David Cockayne Centre for Electron Microscopy.

Size: px
Start display at page:

Download "David Cockayne Centre for Electron Microscopy."

Transcription

1 David Cockayne Centre for Electron Microscopy. Depart of Materials, University of Oxford JEOL JEM-2100 TEM USER GUIDE November 2014 JLH / NPY 1

2 Contents Control panels...4 Tracker Ball...7 Aperture controls....8 Computer control windows...9 Vacuum system diagram. 10 Using the Initial checks Filling the ACD with liquid nitrogen Switching on High Voltage (HT)..15 Switching on electron beam...15 Gun alignment...16 Condenser alignment..16 Condenser astigmatism.16 Objective Lens Reference Voltage 16 Spot Size Condenser aperture selection...17 Alpha angles what they mean Illumination system alignment...18 Specimen airlock.19 Removing specimen from airlock...20 Loading specimen into airlock...21 Specimen holders Aligning Voltage Centre...29 Electron diffraction..30 At the end of your session. 31 Heating the ACD back to room temperature..32 2

3 The JEM 2100 has the following specification: High-tilt polepiece with +/- 40 o tilt available Objective lens spherical aberration coefficient C s = 1.4 mm. Point resolution: 0.25 nm. Lattice resolution better than 0.2 nm. Magnification range ,200,000 X Low mag range: 50 6,000 X Diffraction camera length: mm. STEM facility Oxford Instruments XEDX system Gatan Orius CCD camera 3

4 The JEM 2100 is a sophisticated instrument with a wide range of capabilities. This basic guide will help you to get started with TEM and as you progress, you will then move on to more advanced techniques, including STEM and EDX. It operates at up to 200 kilovolts, the usual settings being 200 or 80 kv. It is controlled by a combination of computer commands via a mouse, and two control panels with buttons and knobs. Note that many of the functions can be operated by either the computer or the control panels. NOT used in this microscope Left hand control panel The buttons: [BEAM] used to switch on the filament [Room Lamp] not used Probe Control selects the mode of operation TEM is for normal imaging & diffraction; EDS has a small probe with high beam current for 4

5 EDX analyses via a stationary probe; NBD has small, (approximately) parallel probe for nanobeam-diffraction; CBD has a very high convergence angle for convergent beam diffraction. Alpha selector is used to optimise parallel illumination, depending on mag range. Spot Size controls the size of a focussed beam on the specimen. DEF/STIG Buttons: Turn on or off the various deflector coils, also condenser & objective stigmators. DEF/STIG knobs on both right and left-hand panels control the DEF/STIG functions as chosen above. As such these are termed multifunction knobs. SHIFT (both panels) knobs (again multi-function knobs) move the beam sideways in X- or Y- directions (BEAM SHIFT). If for example GUN is selected then they become GUN SHIFT. [NTRL] cancels whatever has been input into the selected DEF/STIG coils and returns the chosen DEF/STIG deflector to the engineers basic alignment. NOTE: the aperture controls on this particular instrument are manual, and not operated from this panel. CRS: these increase the power of the linked control by 10X 5

6 Right hand control panel [Z] - up and down adjusts height of specimen. [IMAGE WOBB X & Y] - gives a split image to help set height/focus the specimen [HT WOBB] varies high voltage to assist correction of the HT centre alignment. [F1 to F6] function knobs. Useful ones are F1 to lift viewing screen and F4 for gun alignment. [MAG 1] normal TEM range of magnifications. [MAG 2] sets mag at 40,000X for alignment. MAG/CAM L knob - sets magnification. In diffraction mode, adjusts camera length. 6

7 [Low MAG] very low mag range useful for locating specimen. Note that the objective lens is OFF in this mode and will become thermally unstable, causing specimen drift if left in this mode. Use sparingly. [SA MAG] sets image mode for accurate electron diffraction. [SA DIFF] switches to electron diffraction. [STD FOCUS] sets objective lens strength to optimum value. DIFF FOCUS focusses electron diffraction pattern. OBJ FOCUS focusses the image. Coarse & fine controls, with extracoarse (16X) button. NOTE: The microscope does not use film, so the exposure panel is not used. Tracker Ball controls The specimen X-Y shift is driven by a tracker ball, or by X- and Y- buttons. The FINE control button is not used on the

8 NOTE: in this orientation, the up/down, left/right arrows correspond to the image orientation on the viewing screen. B A C Aperture Controls The Condenser, Objective and Selected Area apertures are all controlled as shown above. Knob A is used to insert/withdraw an aperture, B is used to move it in or out a small distance, and C moves it sideways. The white circles on A indicate which aperture is in position (when white spot is aligned with black dot). Do NOT wind B too far anti-clockwise or it will detach completely! 8

9 Hard X-ray aperture This single aperture is used for minimising background X-rays in EDX spectra. It is very difficult to align, so don t attempt it yourself. If not visible when inserted, ask for help! The menu screen provides access to the main control systems through additional screens. HT is used to switch on high voltage and electron beam. SPC indicates specimen coordinates. Can be used to memorise and recall specimen positions. 9

10 Status Monitor indicates which systems are running useful if a major fault occurs. VAC schematic diagram of vacuum system, indicating pumps, valves and vacuum gauges. 10

11 Green lights - open valves. Orange lights Pirani gauges: PiG1- column vacuum PiG2 gun vacuum PiG3 camera chamber PiG4 specimen airlock PiG5 rotary pump Pink light Penning Gauge PEG1 (column) RP rotary pump DP - diffusion pump SIP ion pump PEG1 window indicates vacuum status Status windows indicate vacuum readings in various gauges 11

12 Operation panel for main (standard) system. Note that most of these are duplicated by buttons & knobs on the two consoles Operation panel [Stage] controls minimum steps sizes for X/Y specimen shift & tilt [NTRL] sets stage coordinates to zero. NOTE: Super Fine control is not used. NOTE: [Aperture] is not used. 12

13 Alignment panel for routine alignments. Under normal circumstances you will not need to Save and Load alignment files. Please do not do this unless you have discussed this in advance with EM Facility staff 13

14 USING THE JEOL JEM-2100 At the start of your session, carry out the following initial checks The microscope should be found in the following standard conditions: Filament OFF High Voltage set at 200 kv and OFF (unless a previous user the same day leaves it on for the next user) MAG at 500,000X Stage coordinates all at zero (neutral) Specimen OUT and non-tilt holder (without the end piece) IN the column Objective and Selector apertures OUT Orius camera OUT EDX & STEM detectors OUT Check that PEG1 Status in VAC diagram is at READY. If it is not READY go for help. Check the log sheet for any messages from previous users 14

