1 GAS DETECTION Cost of Ownership In today's competitive marketplace the initial cost of any capital equipment is important to the end user but of equal importance is the ongoing cost of ownership. In many cases gas detectors are purchased for an immediate safety, operational or legislative need, but little thought is given to keeping them operational and within performance requirements for their life. There are 3 main costs. a) Calibration or performance testing b) Maintenance and replacement costs c) Loss of product or replacement when units are being serviced All of these are factors that need to be considered by the user. If an instrument is to be used to make measurements upon which some action is to be taken, then it is important that it operates with certain accuracy limits. If, for example, an operator has to evacuate a confined space when the Oxygen level is below 16% or 18% then it is important that this measurement is as accurate as possible, otherwise life may be at risk. Similarly if the flammability of the atmosphere is above a particular level then there is a danger to the user. There is, therefore, a need to ensure that the instrument performance remains consistent between checks.
2 Although the level in some cases may not be the absolute judge since the change in reading may be equally important, ie falling Oxygen levels indicate a worsening situation as does increasing flammability levels. There has to be some absolute level to give guidance to the user. Instruments have to be calibrated or tested at regular intervals. If you operate an ISO9000 quality system, test instruments which perform measurements upon which judgements are made have to be calibrated, normally traceable to national standards. Gas detectors are similar. The simplest way is to test the unit regularly in the field with a certified test gas to establish a basic performance level. This can be done before each test, daily, weekly, monthly etc depending upon the risk involved and the performance of the instrument. The simplest test is just a reaction test to see if the instrument works. This could be breathing into an Oxygen instrument to ensure that the O2 level falls or similarly with a CO2 instrument to ensure that the CO2 level rises. We breath out around 4% CO2 so both CO2 and O2 readings are affected. With a flammable instrument the application of the gas from the mains or a lighter will give some indication that the basic unit functions. Apart from time there is little cost and it provides a basic function test but it does not constitute a performance test and is certainly not a calibration. The next level up is to perform a 'bump' test or basic level check. This can be done by the user with some basic instruction and requires the use of a test gas or test gases. Gas is applied to the instrument under controlled conditions and the level of response noted and compared against some agreed criterion. Test gases are available from a number of suppliers with different accuracies. In a bump test the accuracy of the gas is probably the least important since other factors affect the
3 instrument performance, such as temperature, pressure, skill of the operator etc. Hence in bump testing wide parameters are used. A typical gas mixture may have an accuracy of say 2% and could be applied to an instrument with an accuracy of say 3% of Full Scale. This may vary depending upon the ambient temperature and other factors. If the gas is around the mid point of the instrument scale, then depending upon the actual specification of the instrument you may get a variation similar to that illustrated below. Gas = 50%LEL ± 2% ie 48%LEL to 52%LEL Instrument accuracy ± 3%LEL ie 48-3 to = 45% to 55%LEL With no other factors involved this may increase to around 40% - 60%LEL in a practical situation. Provided reasonable limits are used and there are clear instructions, a bump test is a useful tool to ensure calibration but it does add cost in providing the gas and having basic training. If the instrument is calibrated in the field then the same gas can be used. You have to rely on the operator to carry out the calibration and do it regularly and with some skill. Modern gas detectors can indicate to the operator when a calibration is due either by flagging due date or by indicating when a calibration fault has occurred. Many gas detectors have built-in field calibration which automate the process and have built-in safeguards against gross operator error. In many cases they can be pre-programmed to specific gas mixtures and provide internal records of the calibration. Multimix gas reduce the number of cylinders that are required and help reduce cost. Automatic field calibration systems operated by computer based systems are also available to ensure the correct calibration of instruments in the field and can be used to extend the calibration period based on real data. If you find that with, say, monthly calibration checking in a particular application there is no basic change in instrument performance this can be extended to 3 months and if the same performance levels occur they can be extended to 6 months, 12 months etc, hence reducing costs based on real
4 data in your application, rather than some fixed figure decided on by the manufacturer. Many manufacturers have automatic calibration systems with either local computers and data storage or networked systems for large users. The idea is to keep the instruments in use, in the field, whilst still maintaining calibration records and avoiding the return to a specialised repair facility. There are some maintenance costs associated with gas detectors depending upon the application and type of sensors. Flammable detectors using semiconductors may have a life of around 1-5 years, depending on the application and exposure to gas or other compounds. Pellistor based catalytic sensors typically have a life of 5 years+ depending on application. Infra-red sensors have a much longer life typically but the source which could be a lamp type device may have around 3000 hours life. Solid state sources and detectors have a much longer life but still require maintenance if not replacement. Simple modular assemblies with self calibration reduce the skill and cost of replacement. Electrochemical sensors used for Oxygen and Toxic gas measurement such as CO, H2S etc have a typical life of 2-3 years and have to be replaced after this period.
