KARL SUSS MJB3 UV400 Mask Aligner Standard Operating Procedure

Size: px
Start display at page:

Download "KARL SUSS MJB3 UV400 Mask Aligner Standard Operating Procedure"

Transcription

1 KARL SUSS MJB3 UV400 Mask Aligner Standard Operating Procedure Version: 1.0 February 2014 UNIVERSITY OF TEXAS AT ARLINGTON Nanotechnology Research Center (NRC) 1

2 TABLE OF CONTENTS 1 Introduction Scope of Work Description Safety Precautions Requirements Training Hardware, Restrictions Operating Procedure Lamp Powering Up Procedure Mask and Wafer Loading Alignment and Setting Exposure Time Shutdown Procedure

3 1 INTRODUCTION 1.1 Scope These procedures apply to the KARL SUSS MJB3 UV400 mask aligner. All maintenance should follow the procedures set forth in the manufacturer s maintenance and operations manuals. This document is for reference only. Personnel should be trained by authorized staff before operating MJB3 UV400 mask aligner. 1.2 Description The SUSS MJB3 UV400 mask aligner is designed for high resolution photolithography in a laboratory environment. The MJB3 mask aligner offers exceptional flexibility in the processing of irregular shaped samples of different thickness such as sample wafer pieces and glass slides up to standard size 2 diameter wafers. The SUSS MJB3 UV400 is equipped with 350W mercury short-arc lamp providing primary exposure wavelengths of nm. The machines equipped with a high precision stage allowing alignment accuracies to 0.25 microns. The resolution of the MJB3 UV400 is 1-2 microns under optimum conditions. 1.3 Safety Precautions This equipment uses UV LIGHT SOURCE. Be very careful and aware of hazards that stem from the UV light source. The light emanating from the light source is a hazard to vision. DO NOT LOOK INTO THE OUTPUT BEAM of the UV Lamp or reflection of the beam. Permanent damage to the retina of the eye and subsequent blindness can result. If you encounter any hazards, contact NanoFAB staff immediately Prolonged exposure to diffused reflection from the output beam, illuminated surfaces in the beam, exposures of even a few seconds to the direct output beam or to the lamp itself, can cause ultraviolet skin burns or burns to the outer layers of the eye High voltage is used to run the UV lamp; do not attempt to disassemble supply 3

4 1.3.4 If the main lamp housing needs to be opened for service such and a stuck shutter the system that the light source is mounted on and the UV power supply should be powered down and locked out. Then the lamp N 2 cooling should be left on for 30 minutes to properly cool down the inside of the lamp housing before opening it for service. Call Nanofab staff for any maintenance issues This equipment uses a UV light source that is subjected to 20 to 30 atmospheres of internal pressure so there is always the possibility of a lamp exploding due to internal strains or physical abuse.. The lamp contains Mercury. Any vapor released by an exploded lamp would be highly dangerous. In the event of a lamp explosion, clear the area of personnel and contact the NanoFAB staff immediately Before changing the lamp, working on the light source, or performing any system maintenance requiring access to AC electrical power, Staff maintenance needs to turn OFF the power at the FACILITY AC POWER circuit breaker. After the power has been turned off, verify that the output voltage is zero Do not operate the system without proper room ventilation as some Deep UV lamps produce significant levels of OZONE Read any posted NanoFAB Engineering Change Notices (ECN) for any hardware, process or safety changes before running the tool. 4

5 2 REQUIREMENTS 2.1 Training You must be a qualified user on the SUSS MJB3 UV400 ALIGNER. 2.2 Hardware High resolution contact exposure mode printing down to 1-2 um by using primary wavelengths of 350 to 450nm from a 350W mercury short-arc lamp High precision X, Y, and θ alignment stage and microscope manipulator Wafer, glass slides and substrate handling up to dia.50 mm (wafers), 1 x1 (substrates) Special substrate chucks for pieces, III-V materials, thick substrates, hybrids Economical tool for laboratories and small series production. 2.3 Restrictions Glass slides, small sample pieces up to 2 diameter wafer sizes can be aligned and exposed. Small sample pieces need to be fixed to a 2 carrier wafer Only 4 x 4 x ( ) and 3 x 3 x ( ) masks can be used MJB3 Aligner is available 7 days a week / 24 hours a day. 3 OPERATING PROCEDURE 3.1 Lamp Powering Up Procedure/ Mask and Wafer Loading/ Alignment and Exposure The Hg-Arc lamp is peaked and calibrated to 3.0 mw/cm 2 with the UV probe Make sure the power to the aligner is OFF by checking the red Power button light is OFF. Turning ON the Illumination Controller when power to the aligner is ON, will damage the aligner. 5

6 3.1.3 Turn ON the lamp N 2 cooling switch to the up position. The N2 pressure should be 1.5 barr Turn on the green POWER switch on the Illumination Controller under the table by moving the lever to the up position then press the green Lamp button ( for nor more than 1 sec). The lamp should ignite. Press green channel button to toggle between sensor 1 and 2. Sensor 1 is 365nm sensor. Wait at least 10 minutes for the lamp to warm up, the display should read 275 W ( standby idle power on the lamp) Turn the Aligner air pressure switch to the up position. The air pressure should be 2.0 barr. Check the mask vacuum switch is in the UP position (ON), if not move this switch to the up position Turn on the power to the aligner by pressing the red power button. 6

7 3.1.7 If you hear an alarm ( buzzer) sounding press the Nitrogen Loss button to silence the alarm (it was previously pressed by mistake) Move the Contact Lever (1 st ) then the Separation Lever (2 nd ) to the separation position to before loading the mask or substrate. Contact Lever (1 st ) Separation Lever (2 nd ) Loosen the Mask Holder clamping knobs by rotating both a few turns CCW, then slide the Mask holder OUT and place upside down on the table. 7

