE PATIENT & OPERATOR AT ALL TIMES DURING TREATMENT"

Transcription

1 ZX2 High Power Laser WARNING CLASS 3B & 4 LASER EQUIPMENT IS HAZARDOUS TO THE EYE THE RECOMMENDED SAFETY GOGGLES MUST BE WORN BY BOTH THE PATIENT & OPERATOR AT ALL TIMES DURING TREATMENT DO NOT TREAT AROUND THE EYE, CONSULT PAGES 21 & 22, CONTRA- INDICATIONS & PRECAUTIONS BEFORE COMMENCING ANY TREATMENT READ THIS MANUAL BEFORE USE PRESCRIPTION USE ONLY Operating Manual Firmware Rev 1.6 onwards (c) LZR

2 CONTENTS Page Introduction 3 Safety Procedure 4 Quick Start Guide 5 General Care & Use 8 Disposal 8 Advanced Operation Probe Selection 9 Joules Display 9 Beep Settings 10 Temp Settings 10 Tissue Depth Units 11 Contrast 11 Brightness 12 Joules Time Priority 12 Usage Data & Temperature 12 Treatment Time Setting 14 Joules Setting 15 Frequency Setting 15 Tissue Depth 16 Probe Output Test 16 Fault Condition 17 High Temperature 17 Remote Interlock Connector 18 FDA Declaration 19 Maintenance & Servicing 20 Warranty & Repairs 22 Contra-indications 23 Precautions 24 Specifications System General 25 ZX2 Laser Unit 26 Warning Labels 27 Laser Safety Goggles mW Deep Probe W Deep Probe W Multi Probe 31 SP4 Probe 32 Typical Skin Temperature Heating Times 33 (c) LZR

3 INTRODUCTION The ZX2 Laser System is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain or muscle spasm; the temporary increase in local blood circulation; and/or the temporary relaxation of muscle. A selection of probes are available, from single laser to large multi heads. This together with a variable power output ensures, all applications are catered for. Two completely independent probe outputs allow use by two operators at once and/or treatment using two probes at once. This is a professional device, built to withstand years of everyday use, with many advanced features, such as programmable tissue depth and dosage. Please read this operating manual thoroughly before use to ensure safety of both the operator and patient. EYE PROTECTION MUST BE WORN AT ALL TIMES BY BOTH THE OPERATOR AND PATIENT (c) LZR

4 SAFETY PROCEDURE READ THIS BEFORE SWITCHING ON CAUTION - Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous radiation exposure 1. THE PATIENT & OPERATOR MUST WEAR APPROVED SAFETY GOGGLES AT ALL TIMES DURING TREATMENT. 2. DO NOT LEAVE THE KEY IN THE EQUIPMENT 3. DO NOT TREAT THE EYE OR SURROUNDING AREA 4. DO NOT DIRECT LASER RADIATION INTO THE EYES OR FROM A REFLECTED SURFACE 5. THIS APPLIANCE MUST BE EARTHED. ENSURE THE ELECTRICAL SUPPLY HAS A SUITABLE EARTH CONNECTION. 6. DO NOT TREAT AROUND THE EYE, CONSULT PAGES 21 & 22, CONTRA-INDICATIONS & PRECAUTIONS BEFORE COMMENCING ANY TREATMENT 7. THIS DEVICE COMPLIES WITH THE RELEVANT EMC STANDARDS, HOWEVER IT IS POSSIBLE THAT OPERATING IT IN CLOSE PROMIETRY TO A PULSED SHORT WAVE MACHINE COULD DAMAGE THE LASER DIODES, THEREFORE THIS IS NOT RECOMMENDED. 8. AS DEFINED IN IEC , THIS PRODUCT IS CONSIDERED STATIONARY, THEREFORE ONCE INSTALLED, IT SHOULD NOT BE MOVED. 9. THIS DEVICE SHOULD NOT BE INSTALLED IN SUCH A WAY THAT REMOVAL OF THE MAINS PLUG IS OBSTRUCTED. WARNING TO AVOID THE RISK OF ELECTRIC SHOCK, THIS EQUIPMENT MUST ONLY BE CONNECTED TO A SUPPLY MAINS WITH PROTECTIVE EARTH WARNING NO MODIFICATION OF THIS EQUIPMENT IS ALLOWED LASER TREATMENT ROOM SIGNS IEC specifies a laser warning sign should be placed above the treatment area. However it should be noted that some local health authorities will also require a laser warning sign above the entrance door to the treatment room. (c) LZR

5 QUICK START GUIDE This laser therapy device is designed for localised heating of the tissue for therapeutic purposes, which has number of benefits to the patient, notably pain reduction. The infra red radiation penetrates the skin and heats both the surface and deeper tissue. Clinical research suggests therapeutic benefits begin when tissue temperature is heated to between 40 and 45 deg C. The time taken for this can vary considerably depending on ambient temperature, skin temperature at the start of the treatment and the depth of the target tissue. The generally considered useful dosages are 4 to 20 Joules at the target depth. See Appendix A for typical skin temperature heating data. PROBE TEMPERATURE It should be noted that when operating in very high ambient temperatures (25 to 30 deg C) its possible with prolonged use the probe temperature could exceed 48 deg C. Should the probe temperature become uncomfortable for either the patient or operator, either switch to another probe, if one is available, or allow the probe to cool before re-commencing treatment. SKIN COLOUR High power laser therapy devices can cause discomfort and even a minor burn under extreme cases on patients with very dark skin. Patients with dark skin should commence treatment on 30% power, medium skin 60% and light skin 90 to 100%. If the patient does not report any discomfort the power output can be incremented at 10% every 10 seconds, until the warming effect becomes uncomfortable. At this point back off the power 20% and complete the treatment at that power level. The ZX2 laser unit will run 2 probes at once. The LCD screen is split in two halves, one for each probe. Each probe will operate independently, although FUNCTION parameters can only be adjusted when both probes are switched off. Power output can be adjusted independently for each probe while the probe is on. This allows the power to be turned down during treatment if the patient complains that the treatment area is getting too warm. 1. Connect the ZX2 Laser Unit to a suitable mains supply using the mains cord provided. The internal power supply is multi voltage and so will automatically operate on all worldwide standard mains voltages. 2. Insert the key into the keyswitch and rotate to the right to switch on. The power up sequence takes around 20 seconds. (c) LZR

6 3. Connect a probe cable to probe socket 1 or 2. Connect the other end of the probe cable to the probe. The connectors have a key, so rotate them until the keyway is found then push and they should click and lock. To remove a connector, just pull the collar back then pull the connector out of the socket. 4. Press the "MENU" switch. Using the power UP/DOWN switches on the left side of the device, scroll the arrow to "Probe Type 1" or "Probe Type 2", as required. Press "ADJUST". Using the power UP/DOWN switches, scroll the display to select the probe type in use, then press "EXIT" to get back to the main screen. S E T U P > P r o b e T y p e 1 P r o b e T y p e 2 J o u l e s D i s p l a y 5. The display will initially show "STB" (Standby), then after 4 seconds shown "RDY" (Ready). The laser probe can now be switched on by pressing the switch on the probe. During treatment, the power setting of the probe is displayed on the top row. When the output is off, the probe type is displayed m W 6. 1 W M u l t 0 : J 2 : J 2 0 H z 9 0 % C W % O N! 1 ½ C m R D Y 0 C m 6. To stop the treatment, just press the switch on the probe again. 7. The power output can be adjusted both before and during treatment by pressing the appropriate POWER UP/DOWN switches. In the event of an emergency, the emergency stop switch can be pressed. This immediately stops both outputs and requires a power reset to continue. 8. Treatment time, joules, pulse frequency & tissue depth can only be adjusted when both probe outputs are off. To adjust any of these, press the "FUNCTION" switch. The cursor will flash on the parameter to be adjusted. Press the "FUNCTION" switch again to move to the required parameter. To adjust the parameter, press the "ADJUST" switch. The cursor flashes for approximately 4 seconds after the last switch press. (c) LZR

7 S e t T i m e 6. 1 W M u l t 0 : I 2 : I C W % C W % R D Y ½ C m R D Y 0 C m 9. For more details of the menu and programmable functions see "Advanced Operation" below. (c) LZR

8 General Care & Use DO NOT EXPOSE OR IMERESE THE UNIT OR PROBES INTO WATER The laser unit should be wiped clean with a cloth soaked in a 70% alcohol solution. DISCONNECT from the mains supply before cleaning. Cleaning of Probes WARNING Always switch off and remove the key from the laser unit before cleaning or viewing the lens The lens of each probe should be kept clear of impurities at all times so that the optical path of light remains unobstructed. The probes may be cleaned using a medi-swab or similar swab soaked 70% alcohol solution. Use of Probes on Broken Skin & Wounds Where probes are used on broken skin, in order to prevent spread of infection, the probes should not be applied direct to the area to be treated. Instead treat through a suitable transparent barrier, such as Smith & Nephew OpSite Transparent Dressing. This is an adhesive disposable sterilised film which can be applied to probes before treatment and then removed and disposed of afterwards. Disposal All LZR7 products contain small traces of Lead (Pb) and so should not be disposed of in landfill sites. If you have to scrap LZR7 products that you wish to dispose of and you are not sure how to do it in a manner that is safe for the environment, return the equipment to the address on page 20 or to your nearest LZR7 dealer, who will arrange correct disposal/recycling of the equipment. This is a legal requirement in the European Union (EU) under the Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. The Restriction of the use of certain Hazardous Substances (RoHS) Directive 2011/55/EU also becomes a legal requirement in the European Union for Medical Devices from 22 nd July (c) LZR