5 There is increasing use of disposable Oxygen and Toxic Gas detectors which have a life of 2-3 years including batteries and are basically maintenance free but they will need to be tested if the risk warrants it. There is no difference in the need to test, irrespective if the unit is disposable or not. The disposable units do not require batteries and hence this cost is avoided. In other units which do have batteries there is the cost of replacement of either dry cells, usually alkaline, or rechargeable battery packs. The re-chargeable batteries have improved in performance and failure mechanisms with new type batteries, such as Nickel Metal Hydride and Lithium Ion batteries, but there is still the capital cost of chargers and the usually hidden cost of power for these, albeit small. In addition to the cost of replacement sensors there are the costs of skilled calibration and maintenance staff who check the product on a regular basis. This is an overhead cost which applies to all gas detectors. The disposable units are affected in the same way, although there is no repair capability, calibration checking may still be required. In the event of failure then the capital cost may be incurred again but you have a new detector presumably with the manufacturer's warranty. All manufacturers are striving to bring down the cost of maintenance and cost of ownership in many ways. With certified equipment there is a requirement to have the equipment checked for compliance with the certification requirements on a regular basis by a skilled and competent person. This is defined in BS EN and is typically in the order of 2 yearly. CSA, the Canadian Standards Association, insist that any flammable detector approved by them has a statement in the handbook indicating that the flammable detector has to be tested daily or before use, this is a blanket statement irrespective of any performance tests they carry out. The same criterion does not apply to an Oxygen or toxic detector approved by them. This means that there are costs associated with the maintenance and safety of all hazardous area equipment not just gas detectors. If instruments are returned to a competent repairer then they are no longer available in the field for users, hence replacements have to be made available.
6 If an instrument is sent to a central repair facility on a 6 monthly basis and the turnaround time is say 5 working days in 250, eg 50 weeks x 5 days, the instrument is out of service for routine servicing for around 4% of its working year. If you include for failures this figure may increase to 5% to 10% depending upon application. Thus in the capital cost you may have to include a larger number of units for spares. Thus the ability to keep instruments in the field by providing improved reliability, field calibration and maintenance free units can be a vital availability issue as well as replacement parts costs. Many manufacturers will offer complete packages that ensure users always have a working calibrated unit either by means of replacement units, leasing arrangement or imaginative repair operations such as replacements or field service units. The use of automatic calibration and fault diagnosis on fixed systems using computer based systems linked into SCADA or BMS (Building Management Systems) is increasing and helping to reduce the cost of ownership. The use of gas detectors in many industries is increasing but the costs are reducing in an ever more competitive market. Buyers are no longer just looking at a price but are assessing the total cost of ownership as well as technical features. The world gas detection market is addressing the needs of its customers by ever more innovative ways. In the UK CoGDEM members are working together to ensure there are technical standards that enable users to have a minimum requirement for safe gas detection equipment both portable and fixed, whilst at the same time competing in the marketplace to give individual customers the best options for their particular market at the lowest possible cost without compromising safety.