15 FILL COLD TRAP RESERVOIR WITH LIQUID N 2 Obtain LN 2 from the cupboard and remember i) to use protective eyewear & gloves, and ii) lock the door afterwards. Get the code for padlock from one of the EM facility staff following safety training from them. 1. Make sure covers are on the TEM window and binoculars. LN 2 splashes on glass will crack it (VERY expensive!) 2. Insert funnel into the dewar neck 3. Carefully pour LN 2 to fill dewar (use goggles) and insert filler cap 4. After ~15 mins. refill tank it will then remain cold for several hours. SWITCH ON HIGH VOLTAGE (HT) In main control window check that the HT indicator: HT status is Ready Select operating voltage in High Voltage Control window using up/down buttons. Normally used at either 200 or 80 kv. Use 0.5 kv steps. If you are the first user of the day, you must first set the HT to 180 kv then increase the HT slowly to 200kV as per training instructions. Click HT [ON] button. Setting the accelerating voltage appears. No other windows can be accessed while HT is running up. HT icon in main control widow turns green when HT is on. Dark current (beam current without electron emission) should read around 41 µa for 80 kv, and 101 for 200 kv. SWITCH ON ELECTRON BEAM Remember: the beam will not switch on unless the specimen holder is in position and high voltage is on. In High Voltage Control window, click Filament ON, or press the [Beam Switch] button on left panel. Filament will gradually come on. Check emission current should be ~5 6 µa greater than dark current. If this is not the case and you are unsure ask for help. 15

16 GUN ALIGNMENT Set spot size in TEM mode to 2 and alpha 3 (left panel) Set Mag to 40,000X Bring beam to screen centre (Shift X & Y) Engage [GUNA] (F4 on right console) or GUN in the DEF Select menu in Alignment Panel window. Optimise brightness by X & Y DEF/STIG knobs (L & R panels). The current density readout in pa/cm 2 will help you with this. Cancel F4 or GUN to memorise settings. CONDENSER ALIGNMENT: Set MAG to 40,000X and focus beam (Brightness knob) Set Spot size 1 Click DEF Select GUN in Alignment Panel window and centre beam with SHIFT X & Y Set Spot size 5 and focus beam. Engage [BRIGHT TILT] switch (right panel) or click DEF Select CLA in Alignment panel. Then centre beam with SHIFT X & Y. Repeat process until both spot 1 and 5 remain centred. Spot 2 is normally used for TEM a combination of good brightness and beam coherence. CONDENSER ASTIMATISM CORRECTION Focus and centre beam using BRIGHTNESS & SHIFT controls. Expand beam on either side of focus. If beam becomes elliptical or severely elongated correct it using either [COND STIG] (rt panel) or in Alignment panel click CL STIG, and use DEF/STIG knobs (both panels) to make spread beam circular. Cancel COND STIG or CL STIG. SETTING OBJECTIVE LENS REFERENCE CURRENT Because the vertical position of the specimen in the objective lens is critically important, the lens must first be set to its optimum current, and 16

17 the specimen is then raised or lowered into the correct plane. Set MAG at 40,000X. Press [STD FOCUS] switch (Right panel). This now sets the Objective Lens to its optimum voltage/current. SPOT SIZE Choose an appropriate spot size. Spot 1 or 2 will be suitable for most TEM work CONDENSER APERTURE Choose the appropriate aperture #1: (largest) 200 µm for maximum brightness, but low coherence. #2: 70 m, use this for routine TEM work. #3: 50 µm for better coherence, better for HREM, but less bright. #4: 10µm use this for small-probe work, but very dim, and may cause contamination. Centre the aperture in the usual way, Recheck gun tilt and adjust if necessary. ALPHA ANGLE The illumination system consists of three condenser lenses and a condenser mini-lens. The system allows you to select an optimum beam convergence setting for various magnification ranges: with the correct alpha setting you can have a (approximately) parallel beam onto your specimen. Choose an appropriate alpha setting for your work. It also provides a very wide degree of control over small-probe conditions, see below. In TEM mode there are 3 alpha settings available. In the small probe modes there are up to 9 alpha settings available. Alpha 3 is good for lower magnification ranges, i.e. up to 100,000X, (or 2,500X into the GIF) (This will give a "parallel" beam of 2.5 m diameter on specimen. i.e. 10cm on screen at 40,000X) Alpha 2 is good for medium magnification ranges, i.e. from to 100,000X 400,000X. Alpha 1 is suitable for high magnification, i.e. above 400,000X. (This gives a "parallel" beam of 0.5 m diameter on specimen. ~ 2cm on screen at 400,000X) If you have a wrong alpha setting you will find strong distortion in the illumination, particularly near the edges of the beam. 17

18 ILLUMINATION SYSTEM ALIGNMENT Beam position on specimen is controlled by 2 sets of deflector coils which are used to define beam position and tilt. These adjustments (beam tilt purity) are to make sure the beam doesn t tilt when it is shifted, and doesn t shift when tilted. Adjusting TILT X and Y so beam spot doesn t shift when beam is tilted (ESSENTIAL) Set MAG to 40,000 and focus beam to small spot, centred on screen. Engage [BRIGHT TILT] on left panel, or click DEF Select CLA in Alignment panel. Click Compensator Tilt in Alignment panel Click Wobbler Tilt X button in Alignment panel - spot splits into 2 in X direction. Make split spot into one using DEF/STIG X knob. If spot is also split in Y direction click Compensator Angle in Alignment panel, and correct using DEF/STIG X knob. Cancel Compensator Angle. Cancel Wobbler Tilt X button in Alignment panel. If beam is off-centre, engage [BRIGHT TILT] (left panel) and centre with Shift knobs Repeat process for TILT Y Adjusting SHIFT X & Y so beam doesn t tilt when shifted (OPTIONAL) Set up diffraction [SA DIFF] (right panel) and turn brightness knob fully clockwise Adust DIFF FOCUS knob to produce a caustic spot (image) on screen. Click Compensator Shift in Alignment panel Click Wobbler Shift X caustic spot splits into 2 parts Unify split spot by DEF/STIG knob X If spot also splits in Y direction click Compensator Angle in Alignment panel and unify with DEF/STIG knob X. Cancel Compensator Angle Cancel Wobbler Shift X Repeat process for Y- 18