9 ADVANCED OPERATION Menu Functions Probe Selection Press "MENU" from the main screen and you will see the setup menu:- S E T U P > P r o b e T y p e 1 P r o b e T y p e 2 J o u l e s D i s p l a y Use the UP/DOWN POWER switches or the FUNCTION switch to scroll the arrow to the probe to be adjusted, then press ADJUST S E T U P P r o b e T y p e W M u l t i Once the correct selection has been made, press EXIT to return to the main screen, or press MENU to return to the setup screen. NOTE correct probe selection is essential for the joules to be calculated correctly. The device might also go into fault condition if the wrong probe type is selected as the feedback signal from the probe will be incorrect and this will be interpreted as a fault. Joules Display Scroll down to "Joules Display" and press ADJUST S E T U P J o u l e s D i s p l a y T o t a l J o u l e s (c) LZR

10 Using the UP/DOWN POWER switches select either "Total Joules" or "Irradiance J/Cm 2 ". Total Joules gives the total joules the probe will give for the treatment time selected as is denoted by a "J" on the main screen. Irradiance J/Cm 2 is denoted an "I" on the main screen. Note the depth setting is limited to 4.5Cm for Joules and 1Cm for J/Cm 2. Once the correct selection has been made, press EXIT to return to the main screen, or press MENU to return to the setup screen. Beep Settings Scrolling further down will show the 2nd setup menu S E T U P > B e e p S e t t i n g s T e m p S e t t i n g s D e p t h U n i t s Press the ADJUST switch for the Beep Settings menu B e e p S E T U P I n c r e a s e D e c r e a s e 1 0 S e c The beep settings adjust the time of the warning beep when the probe is on. Use the UP/DOWN POWER switches on the left to select the required time. The options are, off, 1 sec, 5 sec, 10 sec. Once the correct selection has been made, press EXIT to return to the main screen, or press MENU to return to the setup screen. Temp Settings Selecting "Temp Settings" from the 2nd menu screen allows the temperature measurement units to be changed S E T U P T e m p S e t t i n g s C e n t i g r a d e (c) LZR

11 Use the UP/DOWN POWER switches to select either Centigrade or Fahrenheit Tissue Depth Units Select "Depth Units" from the 2nd menu screen to bring up the Depth Units Menu S E T U P D e p t h U n i t s C e n t i m e t r e s Use the UP/DOWN POWER switches to select either Centimetres or Inches for the issue depth display. Adjusting the tissue depth to the required depth automatically adjusts either the Total Joules or J/Cm2 setting to compensate for attenuation of tissue. Once the correct selection has been made, press EXIT to return to the main screen, or press MENU to return to the setup screen. Contrast From the 2nd setup menu continue to scroll down to the 3rd setup menu S E T U P > C o n t r a s t B r i g h t n e s s J / T P r i o r i t y Press the ADJUST switch to select "Contrast" S E T U P C o n t r a s t I n c r e a s e D e c r e a s e The contrast of the LCD display can be adjusted by the UP/DOWN POWER switches. Once the correct selection has been made, press EXIT to return to the main screen, or press MENU to return to the setup screen. (c) LZR

12 Brightness Scroll down to "Brightness" and press the ADJUST switch S E T U P B r i g h t n e s s I n c r e a s e D e c r e a s e The brightness of the LCD display can be adjusted by the UP/DOWN POWER switches. Once the correct selection has been made, press EXIT to return to the main screen, or press MENU to return to the setup screen. Joules Time Priority Select J/T Priority from the 3rd setup menu to adjust the joules time priority S E T U P J / T P r i o r i t y J o u l e s Use the UP/DOWN POWER switches to select either Joules or Time priority. Time priority will keep the treatment time the same when the power or depth are adjusted, joules priority will keep the joules or J/Cm 2 the same when the power or depth are adjusted. Once the correct selection has been made, press EXIT to return to the main screen, or press MENU to return to the setup screen. Usage Data and Temperature Pressing the MENU switch twice from the main screen will display the usage screen U S A G E D A T A H O U R S h r 0 2 m i n T E M P 3 9 C 4 6 C M a x F A U L T C O D E N o n e (c) LZR

13 This screen shows the total number of treatment hours, the current internal temperature of the laser unit and the maximum internal temperature that has been recorded. Also displayed are any fault codes that have been recorded. These can be useful in identifying problems with the laser unit, probe cable and probes. If you are having any problems with the device make a note of the fault code number before contacting your LZR7 service engineer. FAULT CODES 1=IC2 Fault detected on power up 2=PHOTODIODE CIRCUIT FAULT 3=PROBE 1 CABLE FAULT 4=PROBE 2 CABLE FAULT 5=TEMP has exceeded 50 deg C 6=F1 present on power up IC2 FAULT 7=F2 present on power up IC2 FAULT 8=Output 1 out of limits (-20%) during treatment OUTPUT POWER FAULT 9=Output 2 out of limits (-20%) during treatment OUTPUT POWER FAULT 10=Output 1 out of limits (+20%) during treatment OUTPUT POWER FAULT 11=Output 2 out of limits (+20%) during treatment OUTPUT POWER FAULT 12=Loss of EEPROM Data - CONTRAST 13=Loss of EEPROM Data - BRIGHTNESS 14=Loss of EEPROM Data - BEEP CODE 15=Loss of EEPROM Data - USAGE MIN 16=Loss of EEPROM Data - USAGE MINx10 17=Loss of EEPROM Data - USAGE HOUR 18=Loss of EEPROM Data - USAGE HOURx10 19=Loss of EEPROM Data - USAGE HOURx100 20=Loss of EEPROM Data - USAGE HOURx =Loss of EEPROM Data - MAX TEMP 22=Loss of EEPROM Data - POWER SETTING 1 23=Loss of EEPROM Data - POWER SETTING 2 24=Loss of EEPROM Data - FREQ SETTING 1 25=Loss of EEPROM Data - FREQ SETTING 2 26=Loss of EEPROM Data - TC1 1MIn 27=Loss of EEPROM Data - TC1 10Sec 28=Loss of EEPROM Data - TC1 1Sec 29=Loss of EEPROM Data - TC2 1MIn 30=Loss of EEPROM Data - TC2 10Sec 31=Loss of EEPROM Data - TC2 1Sec 32=Loss of EEPROM Data - DISPLAY FLAG 33=Loss of EEPROM Data - FAULT CODE 34=Loss of EEPROM Data - DEPTH Cm/Inch (c) LZR

14 35=Loss of EEPROM Data - TEMP F/C 36=Loss of EEPROM Data - Joules J/I 37=Loss of EEPROM Data - Tissue Depth 1 38=Loss of EEPROM Data - Tissue Depth 2 39=Loss of EEPROM Data - CH1 Prog Setting 40=Loss of EEPROM Data - CH2 Prog Setting 41=Loss of EEPROM Data - T/J Priority 42=Loss of EEPROM Data - CH1 Joules setting 43=Loss of EEPROM Data - CH2 Joules setting 44=Loss of EEPROM Data - CH1 Probe Type 45=Loss of EEPROM Data - CH2 Probe Type 46=Interlock operated NOTE fault codes can often occur due to a spike on the mains supply or other electrical noise, this will cause a default value to be stored in the memory. All data is stored in 3 separate memory locations and in the event of any 2 locations not being equal, default values are loaded and the fault code is stored. This does occur from time to time and is not generally anything to worry about, but if loss of data is occurring regularly, consult your LZR7 Engineer. Press EXIT to return to the main screen, or press MENU to return to the setup screen. Treatment Time Setting In the main screen, press the FUNCTION switch. Two fully independent channels are provided, although FUNCTION settings can only be changed when both probes are off. To adjust the 2nd channel, just keep pressing the FUNCTION switch until the 2nd channel parameters are highlighted by the cursor. S e t T i m e 6. 1 W M u l t 0 : I 2 : I C W % C W % R D Y ½ C m R D Y 0 C m Press the ADJUST switch to scroll though preset treatment times. 5 sec, 10, sec, 15 sec, 20 sec, 30 sec, 45 sec, 1min, 1min 15 sec, 1min 30 sec and 2 min are available. For more precise treatment time adjustments, the UP/DOWN POWER switches of the relevant channel can be use to change the setting by 1 second increments so any treatment time from 1 second to 2 minutes is available. Adjustment of the treatment time setting automatically calculates the Joules display. (c) LZR