8 Carefully place your clean mask with the chrome side facing up (glass side in contact with the mask holder) on the mask holder. Make sure the mask is centered and its sides are perpendicular with the mask holder edges. Check to see if the masks die orientation is the correct orientation you need. Then press the Mask Vacuum button to turn ON mask vacuum. Check to see if the mask is holding with your wafer tweezers by gently tapping the mask edge. If you find a mask vacuum problem call staff to check ( to avoid a broken mask) Carefully slide the mask + Mask Holder back into the mask guide rail and tightened the clamping knobs by turning them CW until they are finger tight Gently slide OUT the wafer holder tray and place you coated substrate on the chuck. Slowly slide the wafer holder tray and wafer all the way IN. 8

9 Move Separation Lever (1 st ) then the Contact Lever (2 nd ) to the Contact positions to remove any gap or wedge between the wafer and mask. The green contact light will go ON Separation Lever (1 st ) Contact Lever (2 nd t ) Green Contact light is ON To aligne the substrate to the mask move the separation lever to the Separation position creating a gap. The green contact light will go OFF Turn on the Halogen lamp power supply to the Microscope illuminator to about 4 volts (middle of the scale). ON: Rotate dial CW direction to about 4 volts Adjust the Microscope eyepiece for binocular vision, adjust the microscope manipulator for your field of view and use the focusing knobs to focus on the mask and die alignment markers. 9

10 Align your wafers alignment marks to the mask alignment marks by using X, Y, θ micrometers After the alignment is done move the Separation Lever back to the Contact position The green contact light should go ON. Contact Lever Green Contact light is ON The substrate is now ready for exposure. Set the exposure time on the timer located at the right of the front panel. Your exposure time depends on the exposure Dose (mj /cm2) for your resist type and thickness and Lamp intensity (Mw/cm2). The exposure time is Dose Intensity To set the exposure time 1 st select the time units multiple (seconds, minutes, hours) by rotating the inner black knob to align with the black arrow to the time units multiple you need. Align with the black arrow to the time units you need ( ie. X 10s ) 10

11 Next rotate the Outer plastic dial the time base ( 0 to 3 ) you need. ( i.e. 2 X 10s) = 20 second exposure The exposure time in now set. Press the blue Exposure button to expose the wafer. Turn away from the UV light, do not look at the UV light or you will damage your retinas permanently After the exposure is done move the Separation Lever (1 st ) then the Contact Lever (2 nd ) to the Seperation positions. The green contact light will go OFF Gently slide the wafer transport tray out and remove your wafer for further processing. 11

12 Loosen the Mask Holder clamping knobs by rotating both a few turns CCW, then slide the Mask holder OUT and place upside down on the table. Press the mask vacuum button to turn OFF mask vacuum and remove your mask After your mask is removed replace the mask holder and tightened the clamping knobs finger tight Turn OFF the microscope halogen illuminator power supply Turn OFF the Hg-Arc lamp power supply Turn OFF the MJB3 system power on the front panel. 12

13 Turn OFF the MJB3 air pressure by moving the switch to the down position Leave the mercury lamp N 2 cooling ON for 30 minutes to properly cool the lamp After 30 minutes turn OFF the N2 cooling by moving the switch to the down position Enter the required information in the logbook. 13

14 14

KARL SUSS MJB3 MASK ALIGNER STANDARD OPERATING PROCEDURE

KARL SUSS MJB3 MASK ALIGNER STANDARD OPERATING PROCEDURE KARL SUSS MJB3 MASK ALIGNER STANDARD OPERATING PROCEDURE Purpose of this Instrument: This instrument is for patterning photosensitive polymers with UV light. Location: White Hall 410 Cleanroom Primary

More information

SOP for Karl Suss MJB3 #1 Mask Aligner

SOP for Karl Suss MJB3 #1 Mask Aligner SOP for Karl Suss MJB3 #1 Mask Aligner Rev. 5 (30/11/2016) Safety UV Exposure: The high-energy light produced by the high-pressure Mercury Xenon lamp can cause eye damage and skin burns. Be sure that the

More information

MJB-3 Mask Aligner Property of TAU MNCF August 2012

MJB-3 Mask Aligner Property of TAU MNCF August 2012 MJB-3 Mask Aligner I. Power Up Sequence 1 LOG IN. Turn on Nitrogen & compressed air lines on the back wall. 2 Turn on the vacuum pump. 3 Flip the COMPRESSED AIR toggle on (up) Compressed air is used to

More information

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone:

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone: MJB4 Mask Aligner Operating Procedure Effective Date: 07/12/2012 Author(s): Jiong Hua Phone: 402-472-3773 Email: jhua2@unl.edu 1 1 Introduction 1.1 Key Words Karl Suss MJB4 Mask Aligner, Optical Lithography,

More information

Karl Suss MJB4 Mask Aligner

Karl Suss MJB4 Mask Aligner Karl Suss MJB4 Mask Aligner Tool Manager: Yong Sun ( yongs@princeton.edu; Office 8-8234; Cell 609-917-5076 ) Backup: George Watson ( gwatson@princeton.edu; Office 8-4626; Cell 732-996-2713 ) ******************************************************************************

More information

COBILT CA-800 Mask Aligner Equipment Operation

COBILT CA-800 Mask Aligner Equipment Operation COBILT CA-800 Mask Aligner Equipment Operation For the Micro-Electronics Laboratory At University of Notre Dame Department of Electrical Engineering This user manual is not be removed from room 247A. This