19 Reminder notes on SPECIMEN EXCHANGE Please be very careful when loading & unloading specimens. Take your time and ask for help if unsure A B Specimen Airlock Follow these steps carefully Turn off the beam (if not already off) by either pressing the BEAM button on the left hand control panel or selecting the FILAMENT button in the HT window. Wait until beam if fully off. The HT can stay turned on. Reset the sample holder to its neutral position by double clicking on the Stage Neutral button in the top right hand corner of the left 19

20 monitor (black square with white letters). Wait until X, Y & Z values all are at zero &4 For removal of double-tilt holders, disconnect Y-tilt cable from airlock [B]. 1. Gently pull the holder out until it stops. Do this steadily and without placing any lateral or vertical pressure on the holder. 2. Now rotate the holder ~80 degrees ANTICLOCKWISE. DO NOT pull on the holder while rotating it. 3. Gently slide the holder out (~ 1 cm) until it stops. 4. Carefully rotate the holder the final ~10 degrees anticlockwise. Again DO NOT pull on the holder, as this will let air into the column 5. Switch the airlock switch to AIR and turn on gas line. Wait for the pressure airlock to reach air pressure/ SPECIMEN vacuum display will change to AIR and vacuum level will rise. After ~30s the airlock will be at atmospheric pressure. 6. Carefully remove holder from the airlock 7. Switch off gas line. 6 20

21 See pages for detailed instructions on loading different specimen holders. Before putting the specimen holder into the airlock: Check that the O-rings are clean and free of fluff or dirt. If fibres are found on an O-ring, carefully remove them with a cocktail stick. Do NOT use tweezers! Also, check the brown ceramic ring for dirt or grease any dirt or grease will cause specimen drift. And remember do NOT touch any part of a specimen holder beyond the O-rings with bare skin ALWAYS use neoprene or plastic gloves provided. Do not wipe the O-ring with a tissue or any other material. It has a thin coating of special high-vacuum grease already, so do not add any more &5 3 Wait for green light 1 1. Insert the holder with the small guide pin positioned in the airlock guide slot at 9 o clock [A]. Guide the holder gently into the column until it comes to a stop. Push the end of the holder gently to ensure a good seal and then release it. It should remain in position. If 21

22 using a double-tilt holder, you should hang the Y-cable anticlockwise over the holder body, to prevent twisting the holder Try not to put any sideways or up/down pressure on the holder. Note: Do NOT rotate the specimen holder at this time. 2. Set the airlock switch to PUMP. The orange lamp lights up, indicating that the airlock is pumping. The SPECIMEN vacuum level will begin to drop from 252μAmps. 3. Wait until the green light on the goniometer is lit and the SPECIMEN vacuum indicates VAC READY or the Specimen/PIG4 (Vacuum Window) value is 78μAmps. 4. To insert holder into the column - rotate the holder clockwise until it stops after ~ 10 degrees. 5. The holder will then move in by ~1cm. 6. Rotate the holder clockwise by a further ~80 degrees. The column vacuum will now pull it inwards so do not let go! Gently allow the sample holder to slide in until it reaches a stop. Finally, gently pull holder out a few mm and then return it back into position. This releases any strain on the O-rings. If using a double-tilt holder, now insert the plug into [B]. Now select the correct holder in the Control window (top right-hand corner). This tells the microscope which holder is in use. Switch on electron beam. Find an image of your specimen, and bring it into approximate focus using Z-up or Z-down buttons (right panel). X- or Y- image wobblers (right panel) can be helpful. 22

23 There are three holders available: You must observe the following rules when you use any of them. 1. You must ALWAYS wear plastic gloves when touching specimen holders. 2. After you load your specimen, check both O-rings carefully. and gently remove any dust or dirt particles. 3. Use a lint-free tissue (red and white box) (NOT an ordinary tissue!) to remove any grease from the brown ceramic ring at the end of the specimen rod. This surface connects the rod to the goniometer and if it is in any way greasy or dirty, there will be bad specimen drift. Single-tilt holder. 30 degrees of tilt around the rod axis. Robust and easy to use. Berryllium double-tilt holder. This is the most delicate (and most expensive!) holder, so please handle it carefully. The Be surrounding the sample gives clean, low-background EDX spectra with suitable specimens. This holder must NEVER be used for any magnetic specimens. The small tabs which hold the specimen are in turn held by small screws. Use the correct size of screwdriver, and do NOT overtighten these screws, as this will damage the gimbal (replacement cost: 7K). Strong double-tilt holder. Not available for general use see Sergio Lozano-Perez if you need to use this one. This is built from titanium, and MUST be used if your specimen is magnetic (e.g. steel). Specimens are held in position by a small screw. These screws are VERY expensive (~ 500 each), so be very careful not to lose one. The tilting gimbal pivots on two very small sapphire bearings, which are also very expensive (~ 1,000). You MUST ALWAYS use the correct support pillar when loading specimens. This supports the gimbal and prevents any excessive strain on the bearings. Do NOT over-tighten the retaining screw. 23

24 SINGLE-TILT HOLDER A Bronze clip B Specimen mounting plate Note: the bronze clip has a slot which fits under screw A; screw B acts as a pivot. ALWAYS use the mounting jig as shown here. Make sure the mounting plate is secure and straight in the holder rod after loading your specimen. DON T touch the mounting jig with bare fingers, or you will transfer grease to it. 24

25 LOW BACKGROUND Beryllium DOUBLE-TILT HOLDER Bronze clip Retaining screws Be cover plate Bronze clip Low background holder correctly positioned on mounting jig 25

26 NOTES: 1. This holder is very fragile! 2. The specimen loading side is the bottom side. The other side is the one that faces the X-ray detector. Make sure your specimen grid has the carbon side down when loading your specimen. 3. ALWAYS use the mounting jig to support the gimbal. 4. Place the mounting jig in position in the platform and lower the gimbal very gently onto the mounting jig. Tighten the perspex screw to keep the holder steady. 5. The sequence is: specimen, then the Be spacer ring, then finally the Be cover plate. 6. Use the Be spacer ring on top of your specimen if it is too thin. 6. DO NOT undo the retaining screws completely just enough to release the bronze clips from the locating pins. 7. DO NOT over-tighten the retaining screws. NOTE: the gimbal, spacer and cover plate are made of Be. This is very TOXIC. Handle these with great care! Using the holder: 1. X & Y tilt angles are limited to +/- 30 o with this holder. 2. When using this holder select Beryllium double-tilt holder in the Controller menu. 26