15 Joules Setting Press the FUNCTION switch to scroll to the Joules display S e t J / C m ² 6. 1 W M u l t 0 : I 2 : I C W % C W % R D Y ½ C m R D Y 0 C m The UP/DOWN POWER switches of the relevant channel can be adjusted to increment or decrement the Joules setting by 1. This will automatically recalculate the treatment time required. Note "I" denotes "Irradiance" in J/Cm 2. "J" denotes the total number of Joules. This shows the total dosage for the treatment time shown and will count down to zero along with the treatment time counter. Note the depth setting is limited to 4.5Cm for Joules and 1Cm for J/Cm 2. Due to the high attenuation and irregular beam dispersion at depth, working in J/Cm 2 is not recommended at depths greater than 1Cm, therefore working in Joules is more accurate at depth. Frequency Setting Press the FUNCTION switch to scroll to the frequency display S e t F r e q 6. 1 W M u l t 0 : J 2 : J 2 0 H z 9 0 % C W % R D Y ½ C m R D Y 0 C m Press the ADJUST switch to select the desired frequency. 2.5Hz, 5Hz, 10Hz, 20Hz, 50Hz and 100Hz are available. Note the frequency output is only available for 10 to 90% power settings. When 100% power is selected, CW is shown on the display regardless of the frequency setting. To explain this an understanding of how the power level is adjusted is necessary: To vary the power, the duty cycle is adjusted from 10 to 100%, therefore only the average power is reduced and the peak power remains the same regardless of the power setting. This is a better method of power reduction as it still gives good penetration and still allows the average power to be reduced to adjust localised heating. (c) LZR

16 Tissue Depth Press the FUNCTION switch to scroll to the tissue depth display S e t D e p t h 6. 1 W M u l t 0 : J 2 : J 2 0 H z 9 0 % C W % R D Y 1 ½ C m R D Y 0 C m Press the ADJUST switch to select the desired tissue depth. Tissue depth can be set from zero to 4.5 Cm in half centimetre increments (1Cm max when using J/Cm 2 ), or zero to 2 inches in 0.2 inch increments. Adjusting the tissue depth to the required depth automatically adjusts either the Total Joules or J/Cm 2 setting to compensate for attenuation of tissue. Note the depth setting is limited to 4.5Cm for Joules and 1Cm for J/Cm 2. Due to the high attenuation and irregular beam dispersion at depth, working in J/Cm 2 is not recommended at depths greater than 1Cm, therefore working in Joules is more accurate at depth. Probe Output Test To test the output of a probe, switch the probe on, then press the FUNCTION switch, and place over the "TEST" window. The display will show "TEST" and the measured power reading will be displayed m W 6. 1 W M u l t 0 : J 2 : J 2 0 H z 9 0 % C W % T E S T R D Y 0 C m For an accurate measurement the probe should be adjusted to find the position for the maximum reading. When measuring Multi probes, position the probe head approximately over the aperture of each laser, with care the output of each individual laser can be measured. The photodiode is calibrated to measure the Infra Red lasers. The visible laser in the 6.1W Multi probe can be measured, but not accurately, as this typically shows twice the power it should. Power measurement is best performed on 100% power. Note, the test function will only work when one probe is in use; it is disabled when the output of both probes is on. (c) LZR

17 Fault Condition Should a fault be detected the display will show FAULT m W 6. 1 W M u l t 0 : J 2 : J 2 0 H z 9 0 % C W % F A U L T F A U L T If this occurs, the power will need to be switched off, then back on again to reset the system. Fault conditions can occur for the following reasons:- 1. Operation of the EMERGENCY STOP SWITCH 2. Operating of the REMOTE INTERLOCK CONNECTOR 3. Removal of the probe cable during treatment 4. 20% over or under set power is detected by the internal circuitry If fault conditions persist, checking the fault code as detailed above. Fault codes 3,4 and 8-11 can all indicate a probe cable fault, so change this if you have a spare. If faults persist, contact your LZR7 service engineer. High Temperature The ZX2 unit is rated to run both probe outputs continuously at 50% power. It is safe to run at higher power settings for long periods as long as the probe temperature does not get too hot. The internal circuitry has a high temp cut out that will operate above 50 deg C as well as a software algorithm designed to limit skin and probe temperature to below 45 deg C. If this should occur the following screen will be displayed: W D e e p 6. 1 W M u l t 0 : I 2 : I C W % C W % T E M P - W A I T T E M P - W A I T The temp warning should go off in a minute or so once the temperature has reduced. If this high temp warning is shown repeatedly, check the vents on the lower part of enclosure and the fan opening on the rear of the unit are not obstructed. The fan should operate when a probe is on and at any time when the internal temperature exceeds 45 deg C. (c) LZR

18 Remote Interlock Connector This device features a Remote Interlock Connector that enables the door into the treatment room to be connected to the interlock jack on the rear panel. This will put unit into the Fault condition (and therefore stop the laser emission) in the event of the door being opened during treatment in order to comply with IEC Fit a magnetic contact switch to the entrance door of the treatment room that is closed circuit when the door is shut and open circuit when the door is open. Wire this switch into a standard (mono 2 pin) quarter inch jack plug and plug this into the rear panel "Interlock" jack socket. These components can be supplied and installed by LZR7 Inc. or one of its distributors if required. (c) LZR

19 FDA DECLARATION LZR7 Inc confirms that the ZX2 Laser unit and probes as specified in this operating manual meet and fully comply with the following Federal guidelines, 21 CFR , 21 CFR , 21 CFR This device does not cause radio interference with other equipment and complies with Federal guidelines on Radio Interference as defined in IEC , Cispr 11 Radiated Emissions, EN ESD & EN Radiated Immunity. This device complies with IEC /UL Edition 3 Medical Electrical Equipment, inc ISO14971, IEC :2007 Ed 3 Therapeutic Laser Equipment, IEC :2007 Safety of Laser Products & ISO :2009 Biocompatibility. The ZX2 Laser unit and probes, as detailed in this operating manual, have been cleared for medical use by the FDA as Infrared Heating Devices under 510(k) application K (Pending). Randy Rinkleib Cheif Executive Officer LZR7 Inc Date (c) LZR

20 MAINTANANCE & SERVICING All Medical Devices should be periodically checked for electrical safety and effectiveness i.e. correct output. Electrical safety tests in accordance with IEC should be performed every 12 months and the output of all probes verified on a suitable calibrated optical power meter. Contact your sales & service representative if you require your equipment to be serviced. No user maintenance to be carried out. Please refer all technical servicing to your supplier. However, your supplier will make available such diagrams, components lists and parts and other information which will assist the user's appropriately qualified technical staff to repair the parts of the equipment which are classified by your supplier as repairable. NOTE There are high voltages present inside the ZX2 Laser Unit. Do not attempt to dismantle the equipment without first disconnecting from the mains supply. Replacement of the Mains Cable/Cord The mains cable/cord supplied with this device should be checked at routine service intervals for damage, continuity and correct wiring. Should the cable/cord need replacing the correct orientation of the wiring MUST BE CHECKED. Incorrect wiring poses a risk of electric shock, especially if the earth (E) and live (L) are reversed. Shown below are the correct connections for the plug (on the laser device) and the socket (on the mains cable/cord). The cable plug required is an IEC C13. Cable length should be no more than 2.5m (8Ft) and have a minimum of 10A rating. If the plug connection on the other end is fused, this should be a 5A rating. (c) LZR

21 Notes for MEDICAL ELECTRONICS ENGINEERS When performing electrical safety tests on this unit using a Rigel safety tester or similar high voltage test device please note the following procedures in order to avoid damage to Laser Probes:- All Laser Diodes are easily susceptible to damage from static or high voltage sources. The ZX2 unit grounds both laser drive pins when the probe is off. Therefore it recommended that high voltage tests are only performed on probes when they are plugged into the unit & switched off. Calibration of the Laser Measuring System The ZX2 unit contains a calibrated laser power meter that has built in compensation for probe type, therefore once calibrated the power meter will give accurate readings. Should re-calibration be necessary proceed as follows:- Remove the right hand end plate from the laser unit, by removing the four outermost screws. The adjustment potentiometer (VR1) will now be visible on the upper front panel PCB. Adjust the potentiometer as required to obtain an accurate reading on the power meter. NOTE this adjustment should only be performed by qualified medical service engineers, since the unit will need to be tested for electrical safety once reassembled. (c) LZR

22 WARARANTY LZR7 Inc warrants that the ZX2 laser unit, probes and probe cables are free from manufacturing defect from the date of purchase and will guarantee and undertake to remedy free of charge by repair or replacement any defects in the equipment that occur within 1 year of delivery, which arise under proper use and normal conditions. This warranty does not cover damage caused by negligence or misuse, nor does it cover routine servicing costs. Note dismantling this equipment during the warranty period will invalidate the warranty. Warranty does not cover carriage charges to and from the factory, nor will LZR7 Inc be liable for any loss of business caused by failure of the equipment. LZR7 and its service agents, reserve the right to refuse to service or repair under warranty any equipment supplied without a completed HSG(93)26 form or any equipment that is seriously contaminated with bodily fluids. REPAIRS & SERVICING For routine servicing during the warranty period, consult your supplier, who will arrange for suitably qualified personnel to service this equipment at competitive rates. If you have purchased your equipment through an agent, initially refer any technical queries or servicing requirements to them. Before contacting your service engineer, make a note of the serial number of your laser unit (located on the rear panel) and your date of purchase. It is also helpful to make a note of any fault codes (see "Usage data & Temperature") and the firmware version installed on your device. The firmware version is displayed on power up L Z R 7 I n c Z X 2 H i g h P o w e r L a s e r F i r m w a r e V e r s i o n 1. 1 (c) LZR

