OPTISWIRL 4070 C. Handbook. Vortex-flowmeter

Size: px
Start display at page:

Download "OPTISWIRL 4070 C. Handbook. Vortex-flowmeter"

Transcription

1 OPTISWIRL 4070 C Handbook Vortex-flowmeter

2 CONTENTS OPTISWIRL 4070 C 1 Safety Intended use Certifications Safety instructions from the manufacturer Notes about the documentation Display conventions Safety instructions for the operator INSTRUMENT DESCRIPTION Scope of delivery Instrument versions Versions with flanged connection Sandwich versions Instrument description Nameplate Installation General installation instructions Storage Transport Installation requirements General Liquid measurement Vapor and gas measurement Piping with control valves Preferred installation location Turning the connection housing Turning the display Thermal insulation Inlet and outlet runs Minimum inlet runs Flow straightener Minimum outlet runs Installation General installation instructions Installing instruments in sandwich design Installing instruments in flange design Electrical connections Safety instructions Connecting the signal converter Current output, electrical data Power supply /

3 OPTISWIRL 4070 C CONTENTS Load for HART communication Pulse output, electrical data Connection pulse output Configuring the pulse output Protection category Earthing connections Start-up and control Start-up Start Start-up and control Operation Display and operating elements Operating principles Description of the key functions Switch from measuring mode to menu mode Navigation within the menu structure Changing the settings in the menu Measures in the event of faulty indications Overview of the most important functions and indicators Error messages Service Replacing converter / LCD display Returning the device to the manufacturer General information Form (for copying) to accompany a returned instrument Technical data Functional principle Technical data Dimensions and weights Flow tables Appendix Menu structure Overview of firmware versions Entering values in change mode Selecting characters in change mode Quick Setup menu Tests menu Menu item Setup (Firmware version v1 Basic) Menu item Setup (Firmware version v6 Steam) Menu item Setup (Firmware version v7 Gas) /

4 CONTENTS OPTISWIRL 4070 C 10 KROHNE Product Overview /

5 OPTISWIRL 4070 C SAFETY Intended use The vortex flowmeters are made to measure the flow of gases, vapors and liquids. These flowmeters are particularly suitable for measuring: Clean liquids with low viscosity (< 10 cp) Hydrocarbons with low viscosity (< 10 cp) Water Chemicals with low corrosiveness Saturated steam Superheated steam, including CIP and SIP applications in the food industry Industrial gases The meters are rated for the following flow velocities: From 0.3 to 9.0 m/s for liquids From 3.0 to 80 m/s for gases and vapors 1 1 DN15-3,0 to 45 m/s DN25-3,0 to 70 m/s If the danger of waterhammers can occur in steam net works appropriate condendsate separators have to be installed. If the danger of water cavitations can occur appropriate measures have to meet. WARNING! The operator himself bears the sole responsibillty for the intended use of the flow meter regarding the suitability and the comision resistance of the used materials against the measured fluid. The manufacturer sis not liable for any damage resulting from improper use or use for other than the intended purpose. Do not use any abrasive or highly viscous media < 10 cp. INFORMATION! The sensors are fabricated from stainless steel 316 L (1.4404) or Hastelloy C4. In your project planning, please observe the data given in the corrosion tables. The pressure-bearing parts have been designed and rated for stationary operation taking account of maximum pressure and temperature. External forces and moments, caused e.g. by pipe stresses, have not been considered thereby. Primarily, volumetric flow and temperature are measured, with pressure measurement as an option. From these parameters the flowmeter calculates the mass flow or standard volumetric flow using pre-programmed density data and then issues the measured values via various communication interfaces. 12/

6 1 SAFETY OPTISWIRL 4070 C 1.2 Certifications The flowmeter fulfil the statutory requirements of the following EC directives: Pressure Equipment Directive 97/23/EC Low Voltage Directive 73/23/EEC EMC Directive 89/336/EC As well as EN EMC specification as per EN 61326/A1 NAMUR Recommendations NE 21 and NE 43 The manufacturer certifies successful testing of the product by applying the CE mark. 1.3 Safety instructions from the manufacturer The flowmeter has been built and tested in accordance with the current state of the art, and complies with the relevant safety standards. However, dangers may arise from improper use or use for other than the intended purpose. For this reason, observe all of the safety instructions in this document carefully Notes about the documentation In addition to the safety rules and industrial safety regulations in this documentation, national and regional safety rules and industrial safety regulations must also be observed /

7 OPTISWIRL 4070 C SAFETY Display conventions The following symbols are used to help you navigate this documentation more easily: WARNING! These warning signs must be observed without fail. Even only partial disregarding such warnings can result in serious health damage, damage to the device itself or to parts of the operator s plant. DANGER! This symbol designates safety advice on handling electricity. CAUTION! These warnings must be observed without fail. Even only partial disregarding such warnings can lead to improper functioning of the device. LEGAL NOTICE! This symbol designates information on statutory directives and standards. NOTE! This symbol designates important information for the handling of the device. HANDLING This symbol designates all instructions for actions to be carried out by the operator in the specified sequence. i CONSEQUENCE This symbol designates all important consequences of the previous actions. 1.4 Safety instructions for the operator CAUTION! The flowmeters from may only be installed, commissioned operated and maintained by properly trained and authorized personnel. This document must be read and understood in its entirety by all users prior to installation, commissioning, operation and maintenance of the flowmeter. 12/

8 2 INSTRUMENT DESCRIPTION OPTISWIRL 4070 C 2.1 Scope of delivery INFORMATION! Please check the contents of the consignment for completeness and intactness. QuickStart Figure 2-1: Scope of delivery 1 Device in ordered version 2 Quick Start 3 Certificates, calibration report and parameter datasheet 4 CD with complete documentation 5 Bar magnet 2.2 Instrument versions The devices come in the following versions: Converter with display Blind device Sensor in flanged design Sensor in sandwich design The following designs are available as options: with pressure sensor with isolation valve for the pressure sensor /

9 OPTISWIRL 4070 C INSTRUMENT DESCRIPTION Versions with flanged connection The compact unit consists of a measurement sensor and a signal converter. Together these form a fixed mechanical unit. Figure 2-2: Flanged instrument with display 1 Version with temperature sensor 2 Version with temperature sensor and optional pressure sensor 3 Version with temperature sensor, optional pressure sensor and shutoff valve Sandwich versions Sandwich-type flowmeters have 2 centering rings to facilitate installation. Figure 2-3: Sandwich versions with display 1 Version with temperature sensor 2 Version with temperature sensor and optional pressure sensor 3 Version with temperature sensor, optional pressure sensor and shutoff valve 12/

10 2 INSTRUMENT DESCRIPTION OPTISWIRL 4070 C Instrument description Figure 2-4: Instrument description 1 Signal converter 2 Cable gland grey, standard design 3 Pressure sensor, optional 4 Sensor 5 Shutoff valve, optional 6 Centering ring /

11 OPTISWIRL 4070 C INSTRUMENT DESCRIPTION Nameplate INFORMATION! Before installing the device, make sure that the information on the nameplate corresponds to the ordering data. Figure 2-5: Example of a nameplate 1 Device type 2 Manufacturer 3 Appointed ATEX & DGRL body 4 Connection data: nominal diameter and pressure rating 5 DGRL data 6 Ex data 7 Electrical connection data 8 Manufacturer's website Figure 2-6: Example of a nameplate 1 Serial number 2 Order number 3 Type code 4 Item number 5 fluid data 6 Software variant 7 Tag number 12/

12 3 INSTALLATION OPTISWIRL 4070 C 3.1 General installation instructions CAUTION! Installation, assembly, start-up and maintenance may only be performed by appropriately trained personnel. The regional occupational health and safety directives must always be observed. The following procedures have to be carried out before installing the device! Check the packaging and the device itself for any damage. Check the contents of the consignment for completeness. Compare your order specification with the scope of delivery. INFORMATION! Comply with the application limits of the device with regard to pressure and temperature in accordance with EN and/or ASME B Additional information refer to Technical data on page Storage Store the flowmeter in a dry and dust-free location. Avoid lasting direct exposure to the sun. Store the flowmeter in its original packaging. The permissible storage temperature is from -40 to +80 C for standard meters /

13 OPTISWIRL 4070 C INSTALLATION Transport Use carrying straps for transport and lay them around the two process connections. Flowmeters must not be lifted by the converter housing for transport. Never lift the flowmeter by the pressure sensor. Do not use chains as these could damage the housing. Figure 3-1: Transport instructions CAUTION! There is a risk of injury due to equipment that is not secured properly. The center of gravity of the equipment is often higher than the lug for attaching the lifting belts. During transport, prevent the flowmeter from unintentional slipping or turning. 12/

14 3 INSTALLATION OPTISWIRL 4070 C 3.4 Installation requirements INFORMATION! For accurate volumetric flow measurement the flowmeter needs a completely filled pipe and a fully developed flow profile. Please observe the instructions regarding inlet and outlet pipe runs as well as the installation position. In the case of vibrations on the piping choose the installation site in such a way that the vibrations are at their lowest in transverse direction to the flowmeter. CAUTION! When installing the device in the piping, the following points must be observed: Nominal diameter of connection pipe flange = nominal flange diameter of pipe! Use flanges with smooth holes, e.g. welding neck flanges. Carefully align the holes of the connecting flange and the flowmeter flange. Check the compatibility of the gasket material with the process product. Make sure that the gaskets are arranged concentrically. The flange gaskets must not project into the pipe cross-section. The flanges have to be concentric. There must not be any pipe bends, valves, flaps or other internals in the immediate inlet run. Sandwich-type flowmeters may only be installed using a centering ring. Never install the flowmeter directly behind piston compressors or rotary piston meters. Do not lay signal cables directly next to cables for the power supply General INFORMATION! A sufficient distance around the flowmeter in all directions must be kept free to allow servicing Figure 3-2: Temperature ranges 1 Max. medium temperature see tech. datasheet 2 Max. storage temperature and ambient temperature Ambient temp.: Ex devices C, non Ex devices C Storage temperature: C 3 Protect device from direct exposure to sunlight /