27 STRONG DOUBLE-TILT HOLDER Retaining screw ALWAYS use the mounting jig as shown above, and always use the special tool to tighten the retaining screw. 27

28 NOTES FOR USING STRONG DOUBLE-TILT HOLDER: 1. The specimen should be located 150 m up from the base of the recess in the gimbal. Use spacer rings as shown, with specimen facing UPWARDS. Spacer arrangement for a standard Cu grid Spacer arrangement for a 100 m thick specimen Spacer arrangement for a 150 m thick specimen. 28

29 ALIGNING THE VOLTAGE AXIS This step assumes that the other alignments described have already been carried out. This is the optimum alignment axis for high resolution. Set MAG to 100,000X or greater and focus image. Spread beam to fill screen. Engage [Bright Tilt] and [HT Wobb] (left & right panels) or Click Wobbler HT, and DEF Select CLA in Alignment panel Image will expand and contract around a point. Bring this point to the screen centre by DEF/STIG X & Y knobs. Cancel [HT Wobbler] or Wobbler HT 29

30 ELECTRON DIFRACTION: note that several different modes are available: Selected Area (SAED) in normal TEM mode, where diffraction area is limited by selector aperture Nano-beam (NBD) & Convergent Bam (CBD) special modes where very small beam diameters are used to define diffracting area. SELECTED AREA DIFFRACTION Select [SA MAG] (right panel) and set at desired value. Move object of interest to screen centre Insert selector aperture and focus edge of aperture. Note effective aperture diameters: µm, corresponds to ~ 1600 nm at specimen µm, corresponds to ~ 800 nm at specimen µm, corresponds to ~ 350 nm at specimen µm, corresponds to ~ 180 nm at specimen Press [SA DIFF], and select appropriate camera length using MAG/CAM L knob Reduce brightness by turning Brightness knob anticlockwise Use DIFF FOCUS knob to adjust sharpness of diffraction spots Bring central spot to screen centre with [PLA] button (left console) and DEF/STIG knobs. Cancel [PLA] Beam stopper can be inserted to cover central bright spot. If you are tilting the specimen, you can track any lateral movement by defocussing the diffraction pattern the selected area then appears in the central spot. In this way you can see both the image and its diffraction pattern simultaneously. 30

31 At the end of your session: Check the booking schedule to see if someone will be using the next session. If so, leave high voltage switched ON and the cold trap filled. Filament OFF High Voltage set at 200 kv and OFF unless someone else is using the microscope after your session. MAG at 500,000X Stage coordinates at zero (neutral) Covers on window & binoculars Double-tilt specimen holder OUT and non-tilt holder (without the end piece) IN the column. Make sure you remove your specimen from the holder you were using. Objective and Selector apertures OUT Orius camera OUT EDX & STEM detectors OUT All holders (apart from the non-tilt rod) must be in their plastic sleeves and in their boxes with the lids securely fastened. Specimen loading desk left tidy. Transfer all your data to the Z-drive Leave all PCs ON If no-one is using the next session, heat the cold trap (Anti-Contamination-Device) to return it to room temperature (see p. 30). Sign and fill in the log sheet, indicating which techniques you have used, and any problems. Also, report any problems to em-faults. 31

32 Returning the ACD to room temperature If the ACD is allowed to warm up when the LN 2 is depleted it will release organic vapour into the column and into the vacuum pumps which will degrade their performance. The following procedure is used to bring the ACD up to room temperature in a controlled manner, protecting the vacuum system. 1. Put covers on window and binoculars 2. Remove cap from reservoir 3. Goggles on! 4. Insert ACD heater into dewar, engage the plug into the 2-pin socket 5. Select TEM Controller - Maintenance window ACD & Bake 6. In Bake Out/ACD window select ACD Heat and click On button. This closes valves in the vacuum system and turns off the ion pump. The system is on a timer and will switch off in ~ 20 minutes. Note that you cannot pump out the airlock during this time, so make sure the specimen holder is in position before doing this operation. 32

Scanning Electron Microscope JEOL JSM F

Scanning Electron Microscope JEOL JSM F Scanning Electron Microscope JEOL JSM - 7600F How to Use This Manual You will be promised to obtain successful results if you follow this step by step manual gently and carefully. In order to that, you

More information

TYPICAL OPERATION PROCEDURES

TYPICAL OPERATION PROCEDURES JSM-7500F TYPICAL OPERATION PROCEDURES Reminders! Remember to sign in on the sign up sheet. WEAR GLOVES: always wear gloves when handling your samples, the sample holders and the specimen exchange rod.

More information

LEO SEM SOP Page 1 of 9 Revision 1.4 LEO 440 SEM SOP. Leica Leo Stereoscan 440i

LEO SEM SOP Page 1 of 9 Revision 1.4 LEO 440 SEM SOP. Leica Leo Stereoscan 440i LEO SEM SOP Page 1 of 9 LEO 440 SEM SOP Gun (Filament) Column Manual Valves Chamber Window Chamber Stage Movement Leica Leo Stereoscan 440i 1. Scope 1.1 This document provides the procedure for operating

More information

R I T. Title: Amray 1830 SEM Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 09/29/03 1 SCOPE 2 REFERENCE DOCUMENTS

R I T. Title: Amray 1830 SEM Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 09/29/03 1 SCOPE 2 REFERENCE DOCUMENTS Fabrication Laboratory Revision: A Rev Date: 09/29/03 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Amray 1830 SEM. All users

More information

LEO 1525 FEG SEM Standard Operating Procedures Mar. 6, 2012 For additional assistance, please contact the facility manager.

LEO 1525 FEG SEM Standard Operating Procedures Mar. 6, 2012 For additional assistance, please contact the facility manager. LEO 1525 FEG SEM Standard Operating Procedures Mar. 6, 2012 For additional assistance, please contact the facility manager. Please contact in case of emergency: SEM microscopist: Mr. Eric Miller, 7-0789

More information

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone:

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone: MJB4 Mask Aligner Operating Procedure Effective Date: 07/12/2012 Author(s): Jiong Hua Phone: 402-472-3773 Email: jhua2@unl.edu 1 1 Introduction 1.1 Key Words Karl Suss MJB4 Mask Aligner, Optical Lithography,

More information

SEM LEO 1550 MANUAL. I. Introduction. Generality

SEM LEO 1550 MANUAL. I. Introduction. Generality SEM LEO 1550 MANUAL RESERVATION POLICY: 2 booking slots maximum per day and per person (ie. 1h). 6 booking slots maximum per week and per person (ie. 3h). Reservation names must correspond to the operators.

