1 1 User guide for the NMR spectrometer AVA400Stud Dear User! You are using a modern NMR spectrometer, which has been installed in About spectra are measured here per year. Please respect the interests of other users and read the entire manual before using the spectrometer. Nevertheless, nobody is perfect and we are all learning from mistakes. Please report such cases immediately to the NMR department to avoid spectrometer downtime or even damage. Do not waste your time with trying to delete the traces of your activities. You will not be punished! On the contrary, your courage and responsibility can help tremendously in troubleshooting problems caused by your error. I. Safety precautions The superconducting magnet of NMR equipment produces strong magnetic and electromagnetic fields that can interfere with the function of cardiac pacemakers. Users of pacemakers and other implanted ferromagnetic medical devices are not allowed to enter the NMR room! Pregnant women and persons with non-ferromagnetic medical auxiliaries are required to contact the NMR department before using the spectrometer. The force that the magnetic field exerts on a ferromagnetic object (iron) may be very strong. Loose metal objects in the vicinity of the magnet can make sudden movements and be attracted to the magnet. This especially applies to very small ferromagnetic objects, which can be drawn into the magnet. They can damage the probe, are very difficult to remove, and will block the use of the spectrometer for a long time. The magnetic field can erase the magnetic media of USB-sticks, MP3-players, and mobile phones, disable credit and automated teller machine (ATM) cards, and damage analog watches. Please leave such items at the entrance of the room. The wall thickness of an NMR tube is about of 0.4 mm. Do not apply excessive force on when you close, open or insert the NMR tube into the spinner turbine. In rare cases, the superconducting magnet can spontaneously change from the superconducting to the normal state ("quench"). This transition causes an explosive vaporization of cryogenic substances from the magnet. In such case you are obligated to leave the room immediately and to inform the NMR department.
2 2 II. NMR sample preparation and injection 1. Samples for NMR analysis should be prepared in a defect-free 5 mm NMR tube. The tube cap must be on tightly to prevent evaporation of solvent while waiting in the queue. 2. The use of sealed or J. Young NMR tubes is prohibited. 3. Either 7 or 8 inch NMR tubes can be used and the tubes must have a quality according to the fieldstrength (e. e. 507-PP or higher). 4. Samples must be prepared using deuterated solvents. It is prohibited to use acids as the NMR solvent! 5. The optimal height of the liquid in the NMR tube is about 4.5 cm that corresponds to about 0.7 ml of the solvent. 6. The required concentration of an NMR sample depends on the experiment and the sensitivity of the observed nuclei. If possible, increase the concentration to improve the quality of the spectrum, especially for 13 C experiments. 7. The solution must be homogeneous. Precipitation and air bubbles impair the quality of NMR spectra and waste valuable measuring time. Solvent mixtures are avoided whenever possible. 8. NMR tubes must be clean outside. 9. Always wipe the NMR tube with Kleenex before inserting it into the spinner turbine. 10. Spinner turbines can be touched exclusively with clean, grease-free fingers at the upper edge (barcode). 11. The position of the NMR tube in the spinner turbine must be set using the red depth gauge. Slide carefully the NMR tube into the spinner turbine, plug the NMR tube along with the spinner turbine in the depth gauge and slide the tube gently to the bottom of the depth gauge. If you notice that the NMR-tube is loose inside the spinner then try another spinner or put your sample in another NMR-tube. Not correctly adjusted NMR tubes can lead to costly damage of the NMR spectrometer! 12. Place your NMR sample in the next free holder of the sample changer carousel. Please avoid the area of the light sensor to prevent the malfunction of automation. Use the holder number to register the sample in ICON-NMR. Caution: If the arm of the changer is moving or the carousel is rotating it is vorbidden to put a sample into the carousel or to remove a sample from it! Note the different holder numbers of the inner and outer chains of the carousel (see the schema on the sample changer). Experiments are started according to the required measuring time not the holder number (day experiments from 7 to 19 o clock and night experiments from 19 to 7 o clock).
3 3 13. Please always check the proper adjustment of your NMR tube by comparison with other NMR samples (visual control). 14. Each NMR sample must be accompanied with a perforated note of reasonable size providing the username, the sample composition, and the holder number. III. General rules Please respect the interests of your colleagues. You are not the only ones who are in a hurry. ICON-NMR shows the available measurement time for the daytime and nighttime experiments as well as the expected duration of experiments. On the weekdays (from Monday to Friday) the pending queue of nighttime experiments is terminated at 7:30 o clock. The uncompleted experiments have to be resubmitted for the next night! 1. Warning: In case ICON-NMR is failed to start three consecutive NMR samples (incorrect number holder, inhomogeneous solution, too little solvent), the automation will be stopped for safety reasons and further measurements will not be carried out. 2. Please collect the measured samples as soon as possible.