Verification Of Calibration for Direct-Reading Portable Gas Monitors
U. S. Department of Labor Occupational Safety and Health Administration Directorate of Science, Technology and Medicine Office of Science and Technology Assessment Verification Of Calibration for Direct-Reading
More informationMisuse of Combustible Gas Meters
Misuse of Combustible Gas Meters An incident investigation raised concerns that users of combustible gas meters may be greatly underestimating fire and explosion hazards. Errors result when meters calibrated
More informationGas Detection. Portable Gas Detection for Safety in Confined Spaces
Gas Detection Portable Gas Detection for Safety in Confined Spaces Portable Gas Detection for Safety in Confined Spaces An Overview In today s working environment confined spaces and confined space entry
More informationTOXIC GAS DETECTORS IN THE WORKPLACE UK REGULATIONS AND STANDARDS
SUMMARY TOXIC GAS DETECTORS IN THE WORKPLACE UK REGULATIONS AND STANDARDS In recent articles supplied by the Council of Gas Detection and Environmental Monitoring (CoGDEM) for inclusion in International
More informationBASF Corporation Title: Portable Multi-Gas Monitoring Equipment. Individual Unit Function: Safety Procedure No.: GSS09 Page: 1 of 5
Procedure No.: GSS09 Page: 1 of 5 RECORD OF REVISIONS Date Details of Change Sept. 01 The revision addresses the deletion of the tool room responsibilities and defines the new responsibilities for the
More informationCalibration and Bump Test Requirements for Direct Reading Portable Gas Monitors
: Calibration and Bump Test Requirements for Direct Reading Portable Gas Monitors Manufacturers and regulatory agencies agree the safest and most conservative approach is to perform a functional test by
More informationCalibration Requirements for Direct Reading Confined Space Gas Detectors
: Calibration Requirements for Direct Reading Confined Space Gas Detectors However, the definition of bump test has always been a little slippery. Some manufacturers differentiate between a bump test that
More informationOXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:
OXY Integral INTERCON ENTERPRISES INC Tel: 800 665 6655 Fax: 604 946 5340 E-Mail: sales@intercononline.com Internet: www.intercononline.com Manual Integral 2006 1 INDEX 2-3 PREFACE 4 INTRODUCTION 5 Principle
More informationCalCheck. Instrument User Manual V1.8
CalCheck Instrument User Manual V1.8 Unrivalled Part number: 21518 Detection. Part number: 21518 Declaration of conformity Manufacturer:, The Way, Fowlmere, Cambridge, England. SG8 7UJ Product: CalCheck
More informationSpot the Hazards: Review Confined Space Hazards Missed by Most Safety Professionals, and the PPE Solutions that Meet OSHA Requirements
Spot the Hazards: Review Confined Space Hazards Missed by Most Safety Professionals, and the PPE Solutions that Meet OSHA Requirements Why is Gas Detection Important? Spot the hazard Composition of Fresh
More informationPortable Gas Monitor GX User Maintenance Manual (H4-0050)
H4E-0050 Portable Gas Monitor GX-8000 User Maintenance Manual (H4-0050) Need of Maintenance and Servicing This gas monitor must be maintained in a normal state at all times to prevent accidents due to
More informationModel Description
LEAKHUNTER Plus LEAKHUNTER Plus Model 8066 Description The MATHESON LEAKHUNTER Plus is a truly universal leak detector. Extensive R&D has produced a multi-functional leak detector engineered to perform
More informationSF 6 Product Guide. Gas Analysis Instruments and Accessories
SF 6 Product Guide Gas Analysis Instruments and Accessories Cambridge Sensotec Ltd specialises in the design and manufacture of gas analysis equipment under the Rapidox brand name. Based in the UK, the
More informationHIGH ACCURACY MULTI-GAS MONITORING USING AUTOMATED SELF- CALIBRATION
HIGH ACCURACY MULTI-GAS MONITORING USING AUTOMATED SELF- CALIBRATION SCENTROID is the global leader in sensory technology providing intelligent environmental solutions across a wide range of industries
More informationWhite Paper. Chemical Sensor vs NDIR - Overview: NDIR Technology:
Title: Comparison of Chemical Sensor and NDIR Technologies TSN Number: 25 File:\\MII- SRV1\Metron\Bridge_Analyzers\Customer_Service_Documentation\White_Papers\25_G en EGA NDIR vs Chemical Sensor.docx Created
More informationCalibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.
Advanced Calibration Designs, Inc. Calibration Gas Instrument INSTRUCTION MANUAL Release I www.goacd.com Instruction Manual Gas Generator Release I TABLE OF CONTENTS I. General Description Page 2 II. Start-Up
More informationApplication Note AN-107
SPEC Sensor TM Characterization & Calibration Considerations Scope This document is provided to describe the considerations needed to characterize, calibrate, verify and validate the measurement performance
More informationGASES FIRE & RESCUE SPECIALTY. Gases & Equipment. Medical Gases Liquid & High Pressure Oxygen U.S.P. Nitrogen N.F. Carbon Dioxide U.S.P.
FIRE & RESCUE Gases & Equipment Medical Gases Liquid & High Pressure Oxygen U.S.P. Nitrogen N.F. Carbon Dioxide U.S.P. Zero Air Methane Carbon Monoxide SPECIALTY GASES 061113 Total Support... our business
More informationAdvanced Calibration Designs, Inc. goacd.com. Certified Calibration Gas Solutions
Advanced Calibration Designs, Inc. goacd.com Certified Calibration Gas Solutions Table of Contents We Provide The Ultimate Calibration Gas Solution... 4 How to Buy... 4 CAL2000 Instrument & Accessories...5
More informationGas-Detection Instruments
ALTAIR 5X Multigas Detector Accessories SPARE PARTS/ REPLACEMENTS CALIBRATION 10114835 Battery pack assembly, rechargeable 10114837 Battery pack, alkaline (includes belt clip) 10114839 Battery pack, rechargeable,
More informationUltima. X Series Gas Monitor
Ultima X Series Gas Monitor Safety Manual SIL 2 Certified " The Ultima X Series Gas Monitor is qualified as an SIL 2 device under IEC 61508 and must be installed, used, and maintained in accordance with
More informationHyper-Gas MkII TM Bell Unit
THIS DOCUMENT RELATES ONLY TO THE BELL UNIT OF THE HYPER-GAS MkII TM This document must be read by all working in dive control and all divers before working in a bell which carries a Hyper-Gas MkII TM.