More information

OPERATION OF THE DIMPLER

OPERATION OF THE DIMPLER OPERATION OF THE DIMPLER After thinning your sample to ~80 μm you can now do a dimpling process to thin the center up to 10 μm. When you walk in and use the DIMPLER it should already be calibrated and

More information

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Sensitivity: 1x10 11 A m 2 Applied DC fields: 0 T to 5 T Applied AC fields: 0 G to 3 G (zero-to-peak), 0.01 Hz to 1000 Hz Temperatures

More information

Equipment Standard Operating Procedure Greg Allion and Kimberly Appel

Equipment Standard Operating Procedure Greg Allion and Kimberly Appel Date Created: May 3, 2004 Date Modified: June 1, 2005 MA6/BA6 Mask Aligner Equipment Standard Operating Procedure Greg Allion and Kimberly Appel 1. Purpose 1.1. Photolithography involves transferring a

More information

THERMAL EVAPORATION UNIT (for Al evaporation)

THERMAL EVAPORATION UNIT (for Al evaporation) THERMAL EVAPORATION UNIT (for Al evaporation) System Owner: NeerajPanwar 9619507210 panwar.iitr@gmail.com System Operator: 1. Bhimraj Sable 9930189878 bhimrajsable71@gmail.com Authorized User: 1. Ramesh

More information

[USING THE VITROBOT April 9, Using the Vitrobot

[USING THE VITROBOT April 9, Using the Vitrobot Using the Vitrobot The Vitrobot Mk 3 (aka Mark III, Mark 3, etc.) can control both the local temperature and the local humidity during the process of plunge-freezing a grid for cryoem. It does require

More information

Warnings: Notes: Revised: October 5, 2015

Warnings: Notes: Revised: October 5, 2015 Karl Suss MA6 Mask Aligner Standard Operating Procedure Faculty Supervisor: Prof. Robert White, Mechanical Engineering (x72210) Safety Office: Peter Nowak x73246 (Just dial this directly on any campus

More information

Angstrom E-Beam Instructions. PROCESS CHECKS: the tooling factors of each metal; Ti/Au layer for wire bonding pull test.

Angstrom E-Beam Instructions. PROCESS CHECKS: the tooling factors of each metal; Ti/Au layer for wire bonding pull test. Angstrom E-Beam Instructions Tool Manager: Joe Palmer (jpalmer@princeton.edu) Office: 8-4706; Cell:609-751-1353 Backup: David Barth (dbarth@princeton.edu) Office: 8-4626; Cell: 610-405-8227 PROCESS CHECKS:

More information

Cryo-Evaporator Operation

Cryo-Evaporator Operation Cryo-Evaporator Operation Cara Ricci November 25, 2003 Thin Film Deposition THE UNIVERSITY OF TEXAS AT DALLAS ERIK JOHNSON SCHOOL OF ENGINEERING DOCUMENT NUMBER: FA2003-TF-008 EDITION: 1.1 PAGE: 1 of 28

More information

SPUTTER STATION STANDARD OPERATING PROCEDURE

SPUTTER STATION STANDARD OPERATING PROCEDURE SPUTTER STATION STANDARD OPERATING PROCEDURE Purpose of this Instrument: This instrument is used for deposition of thin metal or oxide films. Source materials supplied by WVU Shared Research Facilities:

More information

Karl Suss MA6 Mask Aligner SOP

Karl Suss MA6 Mask Aligner SOP Page 1 of 11 Karl Suss MA6 Mask Aligner SOP Safety UV Exposure: The high energy light produced by the high pressure Mercury Xenon lamp can cause eye damage and skin burns. Be sure that the light guards

More information

Login to ilab Kiosk. Revised 05/22/2018. Load your sample:

Login to ilab Kiosk. Revised 05/22/2018. Load your sample: Login to ilab Kiosk Load your sample: 1. Check: The analysis chamber pressure is

More information

1.1 Equipment: substrate, wafer tweezers, metal targets 1.2 Personal Protective Equipment: nitrile gloves, safety glasses 1.

1.1 Equipment: substrate, wafer tweezers, metal targets 1.2 Personal Protective Equipment: nitrile gloves, safety glasses 1. Nanomaster NSC-3000 DC Magnetron Sputter Tool Standard Operating Procedure Faculty Supervisor: Prof. Robert White, Mechanical Engineering (x72210) Safety Office: Peter Nowak x73246 (Just dial this directly

More information

SOP for Karl Suss MJB3 #1 Mask Aligner

SOP for Karl Suss MJB3 #1 Mask Aligner SOP for Karl Suss MJB3 #1 Mask Aligner Rev. 5 (30/11/2016) Safety UV Exposure: The high-energy light produced by the high-pressure Mercury Xenon lamp can cause eye damage and skin burns. Be sure that the

More information

Standard Operating Procedure. For. PVD E-Beam

Standard Operating Procedure. For. PVD E-Beam P a g e 1 Standard Operating Procedure For PVD E-Beam P a g e 2 Introduction The PVD Electron-Beam Evaporator (E-Beam) thin film deposition machine uses a magnetically guided and collimated stream of electrons

More information

Operating Procedures for Metal Evaporator I

Operating Procedures for Metal Evaporator I Operating Procedures for Metal Evaporator I Metal Evaporator I is intended as a tool and a training device. Understanding the operation of this equipment should give you a basic knowledge of vacuum and

More information

RIPPLE TANK - with rippler & kit

RIPPLE TANK - with rippler & kit GENERAL DESCRIPTION: RIPPLE TANK - with rippler & kit Cat: SW3430-001 with illuminator, rippler & kit. The ripple tank is used to investigate wave motion in a shallow trough of water to understand how

More information

KARL SUSS MJB3 MASK ALIGNER STANDARD OPERATING PROCEDURE

KARL SUSS MJB3 MASK ALIGNER STANDARD OPERATING PROCEDURE KARL SUSS MJB3 MASK ALIGNER STANDARD OPERATING PROCEDURE Purpose of this Instrument: This instrument is for patterning photosensitive polymers with UV light. Location: White Hall 410 Cleanroom Primary

More information

Operation of the contact mask aligner Canon PPC 210 Projection Print Camera

Operation of the contact mask aligner Canon PPC 210 Projection Print Camera Operation of the contact mask aligner Canon PPC 210 Projection Print Camera Start-up Procedure: 1. Start the grey pump in the service corridor behind the aligner 2. Open the valve for the Nitrogen gas

More information

Angstrom Dielectric Sputterer Operation Manual

Angstrom Dielectric Sputterer Operation Manual Angstrom Dielectric Sputterer Operation Manual I. System overview The Angstrom Dielectric Sputterer (ADS) has a similar interface as the Angstrom metal sputterer. It has two screens, the process screen

More information

MMRC PVD Evaporator Instructions

MMRC PVD Evaporator Instructions Warnings MMRC PVD Evaporator Instructions (Updated Aug 2016) Under PVD conditions, even oils and greases are volatile. WEAR GLOVES, and if you touch your skin or anything else oily or wet, change them.