15 OPTISWIRL 4070 C INSTALLATION Liquid measurement Prohibited installation Figure 3-3: Upper pipe bend CAUTION! Prohibited: Installing the flowmeter in an upper pipe bend 1, because there is a risk of gas bubbles 2 forming. Gas bubbles can lead to pressure surges and inaccurate measurement. Figure 3-4: Downpipe and outlet CAUTION! Prohibited: Installing the device in a downpipe 3 or upstream of a run-out 4. There is the risk of a partially-filled pipe. 12/

16 3 INSTALLATION OPTISWIRL 4070 C Recommended installations CAUTION! It is absolutely necessary to maintain the required inlet runs and outlet runs! If the device is installed in a downpipe, it is absolutely essential to install a riser pipe immediately after it. 2 Installing the device in an inclined riser pipe. 3 Installing the device in a vertical riser pipe. 4 Installing the device in the bottom pipe bend /

17 OPTISWIRL 4070 C INSTALLATION Vapor and gas measurement 1 2 Figure 3-5: Non-recommended installation 1 Lower pipe bends 2 Condensate DANGER! Do not install the instrument in a lower bend 1 because the risk of condensate 2 occurs. Condensate can lead to cavitation and inaccurate measurement. Under certain circumstances the instrument can be destroyed and the measured medium can leake Piping with control valves INFORMATION! To ensure trouble-free and accurate measurement, the manufacturer recommends not installing the flowmeter downstream of a control valve. There is a risk of turbulence occurring that distorts the measuring result. Figure 3-6: Pipeline with control valve 1 Recommended: Installing the flowmeter upstream of the control valve, distance 5 DN 2 Not recommended: Installing the flowmeter directly downstream of control valves, due to vortex formation. 12/

18 3 INSTALLATION OPTISWIRL 4070 C Preferred installation location Figure 3-7: Preferred installation locations 1 Above a horizontal pipe 2 Below a horizontal pipe 3 On a vertical pipe INFORMATION! Depending on installation position you will have to turn round the display and/or the connection housing. Additional information refer to Turning the display on page /

19 OPTISWIRL 4070 C INSTALLATION Turning the connection housing DANGER! All work on the device electrics may only be carried out by appropriately trained personnel. The regional occupational health and safety directives must always be observed. CAUTION! Do not damage the electrical cable by over tightening. Do not undo the electrical plug connection. 1 Figure 3-8: Allen screws on converter housing Switch off the power supply to the flowmeter. Loosen the four screws 1 on the underside of the connection housing. Lift the connection housing and turn it to the required position in 90 stages. Screw the connection housing back on tightly. 12/

20 3 INSTALLATION OPTISWIRL 4070 C Turning the display DANGER! All work on the flowmeter electrics may only be carried out by appropriately trained personnel. The regional occupational health and safety directives must always be observed. INFORMATION! If the flowmeter is installed in a vertical pipe run, you will have to turn the display by 90 ; if installed below a pipe, turn 180. Figure 3-9: Turning the display Turn the display as follows: Switch off the power supply to the flowmeter. Unscrew the cover in front of the display 1 from the connection housing. Pull the display 2 carefully a few centimetres out of the anchor fitting and turn it to the required position 3. Press the display back onto the spacer pins 4, until it clicks firmly into place. Turn the cover with gasket 5 back onto the housing and tighten it by hand /

21 OPTISWIRL 4070 C INSTALLATION Thermal insulation CAUTION! The area above the converter support must not be thermally insulated. The thermal insulation 3 may only extend to the maximum height 1 shown below up to the connecting screws of the sensor Figure 3-10: Thermal insulation on connection piece and signal cable 1 Max. height of insulation up to intermediate piece between sensor and converter 2 Max. thickness of the insulation up to the bend of the pressure pipe 3 Insulation CAUTION! The thermal insulation 3 may only extend as far as the bend in the pressure measurement piping 2. 12/

22 3 INSTALLATION OPTISWIRL 4070 C 3.5 Inlet and outlet runs Minimum inlet runs Figure 3-11: Minimum inlet runs 1 General inlet run without disturbing flow 20 DN 2 Behind a control valve 50 DN 3 After a pipe diameter reduction 20 DN /

23 OPTISWIRL 4070 C INSTALLATION Figure 3-12: Minimum inlet run 1 After a single bend DN 2 After a double bend 2x90 30 DN 3 After a double three-dimensional bend 2x90 40 DN 12/

24 3 INSTALLATION OPTISWIRL 4070 C Flow straightener If, due to the type of installation, the required inlet runs are not available, the manufacturer recommends using flow straighteners. Flow straighteners are installed between two flanges upstream of the flowmeter and shorten the required inlet run. Figure 3-13: Flow straightener 1 Straight inlet run upstream of straightener 2 DN 2 Flow straightener 3 Straight pipe run between flow straightener and flowmeter 8 DN 4 Minimum straight outlet run 5 DN Minimum outlet runs 1 2 Figure 3-14: Minimum outlet runs 1 Upstream of pipe expanders, pipe bends, control valves, etc. 5 DN 2 Upstream of metering points 5 6 DN INFORMATION! The interior of the pipe at the metering points must be free of burrs and other flow impediments. The flowmeter has an internal temperature sensor. External temperature sensor must be located 5 DN behind the flow sensor outlet. Use sensors that are as short as possible to avoid disturbing the flow profile /

25 OPTISWIRL 4070 C INSTALLATION Installation General installation instructions CAUTION! Installation, assembly, start-up and maintenance may only be performed by appropriately trained personnel. The regional occupational health and safety directives must always be observed. The following procedures have to be carried out before installing the flowmeter in the piping! Remove all transport safeguards and protective coverings from the flowmeter. Make sure that the gaskets have the same diameter as the piping. Make sure that the flowmeter is in the correct flow direction. This is indicated by an arrow on the sensor housing. CAUTION! On metering points with varying thermal loads the flowmeters have to be mounted with stress bolts (DIN 2509). Stress bolts and/or nuts and bolts are not included in the scope of delivery. Make absolutely certain that the measuring flanges are sitting concentrically. When preparing the metering point, keep in mind the exact installed length of the flowmeter. Additional information on the dimensions refer to Dimensions and weights on page Figure 3-15: Preparing the metering point 1 Installed length of flowmeter + thickness of gaskets. 12/

26 3 INSTALLATION OPTISWIRL 4070 C CAUTION! The inside diameters of the pipes, of the measuring sensor and the gaskets must be the same! The gaskets must not project into the pipe cross-section. Figure 3-16: Internal diameter 1 Internal diameter connection pipe 2 Internal diameter flange and gasket 3 Internal diameter measuring sensor /

27 OPTISWIRL 4070 C INSTALLATION Installing instruments in sandwich design Figure 3-17: Installation using centering ring 1 Sensor 2 Centering ring 3 Bolts with fixing nuts 4 Hole 5 Hole 6 Gasket Install the flowmeter as follows: Push the first bolt 3 through hole 5 of both flanges. Screw on the nuts with washers at both ends of the bolt 3 but do not tighten them. Mount the second bolt through the holes 4. Place the measured value sensor 1 between the two flanges. Insert the gaskets 6 between measured value sensor 1 and flanges and align them. Check that the flanges are sitting concentrically. Install the remaining bolts, nuts and washers. But do not tighten the nuts yet. Turn the centering ring 2 anti-clockwise and align the flowmeter. Check that the gaskets 6 are concentric; they must not project into the pipe. Now tighten all nuts bit by bit alternately across the diagonal. 12/

28 3 INSTALLATION OPTISWIRL 4070 C Installing instruments in flange design Figure 3-18: Installing devices in flange design 1 Gasket 2 Bolts with fixing nuts Install the flowmeter as follows: Fix the flowmeter to one side of a flange using bolts and fixing nuts 2. Insert the gaskets 1 between sensor and flange and align them. Check that the gaskets are concentric; they must not project into the pipe. Install the gasket, bolts and fixing nuts on the other flange side. Align the meter and the gaskets so that they are concentric. Now tighten all nuts bit by bit alternately across the diagonal /

29 OPTISWIRL 4070 C ELECTRICAL CONNECTIONS Safety instructions DANGER! Only perform work on the electrical connections in the de-energized state. Take note of the voltage data on the nameplate! DANGER! Observe national installation regulations! CAUTION! All work on the flowmeter electrics may only be carried out by appropriately trained personnel. The regional occupational health and safety directives must always be observed. 12/

30 4 ELECTRICAL CONNECTIONS OPTISWIRL 4070 C 4.2 Connecting the signal converter Figure 4-1: Signal converter housing with housing cover 1 Housing cover of the electrical terminal compartment 2 Electrical connection terminals with the housing cover open 3 Terminal A current output - 4 Terminal A+ current output + 5 Terminal B pulse output - 6 Terminal B+ Pulse output + 7 PE terminal in housing 8 PE terminal on connection piece between sensor and converter The following procedures are to be performed: Unscrew the housing cover 1 of the electrical terminal compartment. Thread the connection cable through the leadthrough in the housing. Connect the cable for the current output and the cable for the optional pulse output as shown in the cable terminal diagrams below. To facilitate installation the connection plug can be removed from the meter. The plug and socket are configured in such a way as to prevent reverse polarity. Connect the earthing to terminal 7 alternatively use the PE terminal on the connection piece between sensor and converter 8. Tighten he cable glands. Manually screw on the housing cover with gasket /