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual AB-M Mask Aligner Version 1.1 Page 1 of 18 Contents 1. Picture and Location 2. Process Capabilities 2.1 Cleanliness Standard 2.2 Wafer Chuck Selection 2.3 Mask Holder Selection

More information

Karl Suss MA6 Mask Aligner SOP

Karl Suss MA6 Mask Aligner SOP Page 1 of 11 Karl Suss MA6 Mask Aligner SOP Safety UV Exposure: The high energy light produced by the high pressure Mercury Xenon lamp can cause eye damage and skin burns. Be sure that the light guards

More information

Usage Policies Notebook for Karl Suss MA6 Mid / Deep UV Mask Aligner

Usage Policies Notebook for Karl Suss MA6 Mid / Deep UV Mask Aligner Usage Policies Notebook for Karl Suss MA6 Mid / Deep UV Mask Aligner Revision date September 2014 2 Emergency Plan for Karl Suss MA6 Aligner Standard Operating Procedures for Emergencies Contact information

More information

Warnings: Notes: Revised: January 8,

Warnings: Notes: Revised: January 8, OAI Model 204IR Mask Aligner Standard Operating Procedure Faculty Supervisor: Prof. Robert White, Mechanical Engineering (x72210) Safety Office: Peter Nowak x73246 (Just dial this directly on any campus

More information

Equipment Standard Operating Procedure Greg Allion and Kimberly Appel

Equipment Standard Operating Procedure Greg Allion and Kimberly Appel Date Created: May 3, 2004 Date Modified: June 1, 2005 MA6/BA6 Mask Aligner Equipment Standard Operating Procedure Greg Allion and Kimberly Appel 1. Purpose 1.1. Photolithography involves transferring a

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual Karl Suss MA6 Mask Aligner Version 1.1 Page 1 of 24 Contents 1. Picture and Location 2. Process Capabilities 2.1 Cleanliness Standard 2.2 Substrate Size 2.3 Photo Mask Size 2.4

More information

Warnings: Notes: Revised: October 5, 2015

Warnings: Notes: Revised: October 5, 2015 Karl Suss MA6 Mask Aligner Standard Operating Procedure Faculty Supervisor: Prof. Robert White, Mechanical Engineering (x72210) Safety Office: Peter Nowak x73246 (Just dial this directly on any campus

More information

University of Minnesota Nano Fabrication Center

University of Minnesota Nano Fabrication Center Equipment name: Karl Suss Mask Aligners Badger name: MA6 & Maba6 Revision number: 4 Models: Revisionist: Paul Kimani Location: Bay 2 Date: April 25, 2013 1. Introduction The Karl Suss MA6 and MABA6 are

More information

ABM MASK ALIGNERS. NanoFab 26 March 2009 A Micro Machining & Nanofabrication Facility

ABM MASK ALIGNERS. NanoFab 26 March 2009 A Micro Machining & Nanofabrication Facility ABM MASK ALIGNERS LOCATION: Optical Lithography PRIMARY TRAINER: Stephanie Bozic (2-6724, sbozic@ualberta.ca) SECONDARY TRAINER: Jolene Chorzempa (2-4823, jolenec@ualberta.ca) 1. OVERVIEW The ABM Mask

More information

University of Minnesota Nano Fabrication Center

University of Minnesota Nano Fabrication Center Equipment name: Karl Suss Mask Aligners Badger name: MA6-P Revision number: 0 Model: MA6 Revisionist: Paul Kimani Location: Bay 4 PAN Date: September 17, 2015 1. Introduction The Karl Suss MA6-PAN is a

More information

Quintel Mask Aligner

Quintel Mask Aligner Quintel Mask Aligner (quintel) 1.0 1.0 Title Quintel Q4000 Mask Aligner 2.0 2.0 Purpose The Quintel Q4000 MA (Mask Aligner) is a top and bottom side contact lithography printer with the video-view split

More information

Approved by Principal Investigator Date: Approved by Super User: Date:

Approved by Principal Investigator Date: Approved by Super User: Date: Approved by Principal Investigator Date: Approved by Super User: Date: Standard Operating Procedure BNC OAI 200 Lithographic Mask Aligner (Aligner 3) Version 2011 June 2 I. Purpose This Standard Operating

More information

OAI Model 200 Tabletop Mask Aligner Portland State University

OAI Model 200 Tabletop Mask Aligner Portland State University OAI Model 200 Tabletop Mask Aligner Portland State University WARNING: This machine exposes users to ultraviolet radiation. Do not touch the lens underneath the lamp hood as it may damage the machine and

More information

Operation of the mask aligner MJB-55

Operation of the mask aligner MJB-55 John Paul Adrian Glaubitz Operation of the mask aligner MJB-55 Department of Physics Faculty of Mathematics and Natural Sciences University of Oslo 1 Introduction The mask aligner is an essential tool

More information

Approved by Principal Investigator Date: Approved by Super User: Date:

Approved by Principal Investigator Date: Approved by Super User: Date: Approved by Principal Investigator Date: Approved by Super User: Date: Standard Operating Procedure BNC OAI Lithographic Mask Aligner (Aligner 2) Version 2008 October 31 I. Purpose This Standard Operating

More information

Operation of the contact mask aligner Canon PPC 210 Projection Print Camera

Operation of the contact mask aligner Canon PPC 210 Projection Print Camera Operation of the contact mask aligner Canon PPC 210 Projection Print Camera Start-up Procedure: 1. Start the grey pump in the service corridor behind the aligner 2. Open the valve for the Nitrogen gas

