KARL SUSS MJB3 MASK ALIGNER STANDARD OPERATING PROCEDURE

Size: px
Start display at page:

Download "KARL SUSS MJB3 MASK ALIGNER STANDARD OPERATING PROCEDURE"

Transcription

1 KARL SUSS MJB3 MASK ALIGNER STANDARD OPERATING PROCEDURE Purpose of this Instrument: This instrument is for patterning photosensitive polymers with UV light. Location: White Hall 410 Cleanroom Primary Staff Contact: Harley Hart (412) (M) (304) (O) Office: White Hall 409 Secondary Staff Contact: The Shared Research Facilities are operated for the benefit of all researchers. If you encounter any problems with this piece of equipment, please contact the staff member listed above immediately. There is never a penalty for asking questions. If the equipment is not behaving exactly the way it should, contact a staff member. WARNING: The Mask Aligner emits UV light. Exposure of UV light on unprotected skin can cause severe burns. Looking directly into the UV light source can cause temporary or permanent blindness. WARNING: If at any time during operation, there is a sound of shattering glass and the voltage led display on the lamp power supply drops to zero or displays fail turn off the lamp controller and the Mask Aligner, evacuate all users from the room, and notify a Shared Research Facilities staff member. Do not attempt to open the system or inspect the arc lamp! Mercury vapor may be released if the arc lamp breaks. Opening the lamp housing will expose you to this vapor. ATTENTION: Users are required to wear UV protective glasses when operating the Mask Aligner. These glasses are located on top of the Mask Aligner.

2 INSTRUMENT START UP 1. Reserve the equipment in CORES and login using the CORES kiosk in the cleanroom entry lab. 2. Sign into the logbook at the instrument. 3. Check the compressed air pressure gauge behind the instrument and before the nitrogen generator. Record this value in the logbook. The pressure reading should be between psi see Figure 1. If it is not within this range, contact a Shared Research Facilities staff member. Figure 1: Pressure before the nitrogen generator 4. Check the compressed air pressure gauge after the nitrogen generator behind the instrument that is attached to the optical table. Record this value in the logbook. The pressure reading should be between psi see Figure 2. If it is not within this range, contact a Shared Research Facilities staff member. 5. Turn on the vacuum pump behind the instrument by pressing the black toggle switch to the ON position and listen for the pump to start see Figure 3. This is required for vacuum to occur at the sample holder and at the mask holder. Figure 2: Pressure regulator after the nitrogen generator 1

3 Figure 3: Image of the vacuum pump 6. Make sure the Mask Aligner is powered OFF. Power up the lamp controller by turning on the power switch on the right hand side of the panel. A picture of the lamp controller is shown in Figure 4 and it is located to the right of the Mask Aligner. Once turned on, the lamp controller led panel will cycle through a warm-up procedure and then the display will read rdy for ready. NOTE: The lamp power supply is interlocked with the Mask Aligner. The lamp cannot ignite if the Mask Aligner is on. Figure 4: Lamp Controller in idle state 7. Start the arc lamp by pressing, and holding for a few seconds, the start button. If the arc lamp ignites successfully, the led display on the lamp power supply will show cold. If the lamp does not ignite immediately, it will attempt ignition five times. If the lamp controller does not display the cold message but displays fail or continues to display rdy, then try again. If the lamp does not ignite after the third attempt, please contact a Shared Research Facilities staff member for assistance. 8. Wait 15 minutes for the lamp to warm up. The led display on the lamp power supply should read 275 Watts. WARNING: If the power does not read 275 W do not proceed. Attempting to run the lamp at a higher idle power may cause the lamp to fail or crack. Contact a Shared Research Facilities staff member for assistance. 2

4 9. Power on the Mask Aligner by pressing the red POWER button on the control panel and verify the following system nitrogen pressures on the front of the instrument see Figure 5: Exposure: 4 Bar White Hall Mask Aligner Parallelity: 2 Bar Pressure/Wafer: 1 Bar Figure 5: Nitrogen Pressures on Mask Aligner 10. Turn the microscope power supply toggle switch to the ON position as shown in Figure 6. The microscope lamp will turn ON. Figure 6: Microscope Power Supply MEASURING THE LAMP INTENSITY 1. The output UV intensity should be measured and recorded in the log book when the lamp has been turned on. Make sure that the lamp has warmed up before taking the UV measurement by waiting 15 minutes. 2. Turn on the power to the Karl Suss 1000 UV intensity meter and probe, shown in Figure 7. The probe should have a label on the side marking it as a 365nm probe. Figure 7: Karl Suss 1000 UV Intensity meter with probe 3

5 3. Place the UV probe on the stage and move the stage into contact position by moving the contact lever on the left-hand side of the instrument counterclockwise to the CONTACT position. The CONTACT light will illuminate. Then press the HP/ST button and the VACUUM chamber button both buttons should illuminate - to apply vacuum to the back of the probe. This will hold the sensor in place during exposure see Figure 8. Figure 8: Probe on Stage 4. Set the exposure time to 10 seconds and press the EXPOSURE Button. The mirror housing will move forward and expose the sensor as shown in Figure 9. Figure 9: UV measurement in progress 5. Record the UV meter value and record it in the logbook. Units are in mw/cm This process should be repeated with the photo-mask placed over the 365nm UV probe for experimental calculations to determine the through mask intensity. NOTE: Make sure that the silver disc is not covered by any darkened area of the mask. If any part of the mask covers the silver disc area, then the measured value will be lower than the actual UV intensity. 7. Calculate the exposure time by dividing the desired exposure dosage, which should be in mj/cm 2 by the UV intensity measured through the mask in mw/cm 2. Exposure time (s) = Exposure Energy (mj/cm2 ) UV Intensity (mw/cm 2 ) NOTE: Make sure to record both the measured intensity and calculated time in your lab notebook. The measured intensity will change over time, resulting in a change in exposure time. 4