More informationCOAL MINES TECHNICAL SERVICES, GAS DETECTION RELATED SERVICES FOR MINES RESCUE
COAL MINES TECHNICAL SERVICES, GAS DETECTION RELATED SERVICES FOR MINES RESCUE Peter Mason Manager, Coal Mines Technical Services, a division of Mines Rescue Pty Ltd, Australia ABSTRACT The NSW Mines Rescue
More informationWP Eliminating Oxygen from the Purge Gas and the use of Monitoring Equipment
Part 4 WP - 269 Eliminating Oxygen from the Purge Gas and the use of Monitoring Equipment Even using specialised weld purging equipment does not guarantee defect free welds. Control of the oxygen content
More informationColumbus Instruments
0215-003M Portable O 2 /CO 2 /CH 4 Meter User s Manual Columbus Instruments 950 NORTH HAGUE AVENUE TEL:(614) 276-0861 COLUMBUS, OHIO 43204, USA FAX:(614) 276-0529 1 www.colinst.com TOLL FREE 1-800-669-5011
More informationService Calibration are you doing it properly?
ABB MEASUREMENT & ANALYTICS ARTICLE Service Calibration are you doing it properly? A variety of factors, including wear and tear and non-ideal installation conditions, can cause the performance of an instrument
More informationWorking in Gas Hazard Areas
/ Health & Safety OPERATIONS STANDARDS Ref.: Aperam Safety ST 012 Working in Gas Hazard Areas Issue: 15.05.2012 Vers.: v0 Rev.: AM Safety ST 012 v1.2 08.05.2012 Working in Gas Hazard Areas Controlled Updated
More information!!!!! Are you applying for Recognition of Prior Learning? If yes, please see your instructor. Y N. Suburb State Postcode Postal Address
Course Enrolment Form Confined Gas Test Space Atmosphere Entry - CS 200 MSAPMOHS200A MSAPMPER217A Work Gas Test safely Atmosphere MSAPMPER202A Observe Permit Work MSAPMOHS217A Gas Test Atmospheres Unique
More informationGasLab SAN-102 Micro Oxygen Monitor User Manual
GasLab SAN-102 Micro Oxygen Monitor User Manual The GasLab SAN-102 is a wearable, personal safety meter designed to monitor ambient oxygen levels in real time. It is designed to protect workers in confined
More informationFlammable Environments Guideline
GUIDE Flammable Environments Guideline 2014-0004 The Canadian Association of Petroleum Producers (CAPP) represents companies, large and small, that explore for, develop and produce natural gas and crude
More informationApplications Note: Use of "pentane equivalent" calibration gas mixtures
Introduction The gas that is used to verify accuracy is every bit as important as the detector itself when it comes to worker safety. Choosing (and using) the right mixture is critical to the success of
More informationCPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL
CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL Company Registered Office: Crown House, Stockport, Cheshire, SK13RB No. 850 0700 67 Registered in England and Wales No. 05058855
More informationThis test shall be carried out on all vehicles equipped with open type traction batteries.