23 CONTRA-INDICATIONS DIRECT IRRADIATION OF THE EYES Class 3B & 4 lasers are harmful to the retina. Safety goggles must be worn by both patient and operator at all times. Do not treat around the eye area, even if safety goggles are worn. PREGNANCY Laser is contra-indicated for use over the pregnant uterus. It may be used on the pregnant woman as an adjunct to the other modalities being used for the treatment of back pain or other complaints. CARCINOMA Do not use the laser over any known primary or secondary lesions, tumours or carcinoma. THYROID Laser should not be used over the thyroid gland. IMMUNE SUPPRESSANT DRUGS Laser Therapy is contra-indicated for patients on these drugs. TREATMENT OVER SYMPATHETIC GANGLIA, THE VAGUS NERVES & CARDIAC REGION IN PATIENTS WITH HEART DISEASE Laser Therapy may significantly alter neural function, and is therefore contraindicated over these regions in patients with heart disease. (c) LZR

24 PRECAUTIONS INDIRECT IRRADIATION OF THE EYES Reflected light, or even looking at the bright crescent around the probe where it is in full contact with the skin, can induce eye strain and headache. Therefore safety goggles should be worn by both patient and operator at all times, even when direct viewing of the laser aperture is unlikely. PHOTOSENSITIVITY REACTIONS Some patients are on drugs known to cause photosensitivity reactions. It is unlikely that a combination of laser and drug will trigger this response. We recommend that "at risk" patients or patients with a history of such reactions be "patch tested" for the minimum recommended treatment time, for that probe, with each probe expected to be used on that patient. ANTI COAGULANTS It is possible that probe pressure may cause slight bruising after treatment. N.S.A.I.D. AND STEROIDAL A.I. Patients using topical or systemic steroids or NSAID's (non steroidal antiinflammatory drugs) for pain or skin conditions may experience a mild "flare up" of their symptoms. If such a reaction occurs, laser therapy should be stopped. The reaction should be allowed to settle. Laser treatment can then be recommenced at minimum or half minimum treatment times - if necessary building up to an "effective dose" if the patient does not respond to the lowered treatment times. For patients with persisting treatment reactions, Laser therapy should be discontinued. EPILEPSY Take care when using laser therapy for these patients especially around the head and neck. Cover the area being treated with a towel draped over the probe. REACTIONS TO TREATMENT Some patients may suffer a slight increase in pain - patients should be warned of this prior to laser therapy treatment. In patients with persisting treatment reactions, laser treatment should be discontinued. (c) LZR

25 PINS, METAL PLATES, PLASTICS AND PACEMAKERS ARE NOT CONTRAINDICATED Laser may be safely used over metal implants, plastics and stitches and on patients fitted with a pacemaker. DARK SKIN High power laser therapy devices can cause discomfort and even a minor burn under extreme cases on patients with very dark skin. Patients with dark skin should commence treatment on 30% power, medium skin 60% and light skin 90 to 100%. If the patient does not report any discomfort the power output can be incremented at 10% every 10 seconds, until the warming effect becomes uncomfortable. At this point back off the power 20% and complete the treatment at that power level. SKIN SENSITIVITY 1. Do not use the laser probe if there are any cracks or marks on the lens. 2. Watch for any signs of skin irritations from the use of this device. Discontinue use if signs of skin irritations, redness or itching are observed. (c) LZR

26 SPECIFICATIONS System General Operating Temperature Certified and tested at 23 +/-2 deg C, Recommended limits +5 to +30 deg C Storage Temperature -25 to +70 deg C Relative Humidity Certified and tested at 60% +/-15 Recommended limits 10 to 90% Atmospheric Pressure Certified and tested at 860 to 1060hPA 645 to 795 mm HG Compliance Standards IEC Ed3, ISO14971, IEC :2007, IEC :2007, IEC , Cispr 11, EN , EN , ISO :2009, ISO14971: CFR , 21CFR ; 21CFR ZX2 Laser Unit Voltage Certified and tested at 115 to 230V 50-60Hz Recommended limits 100 to 250V Power Consumption 200VA max Current 2A max Power Supply Classification Class I Insulation Classification Type B Dimensions 175mm(L) x 220mm(W) x 120mm(H) 7(L) x 9(W) x 5(H) Inches Weight 3Kg 6.5lb Mode of Operation Continuous at 50% power, or below Protection IPX0 Display 80 Character Alphanumeric LCD Laser Class Class 3B & 4 Laser Product (probe dependant) Emission Timer 1sec to 2min fully programmable Timer Accuracy ±1second Programming parameters Treatment time Total Joules Joules/Cm 2 Tissue depth Frequency 2.5Hz, 5Hz, 10Hz, 20Hz, 50Hz, 100Hz, CW Duty (power) 10 to 100% in 10% steps (50-100% SP probes) Standby/Ready Delay 4 seconds Output measurement Silicon Photodiode Calibration Accuracy ±20%. (c) LZR

27 Warning Labels Laser warning label on front of device Ratings label on rear of device Serial number label on rear of device, which also indicates date of manufacture (DOM) (c) LZR

28 Laser Safety Goggle Specifications Lens type EN207 Classification ML3 DIR L3 DIR L3 ANSI Classification OD nm OD3+ WARNING: use of laser safety goggles, other than those supplied or meeting the above specification could result in hazardous eye exposure. Should the lenses of the laser safety goggles get excessively scratched, cracked or damaged, they should be discarded and replaced. (c) LZR

29 750mW Deep Probe (Applied Part) Type CW Single Point Total Maximum Power Output 750mW No. of Laser Diodes 1 Wavelength 808nm Aperture 15mm Laser Classification Class 4 NOHD* 90Cm, 3ft Safety goggle requirement OD3 808nm Treatment Time for 4 J/Cm seconds Total Energy delivery per minute 45 Joules, 25 J/Cm 2 Emitter Type Ga Al As Semiconductor Laser Beam Divergence 5º x 12º Spot Size 1.76Cm 2 Power Density (Irradiance) 0.425W/Cm 2 (4,250Wm -2 ) * Nominal Ocular Hazard Distance (c) LZR

30 1.5W Deep Probe (Applied Part) Type CW Single Point Total Maximum Power Output 1.5W No. of Laser Diodes 1 Wavelength 808nm Aperture 15mm Laser Classification Class 4 NOHD* 150Cm, 5ft Safety goggle requirement OD3 808nm Treatment Time for 4 J/Cm seconds Total Energy delivery per minute 90 Joules, 51 J/Cm 2 Emitter Type Ga Al As Semiconductor Laser Beam Divergence 5º x 12º Spot Size 1.76Cm 2 Power Density (Irradiance) 0.85W/Cm 2 (8,500Wm -2 ) * Nominal Ocular Hazard Distance (c) LZR

31 6.1W Multi Probe (Applied Part) Type CW Large area Multi Total Maximum Power Output 6.1W No. of Laser Diodes 5 Wavelength 658nm x1 & 808nm x 4 Aperture 25mm Laser Classification Class 4 NOHD* 150Cm, 5ft Safety goggle requirement OD3 808nm and 658nm Treatment Time for 4 J/Cm seconds Total Energy delivery per minute 366 Joules, 38.5 J/Cm 2 Spot Size 9.35Cm 2 Power Density (Irradiance) 0.642W/Cm 2 (6,420Wm -2 ) 658nm Emitter Emitter Type: Ga Al In P Semiconductor Laser Beam Divergence 9.5º x 17º Maximum Output Power 100mW 808nm Emitter Emitter Type: Ga Al As Semiconductor Laser Beam Divergence 8º x 30º Maximum Output Power 1.5W * Nominal Ocular Hazard Distance (c) LZR

32 SP4 Probe (Applied Part) Type Super Pulsed at 100KHz Total Maximum Power Output 100W peak, 200mW Average No. of Laser Diodes 4 Wavelength 904nm x 4 Aperture 25mm Laser Classification Class 3R NOHD* 30Cm, 1ft Safety goggle requirement OD3 808nm Treatment Time for 4 J/Cm seconds Total Energy delivery per minute 12 Joules, 10.5 J/Cm 2 Emitter Type Ga As Semiconductor Laser Pulsed Beam Divergence 9º x 38º Spot Size 0.28Cm 2 Power Density (Irradiance) 0.176W/Cm 2 (1,760Wm -2 ) * Nominal Ocular Hazard Distance (c) LZR

33 Typical skin temperature heating times (c) LZR

OptonPro. High-power laser up to 7000 mw. Laser therapy natural healing with the power of light

OptonPro. High-power laser up to 7000 mw. Laser therapy natural healing with the power of light GB High-power laser up to 7000 mw Laser therapy natural healing with the power of light Application of high-power light for targeted penetration of high amounts of energy deep into the tissue. Initiates

More information

77-I0PN230 Rev. A. Model 1030 User Manual

77-I0PN230 Rev. A. Model 1030 User Manual 77I0PN230 Rev. A TABLE OF CONTENTS FUNCTION...2 TREATMENT...3 INDICATIONS FOR USE...4 CONTRAINDICATIONS...4 CLEANING...4 MAINTENANCE...4 CLASSIFICATION & SYMBOLS...4 DISINFECTION / STERILIZATION...4 SAFETY

More information

Digital Melting Point Apparatus

Digital Melting Point Apparatus Digital Melting Point Apparatus Heating Plateau Ramping Start/Stop Plateau set Ramp stop Hold User Guide Version 1.1 Heating Viewing tube Sample Chamber IEC power inlet socket Power on/off Temperature

More information

! Warning, refer to accompanying documents.