4 4 IV. Referencing of NMR spectra on AVA400Stud 1 H-Spectra: Use the solvent peak as the reference. 13 C-Spectra: Use the solvent peak as the reference. In D 2 O use the SR-value of -6, P- and 19 F-Spectra: The lock signal should be a reasonable size. The SR-value depends on the solvent: Solvent SR-Wert für 31 P SR-Wert für 19 F SR-Wert für 11 B Aceton-d 6-270,95-397,36-103,08 CDCl 3-110, , CD 2 Cl 2-99,48-37, ,21 CD 3 CN - 242,13-457,29-80,27 C 6 D 6-109, , ,92 D 2 O - 116, , ,79 DMSO-d 6-196,95-162,57-43,07 CD 3 OD* - 225,15* - 611,55* - 66,80* *the SR-value is valid only if the methyl resonance is used to lock the spectrometer!
5 5 V. Instructions for ICON-NMR 1. Use the "Change User" to select your work group. 2. Locate the next free place in the sample changer carousel. The colors indicate the following: Yellow: occupied, experiment done. Rot: occupied, experiment in progress. Pink: occupied, waiting to run. Green: free, experiment may be set up. Note: If all the 120 positions are occupied, do the following: a) Starting from the position 1 locate the first yellow position. X b) If the latter position in the carousel is still occupied by an NMR sample, proceed to the next position labeled in yellow in ICON-NMR. Thus, locate the first position which is labeled in yellow in ICON-NMR and is not occupied by an NMR sample in the carousel. c) Delete the corresponding record in ICON-NMR list with "DELETE". d) Place your sample in this position on the carousel. 3. To submit your experiment make a double click on "STATUS" and fill out the form that appears: Holder: The holder number. Type: Prescribed. Status: See item 2. Disk: Prescribed. Name: Generated automatically. Exp. No.: Generated automatically. Solvent: Define from the list. Experiment: Define from the list. 4. Check whether the holder number in ICON-NMR corresponds to the one in the carousel. Submit your experiment using the button "SUBMIT" at the middle of the left side of the screen.
6 VI. NMR experiments and the required concentrations of samples 6 Title of the automation: NS: Sweep width [ppm]: Minimal concentration [mol/l]: Approximate measurement time (without Shiming!): 1 H rau_sprotonlf_16 NS minutes rau_sprotonlf_64 NS minutes 13 C{ 1 H} rau_sc13cpd_256 NS minutes rau_sc13cpd_1k NS 1k minutes Dept135 rau_sdept135_64 NS minutes rau_sdept135_256 NS minutes Dept90 rau_sdept90_64 NS minutes rau_sdept90_256 NS minutes 31 P{ 1 H } rau_sp31cpd_64 NS minutes 31 P rau_sp31_64 NS minutes 19 F{ 1 H } rau_sf19cpd NS minute 19 F rau_sf19 NS minute HSQC rau_shsqcedetgpsw NS 2 Optimized in F2 in F1: 0 165ppm minutes HMBC rau_shmbcgpsw NS 2 Optimized in F2 in F1: 0 220ppm minutes Cosy rau_scosygpsw NS 1 Sweep optimized minutes Noesy rau_snoesygpsw NS 2, d8=1s Sweep optimized minutes Roesy rau_sroesysw NS 8, p15=0.2s Sweep optimized minutes Tocsy rau_stocsysw NS 8, d9=0.08s Sweep optimized minutes
7 7 VII. 2D-Spectra: Each 2D-experiment must be preceded by a 1 H experiment (rau_sprotonlf_16) based on the same set-up file. The later spectrum is required to define automatically the measuring range for the 2D spectrum. Upon the submission of a 2D-experiment ICON-NMR will generate automatically some parameters in the Setup Form. Please do not change these parameters to avoid a malfunction of automation. HSQC and HMBC: The measuring range of the direct dimension F2 (protons) is automatically optimized using the 1 H NMR spectrum (rau_sprotonlf_16). The measuring range of the indirect dimension F1 ( 13 C) is prescribed. COSY, NOESY, ROESY, and TOCSY: The measuring range of both the direct and indirect dimensions is automatically optimized using the 1 H NMR spectrum (rau_sprotonlf_16). Par. : Parameters. If necessary define d8 and d9 for NOESY- and TOCSY experiments. (Note: the values are in seconds!) Orig./Title: Enter your Status (Student, Bachelor, Master), your name, the substance, and the concentration: e.g.: Bachelor, Max Mustermann, TB45 (10mg/0.8ml CDCl3). Pri.: Display of Day- or Nighttime Experiment, otherwise disabled. Time: Display of the measuring time. (Note: add up to 5 minutes for Lock, Shim and the thermostating of the sample. User: Display of your username.