5.4. Determination of hydrogen emissions page 1 RESS-6-15 5.4.1. This test shall be carried out on all vehicles equipped with open type traction batteries. 5.4.2. The test shall be conducted following
More informationGAS DETECTION: CALIBRATION GAS
ENVIRONMENTAL SURVEY/PURGING MONITORS GAS DETECTION: CALIBRATION GAS Gas Mixtures Non-Reactive Single & Two-Gas Mixtures 98 Reactive Single & Two-Gas Mixtures 98 Non-Reactive Multi Mixtures 99 Reactive
More informationPM105 PRESSURE METER USER S MANUAL
PM105 PRESSURE METER USER S MANUAL Manual revision: 15 March, 2001 SCSMANP015 Copyrighted 2001 Thank you for purchasing a Tek Know pressure meter. The Tek Know products are manufactured by Scan-Sense AS
More informationFlare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1
Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost HERMAN HOLM SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 4C CONFERENCE APRIL 2018 SAN ANTONIO, TX SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC
More informationDescribe Flammable Gas Measurement
Training Module Describe Flammable 100% 90% C 80% No Combustion 70% 60% 50% 40% 30% 20% UEL B Combustion No Combustion 10% 0% LEL A Human Development HDC Human Development All rights reserved. No part
More informationQRAE II Quick Reference Covers Diffusion & Pump Models with Firmware Version 3.12 and higher
QRAE II Quick Reference Covers Diffusion & Pump Models with Firmware Version 3.12 and higher WARNINGS Use only RAE Systems rechargeable battery pack part number 020-3402-000, or alkaline battery pack part
More information973-SF 6 Analyzer. Precise and Stable SF 6 Gas Analyzer REFLECTING YOUR STANDARDS
Precise and Stable SF 6 Gas Analyzer Simultaneous measurement of humidity, SF 6 purity and SO 2 concentration Integrated gas recovery system with automatic pump back Fully automated SF 6 gas testing Fundamental
More informationGMC-I Service GmbH Calibration Center
2016 GMC-I Service GmbH Calibration Center DAkkS, ISO & Factorycalibration www.gmci-service.com From spare parts, reparation, rental service, maintenance of industrial health and safety standards, calibration
More informationGas Detection Technology in Confined Spaces
Session # 621 Gas Detection Technology in Confined Spaces Introduction Andrew Saunders Applications and Training Specialist BW Technologies by Honeywell The danger of toxic gas hazards is a very real and
More informationToolbox Talk. Carbon Monoxide
Carbon Monoxide Toolbox Talk Carbon monoxide defies all five human senses. You can t see it, smell it, touch it or taste it and you certainly can t hear this potentially deadly gas, which has been called
More informationGasSense NDIR User Manual
INDEX INDEX... 1 1. OVERVIEW... 2 2. TECHNICAL DATA... 3 3. SPECIFICATIONS... 4 4. PRODUCT DESCRIPTION... 5 4.1 Mechanical details... 5 4.2 Piping... 7 4.3 Connections... 7 5. INSTALLATION... 10 6. CALIBRATION
More informationO3 3E 1 F Gas Sensor Module
Product Information Pack O3 3E 1 F Gas Sensor Module (Ozone) CONTENTS Product Data Sheet Product Specification 2 Poisoning and Cross Sensitivities 3 Operating Instructions Introduction 4 Electrostatic
More informationIRwin Methane Leak Detector. Innovation makes the difference
IRwin Methane Leak Detector Innovation makes the difference Innovative methane detector system for easy survey of gas pipelines IRwin is a portable methane leak detector for fast and easy survey of gas
More informationEasyLine Continuous Gas Analyzers. So smart, they re simple
EasyLine Continuous Gas Analyzers So smart, they re simple Reliable Economic Powerful EasyLine is both a powerful and affordable line of instruments for the monitoring of gas concentrations in numerous
More informationImprove Process Reliability
Convenient Conductivity Calibration to Meet Pharmacopeia Regulations Pharmacopeias require that conductivity measurement systems be periodically calibrated to ensure their precision. The range of accessories
More informationUSE OF RESIDUAL PRESSURE VALVES
USE OF RESIDUAL PRESSURE VALVES AIGA 063/16 Revision of AIGA 063/14 Asia Industrial Gases Association 3 HarbourFront Place, #09-04 HarbourFront Tower 2, Singapore 099254 Tel : +65 6276 0160 Fax : +65 6274
More informationDilution Manual Cannonball 3 Multi Gas Detector
Dilution Manual Cannonball 3 Multi Gas Detector GasTech Australia Pty Ltd 24 Baretta Rd Wangara Western Australia 6065 Tel 1800 999 902 Fax 1800 999 903 http://www.gastech.com.au 0 THE CANNONBALL 3 PERSONAL
More informationManifolds & pressure reducing sets
Manifolds & pressure reducing sets Automatic Changeover Manifold (Electronic) Providing continuous supply of medical gases Pneumatech Medical Gas Solutions (PMGS) PureGAS Manifolds provide a duty or standby
More informationABB MEASUREMENT & ANALYTICS. Continuous gas analyzers EL3060 The specialists for hazardous areas
ABB MEASUREMENT & ANALYTICS Continuous gas analyzers EL3060 The specialists for hazardous areas The specialists for hazardous areas ABB Analytical possesses decades of experience in the area of explosion
More informationPURA. Pure Gas Dewpoint Transmitter. Users Guide
PURA Pure Gas Dewpoint Transmitter Users Guide Issue January 2003 Page 2 Table of Contents SECTION PAGE 1 Product Overview 3 2 Preparation 3 3 Installation 3 3.1 PURA Premium & OEM Dewpoint Transmitter
More informationUKAS calibration laboratory 8809 Temperature, pressure and electrical signals calibration guide
ABB MEASUREMENT & ANALYTICS PRODUCT GUIDE PG/SMD140 EN REV. B UKAS calibration laboratory 8809 Temperature, pressure and electrical signals calibration guide Measurement made easy UKAS calibration laboratory
More informationOperation Manual. Pro CO O 2 Alarm. Carbon Monoxide and Oxygen Analyzer. Rev
Operation Manual Pro CO O 2 Alarm Carbon Monoxide and Oxygen Analyzer Rev. 08.17 Quick Reference Guide for CO Analyzer READ ENTIRE MANUAL BEFORE USE 1. To switch on, hold the On/Off button until the display
More informationCALIBRATION KIT. Instruction Manual
CALIBRATION KIT Instruction Manual 2 CALIBRATION KIT CALIBRATION KIT 3 Table of Contents General description....4 Getting Started....4 GG-NH3....5 GG-NH3-2%....6 GG-CO....7 GG-CO2...8 GG-CL2...9 GG-H2-EC....