6 LOADING A MASK 1. Mount the photo-mask on one of the mask holders shown in Figure 10a & 10b. The photo-mask should be positioned so that the chrome side of the photo-mask is facing up and manually centered over the mask holder opening. NOTE: If the 3 sample holder is to be used, the mask holder will need to be changed by disconnecting the vacuum line from the photo mask holder for 2 samples and connecting it to the photo mask holder for 3 samples. Figure 10a: Photo-mask holder for 3 sample holders Figure 10b: Photo-mask holder for small pieces and 2 sample holders 2. Press the VACUUM MASK button on the control panel of the Mask Aligner and check that the photo-mask is securely attached to the holder by trying to gently move the mask. NOTE: The mask should be centered in the mask holder. If the mask is placed too close to the front or back edge of the holder, it will interfere with mask holder insertion and potentially damage the mask. 3. Rotate the microscope objective turret so that the 10x objective is over the stage. 5

7 4. Pick up the mask holder, and flip it over. Insert the mask holder into the mask holder grooves on the Mask Aligner see Figure 11. Figure 11: Grooves for mask holder 5. Fully insert the mask holder, and make sure that the holder is flush and level with the Mask Aligner. Tighten the mask holder knobs on the front of the instrument see Figure 12. Figure 12: Mask holder with photomask installed LOADING A WAFER 1. Insert the desired sample holder into the transport slide and verify that the sample holder is secured and that the transport slide post is properly inserted into the notch of the sample holder see Figures 13a & 13b. Figure 13a: 3 Sample Holder Figure 13b: 2 Sample Holder Figure 13c: Small pieces Sample Holder 6

8 2. Place the sample on the sample holder and verify that all vacuum holes are covered see Figure 14 below. Failure to completely cover all holes will lead to loss of sample vacuum during processing. Figure 14: Sample properly loaded onto the small piece sample holder 3. Squeeze the button on the front of the finger grip located at the right-hand side of the transport slide. This will hold the substrate in place. Then gently move the transport slide completely to the left under the mask. ALIGNING A WAFER TO THE MASK 1. The idle position of the variable thickness adjustment knob is 10. The thickness of each user s wafers is different, so this position will need to be adjusted by each user to achieve the proper alignment gap. While viewing through the microscope, gently move the contact lever, which controls the z-axis of the stage, on the left-hand side of the instrument counter-clockwise while watching for the shadow between the sample and the photo-mask to disappear and while feeling for resistance in the contact lever. The preferred variable thickness adjustment knob position is for the shadow to disappear when the contact lever has reached the black line on the side of the instrument see Figure 15 below. If your sample begins to make contact with the photo-mask before reaching the CONTACT position, then increase the value of the variable thickness adjustment knob. If your sample does not make contact with the photo-mask before reaching the CONTACT position, then decrease the value of the variable thickness adjustment knob and repeat. NOTE: Each revolution of variable thickness adjustment knob is equal to 150 microns. 7

9 Figure 15: Contact Lever WARNING: Always be aware of the microscope objectives position in relation to the photo-mask as it is possible to collide the objectives with the photo-mask and/or mask holder. This can result in damage to the objectives and/photo-mask. NOTE: To change the variable thickness adjustment knob value, first unlock the position of the adjustment knob by moving the locking tab counter-clockwise. Once the value is changed, be sure to lock the position by moving the locking tab clockwise as shown in Figure 16. Figure 16: Variable thickness adjustment knob WARNING: Only adjust the variable thickness adjustment knob whenever the CONTACT lever is in the retracted position. Failure to do so can result in damage to the instrument. Users are responsible for resetting the variable thickness adjustment knob back to 10 when their process is completed. 8

10 2. Select an exposure mode by pressing the appropriate buttons. The possible Exposure modes are as follows: Standard Mode (ST): SOFT CONTACT: The ST button and the SOFT CONTACT buttons are illuminated. This brings the substrate into contact with the mask using the only pressure applied from the contact lever. NOTE: Sample carriers without vacuum gaskets are required to use this mode of operation. HARD CONTACT: ST mode button is illuminated on the control panel and the SOFT CONTACT button is not. Prior to exposure vacuum will be removed and nitrogen pressure will press the substrate against the mask. Once exposure is complete the nitrogen pressure will be removed and vacuum will be reapplied to the substrate HP Mode: High Precision is used to achieve the highest possible resolution as this method creates the smallest gap possible between the photo-mask and the substrate. Prior to exposure the vacuum under the substrate is switched OFF and vacuum between the substrate and mask is switched ON. Then after exposure the vacuum between the substrate and mask is switched OFF and vacuum under the substrate is switched back ON. 3. Move the CONTACT LEVER counterclockwise to bring the substrate into contact position. The SEPARATION LEVER cannot be used unless the CONTACT LEVER has the sample in the contact position. 4. Move the SEPARATION LEVER to the front of the instrument. A small flow of nitrogen will be introduced into the vacuum chamber to ensure that the substrate separates from the photo-mask. The SEPARATION indicator light on the control panel will illuminate and the CONTACT light will power off - see Figure 17. Figure 17: Separation Lever in the sample separated position 9

11 NOTE: If the substrate continues to adhere to the photo-mask during SEPARATION, then the nitrogen throttle pressure can be increased. If the substrate shifts during SEPARATION, then the nitrogen throttle pressure needs to be decreased. Please contact a Shared Research Facilities staff member. 5. Focus the microscope using the focus adjustments knobs shown in Figure 18. NOTE: The microscope focusing knob has combined coarse/fine focus adjustment. If the focus knob is turned in only one direction, then coarse adjustment will occur. Fine focus adjustment is automatically selected whenever the slightest turn is made in the opposite direction. The adjustment will switch back to coarse adjustment whenever the fine focus adjustment limits are exceeded. Figure 18: Microscope adjustment knobs 6. Using the X position, Y position, and θ position micrometers shown in Figure 19a, as well as the microscope manipulator shown in Figure 19b, adjust the position of the substrate features to line up with the mask patterns. For best results, use the θ position micrometer to align the substrate rotation first, before adjusting the X and Y positions. 7. Then move the SEPARATION LEVER backwards to move the sample into the contact position. The SEPARATION indicator light will go out and the CONTACT light will illuminate. Figure 19a: X, Y, and Theta Alignment knobs Figure 19b: Microscope Manipulator 10