User guide for the NMR spectrometer AVA300
1 User guide for the NMR spectrometer AVA300 Dear User! You are using a highly frequented NMR spectrometer, which has been installed in 2002. About 26000 spectra are measured here per year. Please respect
More informationPINMRF Checkout Quiz - Bruker/TopSpin Version
PINMRF Checkout Quiz - Bruker/TopSpin Version Please carefully read every question and select the best answer(s). Some questions may have more than one correct answer. To receive full credit you must mark
More informationThings to know before operating an NMR spectrometer André Boltjes, University of Groningen DD NMR Lab v
Things to know before operating an NMR spectrometer André Boltjes, University of Groningen DD NMR Lab v1.0 2013 I. A tour of an NMR lab. Figure 1 shows a picture of our Avance DRX 500. Although each NMR
More informationVariable Temperature Operation
Variable Temperature Operation AVIII 400 MHz Spectrometer November 11, 2012 The CBC NMR Laboratory University of Delaware AVIII 400 User Guide: Variable Temperature Operation 1 INTRODUCTION... 5 1.1 SAFETY...
More informationNMR Service. How to Prepare Samples for NMR
NMR Service How to Prepare Samples for NMR In NMR, unlike other types of spectroscopy, the quality of the sample has a profound effect on the quality of the resulting spectrum. If you follow a few simple
More informationStep By Step Instructions for VT Experiments on the Bruker 300 MHz Spectrometer
Step By Step Instructions for VT Experiments on the Bruker 300 MHz Spectrometer (1) Use the drop down menu to create a new experiment/dataset for acquiring the room temperature 1D spectrum. The name of
More informationShimming with VnmrJ 3.1 Software for Peak NMR Performance
Shimming with VnmrJ 3.1 Software for Peak NMR Performance Application Note Author Ron Crouch Agilent Technologies 13000 Weston Parkway Cary, NC 27513 USA Shimming is the term used to describe the process
More informationDCIF Training Guide Variable Temperature Bruker B400 (last update )
DCIF Training Guide Variable Temperature Bruker B400 (last update 20100421) As with all of the other DCIF tutorials, the conventions used in this guide are as follows: Boldface type indicates commands
More informationConsole Acceptance Tests & Specifications. MERCURYplus NMR Spectrometer Systems Pub. No , Rev. B0902
Console Acceptance Tests & Specifications MERCURYplus NMR Spectrometer Systems Pub. No. 01-999186-00, Rev. B0902 Console Acceptance Tests & Specifications MERCURYplus NMR Spectrometer Systems Pub. No.
More informationTypically NMR Sample Configuration
Typically NMR Sample Configuration = B o Solvent column ~ 3 as long as detection region. Copyright 003-014 UWChemMRF All Rights Reserved. Last updated Sept 15, 014. The following slides are useful as an
More informationDeuterium more than just for locking: Part II. H Topshim. Donna Baldisseri Boston Massachusetts Pre-ENC Workshops 2014, March 22
Deuterium more than just for locking: Part II 2 H Topshim Donna Baldisseri Boston Massachusetts Pre-ENC Workshops 2014, March 22 Topshim Fully automated Reliable Easy to use Customizable Good, rapid results
More informationInstructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems
Corporation Issued: 5/09 P/N: 600104 Rev. 2 Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems The Gas Addition Accessory permits the
More informationBante810 Benchtop Dissolved Oxygen Meter Instruction Manual
Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante810 Benchtop Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante810 benchtop dissolved oxygen
More informationCopyright Velocomp LLP 2006
ibike Pro Cadence Handlebar Mount Owners Manual January 2007 1 Table of Contents Installing the ibike Pro cadence mount on your bicycle Installing the ibike Pro cadence mount on your time trial bike Installing
More informationPulsed-Field Gradient Shimming With VNMR
PFG Shimming With VNMR Page 65 Pulsed-Field Gradient Shimming With VNMR cg fry: created 99.01.27 updated 99.02.01 I. General Discussion Pulsed-field gradients allow analytically accurate, automated adjustments
More informationBante820 Portable Dissolved Oxygen Meter Instruction Manual
Bante820 Portable Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante820 Portable Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante820 portable dissolved oxygen
More informationSuperconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)
Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Sensitivity: 1x10 11 A m 2 Applied DC fields: 0 T to 5 T Applied AC fields: 0 G to 3 G (zero-to-peak), 0.01 Hz to 1000 Hz Temperatures
More informationRead this first. Zetasizer nano series Self installation and Quick start guide MRK825-02
! Read this first Zetasizer nano series Self installation and Quick start guide I N S T R U M E N T S MRK825-02 Zetasizer Nano series Self installation and Quick start guide MAN0383 Issue 1.1 July 2007
More informationUpdated: Variable Temperature
Updated: 2016-11-02 Variable Temperature Sample Preparation Make sure the probe is rated for the ultimate/final temperature Fully degas the sample, particularly if going to low temperature Be aware of
More informationFOURIER TRANSFORM INFRARED SPECTROSCOPY
Purpose and Scope: This document describes the procedures and policies for using the MSE Bruker FTIR Spectrometer. The scope of this document is to establish user procedures. Instrument maintenance and
More informationSOLUBILITY OF A SOLID IN WATER
1516L Experiment 1 SOLUBILITY OF A SOLID IN WATER Objectives In this experiment you will determine the solubility of potassium nitrate (KNO 3 ) in water at various temperatures. You will prepare a plot
More informationASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS Dionex Corporation
ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS 2000 Dionex Corporation Document No. 031277 Revision 03 April 2000 2000 Dionex Corporation All rights reserved worldwide. Printed in the United States of
More informationOPERATION. Estimated kerf width compensation. HPR260 Manual Gas Instruction Manual 4-9
Estimated kerf width compensation The widths in the chart below are for reference. Differences between installations and material composition may cause the specific user results to vary from those shown
More informationUsER manual for Watersens ph -REDOX
UsER manual for Watersens -REDOX Cl 8 1 2 6 3 3 7 7 4 4 4 4 Parts List 1 Redox Probe 1 x 2 PH Probe 1 x 5 Tube Weight 2 x 6 Connection Valve 1 x chlorine 3 Chlorine and Pumps 2 x 7 Dosing Valve 2 x 5 5
More informationSensor for Air Bubble Detection at Liquid Filled Tubes. SONOCHECK Type ABD06.xx. Operating Manual
Sensor for Air Bubble Detection at Liquid Filled Tubes SONOCHECK Type ABD06.xx Operating Manual Manufacturer: Model: Type: SONOTEC Ultraschallsensorik Halle GmbH Air Bubble Detector ABD06.xx SONOTEC Ultraschallsensorik
More informationAIR PRESSURE MASSAGER #Unwind
AIR PRESSURE MASSAGER #Unwind Marketed by: ADJUSTABLE LEG WRAPS 2 MODES THERAPEUTIC RELIEF EASY CLEANING 1year 2 11 Cleaning Make sure to cut off the power when you clean machine. 1. Please wipe the dirty
More informationBante821 Portable Dissolved Oxygen Meter Instruction Manual
Bante821 Portable Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante821 Portable Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante821 portable dissolved oxygen
More informationKAYAK ERGOMETER. Operating Instructions (Rev 3.0)
KAYAK ERGOMETER Operating Instructions (Rev 3.0) Weba Sport und Med.- Artikel GmbH Liesneckgasse 6/1 1210 Vienna Austria Tel.: ++43 1 2723550 Fax: ++43 1 27235504 office@webasport.at www.weba-sport.com
More informationHow to calibrate the formaldehyde sensor within a PPMonitor Wireless Unit:
How to calibrate the formaldehyde sensor within a PPMonitor Wireless Unit: Please read this document thoroughly before attempting to check or adjust the calibration of the formaldehyde sensor within the
More informationApproach. S3 Owner s Manual. May _0B Printed in Taiwan
Approach S3 Owner s Manual May 2012 190-01467-00_0B Printed in Taiwan All rights reserved. Under the copyright laws, this manual may not be copied, in whole or in part, without the written consent of Garmin.
More informationMcKnight Hockey Association
McKnight Hockey Association Electronic Evaluation Tool Manual 2013-2014 Table of Contents Introduction...3 Evaluation Tool...3 Login to OneClickIce...3 Evaluations...4 PROCESS...4 Evaluation Procedure...5
More informationThermo K-Alpha XPS Standard Operating Procedure
Thermo K-Alpha XPS Standard Operating Procedure Quick Guide Draft v.0.1 Procedure overview 1. Vent the loadlock 2. Secure your sample to the stage using clips, check the height of the final assembly. 3.