More informationStandard Operating Procedure Measuring & Testing Equipment
Standard Operating Procedure Measuring & Testing Equipment Purpose: 7.6 Responsibility: 7.6d Scope To have a procedure to ensure that all is controlled, maintained and calibrated to ensure that Product
More informationABB MEASUREMENT & ANALYTICS. Continuous gas analyzers EL3060 The specialists for hazardous areas
ABB MEASUREMENT & ANALYTICS Continuous gas analyzers EL3060 The specialists for hazardous areas The specialists for hazardous areas ABB Analytics possesses decades of experience in the area of explosion
More informationOperation Manual. Pro 4 Warn Analyzer TM Rev 08.17
Operation Manual Pro 4 Warn Analyzer TM Rev 08.17 If you have any questions on this equipment please contact Technical Support at: Nuvair 1600 Beacon Place Oxnard, CA 93033 Phone: 805-815-4044 FAX: 805-486-0900
More informationUTAS EH&S INSTRUCTION No. 036 June 2012 Revision No. 3. Purging of Flammable Gas Equipment and/or Piping (Explosion Hazard)
Purging of Flammable Gas Equipment and/or Piping (Explosion Hazard) APPLICABILITY: This EH&S Instruction applies to UTAS sites worldwide, including contractors working at UTAS facilities. The document
More informationGasLab SAN-102 Micro Oxygen Monitor User Manual
GasLab SAN-102 Micro Oxygen Monitor User Manual The GasLab SAN-102 is a wearable, personal safety meter designed to monitor ambient oxygen levels in real time. It is designed to protect workers in confined
More informationThermo Probe, Inc. TP-5 TP-7 TP-8. Instruction Manual English. Switch. Display. Carry Strap. Probe. Ground Clip. Switch. Display.
Display Thermo, Inc. Instruction Manual English TP-5 Display TP-7 Display TP-8 1-1 Thermo, Inc. Instructions for use- Models TP-5, TP-7, and TP-8 The manual describes basic function and use of Thermo instruments
More informationNeil Kane Managing Director EMT UK
Neil Kane Managing Director EMT UK SF6 Measurement, Analysis, Diagnostics and Leak Detection Our goals at EMT are to supply Maintenance Operatives with the very best tools to protect both personnel and
More informationLaboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.