More information

[needs to be discussed with faculty if any new material to be done]

[needs to be discussed with faculty if any new material to be done] Tool Name: EVO 18 Version 1 Authorized Users: Gayatri, Vishnu kant Materials Used: Sample holder has facility to mount 9 stubs: Si, Glass, Materials allowed : [needs to be discussed with faculty if any

More information

KARL SUSS MJB3 UV400 Mask Aligner Standard Operating Procedure

KARL SUSS MJB3 UV400 Mask Aligner Standard Operating Procedure KARL SUSS MJB3 UV400 Mask Aligner Standard Operating Procedure Version: 1.0 February 2014 UNIVERSITY OF TEXAS AT ARLINGTON Nanotechnology Research Center (NRC) 1 TABLE OF CONTENTS 1 Introduction 3 1.1

More information

General remarks regarding ALS robot system:

General remarks regarding ALS robot system: General remarks regarding ALS robot system: Preparing pucks at home. a) Preparing pins: - for ALS pucks use ONLY the special Hampton pins or spine pins - break off the 2 bottom sections of the standard

More information

Oxygen Contamination

Oxygen Contamination MPMS Application Note 1014-210 Oxygen Contamination This application note describes potential sources for oxygen contamination in the sample chamber and discusses its possible effects. Molecular oxygen,

More information

COBILT CA-800 Mask Aligner Equipment Operation

COBILT CA-800 Mask Aligner Equipment Operation COBILT CA-800 Mask Aligner Equipment Operation For the Micro-Electronics Laboratory At University of Notre Dame Department of Electrical Engineering This user manual is not be removed from room 247A. This

More information

Met One E-BAM Particulate Monitor

Met One E-BAM Particulate Monitor STANDARD OPERATING PROCEDURES Met One E-BAM Particulate Monitor AMBIENT AIR MONITORING PROGRAM for the 130 LIBERTY STREET DECONSTRUCTION PROJECT LOWER MANHATTAN DEVELOPMENT CORPORATION 1 Liberty Plaza

More information

1)! DO NOT PROCEED BEYOND THIS MARK

1)! DO NOT PROCEED BEYOND THIS MARK Operating Instructions for X-ray Photoelectron Spectrometer: Physical Electronics Model 555 XPS/AES (John H. Thomas, III, Ph.D., Electron Spectroscopy) Sample Insertion: figure 1. Sample insertion rod

More information

Linkam Scientific Instruments USER GUIDE DSC600. Temperature Controlled Stage

Linkam Scientific Instruments USER GUIDE DSC600. Temperature Controlled Stage Linkam Scientific Instruments DSC600 Temperature Controlled Stage USER GUIDE Contents Before Setting Up Your Equipment... 3 Important Notice... 4 Warranty... 4 Technical Support... 4 Equipment Maintenance...

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual AB-M Mask Aligner Version 1.1 Page 1 of 18 Contents 1. Picture and Location 2. Process Capabilities 2.1 Cleanliness Standard 2.2 Wafer Chuck Selection 2.3 Mask Holder Selection

More information

EXPERIMENT XI. Careful!! Improper handling of the vacuum line may result in the release of SO 2 which is an irritating and suffocating gas.

EXPERIMENT XI. Careful!! Improper handling of the vacuum line may result in the release of SO 2 which is an irritating and suffocating gas. Chem 366-3 Page XI - 1 EXPERIMENT XI INFRARED SPECTRUM OF SO2 (S&G, 5th ed. Expt 36, 6th ed. Expt. 35) 1. Pre-Lab preparation. The description of this experiment has disappeared from the more recent editions

More information

Issue: H Title: CHA E-Beam Evaporator Page 1 of 7. Table of Contents

Issue: H Title: CHA E-Beam Evaporator Page 1 of 7. Table of Contents Title: CHA E-Beam Evaporator Page 1 of 7 Table of Contents Purpose/Scope... 2 2.0 Reference Documents... 2 3.0 Equipment/Supplies/Material... 2 4.0 Safety... 2 5.0 Set Up Procedures... 2 5.1 PC Logon and

More information

Installation of Your SprayMaster System

Installation of Your SprayMaster System Installation of Your SprayMaster System 1. At the installation site, remove all equipment from the corrugated box and the polyethylene drum and replace the drum lid. Check the picture to identify each

More information

Misaligned Folds Paper Feed Problems Double Feeds Won t Feed FLYER Won t Run iii

Misaligned Folds Paper Feed Problems Double Feeds Won t Feed FLYER Won t Run iii Operator s Manual Table of Contents Operator Safety... 1 Introduction... 2 Unpacking and Setup... 3 Unpacking... 3 Setup... 4 FLYER Overview... 5 FLYER Diagram... 5 Capabilities... 5 Control Panel... 6

More information

icreasepro Creaser Operators Manual

icreasepro Creaser Operators Manual 6-2013 Version 3.0 icreasepro Creaser Operators Manual WWW.MBMCORP.COM 800-223-2508 TABLE OF CONTENTS SPECIFICATIONS.1a SAFETY PROCEDURES/CARE & MAINTENANCE..1b COMPONENT IDENTIFICATION 2 TOUCH SCREEN

More information

Linkam Scientific Instruments USER GUIDE TST350. Temperature Controlled Tensile Stress Testing Stage

Linkam Scientific Instruments USER GUIDE TST350. Temperature Controlled Tensile Stress Testing Stage Linkam Scientific Instruments TST350 Temperature Controlled Tensile Stress Testing Stage USER GUIDE Contents Before Setting Up Your Equipment... 3 Important Notice... 4 Warranty... 4 Technical Support...