More informationStandard Operational Procedure for Capillary Electrophoresis
Page 1 of 14 Standard Operational Procedure for Capillary Electrophoresis *Please ask an experienced operator for training before you use the CE if you can. A demo is most probably more efficient than
More informationInfrared Ear and Head Thermometer Instruction Manual
DMT-316 Infrared Ear and Head Thermometer Instruction Manual Manual Ver.: 2.1 Issuing Date: 2017/02/03 2017. All rights reserved. Thank you for choosing the iprovèn Dual Mode Thermometer. The DMT-316 can
More informationSOLUBILITY OF A SOLID IN WATER
1516L Experiment 2 SOLUBILITY OF A SOLID IN WATER Objectives In this experiment you will determine the solubility of potassium nitrate (KNO 3 ) in water at various temperatures. You will prepare a plot
More informationOperator s Manual. The Bullet Blender Gold BB24-AU, BB5E-AU
Operator s Manual The Bullet Blender Gold BB24-AU, BB5E-AU Congratulations! Congratulations on your purchase of a Bullet Blender Gold by Next Advance, Inc., for lysing, disrupting, and homogenizing your
More informationNMR FACILITY REGISTRATION FORM
NMR FACILITY REGISTRATION FORM Have you registered on-line? Y / N (if NO, please visit this link http://www.nmr.unsw.edu.au/usercorner/registration.htm and follow step 1. *First name: Middle initial: *Surname:
More informationtesto Leakage detector for refrigerants Instruction manual
testo 316-3 Leakage detector for refrigerants Instruction manual 2 1 Contents 1 Contents 1 Contents... 3 2 Safety and the environment... 4 2.1. About this document... 4 2.2. Ensure safety... 4 2.3. Protecting
More informationAscend DNP. User Manual. Innovation with Integrity. Version NMR
300 154 Ascend DNP User Manual Version 01 Innovation with Integrity NMR Copyright by Bruker Corporation All rights reserved. No part of this publication may be reproduced, stored in a retrieval system,
More informationAngstrom E-Beam Instructions. PROCESS CHECKS: the tooling factors of each metal; Ti/Au layer for wire bonding pull test.
Angstrom E-Beam Instructions Tool Manager: Joe Palmer (jpalmer@princeton.edu) Office: 8-4706; Cell:609-751-1353 Backup: David Barth (dbarth@princeton.edu) Office: 8-4626; Cell: 610-405-8227 PROCESS CHECKS:
More informationSS32L Dissolved O 2 Probe
BIOPAC WWW.biopac.com Systems, Inc. Application Note PH-185 SS32L Dissolved O 2 Probe SS32L Dissolved O 2 Probe BIOPAC Software SS32L Specifications BSL PRO v. 3.6.6 BIOPAC Hardware SS32L Dissolved O 2
More informationHawaiienols A D, Highly Oxygenated p-terphenyls from an. Insect-Associated Fungus Paraconiothyrium hawaiiense
Hawaiienols A D, Highly Oxygenated p-terphenyls from an Insect-Associated Fungus Paraconiothyrium hawaiiense Fengxia Ren,, Shenxi Chen,, Yang Zhang, Shuaiming Zhu, Junhai Xiao, Xingzhong Liu, Ruibin Su,*,
More informationNEULOG OXYGEN LOGGER SENSOR GUIDE
NeuLog oxygen logger sensor NUL-205 The oxygen sensor can be used for any science experiment or activity where oxygen levels, dissolved or gaseous, are required. Some fields of study include Chemistry,
More informationTHE MF-400 SERIES. Operating and Service Manual. Series includes all variants of MF-400/401
THE MF-400 SERIES Operating and Service Manual Series includes all variants of MF-400/401 Issue A October 2013 1 TABLE OF CONTENTS 1. Description... 3 2. Installation... 3 3. Operation... 4 4. Special
More informationOPERATION OF LIQUID PARTICLE COUNTER. GP-B P0045 Rev -
1 W. W. Hansen Experimental Physics Laboratory STANFORD UNIVERSITY STANFORD, CALIFORNIA 94305-4085 Gravity Probe B Relativity Mission OPERATION OF LIQUID PARTICLE COUNTER GP-B P0045 Rev - May 23, 1997
More informationBOC: Living healthcare. Manual. LIV IQ BOC Integrated Valve with digital display portable delivery system for Medical Oxygen. BOC: Living healthcare
BOC: Living healthcare Manual LIV IQ BOC Integrated Valve with digital display portable delivery system for Medical Oxygen. BOC: Living healthcare 02 Manual LIV IQ Oxygen Manual LIV IQ Oxygen 03 Contents.
More informationMoveRoll Conveyor Operating and Maintenance Manual
MoveRoll Conveyor Operating and Maintenance Manual 1. Read this first! This manual contains information for protection of personnel in the roll handling area from possible injury and/or equipment damage.
More informationMicrofluidic Demonstration
Microfluidic Demonstration Co-flow of liquids across a microfluidic chip Create a microfluidic controller to flow liquids across a microfluidic chip. 1. Assemble fluidic circuits and attach to breadboard.