31 OPTISWIRL 4070 C ELECTRICAL CONNECTIONS Current output, electrical data Power supply INFORMATION! The supply voltage has to be between 14 VDC and 36 VDC. This is based on the total resistance of the measuring loop. To determine this, add up the resistances of each component in the measuring loop (not including the flowmeter). The required supply voltage can be calculated using the formula below: U B =R ges. 22 ma + 14 V where U B = the minimum supply voltage and R tot = the total measuring loop resistance INFORMATION! The power supply has to be able to supply a minimum of 22 ma. 12/

32 4 ELECTRICAL CONNECTIONS OPTISWIRL 4070 C Load for HART communication INFORMATION! For HART communication a load of at least 230 ohm is required. The maximum load resistance is calculated as follows: R max = U B - 14 V 22 ma 1 + ma 2 HART 3 A A+ B B+ Figure 4-2: HART current output connection 1 Power supply 2 Optional amperemeter 3 Load DANGER! Use a twisted two-core cable to prevent electrical interference from impeding the DC output signal. In some cases a shielded cable may be necessary. The cable shield may only be earthed (grounded) at one place (on the power supply unit) /

33 OPTISWIRL 4070 C ELECTRICAL CONNECTIONS Pulse output, electrical data INFORMATION! When using the pulse output, two separate signal circuits are necessary if the pulse output is utilized together with analog signals. Each signal circuit requires its own power supply. The total resistance must be adapted so that the total current I tot does not exceed 100 ma Connection pulse output ma + 4 HART 3 A A+ B B+ Figure 4-3: Electrical connection pulse output 1 Power supply 2 Optional amperemeter 3 Load 4 e.g. counter 12/

34 4 ELECTRICAL CONNECTIONS OPTISWIRL 4070 C Configuring the pulse output The pulse output is a passive frequency, "open collector" output, isolated from the current interface and the sensor. It can be configured as a high current output or NAMUR output using a jumper on the amplifier board. Figure 4-4: Data pulse output 1 Pulse 2 closed 3 open 4 f<sub>max</sub> = 0.5 Hz ms /

35 OPTISWIRL 4070 C ELECTRICAL CONNECTIONS 4 Jumper in NAMUR setting R i = 900 Ω U max = 36 VDC The maximum frequency of the pulse output is: f max = 0.5 Hz Jumper in high current setting Open: Maximum voltage U max = 36 VDC Closed current I R < 1 ma Closed: Maximum current I max = 100 ma Voltage U < 2 VDC The maximum frequency of the pulse output is: f max = 0.5 Hz Figure 4-5: Jumper settings pulse output 1 Electronic insert 2 Jumper on board 3 Jumper in high current setting 4 Jumper in NAMUR setting 12/

36 4 ELECTRICAL CONNECTIONS OPTISWIRL 4070 C 4.5 Protection category The flowmeter meets all requirements of protection category IP 67. DANGER! After all servicing and maintenance work on the flowmeter, the specified protection category has to be ensured again Figure 4-6: Cable feedthrough Therefore it is essential to observe the following points: Use only original gaskets. They must be clean and free of any damage. Defective gaskets have to be replaced. The electrical cables used must be undamaged and must comply with regulations. The cables must be laid with a loop 1 upstream of the flowmeter to prevent water from getting into the housing. The cable glands 2 have to be tightened firmly. Align the meter in such a way that the cable gland is never pointing up 3. Close the unused cable glands using blanking plugs 4. Do not remove the specified outer sheath from the cable gland /

37 OPTISWIRL 4070 C ELECTRICAL CONNECTIONS Earthing connections The earthing can be done either by connecting the PE terminal in the housing or the PE terminal on the connection piece between sensor and converter. Both of these electrical connections are equally effective from a technical point of view. Figure 4-7: Earth connection 1 Electrical earthing connection on piece between sensor and converter. 2 Electrical earthing connection earth on housing DANGER! The flowmeter has to be earthed properly to achieve accurate measurement. The earthing cabel may not transfer any interference voltage. Do not use this earth cabel to ground any other items of electrical equipment. 12/

38 5 START-UP AND CONTROL OPTISWIRL 4070 C 5.1 Start-up INFORMATION! The device is always preset for the user and his application. Compare the data in the factory settings report with your specifications! The factory settings report is part of the scope of delivery Start After the device is switched on, the display shows the following in sequence Test, the type of meter, and the version number. Afterwards the device performs a self-test and switches to measurement mode. Here all of the parameters preset for the customer are analysed and checked for plausibility, and the last measured value is displayed Start-up and control INFORMATION! The flowmeter is largely maintenance-free. Observe the application limits in respect of temperature and medium. Additional information refer to Technical data on page /

39 OPTISWIRL 4070 C OPERATION Display and operating elements With the cover open the meter is operated by using the mechanical keys on the front; with the cover closed a bar magnet is used. CAUTION! The switching point for the magnetic sensors is located directly under the glass above corresponding symbol. Touch the symbol with the bar magnet only at right angles from the front. Touching it sideways can cause incorrect operation. Figure 6-1: Display and operating elements 1 Enter button (bar magnet) 2 Right button (bar magnet) 3 Up button (bar magnet) 4 Up button (mechanical) 5 Enter button (mechanical) 6 Right button (mechanical) 7 Display 12/

40 6 OPERATION OPTISWIRL 4070 C The mechanical keys and keys for the bar magnet are identical in their function. In this documentation the keys are represented as symbols to describe the operating functions: Mechanical Bar magnet Symbol ^ Table 6-1: Description of the operating keys 6.2 Operating principles Description of the key functions Switch from measurement mode to menu mode Switch to one menu level lower Open menu item and activate change mode In change mode: Move the input cursor one position to the right; after the last digit the input cursor jumps back to the beginning. In measurement mode: Switch between measured values and error messages Change between the menu items within a menu level In changing mode: Changing parameters and settings; running through the available characters; shifting the decimal point to the right. ^ Change one level up at the menu Return to measuring mode with a query whether the data should be accepted Table 6-2: Description of the opperation buttons Switch from measuring mode to menu mode Measuring mode Operation Menu mode kg/h Language /

41 OPTISWIRL 4070 C OPERATION Navigation within the menu structure Navigation within the menu is by means of the and ^ keys. Pressing the key takes you one menu level lower, ^ takes you one menu level higher. If you are already located at the lowest level (function level), you can use the key to go the the change mode, which can be used to set data and values. If you are located at the first level (main menu), you can use the ^ key to exit the menu mode and return to the measuring mode. Measuring operation Main menu Submenu Function Edit ^ Table 6-3: Navigation menu structure ^ ^ ^ ^ Changing the settings in the menu Using the and keys, navigate to the menu in which you wish to change a setting or a value. Use the key to activate change mode in the selected menu. i The current values or settings are displayed. Change the value or setting using the and keys. Confirm the new value or setting using the ^ key. Some menu items include several setting options. These are displayed one after the other by pressing the ^ key. i You return to the main menu. Press the ^ to save the settings you have made. i The following prompt is displayed: "Save Yes". Switch between "Yes" and "No" using the key. Switch between "YES" and "NO" using the key. Quit Yes Quit No ^ The changes are applied, an update is performed and the display switches back to measuring mode. Measuring operation. ^ The changes are discarded and the display switches back to measuring mode. Measuring operation. CAUTION! Each time parameters or settings are changed, the flowmeter carries out an internal plausibility check. If implausible inputs have been made, the indicator remains in the current menu, and the changes are not accepted. 12/

42 6 OPERATION OPTISWIRL 4070 C Example: Changing the default parameter from m 3 /h to l/min Steps Display Step Display m 3 /h 1 3x Language 2 3x Max. Flow 3 m 3 /h unit 4 3x L/min unit 5 ^ L/min 6 ^ Display unit 7 ^ Max. Flow 8 ^ Quit Yes 9 ^ L/min Measures in the event of faulty indications If the indications on the display or the responses to keypad commands are faulty, you have to do a hardware reset. Switch the power supply OFF and ON again. 6.3 Overview of the most important functions and indicators INFORMATION! A complete list of all functions and short descriptions is provided in the appendix. All default parameters and settings are adapted for the specific customer. Level Designation Explanation Language Select the menu language Max. Flow Maximum flow rate The value set is represented by a 20 ma current at the analog output.. If the current value exceeds the preset value, an alarm is indicated Min. Flow Minimum flow rate The set value does not represent the 4 ma value of the current output Timeconst. Time constant, damping value [s] Test I Check current output Test P Pulse pulse output Error Msg Error indicator Table 6-4: The most important functions Yes: Error messages are displayed No: Error messages are suppressed, but a flashing cursor in the top left corner shows that there are error messages /

43 OPTISWIRL 4070 C OPERATION 6 Measured variables Units Volume m 3 /h m 3 /min m 3 /s L/h L/min L/s ft 3 /h ft 3 /min ft 3 /s gal/h gal/min gal/s ImpGal/h ImpGal/min ImpGal/s Mass kg/h kg/min kg/s t/h t/min t/s lb/h lb/min lb/s Standard volume m 3 /h (std.) m 3 /min (std.) m 3 /s (std.) L/h (std.) L/min (std.) L/s (std.) ft 3 /h (std.) ft 3 /min (std.) ft 3 /s (std.) Table 6-5: Meter types, measured flow variables and units Media (equipment type) Liquids Vapors Gases Liquids Vapors Gases Gases 12/