! Warning, refer to accompanying documents. About this Manual To the best of our knowledge and at the time written, the information contained in this document is technically correct and the procedures accurate and adequate to operate this instrument

More information

POWERPRESS UNIT. User Manual. Gradient Pneumatic Sequential Compressor. Neomedic

POWERPRESS UNIT. User Manual. Gradient Pneumatic Sequential Compressor. Neomedic TIMER (MIN) POWERPRESS UNIT 40 70 40 80 90 10 60 100 MIN MAX PRESSURE (mmhg) User Manual Neomedic 9601 Owens mouth Ave. #8 Chatsworth, CA 91311 Toll Free: 866-990-1168 Tel: 818-998-1023 Fax: 818-998-0277

More information

Instruction Manual for Configura Cushionair Portable Pump

Instruction Manual for Configura Cushionair Portable Pump Instruction Manual for Configura Cushionair Portable Pump Fitted with battery powered pump, suitable for Configura Portable chairs V E R S I O N O N E M A Y 2 0 1 6 Contents Introduction 3 Set up of Cushionair

More information

SPECIFICATIONS PARTICLE SENSOR KS-18F Higashimotomachi, Kokubunji, Tokyo , Japan

SPECIFICATIONS PARTICLE SENSOR KS-18F Higashimotomachi, Kokubunji, Tokyo , Japan SPECIFICATIONS PARTICLE SENSOR KS-18F 3-20-41 Higashimotomachi, Kokubunji, Tokyo 185-8533, Japan No. 05083-5E 18-01 Printed in Japan Outline The KS-18F is a sensor which uses the light scattering method

More information

Air Plus User Manual

Air Plus User Manual Air Plus User Manual Alternating pressure mattress for prophylaxis and treatment of pressure sores Indications for use The Air Plus has been developed as a state of the art pressure relieving system for

More information

Anodyne Therapy Professional System

Anodyne Therapy Professional System Anodyne Therapy Professional System Model 480 Manufacturer: Anodyne Therapy, LLC Address: 14105 McCormick Drive Tampa, FL 33626 Tel: 1-800-521-6664 Fax: 1-800-835-4581 Website: www.anodynetherapy.com MR-03805

More information

CJMJ.DPNQBDU. Instruction for use. Neonatal phototherapy device. BILI-COMPACT WY 1816/ Year of manufacture D2017_00

CJMJ.DPNQBDU. Instruction for use. Neonatal phototherapy device. BILI-COMPACT WY 1816/ Year of manufacture D2017_00 Instruction for use Neonatal phototherapy device CJMJ.DPNQBDU D2017_00 GmbH > BILI-COMPACT WY 1816/ Year of manufacture Herrenhöhe 4 D-51515 Kürten/Germany 02207-9698-0 02207-4750 E-mail: info@weyermed.com

More information

A-TONE. Operating Manual Version 2.0

A-TONE. Operating Manual Version 2.0 Part No: OM-ATN-AMT-01 Engineering AMT Engineering Co., Ltd. 434-6, Sangdaewon 1-dong Jungwon-gu, #811 Seongnam-si, Gyeonggi-do, Korea 462-716 Office: +82-31-737-8092 Fax: +82-31-737-8069 A-TONE (I2PL

More information

SPECIFICATIONS PARTICLE SENSOR KS-19F Higashimotomachi, Kokubunji, Tokyo , Japan

SPECIFICATIONS PARTICLE SENSOR KS-19F Higashimotomachi, Kokubunji, Tokyo , Japan SPECIFICATIONS PARTICLE SENSOR KS-19F 3-20-41 Higashimotomachi, Kokubunji, Tokyo 185-8533, Japan No. 13012-3E 17-06 Printed in Japan Outline The KS-19F is a sensor which uses the light scattering method

More information

F-330 Instruction manual. Magic Glove Bio Skin Smoother High Frequency Ultrasound Galvanic

F-330 Instruction manual. Magic Glove Bio Skin Smoother High Frequency Ultrasound Galvanic F-330 Instruction manual Magic Glove Bio Skin Smoother High Frequency Ultrasound Galvanic Warning Never under any circumstances attempt to open or inspect the internal components or accessories of the

More information

SC505 Sof Care Inflator. Operating Instructions and Service Manual

SC505 Sof Care Inflator. Operating Instructions and Service Manual SC505 Sof Care Inflator Gaymar Industries, Inc. 10 Centre Drive Orchard Park, NY 14127 Toll Free +1 800.828.7341 + 1 716.662.2551 Fax +1 800.993.7890 Outside USA +1 716.662.8636 Outside USA Fax +1 716.662.0730

More information

Vevazz. Operation Manual. Vevazz, LLC Rancho California Road. Suite Temecula, CA Phone: (773) Fax: (888)

Vevazz. Operation Manual. Vevazz, LLC Rancho California Road. Suite Temecula, CA Phone: (773) Fax: (888) Vevazz Operation Manual Vevazz, LLC 30520 Rancho California Road Suite 107-46 Temecula, CA 92591 Phone: (773) 665-4005 Fax: (888) 234-2765 URL: www.vevazz.com Warranty & Service Information Proprietary

More information

User Manual PTW-DensiX. D / Di/Zi

User Manual PTW-DensiX. D / Di/Zi User Manual PTW-DensiX D148.131.0/3 2005-05 Di/Zi General Information General Information The product bears the CE-mark "CE-0124" in accordance with the Council Directive 93/42/EEC about Medical Devices

More information

Operating Instructions Part No

Operating Instructions Part No DIGITAL AUTOMATIC TYRE INFLATOR Operating Instructions Part No. 11.0578 Thank you for selecting this Jamec Pem Automatic Tyre Inflator. Please read this manual before carrying out any installation or service

More information

VERTICAL AIR COMPRESSORS

VERTICAL AIR COMPRESSORS VERTICAL AIR COMPRESSORS MODEL NO: VE11C150, VE15C150, VE18C150 PART NO: 2226005, 2226000, 2226015 OPERATION & MAINTENANCE INSTRUCTIONS LS0615 INTRODUCTION Thank you for purchasing this CLARKE Vertical

More information

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road PROPORTIONING VALVE Model 150 INSTRUCTION MANUAL March 2017 IMS Company 10373 Stafford Road Telephone: (440) 543-1615 Fax: (440) 543-1069 Email: sales@imscompany.com 1 Introduction IMS Company reserves

More information

OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE

OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE 1 These instructions generally describe the installation, operation, and maintenance of subject equipment. The manufacturer reserves the right

More information

Operating instructions Safety Rope Emergency Stop Switches ZB0052 / ZB0053 ZB0072 / ZB0073

Operating instructions Safety Rope Emergency Stop Switches ZB0052 / ZB0053 ZB0072 / ZB0073 Operating instructions Safety Rope Emergency Stop Switches UK ZB0052 / ZB0053 ZB0072 / ZB0073 7390878 / 02 03 / 2011 Contents 1 Safety instructions...3 2 Installation / set-up...4 2.1 Applications...4

More information

Digital Inflation Equipment

Digital Inflation Equipment SINGLE DISPLAY Digital Inflation Equipment 89MXA S P E C I F I C A T I O N S I N S T A L L A T I O N O P E R A T I O N S E R V I C E Please read this manual before carrying out any installation or service

More information

Rigel 601 CHECKBOX. Instruction Manual. 348A551 Issue 2.0. April Seaward Electronic Ltd. Issue 2.0

Rigel 601 CHECKBOX. Instruction Manual. 348A551 Issue 2.0. April Seaward Electronic Ltd. Issue 2.0 Rigel 601 CHECKBOX Instruction Manual 348A551 Issue 2.0 April 2006 2006 Seaward Electronic Ltd. Issue 2.0 Limited Warranty & Limitation of Liability Rigel Medical guarantees this product for a period of

More information

Instructions for use. medisun 400 / 700. UV-phototherapy-device. UVB-311nm UVA UVA / UVB 3,3% Lamp configuration:

Instructions for use. medisun 400 / 700. UV-phototherapy-device. UVB-311nm UVA UVA / UVB 3,3% Lamp configuration: Instructions for use UV-phototherapy-device medisun 400 / 700 Lamp configuration: UVB-311nm UVA UVA / UVB 3,3% Manufacturer: Schulze & Böhm GmbH Hermülheimer Str. 10 D-50321 Brühl, Germany Tel : +49-2232-57998-0