12 NOTE: The microscope manipulator moves on X-Y slides which are equipped with pneumatic brakes. The two buttons located on the manipulator unlocks either or both the X and Y brakes. This allows the microscope to be scanned in either the X or Y directions exclusively or in both directions simultaneously. 8. Once the substrate is in contact with the photo-mask, set the timer to the desired setpoint for exposure on the exposure time see Figure 20. Verify that the proper units have been selected. Figure 20: Exposure timer 9. Check that the sample is in desired EXPOSURE mode by checking that the proper CONTROL PANEL lights are illuminated from Step # Say EXPOSE to inform all lab users about the UV hazard. Press the EXPOSURE button on the control panel, the vacuum light will turn off and the mirror will move forward and expose the sample. Turn your eyes away from the instrument until the exposure is complete. 11. Once the exposure is complete, the mirror will retract to its original position. 12. Move the CONTACT LEVER clockwise to the retracted position. This will bring your sample away from the photo-mask. 13. Squeeze the button on the front of the finger grip located at the right-hand side of the transport slide. This will hold the substrate in place. Then gently move the transport slide completely to the right. Then using tweezers remove your sample. 14. If a second sample is to be processed, refer back to step# 1 of this section. 11

13 TURNING OFF THE SYSTEM 1. Before shutting down the system, the user must first remove the mask from the aligner. To remove the mask, loosen the two mask holder knobs and slide holder to the left. Place the holder upside down on the table to the left of the instrument. Press the VACUUM MASK button to OFF on the front of the instrument and then remove the mask and place the mask in the storage container. WARNING: Do not turn off the VACUUM MASK button unless the holder is upside on the table. Failure to do so can lead to mask damage or breakage. 2. Set the variable thickness adjustment knob is set to 10 and ensure the tab on the micrometer is in the lock position. 3. Rotate the microscope objective turret until the 10x objective is over the sample stage. 4. Center the aligner stage by setting the X micrometer to ~5 and the Y micrometer to ~10. The θ micrometer can be adjusted to center the rotational indicator. 5. Turn the microscope power supply toggle switch to the OFF position. The microscope lamp will turn off. 6. Power off the Mask Aligner by pressing the red POWER on the front control panel. 7. Power off the lamp controller by turning off the POWER switch. 8. Turn off the system vacuum pump at the back of the instrument by moving the black toggle switch to the OFF position. 9. Clean up the work area and make sure that all requested parameters have been entered into the log book. 10. Sign out of the CORES system using the barcode scanner in the entry lab. NOTE: If any issues with the instrument were observed, the user must send an to the SRF staff with PROBLEM REPORT in the subject and enter a detailed summation of the issue observed. 12

14 EMERGENCY SHUT-DOWN PROCEDURES If, at any time, the user needs to contact someone for help, call or locate the following staff of the Shared Research Facilities (SRF): Primary Staff Contact: Harley Hart (412) (M) (304) (O) Office: White Hall 409 Secondary Staff Contact: Dr. Kolin Brown (304) (304) (O) Office: ESB G75D If no one is available and the instrument is not acting as expected, the user should do the following: Turn OFF the system power Turn OFF the lamp power supply Turn OFF the vacuum pump Then, if possible, the user should stay by the instrument while trying to contact a Shared Research Facilities staff member. If it becomes necessary to leave the instrument then the user should leave a large, legible note on the MASK ALIGNER stating: The problem (describe what happened and steps taken) When it occurred (date and time) User name and phone number If it becomes necessary to leave the instrument then the user should leave a large, legible note at the MASK ALIGNER stating the instrument is DOWN. If a dangerous situation is evident (smoke, fire, sparks, etc.), or if the sound of shattering glass is heard emitting from the lamp housing, ONLY if it is safe to do so, the user should turn off system, lamp controller, and vacuum pump or unplug the instrument and leave the cleanroom immediately. The user should notify all other cleanroom persons within the cleanroom to evacuate. The user should then contact proper emergency personnel. The contact numbers can be found posted outside of the cleanroom or on the cover of the instrument log book. 13

KARL SUSS MJB3 UV400 Mask Aligner Standard Operating Procedure

KARL SUSS MJB3 UV400 Mask Aligner Standard Operating Procedure KARL SUSS MJB3 UV400 Mask Aligner Standard Operating Procedure Version: 1.0 February 2014 UNIVERSITY OF TEXAS AT ARLINGTON Nanotechnology Research Center (NRC) 1 TABLE OF CONTENTS 1 Introduction 3 1.1

More information

MJB-3 Mask Aligner Property of TAU MNCF August 2012

MJB-3 Mask Aligner Property of TAU MNCF August 2012 MJB-3 Mask Aligner I. Power Up Sequence 1 LOG IN. Turn on Nitrogen & compressed air lines on the back wall. 2 Turn on the vacuum pump. 3 Flip the COMPRESSED AIR toggle on (up) Compressed air is used to

More information

SOP for Karl Suss MJB3 #1 Mask Aligner

SOP for Karl Suss MJB3 #1 Mask Aligner SOP for Karl Suss MJB3 #1 Mask Aligner Rev. 5 (30/11/2016) Safety UV Exposure: The high-energy light produced by the high-pressure Mercury Xenon lamp can cause eye damage and skin burns. Be sure that the

More information

Warnings: Notes: Revised: October 5, 2015

Warnings: Notes: Revised: October 5, 2015 Karl Suss MA6 Mask Aligner Standard Operating Procedure Faculty Supervisor: Prof. Robert White, Mechanical Engineering (x72210) Safety Office: Peter Nowak x73246 (Just dial this directly on any campus

More information

Karl Suss MJB4 Mask Aligner

Karl Suss MJB4 Mask Aligner Karl Suss MJB4 Mask Aligner Tool Manager: Yong Sun ( yongs@princeton.edu; Office 8-8234; Cell 609-917-5076 ) Backup: George Watson ( gwatson@princeton.edu; Office 8-4626; Cell 732-996-2713 ) ******************************************************************************

More information

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone:

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone: MJB4 Mask Aligner Operating Procedure Effective Date: 07/12/2012 Author(s): Jiong Hua Phone: 402-472-3773 Email: jhua2@unl.edu 1 1 Introduction 1.1 Key Words Karl Suss MJB4 Mask Aligner, Optical Lithography,

More information

SPUTTER STATION STANDARD OPERATING PROCEDURE

SPUTTER STATION STANDARD OPERATING PROCEDURE SPUTTER STATION STANDARD OPERATING PROCEDURE Purpose of this Instrument: This instrument is used for deposition of thin metal or oxide films. Source materials supplied by WVU Shared Research Facilities:

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual AB-M Mask Aligner Version 1.1 Page 1 of 18 Contents 1. Picture and Location 2. Process Capabilities 2.1 Cleanliness Standard 2.2 Wafer Chuck Selection 2.3 Mask Holder Selection

More information

Karl Suss MA6 Mask Aligner SOP

Karl Suss MA6 Mask Aligner SOP Page 1 of 11 Karl Suss MA6 Mask Aligner SOP Safety UV Exposure: The high energy light produced by the high pressure Mercury Xenon lamp can cause eye damage and skin burns. Be sure that the light guards

More information

Approved by Principal Investigator Date: Approved by Super User: Date:

Approved by Principal Investigator Date: Approved by Super User: Date: Approved by Principal Investigator Date: Approved by Super User: Date: Standard Operating Procedure BNC OAI 200 Lithographic Mask Aligner (Aligner 3) Version 2011 June 2 I. Purpose This Standard Operating

More information

Warnings: Notes: Revised: January 8,

Warnings: Notes: Revised: January 8, OAI Model 204IR Mask Aligner Standard Operating Procedure Faculty Supervisor: Prof. Robert White, Mechanical Engineering (x72210) Safety Office: Peter Nowak x73246 (Just dial this directly on any campus

More information

Usage Policies Notebook for Karl Suss MA6 Mid / Deep UV Mask Aligner

Usage Policies Notebook for Karl Suss MA6 Mid / Deep UV Mask Aligner Usage Policies Notebook for Karl Suss MA6 Mid / Deep UV Mask Aligner Revision date September 2014 2 Emergency Plan for Karl Suss MA6 Aligner Standard Operating Procedures for Emergencies Contact information

More information

University of Minnesota Nano Fabrication Center

University of Minnesota Nano Fabrication Center Equipment name: Karl Suss Mask Aligners Badger name: MA6-P Revision number: 0 Model: MA6 Revisionist: Paul Kimani Location: Bay 4 PAN Date: September 17, 2015 1. Introduction The Karl Suss MA6-PAN is a

More information

Equipment Standard Operating Procedure Greg Allion and Kimberly Appel

Equipment Standard Operating Procedure Greg Allion and Kimberly Appel Date Created: May 3, 2004 Date Modified: June 1, 2005 MA6/BA6 Mask Aligner Equipment Standard Operating Procedure Greg Allion and Kimberly Appel 1. Purpose 1.1. Photolithography involves transferring a

More information

COBILT CA-800 Mask Aligner Equipment Operation

COBILT CA-800 Mask Aligner Equipment Operation COBILT CA-800 Mask Aligner Equipment Operation For the Micro-Electronics Laboratory At University of Notre Dame Department of Electrical Engineering This user manual is not be removed from room 247A. This

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual Karl Suss MA6 Mask Aligner Version 1.1 Page 1 of 24 Contents 1. Picture and Location 2. Process Capabilities 2.1 Cleanliness Standard 2.2 Substrate Size 2.3 Photo Mask Size 2.4

More information

University of Minnesota Nano Fabrication Center

University of Minnesota Nano Fabrication Center Equipment name: Karl Suss Mask Aligners Badger name: MA6 & Maba6 Revision number: 4 Models: Revisionist: Paul Kimani Location: Bay 2 Date: April 25, 2013 1. Introduction The Karl Suss MA6 and MABA6 are

More information

Quintel Mask Aligner

Quintel Mask Aligner Quintel Mask Aligner (quintel) 1.0 1.0 Title Quintel Q4000 Mask Aligner 2.0 2.0 Purpose The Quintel Q4000 MA (Mask Aligner) is a top and bottom side contact lithography printer with the video-view split

More information

OAI Model 200 Tabletop Mask Aligner Portland State University

OAI Model 200 Tabletop Mask Aligner Portland State University OAI Model 200 Tabletop Mask Aligner Portland State University WARNING: This machine exposes users to ultraviolet radiation. Do not touch the lens underneath the lamp hood as it may damage the machine and

More information

Approved by Principal Investigator Date: Approved by Super User: Date:

Approved by Principal Investigator Date: Approved by Super User: Date: Approved by Principal Investigator Date: Approved by Super User: Date: Standard Operating Procedure BNC OAI Lithographic Mask Aligner (Aligner 2) Version 2008 October 31 I. Purpose This Standard Operating

More information

ABM MASK ALIGNERS. NanoFab 26 March 2009 A Micro Machining & Nanofabrication Facility

ABM MASK ALIGNERS. NanoFab 26 March 2009 A Micro Machining & Nanofabrication Facility ABM MASK ALIGNERS LOCATION: Optical Lithography PRIMARY TRAINER: Stephanie Bozic (2-6724, sbozic@ualberta.ca) SECONDARY TRAINER: Jolene Chorzempa (2-4823, jolenec@ualberta.ca) 1. OVERVIEW The ABM Mask

More information

Operation of the mask aligner MJB-55

Operation of the mask aligner MJB-55 John Paul Adrian Glaubitz Operation of the mask aligner MJB-55 Department of Physics Faculty of Mathematics and Natural Sciences University of Oslo 1 Introduction The mask aligner is an essential tool

More information

Karl Suss Contact Aligner Operation

Karl Suss Contact Aligner Operation Karl Suss Contact Aligner Operation Roger Robbins 6/31/2008 2 nd ed. 3/12/2009 The University of Texas at Dallas Erik Jonsson Engineering School of Engineering TITLE: Karl Suss Contact Aligner Operation

More information

Operation of the contact mask aligner Canon PPC 210 Projection Print Camera

Operation of the contact mask aligner Canon PPC 210 Projection Print Camera Operation of the contact mask aligner Canon PPC 210 Projection Print Camera Start-up Procedure: 1. Start the grey pump in the service corridor behind the aligner 2. Open the valve for the Nitrogen gas

More information

JASCO 810 CD SPECTROPLOARIMETER STANDARD OPERATING PROCEDURE

JASCO 810 CD SPECTROPLOARIMETER STANDARD OPERATING PROCEDURE JASCO 810 CD SPECTROPLOARIMETER STANDARD OPERATING PROCEDURE Purpose of this Instrument: This instrument is for measuring differences in the absorption of lefthanded polarized light versus right-handed

More information

OPERATION OF THE DIMPLER

OPERATION OF THE DIMPLER OPERATION OF THE DIMPLER After thinning your sample to ~80 μm you can now do a dimpling process to thin the center up to 10 μm. When you walk in and use the DIMPLER it should already be calibrated and

More information

R I T. Title: Amray 1830 SEM Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 09/29/03 1 SCOPE 2 REFERENCE DOCUMENTS

R I T. Title: Amray 1830 SEM Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 09/29/03 1 SCOPE 2 REFERENCE DOCUMENTS Fabrication Laboratory Revision: A Rev Date: 09/29/03 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Amray 1830 SEM. All users

More information

5.1.3 Mechanical Hazards Drive assemblies have sufficient power to cause injury. Keep hands, fingers, clothing and tools clear of moving parts.