More information

Karl Suss Contact Aligner Operation

Karl Suss Contact Aligner Operation Karl Suss Contact Aligner Operation Roger Robbins 6/31/2008 2 nd ed. 3/12/2009 The University of Texas at Dallas Erik Jonsson Engineering School of Engineering TITLE: Karl Suss Contact Aligner Operation

More information

Photolithography. Operating Instructions

Photolithography. Operating Instructions Photolithography Operating Instructions The PR used during this laboratory session will be Microposit S1813 (from Shipley). Make sure everyone is following the laboratory protocol. Wear lab coats, safety

More information

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012 JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012 UNIVERSITY OF TEXAS AT ARLINGTON Nanofabrication Research and Teaching Facility TABLE OF CONTENTS 1. Introduction....2 1.1 Scope of Work.....2

More information

icreasepro Creaser Operators Manual

icreasepro Creaser Operators Manual 6-2013 Version 3.0 icreasepro Creaser Operators Manual WWW.MBMCORP.COM 800-223-2508 TABLE OF CONTENTS SPECIFICATIONS.1a SAFETY PROCEDURES/CARE & MAINTENANCE..1b COMPONENT IDENTIFICATION 2 TOUCH SCREEN

More information

7130 Lancer Rear Drive Magnetic Commercial Indoor Cycling Bike

7130 Lancer Rear Drive Magnetic Commercial Indoor Cycling Bike 7130 Lancer Rear Drive Magnetic Commercial Indoor Cycling Bike Owner s Manual Made in Taiwan INDEX IMPORTANT SAFETY INFORMATION... 1 EXPLODED DRAWING... 2 PARTS LIST... 3 ASSEMBLY INSTRUCTION... 4-9 USER

More information

Usage Policies Notebook for Parylene Coating System

Usage Policies Notebook for Parylene Coating System Usage Policies Notebook for Parylene Coating System Revision date September 2014 2 Emergency Plan for Parylene Coating System Standard Operating Procedures for Emergencies Contact information Person Lab

More information

Issue: H Title: CHA E-Beam Evaporator Page 1 of 7. Table of Contents

Issue: H Title: CHA E-Beam Evaporator Page 1 of 7. Table of Contents Title: CHA E-Beam Evaporator Page 1 of 7 Table of Contents Purpose/Scope... 2 2.0 Reference Documents... 2 3.0 Equipment/Supplies/Material... 2 4.0 Safety... 2 5.0 Set Up Procedures... 2 5.1 PC Logon and

More information

OPERATION OF THE DIMPLER

OPERATION OF THE DIMPLER OPERATION OF THE DIMPLER After thinning your sample to ~80 μm you can now do a dimpling process to thin the center up to 10 μm. When you walk in and use the DIMPLER it should already be calibrated and

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual Denton Explorer 14 RF/DC Sputter Version 1.0 Page 1 of 11 Contents 1. Picture and Location 2. Process Capabilities 1. Cleanliness Standard 2. Available for Sputtering Materials

More information

NORDSON MARCH PX-1000 PLASMA ASHER STANDARD OPERATING PROCEDURE Version: 1.0 July 2016

NORDSON MARCH PX-1000 PLASMA ASHER STANDARD OPERATING PROCEDURE Version: 1.0 July 2016 NORDSON MARCH PX-1000 PLASMA ASHER STANDARD OPERATING PROCEDURE Version: 1.0 July 2016 UNIVERSITY OF TEXAS AT ARLINGTON Nanotechnology Research Center TABLE OF CONTENTS 1. Introduction..3 1.1 Scope of

More information

Cryo-Evaporator Operation

Cryo-Evaporator Operation Cryo-Evaporator Operation Cara Ricci November 25, 2003 Thin Film Deposition THE UNIVERSITY OF TEXAS AT DALLAS ERIK JOHNSON SCHOOL OF ENGINEERING DOCUMENT NUMBER: FA2003-TF-008 EDITION: 1.1 PAGE: 1 of 28

More information

SPUTTER STATION STANDARD OPERATING PROCEDURE

SPUTTER STATION STANDARD OPERATING PROCEDURE SPUTTER STATION STANDARD OPERATING PROCEDURE Purpose of this Instrument: This instrument is used for deposition of thin metal or oxide films. Source materials supplied by WVU Shared Research Facilities:

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual Fisher Scientific Isotemp TM Model 281A Vacuum Oven Version 1.1 Page 1 of 9 Contents 1. Picture and Location 2. Process Capabilities 2.1 Cleanliness Standard 2.2 Substrate Size

More information

5.1.3 Mechanical Hazards Drive assemblies have sufficient power to cause injury. Keep hands, fingers, clothing and tools clear of moving parts.

5.1.3 Mechanical Hazards Drive assemblies have sufficient power to cause injury. Keep hands, fingers, clothing and tools clear of moving parts. Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the PE4400. All users are expected to have read and understood this document. It is

More information

Standard Operating Procedure. For. PVD E-Beam

Standard Operating Procedure. For. PVD E-Beam P a g e 1 Standard Operating Procedure For PVD E-Beam P a g e 2 Introduction The PVD Electron-Beam Evaporator (E-Beam) thin film deposition machine uses a magnetically guided and collimated stream of electrons

More information

Model 130M Pneumatic Controller

Model 130M Pneumatic Controller Instruction MI 017-450 May 1978 Model 130M Pneumatic Controller Installation and Operation Manual Control Unit Controller Model 130M Controller is a pneumatic, shelf-mounted instrument with a separate

More information

Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System)

Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System) Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System) Revision date October 2014 2 Emergency Plan for Nano Furnace Standard Operating Procedures for Emergencies Contact information Person

More information

THERMAL EVAPORATION UNIT (for Al evaporation)

THERMAL EVAPORATION UNIT (for Al evaporation) THERMAL EVAPORATION UNIT (for Al evaporation) System Owner: NeerajPanwar 9619507210 panwar.iitr@gmail.com System Operator: 1. Bhimraj Sable 9930189878 bhimrajsable71@gmail.com Authorized User: 1. Ramesh

More information

Unaxis ICP/RIE SOP Revision 8 09/30/16 Page 1 of 5. NRF Unaxis ICP/RIE Etch SOP

Unaxis ICP/RIE SOP Revision 8 09/30/16 Page 1 of 5. NRF Unaxis ICP/RIE Etch SOP Page 1 of 5 NRF Unaxis ICP/RIE Etch SOP Unaxis Shuttlelock Reactive Ion Etcher with Inductively Coupled Plasma Module. Etch Capabilities: SiO2, Si3N4, Al, dielectrics and other commonly used materials.