More information

Instruction Manual BA90 Precision Barometer

Instruction Manual BA90 Precision Barometer Instruction Manual BA90 Precision Barometer halstrup-walcher GmbH Stegener Straße 10 D-79199 Kirchzarten Germany Phone: +49 (0) 76 61/39 63 0 Fax: +49 (0) 76 61/39 63 99 E-Mail: info@halstrup-walcher.com

More information

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021-PR Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021-PR For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4

More information

VERTICAL AIR COMPRESSORS

VERTICAL AIR COMPRESSORS VERTICAL AIR COMPRESSORS MODEL NO: VE15C150, VE18C150, VE25C150 PART NO: 2226010, 2226020, 2226025 OPERATION & MAINTENANCE INSTRUCTIONS LS0715 INTRODUCTION Thank you for purchasing this CLARKE Vertical

More information

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument OxyScan Graphic Operating Instructions UMS Micro-oxygen sensor 501 Microprocessor instrument Introduction Thank you for choosing the UMS Micro Oxygen Sensor 501 - a highly advanced product! Please read

More information

CONSOLE-320 ENGLISH. 230A: CONSOLE-320 with cable data output Item 230B: CONSOLE-320 with cable + wireless radio data output

CONSOLE-320 ENGLISH. 230A: CONSOLE-320 with cable data output Item 230B: CONSOLE-320 with cable + wireless radio data output CONSOLE-320 Item 230A: CONSOLE-320 with cable data output Item 230B: CONSOLE-320 with cable + wireless radio data output Table of contents 1. INTRODUCTION...2 1.1 Power supply...2 1.2 Connections...2 1.3

More information

User manual. MNPG 102 Rev. 04 Edition 04/02/16. Pressotherapy model. Power-Q1000 Premium

User manual. MNPG 102 Rev. 04 Edition 04/02/16. Pressotherapy model. Power-Q1000 Premium User manual MNPG 102 Rev. 04 Edition 04/02/16 Pressotherapy model Power-Q1000 Premium I.A.C.E.R. Srl Via S. Pertini 24/A 30030 Martellago (VE) ITALY Tel. 041.5401356 Fax 041.5402684 e-mail: iacer@iacer.it

More information

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4 Calibration....

More information

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4 Calibration....

More information

Norrsken Family Booklet

Norrsken Family Booklet Section 1: Introduction Low Energy Designs produce efficient and effective LED based lighting products for commercial, retail and industry purposes. Each product may contain specific details on its operation

More information

Datasheet: K-30 ASCII Sensor

Datasheet: K-30 ASCII Sensor Datasheet: K-30 ASCII Sensor The K30 ASCII sensor is a low cost, infrared and maintenance free transmitter module intended to be built into different host devices that require CO2 monitoring data. The

More information

DYNAMIC COMPRESSION THERAPY

DYNAMIC COMPRESSION THERAPY PULSE PRESS DYNAMIC COMPRESSION THERAPY MULTITEC 12 User Manual May 2011 0120 CONTENTS:- SAFETY INSTRUCTIONS ELECTROMAGNETIC INTERFERENCE DESCRIPTION OF CONTROLS MODE SELECTION OPERATING INSTRUCTIONS CARE

More information

Owner s Manual Humiport 10/20

Owner s Manual Humiport 10/20 4201 Lien Rd Madison, WI 53704 Owner s Manual Humiport 10/20 Installation, Operation & Service Instructions Read and Save These Instructions The Phoenix Humiport line of ThermoHygrometers offers the restoration

More information

PLEASE READ CAREFULLY BEFORE INSTALLING OR USING MEGA POOL SAVER MPS 1100

PLEASE READ CAREFULLY BEFORE INSTALLING OR USING MEGA POOL SAVER MPS 1100 MPS-1100 User Manual Mega Pool Saver Ltd PLEASE READ CAREFULLY BEFORE INSTALLING OR USING MEGA POOL SAVER MPS 1100 For further up to date instructions on how to install Mega Pool Saver MPS 1100, please

More information

Smart JVA Module WARNING. For use with AquaLink RS systems with rev R firmware or newer.

Smart JVA Module WARNING. For use with AquaLink RS systems with rev R firmware or newer. INstallation and OPERATION Manual Smart JVA Module For use with AquaLink RS systems with rev R firmware or newer. H0370200 Rev C FOR YOUR SAFETY - This product must be installed and serviced by a contractor

More information

RAM Operation Manual

RAM Operation Manual RAM 4021-1 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021-1 For Your Safety... 2 Description... 2 Setup Mode... 3 Lights/Alarms... 3 Operation... 4 Calibration...

More information

ECHO MANUAL WARNING. L B A ltim e te rs. ECHO is a trademark of LB Altimeters, Denmark

ECHO MANUAL WARNING. L B A ltim e te rs. ECHO is a trademark of LB Altimeters, Denmark ECHO MANUAL L B A ltim e te rs ECHO is a trademark of LB Altimeters, Denmark LB Altimeters operates a policy of continuous development Therefore, we reserve the right to make changes and improvements to

More information

HI 2314 HI 2315 HI 23151

HI 2314 HI 2315 HI 23151 Instruction Manual HI 2314 HI 2315 HI 23151 Multi-Range Conductivity Meters for Laboratories www.hannainst.com Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction

More information

INSTRUCTION MANUAL & PARTS BREAKDOWN

INSTRUCTION MANUAL & PARTS BREAKDOWN 24820 Automatic Tire Inflator 3 Modes in One 2347 Circuit Way, Brooksville, Florida, USA 34604 AME INTERNATIONAL Phone: (+1) - 1 352-799-1111 - Toll Phone: (+1) WWW.AMEINTL.NET 877-775-4AME (4263) Fax:

More information

NovelBright LED Illuminator

NovelBright LED Illuminator NovelBright LED Illuminator Instruction Manual version NS01_11 Instruction Manual This LED illuminator is suitable for research use only. It must be used by specialized personnel that know the health risks

More information

Temperature Controller CC24-7 ULTRA USER S MANUAL Legion Dr. Mason, MI USA October 2010 Ph. (517) Fax (517)

Temperature Controller CC24-7 ULTRA USER S MANUAL Legion Dr. Mason, MI USA October 2010 Ph. (517) Fax (517) Temperature Controller USER S MANUAL Aerotech, Inc. FORM: QM1387 4215 Legion Dr. Mason, MI 48854-1036 USA October 2010 Ph. (517) 676-7070 Fax (517) 676-7078 FOR CUSTOMER USE Enter the serial number located

More information

Instructions for Use

Instructions for Use Instructions for Use medisun HF-144 UV-phototherapy device for hand / foot treatment UVA broadband UVB-311 nm narrowband Timer (System micro-ii) Timer (System MRS) Joule-Controller (System MRS) Manufacturer:

More information

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments Pegas 4000 MF Gas Mixer InstructionManual Contents I Table of Contents Foreword Part I Introduction 1 2 1 System overview... 2 2 Specifications... 3 Part II Installation 4 1 Rear panel connections...

More information

RAM 4021 Operation Manual

RAM 4021 Operation Manual RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For your safety...3 Description...3 Set-up mode...4 Annunciator lights/alarms...4 Operation...5 Calibration...6

More information

Psyclone TM. Protimeter Thermo-Hygrometer. GE Sensing. Operating Manual. INS7800, Rev. A (September 2007)

Psyclone TM. Protimeter Thermo-Hygrometer. GE Sensing. Operating Manual. INS7800, Rev. A (September 2007) GE Sensing Psyclone TM Protimeter Thermo-Hygrometer Operating Manual INS7800, Rev. A (September 2007) Psyclone is a Protimeter product. Protimeter has joined other GE high-technology sensing businesses

More information

RPS900W Redundant Power Supply. Installation Guide.