Laboratory Hardware Custom Gas Chromatography Solutions Custom solutions for your analytical needs. Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions for advanced chromatography
More informationBest Practice for Calibrating LTH Conductivity Instruments
Application Note Best Practice for Calibrating LTH Conductivity Instruments As accurate process measurement becomes an everyday requirement it is vital to be able to calibrate conductivity instruments
More informationPortable Oxygen Concentrator. Give your patients more freedom with mobile oxygen on demand
Portable Oxygen Concentrator Give your patients more freedom with mobile oxygen on demand MAXIMISE YOUR PATIENTS INDEPENDENCE AND PEACE OF MIND You can help many of your oxygen patients to live their lives
More informationElectronic Volume Correctors
Electronic Volume Correctors Ben Manson Landis+Gyr October 14, 2016 PUBLIC Agenda Customer Types and suitable equipment Volume corrector standards / regulations Basics of volume correction Gauge v Absolute
More informationCSA Sample Draw Aspirator Adapter Operator s Manual
30-0951-CSA Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0367 Revision: 0 Released: 4/30/15 www.rkiinstruments.com WARNING Read and understand this instruction manual before operating
More informationB-DETECTION PLUS THE NEXT GENERATION MOBILE AND STATIONARY GAS MEASUREMENT SYSTEMS
B-DETECTION PLUS THE NEXT GENERATION MOBILE AND STATIONARY GAS MEASUREMENT SYSTEMS SPORTS & SAFETY 2 B-DETECTION PLUS BAUER KOMPRESSOREN BECAUSE ROUGHLY ISN T IN YOUR VOCABULARY Even when your compressor
More informationDF-550E PROCESS ANALYSERS APPLICATIONS FEATURES
PROCESS ANALYSERS DF-550E The DF-550E Nano Trace oxygen analyser is a Coulometric sensor based analyser designed to measure oxygen as a contaminant at ultra trace levels in ultra high purity electronic
More informationThe routine maintenance and inspection of local exhaust ventilation (LEV)
The routine maintenance and inspection of local exhaust ventilation (LEV) This document details the standards for inspection and maintenance that should be achieved for all local exhaust ventilation owned
More informationGas Tester Competent Person Course Outline
Gas Tester Competent Person Course Outline Prerequisites: This course shall have no formal pre-requisite. Course Length: 8 hours Course length shall vary depending on the number of delegates. Total course
More informationLaboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.
Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions
More informationUser Manual. GPL 3000 e. gas detector
User Manual GPL 3000 e gas detector Table of contents Application...4 Unit ppm, Vol.%...5 Operation elements GPL 3000 e...6 Gas Sensor...7 Measurement range...8 LED allocation...8 Turning on the GPL 3000
More informationThe New PHD 4 Portable Helium Detector
The New PHD 4 Portable Helium Detector SPECIAL FEATURES: High Sensitivity to Helium Easy to Use Truly Portable Versatile Dependable The system is designed to allow easy replacement of sampling line components
More informationInstruction and Maintenance Manual
Instruction and Maintenance Manual GRYF OXY ZM 02/100/2 E Contact GRYF HB, spol. s r.o. Cechova 314 Havlickuv Brod 580 01 tel.: +420 569 426 627 fax: +420 569 426 627 www.gryf.eu ver.: 11.2.2015 OEM module
More informationCHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV
CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV 16-1 INTRODUCTION 16-1.1 Purpose. The purpose of this chapter is to familiarize divers with the techniques used to mix divers breathing gas. 16-1.2 Scope. This chapter
More informationPRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL
PRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL Company Registered Office: Crown House, Stockport, Cheshire, SK13RB No. 850 0700 67 Registered in England and Wales No. 05058855
More informationOperation and Maintenance Manual
Operation and Maintenance Manual GDS 49 Remote 4 20mA Sensor Transmitter for Toxic Gases GDS Corp. 1245 Butler Road League City, TX 77573 409 927 2980 409 927 4180 (Fax) www.gdscorp.com CAUTION: FOR SAFETY
More informationGAS ANALYSER OXYBABY 6.0 for O 2. or O 2 /CO 2
GAS ANALYSER OXYBABY 6.0 for or Cordless hand held oxygen or combined oxygen and carbon dioxide analyser for checking modified atmospheres in food packs. The ideal instrument for portable, faster and accurate
More informationOperation Manual. O2 Quickstick. Oxygen Analyzer
Operation Manual O2 Quickstick Oxygen Analyzer 01.11 www.nuvair.com If you have any questions on this equipment please contact Technical Support at: Nuvair 2949 West 5 th St. Oxnard, CA 93030 Phone: 805-815-4044
More informationDräger X-dock Frequently Asked Questions
COSTS Why do I save costs when using Dräger X-dock? SOFTWARE What are the benefits of using a central database? What is the difference between the X-dock Manager Standard and the Professional Version?
More informationThe latest automotive systems require innovative leak test methods and fixturing.