More information

MINIBRUTE ANNEAL TUBE STANDARD OPERATION PROCEDURE

MINIBRUTE ANNEAL TUBE STANDARD OPERATION PROCEDURE Arizona State University NanoFab MINIBRUTE ANNEAL TUBE STANDARD OPERATION PROCEDURE Rev A Title: MINIBRUTE ANNEAL TUBE STANDARD OPERATION PROCEDURE Table of Contents Issue: Rev A Contents Table of Contents...1

More information

Revised: June 7, 2017

Revised: June 7, 2017 LC Technologies Thermal Evaporator Standard Operating Procedure Faculty Supervisor: Prof. Robert White, Mechanical Engineering (x72210) Safety Office: Peter Nowak x73246 (Just dial this directly on any

More information

Misaligned Folds Paper Feed Problems Double Feeds Won t Feed FLYER Won t Run iii

Misaligned Folds Paper Feed Problems Double Feeds Won t Feed FLYER Won t Run iii Operator s Manual Table of Contents Operator Safety... 1 Introduction... 2 Unpacking and Setup... 3 Unpacking... 3 Setup... 4 FLYER Overview... 5 FLYER Diagram... 5 Capabilities... 5 Control Panel... 6

More information

Unaxis PECVD. SiH4 (5% in He)

Unaxis PECVD. SiH4 (5% in He) Unaxis PECVD Table of Contents: I: Introduction II: Machine Specifications III: System Components IV: Deposited Materials and Precursor Gases V: Operating Instructions VI: Creating a Recipe VII: Troubleshooting

More information

OPERATION OF LIQUID PARTICLE COUNTER. GP-B P0045 Rev -

OPERATION OF LIQUID PARTICLE COUNTER. GP-B P0045 Rev - 1 W. W. Hansen Experimental Physics Laboratory STANFORD UNIVERSITY STANFORD, CALIFORNIA 94305-4085 Gravity Probe B Relativity Mission OPERATION OF LIQUID PARTICLE COUNTER GP-B P0045 Rev - May 23, 1997

More information

Angstrom E-Beam Instructions. PROCESS CHECKS: the tooling factors of each metal; Ti/Au layer for wire bonding pull test.

Angstrom E-Beam Instructions. PROCESS CHECKS: the tooling factors of each metal; Ti/Au layer for wire bonding pull test. Angstrom E-Beam Instructions Tool Manager: Joe Palmer (jpalmer@princeton.edu) Office: 8-4706; Cell:609-751-1353 Backup: David Barth (dbarth@princeton.edu) Office: 8-4626; Cell: 610-405-8227 PROCESS CHECKS:

More information

Standard Operating Procedure #COE-SOP-0001 Chemical Fume Hood Operation

Standard Operating Procedure #COE-SOP-0001 Chemical Fume Hood Operation Standard Operating Procedure # Chemical Fume Hood Operation Facility: NMSU College of Engineering Laboratories Written by: Juanita Miller, Safety Specialist, miljgh@nmsu.edu (575)-646-1292 Scope: This

More information

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road PROPORTIONING VALVE Model 150 INSTRUCTION MANUAL March 2017 IMS Company 10373 Stafford Road Telephone: (440) 543-1615 Fax: (440) 543-1069 Email: sales@imscompany.com 1 Introduction IMS Company reserves

More information

The SPI Sputter Coater Handbook

The SPI Sputter Coater Handbook The SPI Sputter Coater Handbook Coating of Specimens SPI-Module Sputter Coater with Etch Mode 1. Mount the specimens onto the SEM stub. Keep in mind that many adhesives have high vapor pressure solvents

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual Allwin21 AW610 RTP Page 1 of 18 Contents 1 Picture and Location 2 Process Capabilities 2.1 Cleanliness Standard 2.2 Recipes 2.3 Performance of Allwin21 AW610 RTP 3 Contact List

More information

Instruction Manual Otoflash G171 UV-flash-device for Light Curing

Instruction Manual Otoflash G171 UV-flash-device for Light Curing Instruction Manual Otoflash G171 UV-flash-device for Light Curing Index 1) General Information 2) Initial Operation of Device 3) Important Notes 4) Use with timer 5) Exchanging the flash module 6) Maintenance,

More information

Squeegee Unit for MRS Multi-Recovery System

Squeegee Unit for MRS Multi-Recovery System En Operating Instructions Squeegee Unit for MRS Multi-Recovery System Translation of the original operating instructions Squeegee - MRS Table of Contents Safety Recommendations Technical Data - MRS Booth

More information

GV Standard X-Vent. Setup, Commissioning & Installation Guide

GV Standard X-Vent. Setup, Commissioning & Installation Guide GV Standard X-Vent Setup, Commissioning & Installation Guide Technical experts in the design, manufacture and supply of precision engineered, architectural rooflights for residential and commercial buildings.