RPS900W Redundant Power Supply. Installation Guide. RPS900W Redundant Power Supply Installation Guide www.edge-core.com Installation Guide RPS900W Redundant Power Supply Single DC Output Port with Dual Output Voltages RPS900W E10013-CS-R01 1500000081A

More information

User s Guide Temperature Sensor Converter TSC-599

User s Guide Temperature Sensor Converter TSC-599 User s Guide Temperature Sensor Converter TSC-599 ILX Lightwave Corporation 31950 Frontage Road Bozeman, MT, U.S.A. 59715 U.S. & Canada: 1-800-459-9459 International Inquiries: 406-556-2481 Fax 406-586-9405

More information

T i m i n g S y s t e m s. RACEAMERICA, Inc. P.O. Box 3469 Santa Clara, CA (408)

T i m i n g S y s t e m s. RACEAMERICA, Inc. P.O. Box 3469 Santa Clara, CA (408) RACEAMERICA T i m i n g S y s t e m s Demo Tree Controller Owner s Manual Models 3204D, 3204DW & 3204DX Rev D RACEAMERICA, Inc. P.O. Box 3469 Santa Clara, CA 95055-3469 (408) 988-6188 http://www.raceamerica.com

More information

GE Measurement & Control

GE Measurement & Control GE Measurement & Control Psyclone Protimeter Thermo-Hygrometer Operating Manual INS7800, Rev. C August 2013 Psyclone is a Protimeter product. Protimeter has joined other GE high-technology sensing businesses

More information

In Vivo Scientific, LLC INSTRUCTION MANUAL

In Vivo Scientific, LLC INSTRUCTION MANUAL CO 2 Controller In Vivo Scientific, LLC INSTRUCTION MANUAL CONTENTS CONTENTS...1 ABOUT THIS MANUAL...2 INTRODUCTION...2 Cautions and Warnings...2 Parts List...2 Unpacking...2 INSTRUMENT DESCRIPTION...3

More information

Air Sensor. FCS Ex. Manual. AQ Elteknik AB

Air Sensor. FCS Ex. Manual. AQ Elteknik AB Air Sensor FCS Ex Manual AQ Elteknik AB Air Sensor Model FCS Ex Manual version 2.0 Oct 2010 AQ Elteknik AB AQ Elteknik AB Air Sensor Manual 2 Table of contents 1. Manufacturer information... 4 Manufacture

More information

Digital Inflation Equipment 89XDA 89XDD 89XDB 89XDE.

Digital Inflation Equipment 89XDA 89XDD 89XDB 89XDE. Digital Inflation Equipment 89XDA 89XDD 89XDB 89XDE S P E C I F I C A T I O N S I N S T A L L A T I O N O P E R A T I O N S E R V I C E Please read this manual before carrying out any installation or service

More information

RAM 4021-DPX Operation Manual

RAM 4021-DPX Operation Manual RAM 4021-DPX Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS ABL 4021-DPX / RAM 4021-DPX For Your Safety... 3 Description... 3 Setup Mode... 4 Lights/Alarms... 4 Operation...

More information

AUTOMATIC TIRE INFLATOR # MW-60, MW-60-4WAY & MW-64HP

AUTOMATIC TIRE INFLATOR # MW-60, MW-60-4WAY & MW-64HP USER MANUEL AUTOMATIC TIRE INFLATOR # MW-60, MW-60-4WAY & MW-64HP TIRE EQUIPMENT MANUFACTURER 1.866.409.RACK WWW.MARTINSINDUSTRIES.COM info@martinsindustries.com PARTS Verify that the following components

More information

Digital Inflation Equipment 89XDZ 89XDX. Please read this manual before carrying out any installation or service procedures.

Digital Inflation Equipment 89XDZ 89XDX.  Please read this manual before carrying out any installation or service procedures. Digital Inflation Equipment 89XDX 89XDZ S P E C I F I C A T I O N S I N S T A L L A T I O N O P E R A T I O N S E R V I C E Please read this manual before carrying out any installation or service procedures.

More information

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model User's Manual Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved

More information

Pressure Automated Calibration Equipment

Pressure Automated Calibration Equipment GE Measurement & control Pressure Automated Calibration Equipment Safety Instructions and User Guide - K0447 PACE5000 PACE6000 K0447 Issue No. 9 1 10 1 PACE5000 1 2 3 4 5 PACE6000 2 6 7 8 3 4 5 6 7 8 9

More information

A4s Operation Manual

A4s Operation Manual A4s Operation Manual Safety Instruction Please read this manual carefully, also with related manual for the machinery before use the controller. For installing and operating the controller properly and

More information

USER MANUAL LTH-255A Hand Piece

USER MANUAL LTH-255A Hand Piece USER MANUAL LTH-255A Hand Piece 250 Corporate Blvd.,Suite B Newark, DE 19702, USA Tel 302-709-0408 Fax 302-709-0409 http://www.companiontherapylaser.com LBL000037C USER MANUAL LTH-255A Hand Piece DCN000049

More information

Scoreboard Operator s Instructions MPC Control

Scoreboard Operator s Instructions MPC Control Scoreboard Operator s Instructions MPC Control Some features on the keyboard overlay may not be included on the particular model being operated. Since 1934 Retain this manual in your permanent files 1/21/2011

More information

Installation & Maintenance Instructions TEC-11N TIMER. Analogue electronic timer 02/09

Installation & Maintenance Instructions TEC-11N TIMER. Analogue electronic timer 02/09 Installation & Maintenance Instructions TEC-11N TIMER Analogue electronic timer 02/09 GENERAL OPERATION The TEC-11N timer is produced using SMT (Surface Mounting Technology), ensuring improved performance

More information

OC Panel High Limit Aquastat Kit, Manual Reset p/n

OC Panel High Limit Aquastat Kit, Manual Reset p/n OC Panel High Limit Aquastat Kit, Manual Reset p/n 233202 Instruction Sheet APPLICATION The OC (Option Control) Panel High Limit Aquastat Kit provides electronic temperature sensing in a UL limit-rated

More information

Operating Instructions Part No

Operating Instructions Part No DIGITAL AUTOMATIC TYRE INFLATOR Operating Instructions Part No. 11.0545 Thank you for selecting this Jamec Pem Automatic Tyre Inflator. Please read this manual before carrying out any installation or service

More information

DEEPSEA. Power & Light. Micro SeaLaser. Operator s Manual Ruffin Road San Diego, CA USA T: (858) F: (858)

DEEPSEA. Power & Light. Micro SeaLaser. Operator s Manual Ruffin Road San Diego, CA USA T: (858) F: (858) Micro SeaLaser DEEPSEA Power & Light T: (858) 576-1261 F: (858) 576-0219 4033 Ruffin Road San Diego, CA 92123-1817 USA www.deepsea.com sales@deepsea.com Rev. 08/28/18 Micro SeaLaser Specification Overview

More information

SmarterLights. Guidelines for Use. HPLED Powerheads. By Light Therapy Options, L.L.C. The Dual Care TM. The Pure Amber. The TriPeak TM.

SmarterLights. Guidelines for Use. HPLED Powerheads. By Light Therapy Options, L.L.C. The Dual Care TM. The Pure Amber. The TriPeak TM. SmarterLights HPLED Powerheads By Light Therapy Options, L.L.C. The TriPeak TM The Peak The Pure 415 TM Blue The Pure Green The Pure Amber The Dual Care TM The Pure Red Guidelines for Use 1 Quick Start

More information

WARNING FOR YOUR SAFETY

WARNING FOR YOUR SAFETY ! OWNER S MANUAL Power Alert AquaLink TRI Controller H0389100 REV A WARNING FOR YOUR SAFETY - This product must be installed and serviced by a licensed electrician in accordance with AS/NZ 3000-2007 and

More information

Misaligned Folds Paper Feed Problems Double Feeds Won t Feed FLYER Won t Run iii

Misaligned Folds Paper Feed Problems Double Feeds Won t Feed FLYER Won t Run iii Operator s Manual Table of Contents Operator Safety... 1 Introduction... 2 Unpacking and Setup... 3 Unpacking... 3 Setup... 4 FLYER Overview... 5 FLYER Diagram... 5 Capabilities... 5 Control Panel... 6

More information

INSTALLATION GUIDE Gas Rangetops

INSTALLATION GUIDE Gas Rangetops INSTALLATION GUIDE Gas Rangetops Contents Wolf Gas Rangetops........................... 3 Safety Instructions............................ 4 Gas Rangetop Specifications.................... 5 Gas Rangetop

More information

MP15 Jockey Pump Controller

MP15 Jockey Pump Controller Setup and Operating Instructions MP15 Jockey Pump Controller This manual provides general information, installation, operation, maintenance, and system setup information for Metron Model MP15 Jockey Pump

More information

Type Operating Instructions. 2/2-way solenoid valve 2/2-Wege Magnetventil Electrovanne 2/2 voies.

Type Operating Instructions. 2/2-way solenoid valve 2/2-Wege Magnetventil Electrovanne 2/2 voies. 2/2-way solenoid valve 2/2-Wege Magnetventil Electrovanne 2/2 voies We reserve the right to make technical changes without notice. Technische Änderungen vorbehalten. Sous réserve de modifications techniques.