44 6 OPERATION OPTISWIRL 4070 C 6.4 Error messages Error message Cause Corrective action No Signal No signal from the vortex amplifier Check plug connection At sensor problems contact Service Low Freq Sample frequency too low Contact Service. High Freq Sample frequency too high Contact Service. Low Flow High flow Inv.Config Flow rate lower than minimum flow rate q min Flow rate higher than maximum flow rate q max Configuration data in FRAM (non-volatile memory) is not valid Converter continues to show the current flow rate, but the accuracy of the measurement can be impaired. Correction procedures are depending of the application! If flow exceeds the maximum value, the sensor can be damaged physically. Check entire configuration! If error message persists, contact Service Amp. Fail Pre-amplifier section has failed Contact Service. Chk.Instal Flow signal quality is bad 1. Check q min 2. Excessive pipe vibrations and impeded flow profile Low Signal Vortex signal amplitude too low 1. q min ok? 2. if q min is ok, contact Service Hi.Signal Vortex signal amplitude too high Occurs in high-density media 1. Check q max 2. If q max is ok, contact Service L.Temp.Phy H.Temp.Phy H.Pres.Phy Operating temperature lower than physical limit Operating temperature higher than physical limit Operating pressure higher than physical limit Take corrective action within the process. Take corrective action within the process as quickly as possible, otherwise damage can occur to both sensor and converter. Take corrective action within the process as quickly as possible, otherwise damage can occur to both sensor and converter. Tsens Shrt Temperature sensor short-circuit Shows error on temperature sensor! Contact Service. Tsens Open Temperature sensor open circuit Shows error on temperature sensor! Contact Service. P.Sen.Fail Failure in pressure sensor Shows error at pressure sensor! Contact Service. Table 6-6: Error messages /

45 OPTISWIRL 4070 C SERVICE Replacing converter / LCD display The converter has to be replaced by a converter of the same type. The following parameters have to be observed for it: The item numbers must match: xxx The software version 2 must match. The basic version does not have any software identification The gas version has the designation: "gas" The steam version has the designation: "steam" Figure 7-1: Board, example 1 Connector for display 2 Version number, software identification 3 Connector for sensor 4 Spacer pins 5 Fixing screw (Phillips) 12/

46 7 SERVICE OPTISWIRL 4070 C Perform the following steps: Switch off the power supply Unscrew front cover Detach display from spacer pins. Detach display cable 1. Detach sensor cable 3. Undo three fixing screws 5 (Phillips). Remove converter. Insert new converter. Tighten three screws. Attach sensor cable 3 (caution: the sensor cable must not be above the spacer pin 4 risk of damage). Attach display cable 1. Mount display in required position, exerting uniform pressure over the entire surface. Screw the cover on tightly by hand /

47 OPTISWIRL 4070 C SERVICE Returning the device to the manufacturer General information This device has been carefully manufactured and tested. If installed and operated in accordance with these operating instructions, it will rarely present any problems. CAUTION! Should you nevertheless need to return a device for inspection or repair, please pay strict attention to the following points: Due to statutory regulations on environmental protection and safeguarding the health and safety of our personnel, manufacturer may only handle, test and repair returned devices that have been in contact with products without risk to personnel and environment. This means that the manufacturer can only service this device if it is accompanied by the following certificate (see next section) confirming that the device is safe to handle. CAUTION! If the device has been operated with toxic, caustic, flammable or water-endangering products, you are kindly requested: to check and ensure, if necessary by rinsing or neutralizing, that all cavities are free from such dangerous substances, to enclose a certificate with the device confirming that is safe to handle and stating the product used. 12/

48 7 SERVICE OPTISWIRL 4070 C Form (for copying) to accompany a returned instrument Company: Address: Department: Name: Tel. no.: Fax no.: Manufacturer's order no. or serial no.: The device has been operated with the following medium: This medium is: water-hazardous toxic caustic flammable We checked that all cavities in the device are free from such substances. We have flushed out and neutralized all cavities in the device. We hereby confirm that there is no risk to persons or the environment through any residual media contained in the device when it is returned. Date: Signature: Stamp: /

49 OPTISWIRL 4070 C TECHNICAL DATA Functional principle Vortex flowmeters are used to measure the volumetric flow of gases, vapors and liquids at completly filled pipes. The measuring principle is based on the Karman vortex street. The measuring tube contains a bluff body, behind which vortex shedding occurs. The frequency f of the vortex shedding is proportional to the flow rate v. The nondimensional Strouhal number S describes the relationship between vortex frequency f, width b of bluff body and and the mean flow velocity v: f = S. v b The vortex frequency is recorded at the sensor and evaluated at the converter. v b b b Figure 8-1: Functional principle 12/

50 8 TECHNICAL DATA OPTISWIRL 4070 C 8.2 Technical data Measuring system Application range Operating method / measuring principle Measured value Primary measured value Secondary measured value Flow measurement of liquids and gases Karman vortex street Number of separated vortices Operating and standard volumetric flow, mass flow Measuring accuracy Accuracy Re ± 0.75% for liquids Re ± 1% for gases and vapors < Re < ± 2% for liquids, gases and vapors 1 Repeatability ± 0.1% Stability ± 0.1% over a periode of 1 year Operating conditions Ambient temperature C Ex version C (non Ex version) Storage temperature C Product temperature C Measured substances Liquids gases and vapour Density taken into consideration when rating Viscosity < 10 cp Reynold s number Product pressure limit max. 100 bar, higher pressures on request Inlet conditions Inlet run Outlet run Dimensions and weights 20 x DN 5 x DN see chapter dimensions and weights Materials Sensor Housing /316L Hastelloy C4 Aluminium; /316L in preparation Sensor gasket /316L / FPM; Hastelloy C4 / FFKM 1 Accuracy pressure- and temperature-compensated Re /- 1.5% for gases and vapors; < Re < /- 2.5% for gases and vapors /

51 OPTISWIRL 4070 C TECHNICAL DATA 8 Power supply Ex- version Non Ex- version VDC VDC Current output Measuring range ma Over Range 20.8 ma +/- 1% (105% +/- 1%) Load Minimal 100 Ω; maximal R ges = ((U B - 14 V) / 22 ma) Error signal NAMUR NE 43 (-2.5 +/- 0.5%) /- 1.0% (105% +/- 1.0%) Maximal 22 ma (112.5%) Multidrop mode 4 ma Ddigital output HART Physical layer Device category Pulse output Pulse output Power supply non- Ex Power supply Ex Display and operating interface Local display Operating and display languages Process connections Process connection Flange version Sandwich version FSK Transmitter Pulse frequency max. 0.5 Hz 24 VDC as NAMUR or open < 1 ma, maximal 36 V, closed 100 ma, U < 2 V 24 VDC as NAMUR or open < 1 ma, maximal 30 V, closed 100 ma, U < 2 V 2 lines, 10 characters per line German, English, French EN or ASME flanges DN ; ½...12" DN ; ½...4" Protection category Protection category IP 66/67 Approvals ATEX ATEX II 2G EEx d ia [ia] IIC T6 FM Class 1 Div Pending 12/

52 8 TECHNICAL DATA OPTISWIRL 4070 C 8.3 Dimensions and weights a = 133 mm / 5.24 inch b = 105 mm / 4.13 inch c = 179 mm / 7.05 inch Flange version EN Size Pressure rating Dimensions [mm] Weight [kg] With Without DN PN d D L H l pressure sensor /

53 OPTISWIRL 4070 C TECHNICAL DATA 8 Size Pressure rating Dimensions [mm] Weight [kg] With Without DN PN d D L H l pressure sensor Option: Version with two transmitters Dimension H x 2 Specified weight kg 12/

54 8 TECHNICAL DATA OPTISWIRL 4070 C a = 133 mm / 5.24 inch b = 105 mm / 4.13 inch c = 179 mm / 7.05 inch Flange version ASME B16.5 Size Pressure rating Dimensions [mm] Weight [kg] DN PN d D L H l With pressure sensor Without pressure sensor 1/ / / / / / /

55 OPTISWIRL 4070 C TECHNICAL DATA 8 a = 133 mm / 5.24 inch b = 105 mm / 4.13 inch c = 179 mm / 7.05 inch Flange version ASME B16.5 Size Pressure rating Dimensions [inches] Weight [lb] DN PN d D L H l With pressure sensor Without pressure sensor 1/ / / / / / /

56 8 TECHNICAL DATA OPTISWIRL 4070 C Size Pressure rating Dimensions [inches] Weight [lb] DN PN d D L H l With pressure sensor Without pressure sensor /

57 OPTISWIRL 4070 C TECHNICAL DATA 8 a = 133 mm / 5.24 inch b = 105 mm / 4.13 inch c = 179 mm / 7.05 inch Sandwich version EN Size Pressure rating Dimensions [mm] Weight [kg] DN PN d D L H l With pressure sensor Without pressure sensor Option: Version with two transmitters Dimension H x 2 Specified weight kg 12/

58 8 TECHNICAL DATA OPTISWIRL 4070 C a = 133 mm / 5.24 inch b = 105 mm / 4.13 inch c = 179 mm / 7.05 inch Sandwich version ASME Size Pressure rating Dimensions [inches] Weight [lb] DN PN d D L H l With pressure sensor Without pressure sensor 1/ / / / / / /

59 OPTISWIRL 4070 C TECHNICAL DATA Flow tables Measuring range limits Size Qmin Qmax Qmin Qmax DN to EN DN to ASME B16.5 EN [m 3 / h] EN [m 3 /h] ASME B16.5 [m 3 / h] ASME B16.5 [m 3 /h] Water Air 15 1/ , / Values based on water at 20 C 15 1/ , , , , / Values based on air at 20 C and bar abs Flow rate limits Nominal diameters Minimum flow rates Maximum flow rates Product to EN to ASME [m/s] [m/s] Liquids DN15 DN300 DN 1/2" DN12" 0.5 x ( 998 / ρ ) x ( 998 / ρ ) Gas, vapor DN15 DN300 DN 1/2" DN12" 6 x ( 1.29 / ρ ) x ( 998 / ρ ) Minimum flow rates 0.4m/s - maximum flow rates 10m/s 2 Minimum flow rates 2m/s - maximum flow rates 80m/s 3 Minimum flow rates 2m/s - maximum flow rates 80m/s; DN15-45 m/s and DN25-70 m/s ρ = operating density [kg/m 3 ] 12/

VersaFlow Vortex 100 Vortex Flow Meter Technical Datasheet. Specification. The all-in-one solution. Highlights. Industries.