More informationCONSUMER MODEL INSTALLATION GUIDE
CONSUMER MODEL INSTALLATION GUIDE System requirements Windows System Requirements To use your TOMI and its software, your system should have: A Microsoft Windows compatible PC with a Pentium IV processor
More informationF-330 Instruction manual. Magic Glove Bio Skin Smoother High Frequency Ultrasound Galvanic
F-330 Instruction manual Magic Glove Bio Skin Smoother High Frequency Ultrasound Galvanic Warning Never under any circumstances attempt to open or inspect the internal components or accessories of the
More informationBEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION
BEST KNOWN METHODS Transpector XPR3 Gas Analysis System DESCRIPTION The Transpector XPR3 is a third-generation, quadrupole-based residual gas analyzer that operates at PVD process pressures and is the
More informationUser Manual 1 P a g e Rev. V1.6-EN 11/08/2014
User Manual 1 P a g e Rev. V1.6-EN 11/08/2014 Copyright Disclaimer Trademarks and patents Intended use Contact info 2011 Inflotrolix, Inc. This document may not be copied in whole or in part or otherwise
More informationMini-Pro CO2 Sensor User s Manual - ANALOG MODEL -
Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL - Table of Contents 1. Introduction 3 2. Instrument Setup 4 2.1 Instrument Checklist 4 2.2 Optional Accessories 4 2.3 Gas Concentration Ranges Available
More informationPurpose and Scope: Responsibilities: Warnings and precautions: Background Beers law
Purpose and Scope: This document describes the procedures and policies for using the MSE Bruker FTIR Spectrometer. The scope of this document is to establish user procedures. Instrument maintenance and
More informationBruker Sample Transport
Bruker BioSpin Bruker Sample Transport BST Installation and Technical Manual Version 002 think forward NMR Spectroscopy Copyright by Bruker BioSpin NMR GmbH All rights reserved. No part of this publication
More informationCryoProbe Prodigy. Bruker User Manual. Innovation with Integrity. Version NMR
CryoProbe Prodigy Bruker User Manual Version 001 Innovation with Integrity NMR Copyright by Bruker Corporation All rights reserved. No part of this publication may be reproduced, stored in a retrieval
More informationSchool of Chemistry SOP For Operation Of Glove Boxes
School SOP for Operation of Glove Boxes: The following SOP provides guidelines on how to adequately and safely operating a standard laboratory glove box and its associated devices. General Information:
More informationQuick Reference Guide
HematologyAnalyzer Quick Reference Guide For Veterinary Use Only Running a Blank Blanks must be run every 12 hours, after a Soak Cleaning has been performed, after a reagent pack change, and after the
More informationHI 2314 HI 2315 HI 23151
Instruction Manual HI 2314 HI 2315 HI 23151 Multi-Range Conductivity Meters for Laboratories www.hannainst.com Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction
More informationAscend. User Manual. Innovation with Integrity. Version NMR
400 54 Ascend User Manual Version 04 Innovation with Integrity NMR Copyright by Bruker Corporation All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted,
More informationBI-680 Online Dissolved Oxygen Controller Instruction Manual
BI-680 Online Dissolved Oxygen Controller Instruction Manual BANTE INSTRUMENTS CO., LTD BI-680 Online Dissolved Oxygen Controller 1 Introduction Thank you for selecting the BI-680 online dissolved oxygen
More information[USING THE VITROBOT April 9, Using the Vitrobot
Using the Vitrobot The Vitrobot Mk 3 (aka Mark III, Mark 3, etc.) can control both the local temperature and the local humidity during the process of plunge-freezing a grid for cryoem. It does require
More informationUser s Guide. CO2 Controller for miniature incubators
CO2 Control User s Guide CO2 Controller for miniature incubators Precise CO2 Control throughout the experiment Media ph control Compatible with any perfusion system Miniature incubators for any microscope
More informationGenerating Calibration Gas Standards
Technical Note 1001 Metronics Inc. Generating Calibration Gas Standards with Dynacal Permeation Devices Permeation devices provide an excellent method of producing known gas concentrations in the PPM and
More informationPH01 Perfusion Cannula Manual
PH01 Perfusion Cannula Manual Information in this document is subject to change without notice. No part of this document may be reproduced or transmitted without the express written permission of Multi
More informationInfrared Thermometer Instruction manual
Dual Mode Infrared Thermometer Instruction manual DMT-489 Manual Ver.: 1.0 Issuing Date: 2016/09/26 Thank you for choosing the iprovèn Dual Mode Digital Infrared Thermometer. With its unique technology,
More informationPowermax machine-side reference guide For mechanized applications with Powermax1000, Powermax1250 and Powermax1650