5.1.3 Mechanical Hazards Drive assemblies have sufficient power to cause injury. Keep hands, fingers, clothing and tools clear of moving parts. Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the PE4400. All users are expected to have read and understood this document. It is

More information

March CS-1701F Reactive Ion Etcher

March CS-1701F Reactive Ion Etcher March CS-1701F Reactive Ion Etcher Standard Operating Procedure Faculty Supervisor: Prof. Robert White, Mechanical Engineering (x72210) Safety Office: Peter Nowak x73246 (Just dial this directly on any

More information

Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System)

Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System) Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System) Revision date October 2014 2 Emergency Plan for Nano Furnace Standard Operating Procedures for Emergencies Contact information Person

More information

O P E R ATING INSTRUCTIONS FOR MODEL SPR-45 Automatic Screen and Stencil Printer

O P E R ATING INSTRUCTIONS FOR MODEL SPR-45 Automatic Screen and Stencil Printer O P E R ATING INSTRUCTIONS FOR MODEL SPR-45 Automatic Screen and Stencil Printer TABLE OF CONTENTS I. SPECIFICATIONS...3. II. SAFETY INSTRUCTIONS...4. III. INSTALLATION...5. IV. SET-UP...6. V. SYSTEM OPERATION...9.

More information

Photolithography. Operating Instructions

Photolithography. Operating Instructions Photolithography Operating Instructions The PR used during this laboratory session will be Microposit S1813 (from Shipley). Make sure everyone is following the laboratory protocol. Wear lab coats, safety

More information

Oerlikon Sputtering Evaporator SOP

Oerlikon Sputtering Evaporator SOP Oerlikon Sputtering Evaporator SOP Short UNT Cleanroom 1. Taking out sample holder from Transport Chamber : Log in FOM to access the software Go to the software and log in with user1 and password user1

More information

Operating Procedures for Metal Evaporator I

Operating Procedures for Metal Evaporator I Operating Procedures for Metal Evaporator I Metal Evaporator I is intended as a tool and a training device. Understanding the operation of this equipment should give you a basic knowledge of vacuum and

More information

Approved by Principal Investigator Date: Approved by Super User: Date:

Approved by Principal Investigator Date: Approved by Super User: Date: Approved by Principal Investigator Date: Approved by Super User: Date: Standard Operating Procedure BNC Commonwealth Dual Ion Beam Deposition System (CDIBS) Version 2010 February 14 I. Purpose This Standard

More information

Standard Operating Procedure. For. PVD E-Beam

Standard Operating Procedure. For. PVD E-Beam P a g e 1 Standard Operating Procedure For PVD E-Beam P a g e 2 Introduction The PVD Electron-Beam Evaporator (E-Beam) thin film deposition machine uses a magnetically guided and collimated stream of electrons

More information

STANDARD OPERATING PROCEDURE: ALLIED HIGH TECH MULTIPREP POLISHING SYSTEM MATERIAL REMOVAL

STANDARD OPERATING PROCEDURE: ALLIED HIGH TECH MULTIPREP POLISHING SYSTEM MATERIAL REMOVAL STANDARD OPERATING PROCEDURE: ALLIED HIGH TECH MULTIPREP POLISHING SYSTEM MATERIAL REMOVAL Credit: Based on Allied High Tech manual v.7.3 (www.alliedhightech.com) Purpose of this Instrument: Essential

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual Oxford Plasmalab 80 Plus Plasma Etcher Page 1 of 24 Contents 1. Picture and Location 2. Process Capabilities 2.1 Cleanliness Standard 2.2 Available Etching Materials 2.3 Performance

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual Denton Explorer 14 RF/DC Sputter Version 1.0 Page 1 of 11 Contents 1. Picture and Location 2. Process Capabilities 1. Cleanliness Standard 2. Available for Sputtering Materials

More information

Angstrom Dielectric Sputterer Operation Manual

Angstrom Dielectric Sputterer Operation Manual Angstrom Dielectric Sputterer Operation Manual I. System overview The Angstrom Dielectric Sputterer (ADS) has a similar interface as the Angstrom metal sputterer. It has two screens, the process screen

More information

Revised: June 7, 2017

Revised: June 7, 2017 LC Technologies Thermal Evaporator Standard Operating Procedure Faculty Supervisor: Prof. Robert White, Mechanical Engineering (x72210) Safety Office: Peter Nowak x73246 (Just dial this directly on any

More information

OPERATION OF LIQUID PARTICLE COUNTER. GP-B P0045 Rev -

OPERATION OF LIQUID PARTICLE COUNTER. GP-B P0045 Rev - 1 W. W. Hansen Experimental Physics Laboratory STANFORD UNIVERSITY STANFORD, CALIFORNIA 94305-4085 Gravity Probe B Relativity Mission OPERATION OF LIQUID PARTICLE COUNTER GP-B P0045 Rev - May 23, 1997

More information

Cryo-Evaporator Operation

Cryo-Evaporator Operation Cryo-Evaporator Operation Cara Ricci November 25, 2003 Thin Film Deposition THE UNIVERSITY OF TEXAS AT DALLAS ERIK JOHNSON SCHOOL OF ENGINEERING DOCUMENT NUMBER: FA2003-TF-008 EDITION: 1.1 PAGE: 1 of 28

More information

Model 130M Pneumatic Controller

Model 130M Pneumatic Controller Instruction MI 017-450 May 1978 Model 130M Pneumatic Controller Installation and Operation Manual Control Unit Controller Model 130M Controller is a pneumatic, shelf-mounted instrument with a separate