VersaFlow Vortex 100 Vortex Flow Meter Technical Datasheet. Specification. The all-in-one solution. Highlights. Industries. VersaFlow Vortex 100 Vortex Flow Meter Technical Datasheet 34-VF-03-05 September 3rd, 2008 Specification The all-in-one solution The VERSAFLOW is the only vortex flowmeter with integrated pressure and

More information

DK48 - DK800. Quick Start. Variable area flowmeter

DK48 - DK800. Quick Start. Variable area flowmeter Quick Start Variable area flowmeter General safety notes You can find the newest and/or additional information in the handbook, the datasheet, special manuals, certificates and the download center at www.krohne.com.

More information

OPTISWIRL 4070 Handbook

OPTISWIRL 4070 Handbook OPTISWIRL 4070 Handbook Vortex flowmeter KROHNE : IMPRINT ::::::::::::::::::::::::::::::::::::::: All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without the

More information

GA24 Technical Datasheet

GA24 Technical Datasheet GA24 Technical Datasheet Variable-area flowmeter Sturdy construction for several applications Local indication without auxiliary power Replaceable mounting parts KROHNE CONTENTS GA24 1 Product features

More information

VERSAFLOW VORTEX Handbook. Vortex flowmeter

VERSAFLOW VORTEX Handbook. Vortex flowmeter VERSAFLOW VORTEX Handbook Vortex flowmeter : IMPRINT :::::::::::::::::::::::::::::::::::::::::: V Notices and Trademarks Copyright 2013 by Honeywell WARRANTY/REMEDY Honeywell warrants goods of its manufacture

More information

Pro-V Multivariable Flowmeter Model M22 In-line Vortex

Pro-V Multivariable Flowmeter Model M22 In-line Vortex Pro-V Multivariable Flowmeter Model M22 In-line Vortex Pro-V TM Advantage: Volumetric or mass flow monitoring of most liquids, gases, and steam Multivariable meter delivers mass flow, temperature, pressure,

More information

7000 Series OPTIMASS Mass Flowmeters. Titanium Hastelloy Stainless Steel

7000 Series OPTIMASS Mass Flowmeters. Titanium Hastelloy Stainless Steel KROHNE 04/2003 7.02445.71.00 GR 7000 Series OPTIMASS Mass Flowmeters with single straight measuring tube Titanium Hastelloy Stainless Steel One tube no limits No limits with tube material No limits with

More information

Vortex Flow Meter Wafer or Flange Connection. - Steam - Liquid - Gas

Vortex Flow Meter Wafer or Flange Connection. - Steam - Liquid - Gas Vortex Flow Meter Wafer or Flange Connection - Steam - Liquid - Gas Working Principle & Circuit Diagram Working Principle When a column body placed in flowing fluids in pipe, a series of vortices will

More information

DK46 - DK800 Supplementary instructions

DK46 - DK800 Supplementary instructions DK46 - DK800 Supplementary instructions Variable area flowmeter Device category II2G with electrical internals Additional Ex manual KROHNE CONTENTS DK46 - DK800 1 Safety instructions 3 1.1 General... 3

More information

C-Flow Coriolis Mass Flow Meters

C-Flow Coriolis Mass Flow Meters Coriolis Mass Flow Meters Ready for Take Off The New Compact C-Flow Küppers Elektromechanik GmbH The C-Flow The C-Flow Coriolis Mass Meters consist of two components: KCE Transmitter KCM Transducer 2 KEM

More information

For Multi-Parameter Meters see mvx

For Multi-Parameter Meters see mvx Design Features VORTEX IN-LINE FLOW METERS For Multi-Parameter Meters see m Principles of Operation n No moving parts to wear or fail. n Electronics can be remotely mounted up to 30.5 m (0 ft). n No fluid

More information

For Multi-Parameter Meters see mvx

For Multi-Parameter Meters see mvx VORTEX IN-LINE FLOW METERS Design Features For Multi-Parameter Meters see mvx Principles of Operation VX No moving parts to wear or fail. Electronics can be remotely mounted up to 30.5 m (0 ft). No fluid

More information

3000 Series OPTIMASS Mass Flowmeter. Stainless Steel Hastelloy

3000 Series OPTIMASS Mass Flowmeter. Stainless Steel Hastelloy KROHNE 07/2005 7.02446.25.00 GR 3000 Series OPTIMASS Mass Flowmeter with single Z-shaped measuring tube Stainless Steel Hastelloy Status: 10/99 One tube no limits No limits with tube material No limits

More information

OPTIBAR P 1010 C Technical Datasheet

OPTIBAR P 1010 C Technical Datasheet Technical Datasheet Pressure transmitter with recessed diaphragm for general applications Wide variety thanks to modular design Measuring ranges up to 250 bar / 3750 psi High overload and temperature stability

More information

Flow Measurement SITRANS F X

Flow Measurement SITRANS F X Overview Using a vortex flowmeter with integrated pressure and temperature compensation such as the allows you not only to lower installation costs but also increase the measuring accuracy of the measuring

More information

Technical Data Sheet. OPTIMASS 7000 Custody Transfer Mass Flowmeter

Technical Data Sheet. OPTIMASS 7000 Custody Transfer Mass Flowmeter Technical Data Sheet KROHNE 12/2004 7.02516.21.00 CH/GR OPTIMASS 7000 Custody Transfer Mass Flowmeter with single straight measuring tube OIML R117 Pattern Approval Accuracy classes 0.3 / 0.5 / 1.0 / 1.5

More information

Flow Measurement SITRANS F X

Flow Measurement SITRANS F X SITRANS FX00 Overview The main applications of SITRANS FX00 can be found in the following sectors: Chemical Petrochemical Oil & Gas Power plants -Air - Heating - Cooling -Chilling Food & beverage - Pharmaceutical

More information

INSTRUCTION MANUAL Pressure Relief Device LPT

INSTRUCTION MANUAL Pressure Relief Device LPT INSTRUCTION MANUAL Pressure Relief Device LPT 5COV475800 LPT REV00 CONTENT: 1 SAFETY 1.1 Safety instructions 2 1.2 Specified applications 2 1.3 Safety notes on the equipment operation 2 2 PRESSURE RELIEF

More information

Vortex flowmeters. Product family introduction Principle of operation Product review Applications Key product features

Vortex flowmeters. Product family introduction Principle of operation Product review Applications Key product features Vortex flowmeters introduction Product review s Key product features This document should not be duplicated, used, distributed, or disclosed for any purpose unless authorized by Siemens. Page 1 Vortex

More information

VA40 - VA45 Technical Datasheet

VA40 - VA45 Technical Datasheet VA40 - VA45 Technical Datasheet Variable area flowmeter Local indication without auxiliary power Adaptable to meet customers' requirements Replaceable mounting parts KROHNE CONTENTS VA40 - VA45 1 Product

More information

VG5 Supplementary instructions

VG5 Supplementary instructions VG5 Supplementary instructions 2-wire / 10 GHz Radar (FMCW) Level Meter Supplementary Instructions for ATEX applications HYCONTROL CONTENTS VG5 1 General safety information 4 1.1 Scope of the document...

More information

Pro-V Multivariable Flowmeter

Pro-V Multivariable Flowmeter Pro-V Multivariable Flowmeter Model M23 Insertion Meter M23-V flowmeter provides cost-effective volumetric flow monitoring solution for most liquids M23-VT incorporates temperature sensing to provide a

More information

Mounting and Operating Instructions EB 3007 EN. Self-operated Pressure Regulators. Differential Pressure Regulators (opening) Type Type 42-25

Mounting and Operating Instructions EB 3007 EN. Self-operated Pressure Regulators. Differential Pressure Regulators (opening) Type Type 42-25 Self-operated Pressure Regulators Differential Pressure Regulators (opening) Type 42-20 Type 42-25 Type 42-20 Differential Pressure Regulator Type 42-25 Differential Pressure Regulator Mounting and Operating

More information

Integral type Differential pressure flowmeter VNT Series

Integral type Differential pressure flowmeter VNT Series Integral type Differential pressure flowmeter VNT Series OUTLINE VH series Wafer-Cone differential pressure flowmeter and high precision differential pressure transmitter are integrated into one flowmeter.