Leak Testing Auto Parts The latest automotive systems require innovative leak test methods and fixturing. In the past, automobile manufacturers only required basic forms of leak testing to check standard
More informationENVIRONMENTAL AND POLLUTANTS GAS ANALYZERS
XIX IMEKO World Congress Fundamental and Applied Metrology September 6, 9, Lisbon, Portugal ENVIRONMENTAL AND POLLUTANTS GAS ANALYZERS Ana Madeira, Florbela A. Dias and Eduarda Filipe 3 Instituto Português
More informationSWMS No 009 Samark North Qld Pty Ltd ABN : Electrical Contractors License C37121 Page 2 of 4
Samark North Qld Pty Ltd SWMS Manual SWMS No 009 Samark North Qld Pty Ltd ABN : 99 082 98 87 Electrical Contractors License C72 Page 2 of Safe Work Method Statement Title: SWMS:009 Confined Space Work
More informationOxygen Measurement Systems High Reliability and Wide Application Coverage
Introduction High Process Efficiency Under Any Condition Oxygen Measurement Systems High Reliability and Wide Application Coverage METTLER TOLEDO provides sensors to measure dissolved oxygen (DO) in demanding
More informationOxygas P-500 User Handbook
Oxygas P-500 User Handbook OXYGAS P-500 USER HANDBOOK Issue 2 11/06/13 Part Number: 42645 GMI welcomes comments on all our publications. Your comments can be of great value in helping us to improve our
More informationOxygen & Binary Gas Measurement and Analysis
Oxygen & Binary Gas Measurement and Analysis Protect processes, reduce costs and maintain quality Oxygen & Binary Gas Measurement and Analysis A range of quality process analyzers from experts in control
More informationSupplementary Operator s Manual 42/24-14 EN Rev. 3
Advance Optima Continuous Gas Analyzers AO2000 Series with Uras26, Magnos206, Caldos25, Caldos27 Designed per Category 3G for Measurement of Flammable Gases ( Safety Concept ) and Non-flammable Gases Supplementary
More informationTechnical Manual. Sensepoint XCD Gas Detector
Technical Manual Sensepoint XCD Gas Detector 19 Cross Interference and Cross Calibration 19.1 Cross Interference Table for Toxic and Oxygen This table shows the relative cross sensitivity of the Sensepoint
More informationTC65M/15M TEMPERATURE CALIBRATOR USER S MANUAL
TC65M/15M TEMPERATURE CALIBRATOR USER S MANUAL Thank you for purchasing a Scan-Sense temperature calibrator. The products are manufactured by Scan-Sense AS in accordance with our high quality standards
More informationKeeping It Simple. Gas Mixtures for Stack Emissions Monitoring.
Keeping It Simple Gas Mixtures for Stack Emissions Monitoring www.uk.airliquide.com Stack Emissions Monitoring The Stack Emissions Monitoring market has undergone substantial legislative and quality changes
More informationRAM Operation Manual
RAM 4021-1 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021-1 For Your Safety... 2 Description... 2 Setup Mode... 3 Lights/Alarms... 3 Operation... 4 Calibration...
More informationPRODUCT CONFORMITY CERTIFICATE
PRODUCT CONFORMITY CERTIFICATE This is to certify that the Endura AZ20 / AZ30 measuring system Manufactured by: ABB Ltd Oldends Lane Stonehouse Gloucestershire GL10 3TA has been assessed by Sira Certification
More informationAdditel 761 Automated Pressure Calibrators
Automated Calibrators UPDATED Fully automated pressure calibrator with built-in pressure generator / controller to as high as 600 psi (40 bar) or as low as 0.01 Pa (0.00004 inh 2 O) accuracy Dual pressure
More informationDocumentation Beats Confrontation in Gas Detection
Session No. 770 Documentation Beats Confrontation in Gas Detection Andrew Saunders, CQT Applications and Training Specialist BW Technologies by Honeywell Introduction Deaths of workers in confined spaces
More informationDepartment Of Facilities Management. Occupational Health and Safety. Confined Space Entry Program
Department Of Facilities Management Occupational Health and Safety Confined Space Entry Program April 2015 1 1.0 INTRODUCTION The Department of Facilities Management has identified numerous confined spaces
More informationSensepoint XCD. Calibration Guide
Sensepoint XCD Calibration Guide Sensepoint XCD Calibration Calibration Handbook It is recommended to periodically carry out a gas response check on the Sensepoint XCD to ensure correct operation. This
More informationEN ISO/IEC Technical Requirements
EN ISO/IEC 17025 - Technical Requirements Wolfgang Garche Environmental Protection Agency Department Air Quality Monitoring, Information, Assessment 1 What means accreditation in accordance to EN ISO/IEC
More informationStability and traceability of calibration gases for EN 14181
Stability and traceability of calibration gases for EN 14181 Stephen Mellor Especialista de Produto Product Specialist Analytical, Air Products Plc, United Kingdom DOSSIER EMISSÕES E MONITORIZAÇÃO ESTABILIDADE
More informationAUTOMATIC HOSE TEST UNIT, TYPE SPU
VALVES AND FITTINGS UP TO 14,000 BAR TEST AND CONTROL EQUIPMENT H IGH PRESSURE TECHNOLOGY AUTOMATIC HOSE TEST UNIT, TYPE SPU Pressure range from 1 up to 10,000 bar User-friendly touch panel operation HIGH-PRESSURE
More information