Powermax machine-side reference guide For mechanized applications with Powermax1000, Powermax1250 and Powermax1650 This Powermax machine-side reference guide is a supplement to your Operator Manual and
More informationOPERATION MANUAL FOR VP 710E-3HM SERIES ALLIGATOR TUMBLE STIRRERS *US Patent # 6,176,609
TECHNICAL NOTE 299 OPERATION MANUAL FOR VP 710E-3HM SERIES ALLIGATOR TUMBLE STIRRERS *US Patent # 6,176,609 WARNING!!!!! Be advised that the Magnetic Tumble Stirrer has very strong magnetic fields coming
More informationOxygen Sensor with Electrode
imagine explore learn Oxygen Sensor with Electrode Product Number: ENOXY-A222 Overview The Oxygen sensor is a maintenance-free galvanic oxygen electrode, capable of measuring Oxygen both in the air and
More informationAssembly Drawing: W-311B-A01, or as applicable Parts List: W-311B-A01-1, or as applicable Special Tools: , , &
REDQ Regulators Model 411B Barstock Design Powreactor Dome Regulator OPERATION AND MAINTENANCE Contents Scope..............................1 Installation..........................1 General Description....................1
More information7130 Lancer Rear Drive Magnetic Commercial Indoor Cycling Bike
7130 Lancer Rear Drive Magnetic Commercial Indoor Cycling Bike Owner s Manual Made in Taiwan INDEX IMPORTANT SAFETY INFORMATION... 1 EXPLODED DRAWING... 2 PARTS LIST... 3 ASSEMBLY INSTRUCTION... 4-9 USER
More informationUsing the Akta Prime plus October 22, 2012
Some starting precautions: 1. Vacuum filter all buffers. Removes any large particles/debris that may clog your column De-gases the buffers 2. Clarify lysates first by centrifugation and then filtration
More informationBASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1
BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1 The front door of the main body of the instrument may be open when you arrive. Take the sample chamber and slide it into position with the
More informationLABORATORY TECHNIQUES. Pouring Liquids
LABORATORY TECHNIQUES Working in the chemistry laboratory you will be handling potentially dangerous substances and performing unfamiliar tasks. This section provides you with a guide to the safe laboratory
More informationSeton Hall University. 200 MHz MR. SOP manual
Seton Hall University 200 MHz MR SOP manual NMR Information: Varian Gemini 2000 NMR 1 H Frequency = 200 MHz Original purchase date: 1989 Donated from Merck & Co., Inc., 1998 Software upgrade: 2008 To schedule
More informationHemoCue Hb Procedure Template
HemoCue Hb 201 + Procedure Template PURPOSE The HemoCue Hb 201 + System is used for the quantitative determination of hemoglobin in blood using a specially designed analyzer, HemoCue Hb 201 +, and specially
More informationINSTRUCTION MANUAL. Automatic Blood Pressure Monitor with Fit Cuff. =Fit Cuff=!"#$% IA1B. Model
IA1B INSTRUCTION MANUAL Automatic Blood Pressure Monitor with Fit Cuff =Fit Cuff=!"#$% Model IA1B Contents Introduction... 2 Notes on Safety... 3 Know Your Unit... 5 Quick Reference Guide... 7 Initial
More information6800 Maintenance Instruction System Flush Procedure
Equipment Required FA74005 FA65318 FA900005 FA900003 Damper Drain Tube 6800 Cover Removal Tool Beaker 0.25 Litre Solvent Cleaning Bottle FA940021 Syringe Polypropylene 50 ml as required FA999045 Gloves
More informationEC214 EC215 - EC215R Bench Conductivity Meters
Instruction Manual EC214 EC215 - EC215R Bench Conductivity Meters http://www.hannainst.com These Instruments are in Compliance with the CE Directives Dear Customer, Thank you for choosing a Hanna Instruments
More informationFourier Systems. General Safety Considerations User Manual. Innovation with Integrity NMR
Fourier Systems General Safety Considerations User Manual Version 001 Innovation with Integrity NMR Copyright by Bruker Corporation All rights reserved. No part of this publication may be reproduced, stored
More informationSPECIFICATIONS APCEPH1
APCEPH1 ph CONTROLLER SPECIFICATIONS APCEPH1 Input voltage 120 Volts AC Maximum amperage 14.5 amps @ 120 VAC ph Accuracy +/- 0.2 ph ph Control range Adjustable 4.5 8.5 ph Weight < 1 lbs Dimensions 3" x
More information5.1 WATCH MOVEMENTS QUARTZ MODELS
5.1 WATCH MOVEMENTS QUARTZ MODELS MAIN COMPONENTS A Hour Hand B Minute Hand C Second Hand D Crown Description of the display and control buttons Setting the time 1. Pull out the crown to position II (the
More informationAscend Aeon. User Manual. Innovation with Integrity. Version NMR
500 89 Ascend Aeon User Manual Version 00 Innovation with Integrity NMR Copyright by Bruker Corporation All rights reserved. No part of this publication may be reproduced, stored in a retrieval system,
More informationBS ml Bottle Sampling Unit
BS500 500ml Bottle Sampling Unit User Guide www.mpfiltri.co.uk 200.307-EN Covers Model Numbers BS500 SAFETY WARNING Hydraulic systems contain dangerous fluids at high pressures and temperatures. Installation,
More informationThe Oxygen Sensor can be connected to all types of einstein Tablets, einstein LabMate, and einstein LabMate +.