More information

GA24. GA24 Handbook. Variable-area flowmeter. KROHNE 02/ MA GA24 R01 en

GA24. GA24 Handbook. Variable-area flowmeter. KROHNE 02/ MA GA24 R01 en GA24 Handbook Variable-area flowmeter KROHNE : IMPRINT :::::::::::::::::::::::::::::::::: All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without the prior written

More information

Vortex flow sensor VA Di Ex-d with integrated, configurable transducer UVA in a flameproof enclosure for applications in explosive atmospheres

Vortex flow sensor VA Di Ex-d with integrated, configurable transducer UVA in a flameproof enclosure for applications in explosive atmospheres Vortex flow sensor VA Di Ex-d with integrated, configurable transducer UVA in a flameproof enclosure for applications in explosive atmospheres Drawing 1 VA Di ZG1 Ex-d Measured variables actual flow rate

More information

Certified according to DIN EN ISO 9001 Technical Datasheet TRICOR Series Mass Flow Meters

Certified according to DIN EN ISO 9001 Technical Datasheet TRICOR Series Mass Flow Meters www.kem-kueppers.com info@kem-kueppers.com Certified according to DIN EN ISO 9001 Technical Datasheet TRICOR Series Mass Flow Meters Description The Tricor Mass Flow Meters measure simultaneously mass

More information

T EK-COR 1100A Coriolis Mass Flowmeter

T EK-COR 1100A Coriolis Mass Flowmeter T EK-COR 1100A Coriolis Mass Flowmeter Instruction Manual Document Number: IM-1100A NOTICE Read this manual before working with the product. For personal and system safety, and for optimum product performance,

More information

Coriolis Mass Flow Meter

Coriolis Mass Flow Meter Coriolis Mass Flow Meter TMFW Series Mini Type Coriolis Mass Flow Meter Xi an Tosilon Automation Co., Ltd No.299, Daqing Rd, Lianhu District, Xi'an Shaanxi, China Tel: +86-29-8823 8550 info@tosilon.com;

More information

Vortex Flow Sensor VA Di Ex-d for flow measurement in gases; with integrated, configurable transducer UVA in Ex-d housing

Vortex Flow Sensor VA Di Ex-d for flow measurement in gases; with integrated, configurable transducer UVA in Ex-d housing Vortex Flow Sensor VA Di Ex-d Vortex Flow Sensor VA Di Ex-d for flow measurement in gases; with integrated, configurable transducer UVA in Ex-d housing Drawing 1 Measurable variables actual flow rate actual

More information

SITRANS F I Gardex. Flowmeter. SITRANS F I Intra/Prima. Flap Flowmeter. Contents. English. General 1. SITRANS F I Gardex 2. SITRANS F I Intra/Prima 3

SITRANS F I Gardex. Flowmeter. SITRANS F I Intra/Prima. Flap Flowmeter. Contents. English. General 1. SITRANS F I Gardex 2. SITRANS F I Intra/Prima 3 Contents SITRANS F I Gardex Flowmeter English General 1 SITRANS F I Gardex 2 SITRANS F I Intra/Prima 3 SITRANS F I Intra/Prima Flap Flowmeter Instructions A5E00129182 Version 02 09/2003 SITRANS F ED Automatic

More information

Magnetic level switch type MR783 Instruction Manual

Magnetic level switch type MR783 Instruction Manual Magnetic level switch Magnetic level switch type MR783 1. DESCRIPTION page 3 1.1 Operation page 3 1.2 Application page 3 1.3 Description page 3 2. SPECIFICATIONS page 3 2.1 Service Conditions page 4 2.2

More information

Pressure Measurement Single-range transmitters for general applications

Pressure Measurement Single-range transmitters for general applications Siemens A 207 Overview Application The SITRANS P Compact pressure transmitter is designed for the special requirements of the food, pharmaceutical and biotechnology industries. The use of high-grade materials

More information

Measurement accessories METPOINT OCV for the measurement in systems up to 40 bar

Measurement accessories METPOINT OCV for the measurement in systems up to 40 bar EN - english Instructions for installation and operation Measurement accessories METPOINT OCV for the measurement in systems up to 40 bar Dear customer, Thank you for deciding in favour of the METPOINT

More information

SITRANS P measuring instruments for pressure

SITRANS P measuring instruments for pressure SITRANS P measuring instruments for pressure Z series for gage pressure Siemens AG 008 Overview Design The main components of the pressure transmitter are: Brass housing with silicon measuring cell and

More information

The affordable consumption counter for compressed air and gases R 3 /4 (DN 20)

The affordable consumption counter for compressed air and gases R 3 /4 (DN 20) VA 420 new The affordable consumption counter for compressed air and gases R 1 /4 (DN 8) DN 15 R 3 /4 (DN 20) DN 25 R 1 1 /4 (DN 32) DN 40 R 2 (DN 50) With mit und and ohne without Flansch flange Intelligent

More information

RMG Meßtechnik GmbH. Publication No E

RMG Meßtechnik GmbH. Publication No E Electronic Turbine Meter TERZ 94 RMG Meßtechnik GmbH Publication No. 3.174-E P.O.ox 280 35502 utzbach (Germany) Tel.: +49 (0)6033 897-0 Fax: +49 (0)6033 897-130 E-mail: messtechnik@rmg.de Internet: http://www.rmg.de

More information

Mounting and Operating Instructions EB 3007 EN. Self-operated Pressure Regulators. Type Type Differential Pressure Regulators (opening)

Mounting and Operating Instructions EB 3007 EN. Self-operated Pressure Regulators. Type Type Differential Pressure Regulators (opening) Self-operated Pressure Regulators Type 42-20 Type 42-25 Differential Pressure Regulators (opening) Translation of original instructions Type 42-20 Differential Pressure Regulator Type 42-25 Differential

More information

Flow VA 520. incl. temperature measurement. Intelligent solutions for ac- for compressed air and gases R 1/4 (DN 8) DN 15 R 3/4 (DN 20) DN 25

Flow VA 520. incl. temperature measurement. Intelligent solutions for ac- for compressed air and gases R 1/4 (DN 8) DN 15 R 3/4 (DN 20) DN 25 VA 520 incl. temperature R 1/4 (DN 8) DN 15 R 3/4 (DN 20) DN 25 R 1 1/4 (DN 32) DN 40 R 2 (DN 50) Intelligent solutions for ac- for compressed air and gases work according to the approved calorimetric

More information

Flow VA 520. incl. temperature measurement. Intelligent solutions for ac- for compressed air and gases R 1/4 (DN 8) DN 15 R 3/4 (DN 20) DN 25

Flow VA 520. incl. temperature measurement. Intelligent solutions for ac- for compressed air and gases R 1/4 (DN 8) DN 15 R 3/4 (DN 20) DN 25 VA 520 incl. temperature R 1/4 (DN 8) DN 15 R 3/4 (DN 20) DN 25 R 1 1/4 (DN 32) DN 40 R 2 (DN 50) Intelligent solutions for ac- for compressed air and gases work according to the approved calorimetric

More information

Level transmitter LT 100

Level transmitter LT 100 KROHNE 09/2001 7.02363.21.00 GR/PRINTO Level transmitter LT 100 Status: 10/99 Variable area flowmeters Vortex flowmeters Flow controllers Electromagnetic flowmeters Ultrasonic flowmeters Mass flowmeters

More information

FTC130 Transmitter. Operating Manual

FTC130 Transmitter. Operating Manual Seite 1 von 11 FTC130 Transmitter Fast Thermal Conductivity Analyzer Operating Manual 1.09KD180724CWI1V05 Messkonzept GmbH Seite 2 von 11 1. Intended Use... 3 2. Description... 4 3. Measuring gases and

More information

Operating Instructions. Ball valve fitting according to ZB For pressure transmitter VEGABAR 82. Document ID: 50027

Operating Instructions. Ball valve fitting according to ZB For pressure transmitter VEGABAR 82. Document ID: 50027 Operating Instructions Ball valve fitting according to ZB 2553 For pressure transmitter VEGABAR 82 Document ID: 50027 Contents Contents 1 About this document 1.1 Function... 3 1.2 Target group... 3 1.3

More information

Mounting and operating instructions EB 2530 EN. Self-operated Pressure Regulator. Pressure Reducing Valve Type M 44-2

Mounting and operating instructions EB 2530 EN. Self-operated Pressure Regulator. Pressure Reducing Valve Type M 44-2 Self-operated Pressure Regulator Pressure Reducing Valve Type M 44-2 Type M 44-2, connection G 1 4, K VS = 0.15 Type M 44-2, connection G 1, K VS = 6 Fig. 1 Type M 44-2 Pressure Reducing Valve Mounting

More information

2 Overview. SITRANS P measuring instruments for pressure. Transmitters for gage and absolute pressure. Z series for gage pressure

2 Overview. SITRANS P measuring instruments for pressure. Transmitters for gage and absolute pressure. Z series for gage pressure Z series for gage pressure Overview Design The main components of the pressure transmitter are: Brass housing with silicon measuring cell and electronics plate Process connection Electrical connection

More information

Miniature flowmeters DK 46, DKR 46 DK 47, DK 48, DK 800

Miniature flowmeters DK 46, DKR 46 DK 47, DK 48, DK 800 KROHNE 1/22 7.236.22. GR Miniature flowmeters DK 46, DKR 46 DK 47, DK 48, DK 8 Variable area flowmeters Vortex flowmeters Flow controllers Electromagnetic flowmeters Ultrasonic flowmeters Mass flowmeters

More information

Vortex Flowmeter EF73

Vortex Flowmeter EF73 () 26 May 2011 ISO 9001/ ISO 14001 Kakogawa, Japan is approved by LRQA LTD. to ISO 9001/14001 Vortex Flowmeter EF73 Copyright 2009 by TLV CO., LTD. All rights reserved Rev 2-2009 Contents Contents... 2