More information

Approved by Principal Investigator Date: Approved by Super User: Date:

Approved by Principal Investigator Date: Approved by Super User: Date: Approved by Principal Investigator Date: Approved by Super User: Date: Standard Operating Procedure BNC Commonwealth Dual Ion Beam Deposition System (CDIBS) Version 2010 February 14 I. Purpose This Standard

More information

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1 BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1 The front door of the main body of the instrument may be open when you arrive. Take the sample chamber and slide it into position with the

More information

O P E R ATING INSTRUCTIONS FOR MODEL SPR-45 Automatic Screen and Stencil Printer

O P E R ATING INSTRUCTIONS FOR MODEL SPR-45 Automatic Screen and Stencil Printer O P E R ATING INSTRUCTIONS FOR MODEL SPR-45 Automatic Screen and Stencil Printer TABLE OF CONTENTS I. SPECIFICATIONS...3. II. SAFETY INSTRUCTIONS...4. III. INSTALLATION...5. IV. SET-UP...6. V. SYSTEM OPERATION...9.

More information

ATOC Meteorological Tower (6 meter-4 level) Guide

ATOC Meteorological Tower (6 meter-4 level) Guide ATOC Meteorological Tower (6 meter-4 level) Guide SETUP Outline 1. Siting 2. Components 3. Transportation 4. The Tower 5. The Cables and Data Logger Siting Selecting an appropriate site for the weather

More information

SCS Field Revision No. 1.0 June 2001 STANDARD OPERATING PROCEDURE (SOP) FOR FIELD OPERATION OF THE SEQUENTIAL CYCLONE SAMPLER (SCS)

SCS Field Revision No. 1.0 June 2001 STANDARD OPERATING PROCEDURE (SOP) FOR FIELD OPERATION OF THE SEQUENTIAL CYCLONE SAMPLER (SCS) STANDARD OPERATING PROCEDURE (SOP) FOR FIELD OPERATION OF THE SEQUENTIAL CYCLONE SAMPLER (SCS) Prepared by: Vincent A. Dutkiewicz Date: Reviewed by: Volker A. Mohnen (PMTACS-NY QA Officer) Date: Approved

More information

The SPI Sputter Coater Handbook

The SPI Sputter Coater Handbook The SPI Sputter Coater Handbook Coating of Specimens SPI-Module Sputter Coater with Etch Mode 1. Mount the specimens onto the SEM stub. Keep in mind that many adhesives have high vapor pressure solvents

More information

2 Switching off ECRIS 3. 4 Acquire Beam Spectrum 4. 6 Vent Instrument Chamber 5. 7 Pump down Chamber 5. 8 Baking out Chamber 5.

2 Switching off ECRIS 3. 4 Acquire Beam Spectrum 4. 6 Vent Instrument Chamber 5. 7 Pump down Chamber 5. 8 Baking out Chamber 5. Contents 1 Starting the ECR ion source (ECRIS) 2 2 Switching off ECRIS 3 3 Starting Mefisto Beam Analyzer 4 4 Acquire Beam Spectrum 4 5 Open Valve ECRIS/Chamber (Gate valve) 4 6 Vent Instrument Chamber

More information

Regulators repair and maintenance. XS Compact 2nd stage. January 2014 Rev XSC /3 Ed. C /14 1

Regulators repair and maintenance. XS Compact 2nd stage. January 2014 Rev XSC /3 Ed. C /14 1 XS Compact 2nd stage January 2014 Rev XSC /3 Ed. C /14 1 XS Compact 2nd stage WARNING! This manual is intended for use by expert technicians who have already received training in equipment repairs and

More information

Thermo K-Alpha XPS Standard Operating Procedure

Thermo K-Alpha XPS Standard Operating Procedure Thermo K-Alpha XPS Standard Operating Procedure Quick Guide Draft v.0.1 Procedure overview 1. Vent the loadlock 2. Secure your sample to the stage using clips, check the height of the final assembly. 3.

More information

QUARTZ HOUSING COATING PROCEDURE

QUARTZ HOUSING COATING PROCEDURE W. W. Hansen Experimental Physics Laboratory STANFORD UNIVERSITY STANFORD, CALIFORNIA 94305-4085 Gravity Probe B Relativity Mission QUARTZ HOUSING COATING PROCEDURE GP-B P0044 Rev- November 18, 1997 Prepared

More information

MIMS PROCEDURES. Updated 10 Sept 2008

MIMS PROCEDURES. Updated 10 Sept 2008 Updated 10 Sept 2008 MIMS PROCEDURES A. OPERATING PROCEDURES...2 1. Tuning...2 Coarse Tuning...2 Fine Tuning...2 Implantation...2 2. Balancing Trolleys...3 3. Procedure for Bringing Up New Cesium Source...4

More information

Revised: June 7, 2017

Revised: June 7, 2017 LC Technologies Thermal Evaporator Standard Operating Procedure Faculty Supervisor: Prof. Robert White, Mechanical Engineering (x72210) Safety Office: Peter Nowak x73246 (Just dial this directly on any

More information

Step By Step Instructions for VT Experiments on the Bruker 300 MHz Spectrometer

Step By Step Instructions for VT Experiments on the Bruker 300 MHz Spectrometer Step By Step Instructions for VT Experiments on the Bruker 300 MHz Spectrometer (1) Use the drop down menu to create a new experiment/dataset for acquiring the room temperature 1D spectrum. The name of

More information

PARTS LIST / TECHNICAL GUIDE

PARTS LIST / TECHNICAL GUIDE PARTS LIST / TECHNICAL GUIDE Double Retrograde Chronograph Cal. 7T85 [SPECIFICATIONS] Item Cal. No. Movement 7T85A 55 50 45 7T85A 40 35 3 hands (hour, minute, and small second hands) Calendar (date: date

More information

MJB-3 Mask Aligner Property of TAU MNCF August 2012

MJB-3 Mask Aligner Property of TAU MNCF August 2012 MJB-3 Mask Aligner I. Power Up Sequence 1 LOG IN. Turn on Nitrogen & compressed air lines on the back wall. 2 Turn on the vacuum pump. 3 Flip the COMPRESSED AIR toggle on (up) Compressed air is used to

More information

Using the Akta Prime plus October 22, 2012

Using the Akta Prime plus October 22, 2012 Some starting precautions: 1. Vacuum filter all buffers. Removes any large particles/debris that may clog your column De-gases the buffers 2. Clarify lysates first by centrifugation and then filtration