More information

Approved by Principal Investigator Date: Approved by Super User: Date:

Approved by Principal Investigator Date: Approved by Super User: Date: Approved by Principal Investigator Date: Approved by Super User: Date: Standard Operating Procedure BNC Commonwealth Dual Ion Beam Deposition System (CDIBS) Version 2010 February 14 I. Purpose This Standard

More information

March CS-1701F Reactive Ion Etcher

March CS-1701F Reactive Ion Etcher March CS-1701F Reactive Ion Etcher Standard Operating Procedure Faculty Supervisor: Prof. Robert White, Mechanical Engineering (x72210) Safety Office: Peter Nowak x73246 (Just dial this directly on any

More information

NRF Suss Delta 80 Spinner SOP Revision /14/2016 Page 1 of 11. Suss Delta 80 Spinner SOP

NRF Suss Delta 80 Spinner SOP Revision /14/2016 Page 1 of 11. Suss Delta 80 Spinner SOP Page 1 of 11 Note: latest updates are blue. Table of Contents Suss Delta 80 Spinner SOP 1.0 Safety 2.0 Quality Controls and Calibration 3.0 Equipment Uses and Restrictions 4.0 Equipment Specifications

More information

MoveRoll Conveyor Operating and Maintenance Manual

MoveRoll Conveyor Operating and Maintenance Manual MoveRoll Conveyor Operating and Maintenance Manual 1. Read this first! This manual contains information for protection of personnel in the roll handling area from possible injury and/or equipment damage.

More information

MAGNETIC CYCLING TRAINER SF-B0419 USER MANUAL

MAGNETIC CYCLING TRAINER SF-B0419 USER MANUAL MAGNETIC CYCLING TRAINER SF-B049 USER MANUAL IMPORTANT: Read all instructions carefully before using this product. Retain owner s manual for future reference. For customer service, please contact: support@sunnyhealthfitness.com

More information

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT ATTENTION SYRINGE REPLACEMENT Please read through the document completely before starting any repairs. Refer to the proper section in the service manual for complete removal and replacement procedures.

More information

[needs to be discussed with faculty if any new material to be done]

[needs to be discussed with faculty if any new material to be done] Tool Name: EVO 18 Version 1 Authorized Users: Gayatri, Vishnu kant Materials Used: Sample holder has facility to mount 9 stubs: Si, Glass, Materials allowed : [needs to be discussed with faculty if any

More information

WARNING NOTES TABLE OF CONTENTS

WARNING NOTES TABLE OF CONTENTS NOTES TABLE OF CONTENTS 1.0 Introduction.................................................. 2 1.1 Unpacking 1.2 Complete Kit 2.0 Operating Procedures.......................................... 3 2.1 Memory

More information

EE 432 Lab 3 PMOS source/drain lithography and diffusion

EE 432 Lab 3 PMOS source/drain lithography and diffusion EE 432 Lab 3 PMOS source/drain lithography and diffusion Group Leader: Yue Zhang Group Numbers: Yueyi Jiao, Yin Huang, Lafit Masud TA: Andy Hoyt Section: NO. 5 Introduction: In this lab, we will perform

More information

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Sensitivity: 1x10 11 A m 2 Applied DC fields: 0 T to 5 T Applied AC fields: 0 G to 3 G (zero-to-peak), 0.01 Hz to 1000 Hz Temperatures

More information

COMELEC C-30-S Parylene Coating System

COMELEC C-30-S Parylene Coating System COMELEC C-30-S Parylene Coating System Comelec C-30-S Parylene deposition system Introduction Parylene is a polymere deposited at room temperature in a vacuum chamber (few µb). Parylene coating is perfectly

More information

Unifilm Technology PVD-300 Sputter Deposition Operation Instructions

Unifilm Technology PVD-300 Sputter Deposition Operation Instructions Unifilm Technology PVD-300 Sputter Deposition Operation Instructions Contributors: Devin Brown, Kevin Klein, Ben King, Eric Woods Anything that is BOLD UNDERLINED ITALICS means that you should press that

More information

Usage Policies Notebook for Quintel 2001 Mask Aligner

Usage Policies Notebook for Quintel 2001 Mask Aligner Usage Policies Notebook for Quintel 2001 Mask Aligner Revision date September 2014 2 Emergency Plan for Quintel 2001 Mask Aligner Standard Operating Procedures for Emergencies Contact information Person

More information

VAPOMATIC ANESTHETIC VAPORIZER

VAPOMATIC ANESTHETIC VAPORIZER VAPOMATIC ANESTHETIC VAPORIZER INSTRUCTION MANUAL SOLD FOR VETERINARY USE ONLY BY: A.M. BICKFORD, INC. 12318 BIG TREE ROAD WALES CENTER, NY 14169 1-800-795-3062 Fax: 716-652-2046 www.ambickford.com sales@ambickford.com

More information

Safety Tips for Using Germicidal Lamps

Safety Tips for Using Germicidal Lamps Safety Tips for Using Germicidal Lamps What Are Germicidal Lamps? Germicidal lamps emit radiation in the UV-C portion of the ultraviolet (UV) spectrum, which includes wavelengths between 100 and 280 nanometers

More information

Product Description TUV PL-L

Product Description TUV PL-L Product Description TUV PL-L TUV PL-L lamps are compact UVC (germicidal) lamps used in residential water and air disinfection units. The compact size of the lamp allows for a small system design and design

More information

User Manual. Heads-Up Display (HUD) DiveCAN. Mechanical Button Version

User Manual. Heads-Up Display (HUD) DiveCAN. Mechanical Button Version User Manual Heads-Up Display (HUD) Mechanical Button Version DiveCAN Table of Contents 1. Introduction...4 1.1 Features...4 2. Physical Description...5 3. Reading the PPO2...6 3.1 Modified Smither s Code...7

More information

602 STRINGING MACHINE OWNER'S MANUAL

602 STRINGING MACHINE OWNER'S MANUAL PROGRESSION 602 STRINGING MACHINE OWNER'S MANUAL AL Issue 1- April 2000 Copyright 2000 GAMMA Sports - All Rights Reserved PROGRESSION 602 STRINGING MACHINE TABLE OF CONTENTS PAGE 1... WARRANTY PAGE 2...