More information

FLOW CONDITIONER «EMIS-VEKTA 1200» EMIS-VEKTA 1200.UM.PS v User manual Passport. «EMIS», CJSC Russia, Chelyabinsk

FLOW CONDITIONER «EMIS-VEKTA 1200» EMIS-VEKTA 1200.UM.PS v User manual Passport.   «EMIS», CJSC Russia, Chelyabinsk EMIS-VEKTA 1200.UM.PS 27.11.2012 v1.0.5 FLOW CONDITIONER User manual Passport www.emis-kip.ru «EMIS», CJSC Russia, Chelyabinsk Disclaimer "EMIS", CJSC reserves the right to make changes in the design and

More information

Vortex FlowmeterManual

Vortex FlowmeterManual Vortex Flowmeter Series MTGB Vortex FlowmeterManual Dalian Metern Measurement and Control System Co.,Ltd CONTENTS 1.GENERAL INFORMATION... 3 1.1 Product Description... 3 2.Technical Data... 4 3.Model and

More information

Variable Area Flow Meters

Variable Area Flow Meters PDF Published May, 04 Variable Area Flow Meters Chemline F Series Variable Area Flow Meters are ideal for plastic piping. They provide a combination of accurate visual flow rate measurement and clear flow

More information

ABB MEASUREMENT & ANALYTICS DATA SHEET. WaterMaster FEW530 Electromagnetic water meter

ABB MEASUREMENT & ANALYTICS DATA SHEET. WaterMaster FEW530 Electromagnetic water meter ABB MEASUREMENT & ANALYTICS DATA SHEET WaterMaster FEW530 Electromagnetic water meter 2 WATE RMASTE R F E W 530 ELECT ROMAGNE TIC WATE R ME TE R DS/FE W530/FE T3 -E N RE V. A Measurement made easy Cost

More information

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument OxyScan Graphic Operating Instructions UMS Micro-oxygen sensor 501 Microprocessor instrument Introduction Thank you for choosing the UMS Micro Oxygen Sensor 501 - a highly advanced product! Please read

More information

MASS-STREAM TM Instruction Manual. D-6200 Analog Mass Flow Meters / Controllers

MASS-STREAM TM Instruction Manual. D-6200 Analog Mass Flow Meters / Controllers MASS-STREAM TM Instruction Manual D-6200 Analog Mass Flow Meters / Controllers Doc. No.: 9.17.112A Date: 17-03-2017 ATTENTION Before installing and operating the instrument it is strongly recommended that

More information

Target disk flowmeter Series DP. Instructions manual DP65 DP500. R-MI-DP Rev.: 4 English version

Target disk flowmeter Series DP. Instructions manual DP65 DP500. R-MI-DP Rev.: 4 English version Target disk flowmeter Series DP Instructions manual DP65 DP500 R-MI-DP Rev.: 4 English version PREFACE Thank you for choosing the flowmeter series DP from Tecfluid S.A. This instruction manual allows the

More information

Installation, operating and maintenance Instructions for Seemag bypass level indicator

Installation, operating and maintenance Instructions for Seemag bypass level indicator Issue: S Date: 05-09-14 Type G35 General information The Seetru bypass magnetic level indicator, abbreviate SEEMAG, serves to show the filling level of fluids in tanks, basins, tubes etc. The Seemag operates

More information

Liquid level instruments BM 24 /BM 51

Liquid level instruments BM 24 /BM 51 KROHNE 09/2001 7.02362.21.00 GR/PRINTO Liquid level instruments / Status: 10/99 Variable area flowmeters Vortex flowmeters Flow controllers Electromagnetic flowmeters Ultrasonic flowmeters Mass flowmeters

More information

Differential Pressure Regulator Type Type 45-6 (0.1 to 1 bar, DN 15) Mounting and Operating Instructions EB 3226 EN

Differential Pressure Regulator Type Type 45-6 (0.1 to 1 bar, DN 15) Mounting and Operating Instructions EB 3226 EN Differential Pressure Regulator Type 45-6 Type 45-6 (0.1 to 1 bar, DN 15) Mounting and Operating Instructions EB 3226 EN Edition March 2008 Contents Contents Page 1 Design and principle of operation...................

More information

Pro-V Multivariable Flowmeter Model M22 In-line Vortex

Pro-V Multivariable Flowmeter Model M22 In-line Vortex Pro-V Multivariable Flowmeter Model M22 In-line Vortex Pro-V TM Advantage: Volumetric or mass flow monitoring of most liquids, Multivariable meter delivers mass flow, temperature, pressure, and density

More information

Operating Instructions DB41

Operating Instructions DB41 PKP Process Instruments Inc. 10 Brent Drive Hudson, MA 01749 Tel: +1-978-212-0006 Fax: +1-978-568-0060 PKP Prozessmesstechnik GmbH Borsigstrasse 24 D-65205 Wiesbaden-Nordenstadt Tel: 06122 / 7055-0 Fax:

More information

Torque Tube TB300 Digital Transmitters

Torque Tube TB300 Digital Transmitters Page 1 of 7 7E.300-E Issue 4-2009 Description Series TB300 torque tube liquid level instruments utilize the buoyancy exerted on a displacer when immersed in a liquid. The buoyancy on the displacer is proportional

More information

1 Overview. Pressure Measurement Single-range transmitters for general applications. 1/22 Siemens FI US Edition

1 Overview. Pressure Measurement Single-range transmitters for general applications. 1/22 Siemens FI US Edition Siemens AG 07 SITRANS LH00 Transmitter for hydrostatic level Overview Function U const. p U I EM The pressure transmitter SITRANS LH00 is a submersible sensor for hydrostatic level measurement. The pressure

More information

Monitoring device NTS 30

Monitoring device NTS 30 P R O D U C T I N F O R M A T I O N Electronic level- and temperature monitoring device Monitoring device NTS 30 Contents Description Description... 2 Operating... 3 Design and electrical data... 4 Circuit

More information

TRZ 03-K Volumeter. Proven Technology. Superior Performance. TRZ 03-K Volumeter is a Turbine Gas Flowmeter for Secondary Gas Metering.

TRZ 03-K Volumeter. Proven Technology. Superior Performance. TRZ 03-K Volumeter is a Turbine Gas Flowmeter for Secondary Gas Metering. TRZ 0-K Volumeter TRZ 0-K Volumeter is a Turbine Gas Flowmeter for Secondary Gas Metering. TRZ 0-K Turbine Gas Flowmeter continues the tradition of Honeywell s field-proven turbine meter technology, delivering

More information

Electronic Turbine Meter TERZ 94

Electronic Turbine Meter TERZ 94 Electronic Turbine Meter TERZ 9 PRODUT INFORMTION Reliable Measurements of Gas EETRONI TURINE METER TERZ 9 Methode of operation, construction 2 Methode of operation The TERZ 9 electronic turbine meter

More information

ATS430 turbidity sensor Retractable insertion assembly

ATS430 turbidity sensor Retractable insertion assembly A MEASUREMENT & ANALYTICS INSTRUCTION ATS430 turbidity sensor Retractable insertion assembly Measurement made easy 1 Introduction This publication details installation procedures for the retractable insertion

More information

Mounting and Operating Instructions EB 2172 EN. Series 43 Temperature Regulators. Type 43-5 and Type Type Type 43-6

Mounting and Operating Instructions EB 2172 EN. Series 43 Temperature Regulators. Type 43-5 and Type Type Type 43-6 Series 43 Temperature Regulators Type 43-5 and Type 43-7 Type 43-6 Type 43-6 Type 43-7 with flanged valve body, DN 32 to 50 Type 43-5 Type 43-7 with welding ends Mounting and Operating Instructions EB

More information

WHEATLEY Series 500 Swing Check Valve

WHEATLEY Series 500 Swing Check Valve Document Number: TC003001-13 Revision: 02 WHEATLEY Series 500 Swing Check Valve Installation, Operation, and Maintenance Manual TABLE OF CONTENTS BILL OF MATERIALS...3 SCOPE...5 INSTALLATION AND OPERATION

More information

x act i Precision- Pressure Transmitter for Food Industry, Pharmacy and Biotechnology Stainless Steel Sensor

x act i Precision- Pressure Transmitter for Food Industry, Pharmacy and Biotechnology Stainless Steel Sensor Precision- Pressure Transmitter for Food ndustry, Pharmacy and Biotechnology Stainless Steel Sensor accuracy according to EC 60770: 0. % FSO Nominal pressure from 0... 400 mbar up to 0... 40 bar Output

More information

Level switches Series LD Vibrating fork level switch for liquids and solids

Level switches Series LD Vibrating fork level switch for liquids and solids Level switches Series LD Vibrating fork level switch for liquids and solids Robust and compact Suitable for liquids (model LD61) and solids (model LD60) No moving parts, low maintenance Corrosion resistant

More information

Engineering Data Sheet

Engineering Data Sheet Page 1 of 6 CE MARKING AND THE PRESSURE EQUIPMENT DIRECTIVE 97/23/EC Valves must be installed into a well designed system and it is recommended that the system be inspected in accordance with the appropriate

More information

T EK-COR 1100A. Coriolis Mass Flowmeter. FLOW. Technology Solutions

T EK-COR 1100A. Coriolis Mass Flowmeter.   FLOW. Technology Solutions Technology Solutions T EK-COR 1100A Coriolis Mass Flowmeter FLOW www.tek-trol.com Flow Level Temperature Pressure Valves Analyzers Accessories TekValSys Introduction There can often be more than one type