More information

Helium Level Measurement Unit

Helium Level Measurement Unit Helium Level Measurement Unit HLMU User Manual Version 002 The information in this manual may be altered without notice. BRUKER BIOSPIN accepts no responsibility for actions taken as a result of use of

More information

DUAL ACTION SANDER MODEL NO: CAT121 OPERATION & MAINTENANCE INSTRUCTIONS PART NO: LS0211

DUAL ACTION SANDER MODEL NO: CAT121 OPERATION & MAINTENANCE INSTRUCTIONS PART NO: LS0211 DUAL ACTION SANDER MODEL NO: CAT121 PART NO: 3110877 OPERATION & MAINTENANCE INSTRUCTIONS LS0211 INTRODUCTION Thank you for purchasing this CLARKE product. Before attempting to use this product, please

More information

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer July 2011 OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer WARNING! Before operating this product, read and understand this Operator s Manual. Become familiar with the potential hazards of this unit. Contact

More information

HUNTER Beeper Collar Manual

HUNTER Beeper Collar Manual HUNTER 2000- Beeper Collar Manual For a better performance: Use only Duracell batteries Each adjustment affects the duration of the batteries The battery contacts must be kept clean The thread of the Beeper

More information

Lucid System In-Service Training

Lucid System In-Service Training Lucid System In-Service Training DC-00172 Indications for Use The Lucid M1 System is a medical ultrasound system intended for use as an adjunct to the standard clinical practices for measuring and displaying

More information

Tightness controls TC 1 3 and TC 4

Tightness controls TC 1 3 and TC 4 Tightness controls TC 1 3 and TC 4 Test of both safety valves Short test period thanks to logical decision-making in the program sequence Adjustable test period which can be adapted to different systems

More information

User Manual Article number: I

User Manual Article number: I www.fluidoirrigation.nl User Manual Fluido Irrigation Fluid Warmer Article number: 651230-I INT/R347-EN/0-01/11 Foreword...3 Disclaimer...3 1 Temperature Management...4 2 Fluid Warming...4 3 Fluido Irrigation

More information

Automatic Valve Proving Control

Automatic Valve Proving Control ISO 9001 Automatic Valve Proving Control 7 696 LDU11 UL recognized FM approved Features Performs leak test of the gas shut-off valves before start-up and/or immediately after burner shut-down No inlet

More information

METEOR LSR36500 LIGHTING UNDERWATER POOL LIGHT. Operating Instructions

METEOR LSR36500 LIGHTING UNDERWATER POOL LIGHT. Operating Instructions METEOR LSR36500 UNDERWATER POOL LIGHT Operating Instructions LIGHTING CONTENTS Preface...3 Features...3 Precautions...3 Components and Connectors...4 Installation...5 Controls...7 Appearance...8 Technical

More information

MODEL 100 NITROGEN INFLATION CART

MODEL 100 NITROGEN INFLATION CART MODEL 100 NITROGEN INFLATION CART Installation & Operation Information Branick Industries, Inc. 4245 Main Avenue P.O. Box 1937 Fargo, North Dakota 58103 REV120106 P/N: 81-0113 TABLE OF CONTENTS SAFETY

More information

APA software instruction manual

APA software instruction manual 1. Starting the program In order to start the control software for the APA device press APAxx shortcut located on the desktop of the supplied computer. XX corresponds to the current software version. When

More information

ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS Dionex Corporation

ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS Dionex Corporation ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS 2000 Dionex Corporation Document No. 031277 Revision 03 April 2000 2000 Dionex Corporation All rights reserved worldwide. Printed in the United States of

More information

MODELS (2 foot Version) (4 foot Version)

MODELS (2 foot Version) (4 foot Version) INSTALLATION AND MAINTENANCE MANUAL DUROSITE LED LINEAR FIXTURE Document No: 9100-127-2571-99 Rev A April, 2016 MODELS (2 foot Version) (4 foot Version) ALL LAE ALL LBE These instructions contain important

More information

A4 Operation Manual. Fig.1-1 Controller Socket Diagram

A4 Operation Manual. Fig.1-1 Controller Socket Diagram A4 Operation Manual Safety Instruction Please read this manual carefully, also with related manual for the machinery before use the controller. For installing and operating the controller properly and

More information

Control units. Installation and operating instructions Series: DP-Control

Control units. Installation and operating instructions Series: DP-Control Control units Installation and operating instructions Series: DP-Control Table of Contents 1 Manual introduction 1.1 Preface... 4 1.2 Icons and symbols... 4 2 Identification, service and technical support

More information

Air Sensor. SAC Ex. Manual ATEX. AQ Elteknik AB

Air Sensor. SAC Ex. Manual ATEX. AQ Elteknik AB Air Sensor SAC Ex Manual AQ Elteknik AB ATEX Air Sensor SAC Ex ATEX Certified model Manual version 2.1 April 2014 AQ Elteknik AB 2 Table of contents 1. Manufacturer information... 4 Manufacture Declaration

More information

Table of Contents. Sensor Calibration and Troubleshooting CDS4000 CO 2. Introduction 1. Handling Information. Calibration 2.

Table of Contents. Sensor Calibration and Troubleshooting CDS4000 CO 2. Introduction 1. Handling Information. Calibration 2. FANs 216, 1628.3 Technical Bulletin CDS4000 Issue Date 0797 CDS4000 CO 2 Sensor Calibration and Troubleshooting Table of Contents Introduction 1 Handling Information 1 Calibration 2 Preparation 2 Cautions

More information

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel TANK MANAGER FOR TWO TANKS OPERATING MANUAL 10/31/11 C-More T6C L color touch panel 1 TABLE OF CONTENTS GENERAL...3 INSTALLATION...4 STONE TEST PROCEDURE...7 OPERATIONAL SUMMARY...7 AUTO CARBONATION...10

More information

Operating instructions Pressure sensor. PN422x / / 2013

Operating instructions Pressure sensor. PN422x / / 2013 Operating instructions Pressure sensor PN422x 80002606 / 00 11 / 2013 Contents 1 Preliminary note...3 1.1 ymbols used...3 2 afety instructions...3 3 Functions and features...3 4 Function...5 4.1 witching

More information

Verdi G-Series. High-Power Pumps for Ti:Sapphire Lasers and Amplifiers FEATURES

Verdi G-Series. High-Power Pumps for Ti:Sapphire Lasers and Amplifiers FEATURES Verdi G-Series High-Power Pumps for Ti:Sapphire Lasers and Amplifiers The Verdi G is a high performance CW laser providing up to 20W at 532 nm, ideal for pumping Ti:Sapphire lasers and amplifiers, for

More information

Dual Solenoid Gas Valve Installation

Dual Solenoid Gas Valve Installation Installation IMPORTANT: These instructions are intended as a guide for qualified personnel installing or servicing FLYNN Gas Products. Carefully follow all instructions in this bulletin and all instructions

More information

PRODUCT SUPPORT MANUAL YACHT BEAM 6M Searchlight Remote Control System with Joystick Remote Control Panel

PRODUCT SUPPORT MANUAL YACHT BEAM 6M Searchlight Remote Control System with Joystick Remote Control Panel PRODUCT SUPPORT MANUAL YACHT BEAM 6M Searchlight Remote Control System with Joystick Remote Control Panel The Yacht Group - 4545 Ponce de Leon Blvd. Coral Gables FL 33146 Tel. 305.667.5811 - Fax. 305.663.5551

More information

USER GUIDE. Fog Generator. ATMOSPHERE APS TM Series Haze Generator. MAX APS TM Series. No: , Revision A/a

USER GUIDE. Fog Generator. ATMOSPHERE APS TM Series Haze Generator. MAX APS TM Series. No: , Revision A/a ATMOSPHERE APS TM Series Haze Generator MAX APS TM Series Fog Generator USER GUIDE No: 120815, Revision A/a MADE IN CANADA FABRIQUÉ AU CANADA Les Générateurs de Brouillard Ltée / Fog Generators Ltd 10301

More information

JOLLY2. Installation user s manual. 6 different operating modes selectable. version 3.3. DATA TO BE FILLED OUT BY THE INSTALLER (Page 1)

JOLLY2. Installation user s manual. 6 different operating modes selectable. version 3.3. DATA TO BE FILLED OUT BY THE INSTALLER (Page 1) ENGLISH ENGLISH ENGLISH ENGLISH Installation user s manual Warning! electrical scheme modified JANUARY 2005 version 3.3 JOLLY2 DATA TO BE FILLED OUT BY THE INSTALLER (Page 1) 6 different operating modes

More information

Troubleshooting Guide: 640 Pediatric Exam Table with Midmark Scale

Troubleshooting Guide: 640 Pediatric Exam Table with Midmark Scale Troubleshooting Guide: 640 Pediatric Exam Table with Midmark Scale Contents Description Refer To: Scale Troubleshooting Chart Troubleshooting Error Codes Error Messages Adjustments / Repair Procedures

More information

For Models: 55A00-2 (HAI UPB 1500W Wall Switch Dimmer) 55A00-3 (HAI UPB 2400W Wall Switch Dimmer) 37A00-1 (HAI Auxiliary Switch)

For Models: 55A00-2 (HAI UPB 1500W Wall Switch Dimmer) 55A00-3 (HAI UPB 2400W Wall Switch Dimmer) 37A00-1 (HAI Auxiliary Switch) HAI UPB High Power Wall Switch Dimmer and HAI Auxiliary Switch Installation and Operating Instructions For Models: 55A00-2 (HAI UPB 1500W Wall Switch Dimmer) 55A00-3 (HAI UPB 2400W Wall Switch Dimmer)

More information

VENTI-O 2 plus. O 2 switching valve for: VENTImotion from serial number 4000 VENTIlogic from serial number 4000 VENTImotion 2

VENTI-O 2 plus. O 2 switching valve for: VENTImotion from serial number 4000 VENTIlogic from serial number 4000 VENTImotion 2 VENTI-O 2 plus O 2 switching valve for: VENTImotion from serial number 4000 VENTIlogic from serial number 4000 VENTImotion 2 Device description and operating instructions Overview VENTI-O 2 plus with therapy

More information