The Oxygen sensor is a maintenance- free galvanic oxygen electrode capable of measuring Oxygen in solutions. With built-in temperature compensation, this sensor provides an easy way for students to receive
More informationDOscan10 Pocket Dissolved Oxygen Tester Instruction Manual
DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual BANTE INSTRUMENTS CO., LTD DOscan10 Pocket Dissolved Oxygen Tester 1 Thank you for selecting the DOscan10 pocket dissolved oxygen tester. This
More informationWaste Oil Extraction Unit
75 l, with Visual inspection reservoir 95 l, with Visual inspection reservoir Operating Instructions Contents: 1. General Information 1.1 Intended use 1.2 Construction and description of operation 1.3
More informationProduct instruction manual Easymount Sign Laminators EM-S1400C, EM-S1600C, EM-S1400H, EM-S1600H
Product instruction manual Easymount Sign Laminators EM-S1400C, EM-S1600C, EM-S1400H, EM-S1600H The Easymount Sign has been designed to be user friendly, however we strongly recommend you take a few minutes
More informationUser s Guide. Vacuum Controller for liquid delivery systems
Flow Control User s Guide Vacuum Controller for liquid delivery systems Precise Vacuum Control throughout the experiment Flow control Compatible with any perfusion system Ideal for Small Volume Delivery
More informationInstruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text.
Hach LDO Sensor Instruction Sheet 00745589 Safety Precautions Please read this entire instruction sheet before operating this sensor. Pay particular attention to all danger and caution statements. Failure
More informationicreasepro Creaser Operators Manual
6-2013 Version 3.0 icreasepro Creaser Operators Manual WWW.MBMCORP.COM 800-223-2508 TABLE OF CONTENTS SPECIFICATIONS.1a SAFETY PROCEDURES/CARE & MAINTENANCE..1b COMPONENT IDENTIFICATION 2 TOUCH SCREEN
More information3 Post Pressure Fit System Owner s Manual
3 Post Pressure Fit System Owner s Manual Use and Care Trouble Shooting Warranty Information Table of Contents 3 Post Pressure Fit System Introduction... 3 Overview of the 3 Post Pressure Fit System...
More informationPegas 4000 MF Gas Mixer InstructionManual Columbus Instruments
Pegas 4000 MF Gas Mixer InstructionManual Contents I Table of Contents Foreword Part I Introduction 1 2 1 System overview... 2 2 Specifications... 3 Part II Installation 4 1 Rear panel connections...
More informationSetting a New Standard in Flash Chromatography Performance
Setting a New Standard in Flash Chromatography Performance What s New Faster Flow rates, up to 300 ml/min Higher pressures, up to 300 psi (20 bar) Bigger touchscreen, 12 or 15 inches Wider Dynamic Range
More informationtubepress press procedure
tubepress press procedure tubepress 15mm up to 35mm tubepress 15 mm up to 35 mm 1. The tube ends must be clean with no scratches or grooves. Remove end caps. Cut tube to length. 2. Deburr tube inside and
More informationNew Cortex Software User Guide
New Cortex Software User Guide Initial Set Up Turn on the Cortex and leave to warm up for approximately 15 minutes before calibrating Load MSS Toolbox from the desktop From the toolbox menu select Calibration
More informationCRYOGENIC TANK SERVICES - CHART RECORDER MANUAL
CRYOGENIC TANK SERVICES - CHART RECORDER MANUAL Pressure Chart Recorder / Temperature Chart Recorder / Dual Recorder Portable Mechanical Chart Recorder for the accurate measurement and recording of Pressure
More informationOptical Dissolved Oxygen Meter
Optical Dissolved Oxygen Meter HI764113 Rugged Optical Dissolved Oxygen Probe for Fresh and Saltwater Applications Digital, weighted probe No membranes No electrolytes No oxygen consumption No flow dependence
More informationBekkTech s Procedures For Performing In Plane Membrane Conductivity Testing
BekkTech s Procedures For Performing In Plane Membrane Conductivity Testing Important Equipment and Software Considerations in Conductivity Testing Work Performed Under US DOE Contract # DE FC36 06GO16028
More informationBNL. User Manual. Innovation with Integrity. Version. Accessories
BNL User Manual Version 04 Innovation with Integrity Accessories Copyright by Bruker Corporation All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted,
More information