More information

SG36KTL-M Quick Installation Guide. 1 Unpacking and Inspection

SG36KTL-M Quick Installation Guide. 1 Unpacking and Inspection SG36KTL-M Quick Installation Guide This guide provides a general instruction of the installation procedures of SG36KTL-M. In no case shall this guide substitute for the user manual or related notes on

More information

Pressure Measurement Single-range transmitters for general applications

Pressure Measurement Single-range transmitters for general applications Siemens AG 07 Overview Function U const. p U I EM Sensor Connection for auxiliary power supply Vent pipe with humidity filter Protective conductor connection/ Equipotential bonding Diaphragm The pressure

More information

Operating instructions Safety Rope Emergency Stop Switches ZB0052 / ZB0053 ZB0072 / ZB0073

Operating instructions Safety Rope Emergency Stop Switches ZB0052 / ZB0053 ZB0072 / ZB0073 Operating instructions Safety Rope Emergency Stop Switches UK ZB0052 / ZB0053 ZB0072 / ZB0073 7390878 / 02 03 / 2011 Contents 1 Safety instructions...3 2 Installation / set-up...4 2.1 Applications...4

More information

SG33KTL-M Quick Installation Guide. 1 Unpacking and Inspection

SG33KTL-M Quick Installation Guide. 1 Unpacking and Inspection SG33KTL-M Quick Installation Guide This guide provides a general instruction of the installation procedures of SG33KTL-M. In no case shall this guide substitute for the user manual or related notes on

More information

Flow metering in gases Introduction to COMBIMASS family

Flow metering in gases Introduction to COMBIMASS family Flow metering in gases Introduction to COMBIMASS family - physical principle, basic information - different types of instruments, pricing & delivery time Sales Training Germany October 17 th to 19 th,

More information

ACV-10 Automatic Control Valve

ACV-10 Automatic Control Valve ACV-10 Automatic Control Valve Installation, Operation & Maintenance General: The Archer Instruments ACV-10 is a precision automatic feed rate control valve for use in vacuum systems feeding Chlorine,

More information

Flow Measurement SITRANS F C

Flow Measurement SITRANS F C Overview The flow measuring principle is based on the Coriolis Effect. The FCS400 sensor s measuring tubes are energized by an electromechanical driver circuit which oscillates them at their resonance

More information

1 Overview. Pressure Measurement Transmitters for basic requirements. 1/16 Siemens FI SITRANS P220 for gauge pressure

1 Overview. Pressure Measurement Transmitters for basic requirements. 1/16 Siemens FI SITRANS P220 for gauge pressure Siemens AG 204 Overview The pressure transmitter SITRANS P220 measures the gauge pressure of liquids, gases and vapors. Stainless steel measuring cell, fully welded Measuring ranges 2.5 to 600 bar (36.3

More information

TransPort PT878GC Panametrics Portable Gas Ultrasonic Flowmeter. GE Sensing. Applications. Features

TransPort PT878GC Panametrics Portable Gas Ultrasonic Flowmeter. GE Sensing. Applications. Features Applications The TransPort PT878GC clamp-on gas flowmeter is a complete ultrasonic flow metering system for measurement of most gases, including: Natural gas Compressed air Fuel gases Erosive gases Corrosive

More information

Dual Solenoid Gas Valve Installation

Dual Solenoid Gas Valve Installation Installation IMPORTANT: These instructions are intended as a guide for qualified personnel installing or servicing FLYNN Gas Products. Carefully follow all instructions in this bulletin and all instructions

More information

Electronic gas volume corrector model DGVC-04

Electronic gas volume corrector model DGVC-04 Electronic gas volume corrector model DGVC-04 1. Introduction The following user s manual gives information about the installation, configuration, usage and storage of the electronic gas volume corrector

More information

Level switches Series LD Vibrating fork level switch for liquids and solids

Level switches Series LD Vibrating fork level switch for liquids and solids Level switches Series LD Vibrating fork level switch for liquids and solids Robust and compact Suitable for liquids (model LD61) and solids (model LD60) No moving parts, low maintenance Corrosion resistant

More information

Model PDT Dewpoint Transmitter

Model PDT Dewpoint Transmitter Model PDT Dewpoint Transmitter Instruction Manual Alpha Moisture Systems Alpha House 96 City Road Bradford BD8 8ES England Tel: +44 1274 733100 Fax: +44 1274 733200 email: mail@amsytems.co.uk web: www.amsystems.co.uk

More information

Sensor for Air Bubble Detection at Liquid Filled Tubes. SONOCHECK Type ABD06.xx. Operating Manual

Sensor for Air Bubble Detection at Liquid Filled Tubes. SONOCHECK Type ABD06.xx. Operating Manual Sensor for Air Bubble Detection at Liquid Filled Tubes SONOCHECK Type ABD06.xx Operating Manual Manufacturer: Model: Type: SONOTEC Ultraschallsensorik Halle GmbH Air Bubble Detector ABD06.xx SONOTEC Ultraschallsensorik

More information

Features. Description

Features. Description Features Fast response flow meter ideal for inert gas and liquid mass flow measurement applications Smart electronics permit field adjustment of critical flow meter settings Field validation of flow meter

More information

Differential Pressure Transmiter

Differential Pressure Transmiter Differential Pressure Transmiter Description The is an economical alternative to established differential pressure transmitters. It combines state of the art electronics and a high performance sensor;

More information

SonoMeter 30 Energy Meters

SonoMeter 30 Energy Meters Data Sheet SonoMeter 30 Energy Meters Description The Danfoss SonoMeter 30 is a range of ultrasonic, compact energy meters intended for measuring energy consumption in heating and cooling applications

More information

Instruction Manual Contact Pressure Vacuum Gauge

Instruction Manual Contact Pressure Vacuum Gauge MS10 Instruction Manual Contact Pressure Vacuum Gauge Table of Contents 1. Safety Instructions 2. Intended Applications 3. Product Description and Functions 4. Installation 5. Commissioning 6. Maintenance

More information

T EK-THERMAL 1700B Thermal Mass Flowmeter

T EK-THERMAL 1700B Thermal Mass Flowmeter T EK-THERMAL 1700B Thermal Mass Flowmeter Instruction Manual Document Number: IM-1700B NOTICE Read this manual before working with the product. For personal and system safety, and for optimum product performance,

More information

Instruction Manual Differential Pressure Transmitter

Instruction Manual Differential Pressure Transmitter DE61 Instruction Manual Differential Pressure Transmitter Table of Contents 1. Safety Instructions 2. Intended Applications 3. Product Description and Functions 4. Installation 5. Commissioning 6. Maintenance

More information

Installation and Operation Manual

Installation and Operation Manual Manual Static pressure transducer with controller Differential static pressure transducer with analog output and optional PI control mode Large diaphragm element with differential transformer Transducer

More information

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars.

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars. Simply a question of better measurement SCHMIDT Flow Sensor SS 20.261 The cost-effective alternative in pressurised systems up to 10 bars. Compressed air technology Industrial processes A cost analysis

More information

OPERATING INSTRUCTIONS MANUAL FOR QACV PNEUMATIC DOSING PUMP

OPERATING INSTRUCTIONS MANUAL FOR QACV PNEUMATIC DOSING PUMP This operating instructions contains safety information that if ignored can endanger life or result in serious injury. They are indicated by this icon. Use of this pump with radioactive chemicals is forbidden!

More information

KMT. Thermal Mass Flow Meter. for Compressed Air and Gases. Measuring range: m3n/h m3 N

KMT. Thermal Mass Flow Meter. for Compressed Air and Gases. Measuring range: m3n/h m3 N Thermal Mass Flow Meter for Compressed Air and Gases measuring monitoring analysing KMT Measuring range: 0.32... 63 m3n/h... 3.5... 1400 m3 N /h Accuracy: ±2,5% of reading + 0,15% of full scale p max :

More information

Wafer type. the product

Wafer type. the product . VFM5 Vortex Flow Meter Flanged type Wafer type Insertion type VFM5 is a powerful flow meter utilizing Karman vortex theory, which can meet the requirement of measuring the flow rate of various fluids

More information

COMPAFLOW. Compressed Air. Volumetric flow. Gas. Mass flow. Steam. Net volumetric flow. Liquid

COMPAFLOW. Compressed Air. Volumetric flow. Gas. Mass flow. Steam. Net volumetric flow. Liquid Volumetric flow Compressed Air Mass flow Gas Net volumetric flow Steam Liquid Universal compact orifice flow meter combines a compact orifice flow sensor and sotiphicated converter One-piece flow sensor,

More information

Operating Instructions of BROSA tension load cells, types 0111, 0113

Operating Instructions of BROSA tension load cells, types 0111, 0113 Operating Instructions of BROSA tension load cells, types 0, 03. Description of BROSA tension load cells. Structure and function The BROSA type 0 and 03 tension load cells transmit and measure the tensile

More information

Operating Instructions

Operating Instructions Operating Instructions Light-metal Ex d enclosures / flameproof enclosure > 8265/0 Empty enclosure > 8265/4 Control panel, integrated in Ex e enclosure > 8265/5 Control panel Table of Contents 1 Table

More information

TransPort PT878GC Panametrics Ultrasonic Portable Gas Flowmeter. GE Infrastructure Sensing. Features. Applications

TransPort PT878GC Panametrics Ultrasonic Portable Gas Flowmeter. GE Infrastructure Sensing. Features. Applications Applications The TransPort PT878GC clamp-on gas flowmeter is a complete ultrasonic flow metering system for measurement of most gases, including: Natural gas Compressed air Fuel gases Erosive gases Corrosive

More information