More information

Arizona State University Center for Solid State Electronics Research Issue: E Title: Heat Pulse 610 Operating Procedure Page 1 of 7

Arizona State University Center for Solid State Electronics Research Issue: E Title: Heat Pulse 610 Operating Procedure Page 1 of 7 Title: Heat Pulse 610 Operating Procedure Page 1 of 7 Table of Contents 1.0 Purpose/Scope... 2 2.0 Reference Documents... 2 2.1 Heat Pulse Operations Manual... 2 2.2 Solaris software Operations Manual...

More information

Operating Instructions and Parts Manual 24529

Operating Instructions and Parts Manual 24529 Please read and save these instructions. Read through this owner s manual carefully before using product. Protect yourself and others by observing all safety information, warnings, and cautions. Failure

More information

Version 1.0. icrease pro + Serial Number. Date

Version 1.0. icrease pro + Serial Number. Date icrease pro + 6-2016 Version 1.0 Serial Number Date This page is intentionally blank CONTENTS SPECIFICATIONS. 1a SAFETY PROCEDURES/CARE & MAINTENANCE..... 1b COMPONENT IDENTIFICATION.... 2 TOUCH SCREEN

More information

OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE

OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE 1 These instructions generally describe the installation, operation, and maintenance of subject equipment. The manufacturer reserves the right

More information

O P E R A T O R S M A N U A O P E R A T O R S M A N U A L L

O P E R A T O R S M A N U A O P E R A T O R S M A N U A L L Clam Basic Series O P E R A T O R S M A N U A O P E R A T O R S M A N U A L L Safety Instructions When using your heat press, basic precautions should always be followed, including the following: 1. 2.

More information

1. Features CONTENTS. English

1. Features CONTENTS. English 2 CONTENTS 1. Features... 3 2. Names of Components... 4 3. Before Using... 6 4. Setting the Reference Position... 8 5. Confirming the Time with the Electronic Tone (Minute Repeater)... 11 6. Setting the

More information

LSX Low Smoke Convertor Operations Manual

LSX Low Smoke Convertor Operations Manual LSX Low Smoke Convertor Operations Manual V.05.25.05 Table of Contents Page 1 Page 2 Page 3 Page 4 Page 5 Page 5 Page 6 Page 7 Page 8 Introduction Warning Technical Specifications Operating Procedure Control

More information

Spectrum AquaLight (SAL ) Color Underwater Light Owner s Manual

Spectrum AquaLight (SAL ) Color Underwater Light Owner s Manual Spectrum AquaLight (SAL ) Color Underwater Light Owner s Manual IMPORTANT SAFETY INSTRUCTIONS READ AND FOLLOW ALL INSTRUCTIONS SAVE THESE INSTRUCTIONS Table of Contents Section I. Installing Light Fixture

More information

User Manual GRI- 1500Li

User Manual GRI- 1500Li User Manual GRI- 1500Li Your Cart Tek caddy cart was thoroughly quality control checked and road tested before being shipped to your address. We do everything possible to assure that your caddy is in perfect

More information

RIPPLE TANK - with rippler & kit

RIPPLE TANK - with rippler & kit GENERAL DESCRIPTION: RIPPLE TANK - with rippler & kit Cat: SW3430-001 with illuminator, rippler & kit. The ripple tank is used to investigate wave motion in a shallow trough of water to understand how

More information

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars.

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars. Simply a question of better measurement SCHMIDT Flow Sensor SS 20.261 The cost-effective alternative in pressurised systems up to 10 bars. Compressed air technology Industrial processes A cost analysis

More information

Side-of-Pole Mount for 1 Module (SPM1) For Module Types A & B

Side-of-Pole Mount for 1 Module (SPM1) For Module Types A & B Side-of-Pole Mount for 1 Module (SPM1) For Module Types A & B ASSEMBLY INSTRUCTIONS step-by-step assembly and installation Version 1, Rev A PCN 080311-2 SP3363-1 Side-of-Pole Mount for 1 Module (SPM1)

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual ARC12M Sputter Copyright 11.2015 by Hong Kong University of Science & Technology. All rights reserved. Page 1 Contents 1. Picture and Location 2. Process Capabilities 2.1 Cleanliness

More information

1.1 Equipment: substrate, wafer tweezers, metal targets 1.2 Personal Protective Equipment: nitrile gloves, safety glasses 1.

1.1 Equipment: substrate, wafer tweezers, metal targets 1.2 Personal Protective Equipment: nitrile gloves, safety glasses 1. Nanomaster NSC-3000 DC Magnetron Sputter Tool Standard Operating Procedure Faculty Supervisor: Prof. Robert White, Mechanical Engineering (x72210) Safety Office: Peter Nowak x73246 (Just dial this directly

More information

Nanofabrication Facility: ECR Etcher SOP Rev. 01b, March 06. Standard Operating Procedure for PlasmaQuest ECR II Etching

Nanofabrication Facility: ECR Etcher SOP Rev. 01b, March 06. Standard Operating Procedure for PlasmaQuest ECR II Etching Standard Operating Procedure for PlasmaQuest ECR II Etching Authors: Rev. 00: Al Schmalz, Vighen Pacradouni and Jeff Young, December 21, 1998 Rev. 01: Dr. Andras G. Pattantyus-Abraham, May 24, 2005 Rev.