More information

Issue: H Title: CHA E-Beam Evaporator Page 1 of 7. Table of Contents

Issue: H Title: CHA E-Beam Evaporator Page 1 of 7. Table of Contents Title: CHA E-Beam Evaporator Page 1 of 7 Table of Contents Purpose/Scope... 2 2.0 Reference Documents... 2 3.0 Equipment/Supplies/Material... 2 4.0 Safety... 2 5.0 Set Up Procedures... 2 5.1 PC Logon and

More information

Login to ilab Kiosk. Revised 05/22/2018. Load your sample:

Login to ilab Kiosk. Revised 05/22/2018. Load your sample: Login to ilab Kiosk Load your sample: 1. Check: The analysis chamber pressure is

More information

Nanofabrication Facility: PECVD SOP Rev. 00, April 24

Nanofabrication Facility: PECVD SOP Rev. 00, April 24 Author: Charlie Yao & Mario Beaudoin Email: charlieyao@gmail.com; Beaudoin@physics.ubc.ca Phone: 604-822-1853(MB). Purpose This document outlines the standard operation for the Trion Plasma Enhanced Chemical

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual Fisher Scientific Isotemp TM Model 281A Vacuum Oven Version 1.1 Page 1 of 9 Contents 1. Picture and Location 2. Process Capabilities 2.1 Cleanliness Standard 2.2 Substrate Size

More information

Product Information News

Product Information News Product Information News MMR SECOND STAGE REGULATOR VALVE CORE MSA has changed its recommended rebuilding of the MMR Second Stage Regulator s bypass sleeve assembly during the annual inspection. Based

More information

M3-LED. Operator s Manual. Operator s Manual MADE IN THE USA USA

M3-LED. Operator s Manual. Operator s Manual MADE IN THE USA USA M3-LED Operator s Manual Operator s Manual MADE IN THE USA USA Section TABLE OF CONTENTS Page Warnings and Cautions... 1 M3-LED Parts Diagram... 4 Mounting... 6 Dismounting... 8 Switch Operation... 10

More information

Usage Policies Notebook for Parylene Coating System

Usage Policies Notebook for Parylene Coating System Usage Policies Notebook for Parylene Coating System Revision date September 2014 2 Emergency Plan for Parylene Coating System Standard Operating Procedures for Emergencies Contact information Person Lab

More information

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012 JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012 UNIVERSITY OF TEXAS AT ARLINGTON Nanofabrication Research and Teaching Facility TABLE OF CONTENTS 1. Introduction....2 1.1 Scope of Work.....2

More information

Nanofabrication Facility: ECR Etcher SOP Rev. 01b, March 06. Standard Operating Procedure for PlasmaQuest ECR II Etching

Nanofabrication Facility: ECR Etcher SOP Rev. 01b, March 06. Standard Operating Procedure for PlasmaQuest ECR II Etching Standard Operating Procedure for PlasmaQuest ECR II Etching Authors: Rev. 00: Al Schmalz, Vighen Pacradouni and Jeff Young, December 21, 1998 Rev. 01: Dr. Andras G. Pattantyus-Abraham, May 24, 2005 Rev.

More information

University of MN, Minnesota Nano Center Standard Operating Procedure

University of MN, Minnesota Nano Center Standard Operating Procedure Equipment Name: Image Reversal Oven Badger name: ir-oven Revision #: 2 Model: YES 310 Revisionist: Paul Kimani Location: Bay 2 Date: October 29, 2013 1. Description The Yield Engineering Systems YES-310

More information

Misaligned Folds Paper Feed Problems Double Feeds Won t Feed FLYER Won t Run iii

Misaligned Folds Paper Feed Problems Double Feeds Won t Feed FLYER Won t Run iii Operator s Manual Table of Contents Operator Safety... 1 Introduction... 2 Unpacking and Setup... 3 Unpacking... 3 Setup... 4 FLYER Overview... 5 FLYER Diagram... 5 Capabilities... 5 Control Panel... 6

More information

Angstrom E-Beam Instructions. PROCESS CHECKS: the tooling factors of each metal; Ti/Au layer for wire bonding pull test.

Angstrom E-Beam Instructions. PROCESS CHECKS: the tooling factors of each metal; Ti/Au layer for wire bonding pull test. Angstrom E-Beam Instructions Tool Manager: Joe Palmer (jpalmer@princeton.edu) Office: 8-4706; Cell:609-751-1353 Backup: David Barth (dbarth@princeton.edu) Office: 8-4626; Cell: 610-405-8227 PROCESS CHECKS:

More information

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Sensitivity: 1x10 11 A m 2 Applied DC fields: 0 T to 5 T Applied AC fields: 0 G to 3 G (zero-to-peak), 0.01 Hz to 1000 Hz Temperatures

More information

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1 BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1 The front door of the main body of the instrument may be open when you arrive. Take the sample chamber and slide it into position with the

More information

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems Corporation Issued: 5/09 P/N: 600104 Rev. 2 Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems The Gas Addition Accessory permits the

More information

LEO SEM SOP Page 1 of 9 Revision 1.4 LEO 440 SEM SOP. Leica Leo Stereoscan 440i

LEO SEM SOP Page 1 of 9 Revision 1.4 LEO 440 SEM SOP. Leica Leo Stereoscan 440i LEO SEM SOP Page 1 of 9 LEO 440 SEM SOP Gun (Filament) Column Manual Valves Chamber Window Chamber Stage Movement Leica Leo Stereoscan 440i 1. Scope 1.1 This document provides the procedure for operating

More information

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure PPMS Service Note 1099-412 In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure Introduction: Loss of electricity to the PPMS EverCool II should not cause damage to

More information

Nordiko Metal Sputtering System Standard Operating Procedure

Nordiko Metal Sputtering System Standard Operating Procedure Nordiko Metal Sputtering System Standard Operating Procedure Specifications : Target Size Gases used in the system Base pressure Sputtering pressure Substrates used Substrate size : 2 inch or 4 inch :

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual Allwin21 AW610 RTP Page 1 of 18 Contents 1 Picture and Location 2 Process Capabilities 2.1 Cleanliness Standard 2.2 Recipes 2.3 Performance of Allwin21 AW610 RTP 3 Contact List

More information

1.1 Equipment: substrate, wafer tweezers, metal targets 1.2 Personal Protective Equipment: nitrile gloves, safety glasses 1.