More information

1)! DO NOT PROCEED BEYOND THIS MARK

1)! DO NOT PROCEED BEYOND THIS MARK Operating Instructions for X-ray Photoelectron Spectrometer: Physical Electronics Model 555 XPS/AES (John H. Thomas, III, Ph.D., Electron Spectroscopy) Sample Insertion: figure 1. Sample insertion rod

More information

OxyCheq Expedition-X Oxygen Analyzer. Operator s Manual

OxyCheq Expedition-X Oxygen Analyzer. Operator s Manual OxyCheq Expedition-X Oxygen Analyzer Operator s Manual OxyCheq 3528 Russell Road Marianna, Florida 32446 USA Tel: 850.482.0385 Email: oxy@oxycheq.com Web: http://oxycheq.com CONTENTS 1.0 Introduction..

More information

M3-LED. Operator s Manual. Operator s Manual MADE IN THE USA USA

M3-LED. Operator s Manual. Operator s Manual MADE IN THE USA USA M3-LED Operator s Manual Operator s Manual MADE IN THE USA USA Section TABLE OF CONTENTS Page Warnings and Cautions... 1 M3-LED Parts Diagram... 4 Mounting... 6 Dismounting... 8 Switch Operation... 10

More information

600 / 600FC OWNER'S MANUAL

600 / 600FC OWNER'S MANUAL PROGRESSION 600 / 600FC OWNER'S MANUAL Issue 2 / Version E - Dec. 10, 1997 Copyright 1997 GAMMA Sports - All Rights Reserved PROGRESSION 600 / 600FC OWNER'S MANUAL TABLE OF CONTENTS PAGE 1... WARRANTY

More information

INSTRUCTION MANUAL. January 23, 2003, Revision 0

INSTRUCTION MANUAL. January 23, 2003, Revision 0 INSTRUCTION MANUAL Model 810A In-Vitro Test Apparatus for 310B Muscle Lever January 23, 2003, Revision 0 Copyright 2003 Aurora Scientific Inc. Aurora Scientific Inc. 360 Industrial Parkway S., Unit 4 Aurora,

More information

Gas Check. Short testing time, with automatic start-up if no leakage was detected from previous operation.

Gas Check. Short testing time, with automatic start-up if no leakage was detected from previous operation. Quality Products at Competitive Prices Gas Check Automatic testing of the down-stream installation to check that all valves are closed and that pipework and fittings are gas tight before each system start-up.

More information

O P E R ATING INSTRUCTIONS FOR MODEL SPR-45 Automatic Screen and Stencil Printer

O P E R ATING INSTRUCTIONS FOR MODEL SPR-45 Automatic Screen and Stencil Printer O P E R ATING INSTRUCTIONS FOR MODEL SPR-45 Automatic Screen and Stencil Printer TABLE OF CONTENTS I. SPECIFICATIONS...3. II. SAFETY INSTRUCTIONS...4. III. INSTALLATION...5. IV. SET-UP...6. V. SYSTEM OPERATION...9.

More information

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y INSTALLATION and INSTRUCTION MANUAL for QUALITY AIR BREATHING SYSTEMS Model 50 Systems Outfitted with ABM-725 Monitor M A R T E C H S E R V I C E S C O M P A N Y OFFICE: (507) 843-4700 P.O. BOX 7079 Toll

More information

WELCOME TO THE REVOLUTION

WELCOME TO THE REVOLUTION USER GUIDE WELCOME TO THE REVOLUTION THANK YOU FOR CHOOSING THE GCQUAD We listened to what you wanted - and created the most accurate, versatile and game-enhancing ball and club analysis solution available

More information

INSTRUCTION MANUAL TIPPING BUCKET RAIN GAUGE. MODEL TB3/0.1mm

INSTRUCTION MANUAL TIPPING BUCKET RAIN GAUGE. MODEL TB3/0.1mm INSTRUCTION MANUAL TIPPING BUCKET RAIN GAUGE MODEL TB3/0.1mm QUALITY SYSTEM ISO 9001 2008 CERTIFIED HYQUEST SOLUTIONS PTY LTD PO BOX 332, LIVERPOOL B.C NSW 1871, AUSTRALIA Phone:(Int.) 612 9601 2022 Fax:

More information

QUARTZ HOUSING COATING PROCEDURE

QUARTZ HOUSING COATING PROCEDURE W. W. Hansen Experimental Physics Laboratory STANFORD UNIVERSITY STANFORD, CALIFORNIA 94305-4085 Gravity Probe B Relativity Mission QUARTZ HOUSING COATING PROCEDURE GP-B P0044 Rev- November 18, 1997 Prepared

More information

OEM Manual. MODEL ½ Digit DRUM SCALE

OEM Manual. MODEL ½ Digit DRUM SCALE OEM Manual MODEL 4020-3 ½ Digit DRUM SCALE Scaletron Industries, Ltd. Bedminster Industrial Park 53 Apple Tree Lane P.O. Box 365 Plumsteadville, PA 18949 USA Toll Free: 1-800-257-5911 (USA & Canada) Phone:

More information

DIP STATION WITH SAFETY CONNECTOR SF-BH6507 USER MANUAL

DIP STATION WITH SAFETY CONNECTOR SF-BH6507 USER MANUAL DIP STATION WITH SAFETY CONNECTOR SF-BH6507 USER MANUAL IMPORTANT: Read all instructions carefully before using this product. Retain owner s manual for future reference. For customer service, please contact:

More information