1.1 Equipment: substrate, wafer tweezers, metal targets 1.2 Personal Protective Equipment: nitrile gloves, safety glasses 1. Nanomaster NSC-3000 DC Magnetron Sputter Tool Standard Operating Procedure Faculty Supervisor: Prof. Robert White, Mechanical Engineering (x72210) Safety Office: Peter Nowak x73246 (Just dial this directly

More information

Powder Measure. Care & Usage Instructions. Instructions # Product # Revision: A

Powder Measure. Care & Usage Instructions. Instructions # Product # Revision: A Powder Measure Care & Usage Instructions Instructions #1048578 Product #909292 Revision: A 1 INFORMATION AND SAFETY INSTRUCTIONS Ammunition reloading can be dangerous if done improperly and should not

More information

THERMAL EVAPORATION UNIT (for Al evaporation)

THERMAL EVAPORATION UNIT (for Al evaporation) THERMAL EVAPORATION UNIT (for Al evaporation) System Owner: NeerajPanwar 9619507210 panwar.iitr@gmail.com System Operator: 1. Bhimraj Sable 9930189878 bhimrajsable71@gmail.com Authorized User: 1. Ramesh

More information

3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS. Item #31289

3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS. Item #31289 3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS Item #31289 The EASTWOOD 3 GALLON, OILLESS PANCAKE COMPRESSOR, with an Integral Air Regulator, efficiently supplies all compressed air requirements for

More information

SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA

SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA Page 1 of 9 SSI Solaris 150 RTA The Solaris 150 RTA is a rapid thermal annealing system capable of handling sample sizes up to 100mm (4 diameter) or smaller. The system can anneal in N 2 and Forming gas

More information

Eco Snow CO2 Clean System Operation

Eco Snow CO2 Clean System Operation Eco Snow CO2 Clean System Operation This equipment is to be operated only by individuals that have been trained and qualified by the MNFC staff. Do NOT use or attempt to use this tool unless you are qualified

More information

The SPI Sputter Coater Handbook

The SPI Sputter Coater Handbook The SPI Sputter Coater Handbook Coating of Specimens SPI-Module Sputter Coater with Etch Mode 1. Mount the specimens onto the SEM stub. Keep in mind that many adhesives have high vapor pressure solvents

More information

Plasma Asher: March PX-500 User guide (May-30, 2017)

Plasma Asher: March PX-500 User guide (May-30, 2017) Plasma Asher: March PX-500 User guide (May-30, 2017) This is a highly versatile plasma etch tool that can etch using a direct plasma configuration (Oxygen plasma cleaner), a downstream plasma (Remote plasma),

More information

Operator: Save these instructions for future use!

Operator: Save these instructions for future use! WHITE-RODGERS 6C8-5 Combination Gas Control INSTALLATI INSTRUCTIS Operator: Save these instructions for future use FAILURE TO READ AND FOLLOW ALL INSTRUCTIS CAREFULLY BEFORE INSTALLING OR OPERATING THIS

More information

Cantilever Brake. Dealer's Manual. ROAD MTB Trekking. City Touring/ Comfort Bike

Cantilever Brake. Dealer's Manual. ROAD MTB Trekking. City Touring/ Comfort Bike (English) DM-RCBR001-00 Dealer's Manual ROAD MTB Trekking City Touring/ Comfort Bike URBAN SPORT E-BIKE Cantilever Brake BR-CX70 BR-CX50 BL-4700 BL-4600 BL-R780 BL-R3000 ST-7900 ST-6700 ST-5700 ST-4600

More information

OPERATING INSTRUCTIONS FOR THE MODEL 85B

OPERATING INSTRUCTIONS FOR THE MODEL 85B OPERATING INSTRUCTIONS FOR THE MODEL 85B SAFETY PRECAUTIONS FOR THE MODEL 85B System Under Pressure: Shut off air supply and disconnect air hose before disassembling or disconnecting parts. Flying Debris:

More information

[needs to be discussed with faculty if any new material to be done]

[needs to be discussed with faculty if any new material to be done] Tool Name: EVO 18 Version 1 Authorized Users: Gayatri, Vishnu kant Materials Used: Sample holder has facility to mount 9 stubs: Si, Glass, Materials allowed : [needs to be discussed with faculty if any

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual ARC12M Sputter Copyright 11.2015 by Hong Kong University of Science & Technology. All rights reserved. Page 1 Contents 1. Picture and Location 2. Process Capabilities 2.1 Cleanliness

More information

Installation Instructions

Installation Instructions Page 5750-S-1 Installation Instructions General These mounting instructions for Circular INCINO- PAK Burners are in addition to the specific instructions offered for other Maxon component items: Shut-Off

More information

FAULT CODE TROUBLESHOOTING INDEX

FAULT CODE TROUBLESHOOTING INDEX FAULT CODE TROUBLESHOOTING INDEX 1. Display indicates Change Filters 2. Display indicates Drip Tray Full Continuous Alarm will Sound 3. Display indicates Cold Fault 4. Display indicates Hot Fault 5. Display

More information

Unifilm Technology PVD-300 Sputter Deposition Operation Instructions

Unifilm Technology PVD-300 Sputter Deposition Operation Instructions Unifilm Technology PVD-300 Sputter Deposition Operation Instructions Contributors: Devin Brown, Kevin Klein, Ben King, Eric Woods Anything that is BOLD UNDERLINED ITALICS means that you should press that

More information

[USING THE VITROBOT April 9, Using the Vitrobot

[USING THE VITROBOT April 9, Using the Vitrobot Using the Vitrobot The Vitrobot Mk 3 (aka Mark III, Mark 3, etc.) can control both the local temperature and the local humidity during the process of plunge-freezing a grid for cryoem. It does require

More information

DM-MARD (English) Dealer's Manual. ROAD MTB Trekking. City Touring/ Comfort Bike REAR DERAILLEUR XTR RD-M9100 RD-M9120

DM-MARD (English) Dealer's Manual. ROAD MTB Trekking. City Touring/ Comfort Bike REAR DERAILLEUR XTR RD-M9100 RD-M9120 (English) DM-MARD001-00 Dealer's Manual ROAD MTB Trekking City Touring/ Comfort Bike URBAN SPORT E-BIKE REAR DERAILLEUR XTR RD-M9100 RD-M9120 CONTENTS CONTENTS...2 IMPORTANT NOTICE...3 TO ENSURE SAFETY...4

More information

Basic ICP Operating Procedures

Basic ICP Operating Procedures Center for High Technology Materials 2 February, 2009 University of New Mexico Created by Beth Fuchs Basic ICP Operating Procedures INTRODUCTION: The ICP is an inductively coupled plasma etching system,

More information

SPECIFICATIONS ATTENTION

SPECIFICATIONS ATTENTION VPS 504 S06 Installation Manual - P/N 80122 - Ed. 01/09 VPS 504 S06 and S05 Valve Proving System Installation Instructions VPS 1 6 Gases Natural gas, air and other inert gases. NOT suitable for butane

More information

CircuFlow Quick Setup Guide

CircuFlow Quick Setup Guide CircuFlow 5200 Quick Setup Guide LB04.0001 Rev B 20140314 TABLE OF CONTENTS Setup Therapy Setup Gradient Mode Pressure Mode Lock Device Unlock Device Treatment 03 04 05 08 11 12 13 02 SETUP 01 01 Open

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual Branson IPC 3000 O 2 Asher Copyright 2014 by Hong Kong University of Science & Technology. All rights reserved. Page 1 Contents 1. Picture and Location 2. Process Capabilities

More information

March Asher Operation

March Asher Operation March Asher Operation Roger Robbins 7/31/2006 The University of Texas at Dallas Erik Jonsson Engineering School of Engineering TITLE: March Asher Operation Page 1 of 13 March Asher Operation Roger Robbins

More information

Equipment Operating Procedure Glove Box

Equipment Operating Procedure Glove Box Equipment Operating Procedure Glove Box Page 1 0.0 Changing the Compressed Gas Cylinder 1. Complete Compressed Gas Cylinder training from EHS website before manually exchanging gas cylinders. In order

More information

OPERATION AND FUNCTION

OPERATION AND FUNCTION CHAPTER 3 OPERATION AND FUNCTION This chapter discusses the operation of the M249 automatic rifle. This includes loading, unloading, cycle of functioning, adjusting the sight, and using the bipod. 3-1.

More information

MoveRoll Conveyor Operating and Maintenance Manual

MoveRoll Conveyor Operating and Maintenance Manual MoveRoll Conveyor Operating and Maintenance Manual 1. Read this first! This manual contains information for protection of personnel in the roll handling area from possible injury and/or equipment damage.

More information

Powermax machine-side reference guide For mechanized applications with Powermax1000, Powermax1250 and Powermax1650

Powermax machine-side reference guide For mechanized applications with Powermax1000, Powermax1250 and Powermax1650 Powermax machine-side reference guide For mechanized applications with Powermax1000, Powermax1250 and Powermax1650 This Powermax machine-side reference guide is a supplement to your Operator Manual and

More information

NCTU NFC Lithography-Related Tools Operation. Manual

NCTU NFC Lithography-Related Tools Operation. Manual NCTU NFC Lithography-Related Tools Operation Manual A. General Procedures HMDS priming Resist coating Soft bake Exposure Development After development inspection (ADI) Hard bake Record sheet filling. B.

More information

1)! DO NOT PROCEED BEYOND THIS MARK

1)! DO NOT PROCEED BEYOND THIS MARK Operating Instructions for X-ray Photoelectron Spectrometer: Physical Electronics Model 555 XPS/AES (John H. Thomas, III, Ph.D., Electron Spectroscopy) Sample Insertion: figure 1. Sample insertion rod

More information

Title: Xactix XeF2 Etcher Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 03/23/2016

Title: Xactix XeF2 Etcher Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 03/23/2016 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Xactix XeF2 Etcher. All users are expected to have read and understood this document.

More information

User Manual for the Mars Calibration Bench

User Manual for the Mars Calibration Bench User Manual for the Mars Calibration Bench Fall 2013 Table of Contents Table of Contents Table of Contents... iii Introduction... v Chapter 1: The Mars Calibration Bench... 1 What Is the Mars Calibration

More information

RC 195 Receiver-Controller

RC 195 Receiver-Controller Document No. 129-082 RC 195 Receiver-Controller Product Description The POWERS RC 195 Receiver-Controller is a pneumatic instrument that receives one, two or three pneumatic inputs. It produces a pneumatic

More information

Teledyne Oil & Gas North Williamson Blvd. Daytona Beach, FL USA (phone) (fax)

Teledyne Oil & Gas North Williamson Blvd. Daytona Beach, FL USA (phone) (fax) PER CO # 33537 Approved By: M. Bartell Originator: Checked By: S. Jackson B. Cornelius 1 of 8 ODN-084-23 23 TITLE: Installation and Operation Manual for Dry- Teledyne Oil & Gas 1026 North Williamson Blvd.

More information

OXFORD PLASMALAB 80PLUS (CLOEY)

OXFORD PLASMALAB 80PLUS (CLOEY) Arizona State University NanoFab OXFORD PLASMALAB 80PLUS (CLOEY) Rev D Table of Contents Contents Table of Contents...1 1. Purpose / Scope...2 2. Reference Documents...2 3. Equipment / Supplies / Material...2

More information

User Manual GRI- 1500Li

User Manual GRI- 1500Li User Manual GRI- 1500Li Your Cart Tek caddy cart was thoroughly quality control checked and road tested before being shipped to your address. We do everything possible to assure that your caddy is in perfect

More information

DM-MBRD (English) Dealer's Manual. ROAD MTB Trekking. City Touring/ Comfort Bike. Rear Derailleur SLX RD-M7000 DEORE RD-M6000

DM-MBRD (English) Dealer's Manual. ROAD MTB Trekking. City Touring/ Comfort Bike. Rear Derailleur SLX RD-M7000 DEORE RD-M6000 (English) DM-MBRD001-04 Dealer's Manual ROAD MTB Trekking City Touring/ Comfort Bike URBAN SPORT E-BIKE Rear Derailleur SLX RD-M7000 DEORE RD-M6000 CONTENTS IMPORTANT NOTICE... 3 TO ENSURE SAFETY... 4

More information

Introduction. Part one: Identify the Hydraulic Trainer Components

Introduction. Part one: Identify the Hydraulic Trainer Components The University Of Jordan School of Engineering Mechatronics Engineering Department Fluid Power Engineering Lab Experiments No.4 Introduction to Hydraulic Trainer Objective: Students will be able to identify

More information