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual ARC12M Sputter Copyright 11.2015 by Hong Kong University of Science & Technology. All rights reserved. Page 1 Contents 1. Picture and Location 2. Process Capabilities 2.1 Cleanliness

More information

Manual Actuated Boiler Blowdown Valves

Manual Actuated Boiler Blowdown Valves Manual Actuated Boiler Blowdown Valves Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation 4. Operation 5. Maintenance 6. Spare parts p.1 1. Safety

More information

Powermax machine-side reference guide For mechanized applications with Powermax1000, Powermax1250 and Powermax1650

Powermax machine-side reference guide For mechanized applications with Powermax1000, Powermax1250 and Powermax1650 Powermax machine-side reference guide For mechanized applications with Powermax1000, Powermax1250 and Powermax1650 This Powermax machine-side reference guide is a supplement to your Operator Manual and

More information

FAULT CODE TROUBLESHOOTING INDEX

FAULT CODE TROUBLESHOOTING INDEX FAULT CODE TROUBLESHOOTING INDEX 1. Display indicates Change Filters 2. Display indicates Drip Tray Full Continuous Alarm will Sound 3. Display indicates Cold Fault 4. Display indicates Hot Fault 5. Display

More information

University of Minnesota Nano Fabrication Center

University of Minnesota Nano Fabrication Center Equipment name: Karl Suss Mask Aligners Badger name: MA6 & Maba6 Revision number: 4 Models: Revisionist: Paul Kimani Location: Bay 2 Date: April 25, 2013 1. Introduction The Karl Suss MA6 and MABA6 are

More information

Karl Suss MJB4 Mask Aligner

Karl Suss MJB4 Mask Aligner Karl Suss MJB4 Mask Aligner Tool Manager: Yong Sun ( yongs@princeton.edu; Office 8-8234; Cell 609-917-5076 ) Backup: George Watson ( gwatson@princeton.edu; Office 8-4626; Cell 732-996-2713 ) ******************************************************************************

More information

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012 JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012 UNIVERSITY OF TEXAS AT ARLINGTON Nanofabrication Research and Teaching Facility TABLE OF CONTENTS 1. Introduction....2 1.1 Scope of Work.....2

More information

OAI Model 200 Tabletop Mask Aligner Portland State University

OAI Model 200 Tabletop Mask Aligner Portland State University OAI Model 200 Tabletop Mask Aligner Portland State University WARNING: This machine exposes users to ultraviolet radiation. Do not touch the lens underneath the lamp hood as it may damage the machine and

More information

COMELEC C-30-S Parylene Coating System

COMELEC C-30-S Parylene Coating System COMELEC C-30-S Parylene Coating System Comelec C-30-S Parylene deposition system Introduction Parylene is a polymere deposited at room temperature in a vacuum chamber (few µb). Parylene coating is perfectly

More information

AQUARIUS. Operating Instructions for the. Type 70 Electrolytic Gas Soldering / Welding Unit. Table of Contents

AQUARIUS. Operating Instructions for the. Type 70 Electrolytic Gas Soldering / Welding Unit. Table of Contents AQUARIUS Operating Instructions for the Type 70 Electrolytic Gas Soldering / Welding Unit Table of Contents 1. Important General Remarks 2 2. Safety Instructions 2 3. Description of the Soldering Unit

More information

Model 130M Pneumatic Controller

Model 130M Pneumatic Controller Instruction MI 017-450 May 1978 Model 130M Pneumatic Controller Installation and Operation Manual Control Unit Controller Model 130M Controller is a pneumatic, shelf-mounted instrument with a separate

More information

Maintenance and Repair. AirHawk II Air Mask. Second Stage Regulator. Order No.: /02 Prnt. Spec (I) MSAsafety.

Maintenance and Repair. AirHawk II Air Mask. Second Stage Regulator. Order No.: /02 Prnt. Spec (I) MSAsafety. Maintenance and Repair Second Stage Regulator Order No.: 10104241/02 Prnt. Spec. 10000005389(I) MSAsafety.com WARNING! Read this manual carefully before servicing the device. The device will perform as

More information

UNITY 2. Uni-DAAMS Manual. Version 1.3 (Changes to Section 1.1) March 2010

UNITY 2. Uni-DAAMS Manual. Version 1.3 (Changes to Section 1.1) March 2010 UNITY 2 Uni-DAAMS Manual Version 1.3 (Changes to Section 1.1) March 2010 1. Introduction...2 1.1. Installing the DAAMS tube ovens...2 1.2. Swapping between tube types...5 1.3. Changing tube oven o-rings...6

More information

Usage Policies Notebook for Karl Suss MA6 Mid / Deep UV Mask Aligner

Usage Policies Notebook for Karl Suss MA6 Mid / Deep UV Mask Aligner Usage Policies Notebook for Karl Suss MA6 Mid / Deep UV Mask Aligner Revision date September 2014 2 Emergency Plan for Karl Suss MA6 Aligner Standard Operating Procedures for Emergencies Contact information

More information

Armfield Distillation Column Operation Guidelines

Armfield Distillation Column Operation Guidelines Armfield Distillation Column Operation Guidelines 11-2016 R.Cox Safety SAFETY GLASSES ARE REQUIRED WHEN OPERATING THE DISTILLATION COLUMN Wear gloves when mixing alcohol feedstock The column will become

More information

Linkam Scientific Instruments USER GUIDE. LTS420 and LTS120. Temperature Controlled Stage

Linkam Scientific Instruments USER GUIDE. LTS420 and LTS120. Temperature Controlled Stage Linkam Scientific Instruments LTS40 and LTS10 Temperature Controlled Stage USER GUIDE Contents Before Setting Up your Equipment... 3 Important Notice... 4 Warranty... 4 Technical Support... 4 Equipment

More information

Cleaning rod: spring steel, stainless steel or carbon fibre cleaning rod - only use a one-piece rod. Avoid using snakes.

Cleaning rod: spring steel, stainless steel or carbon fibre cleaning rod - only use a one-piece rod. Avoid using snakes. Telemark Biathlon Where performance and precision come together http://telemarkbiathlon.com Rifle Cleaning Date : July 19, 2013 Anschutz Rifle Manual - Click Here Izhmash 7-3 Rifle Manual - still looking

More information

Bedienungsanleitung / Operating Instructions / Instrucciones de uso / Mode d emploi / Istruzioni d'uso

Bedienungsanleitung / Operating Instructions / Instrucciones de uso / Mode d emploi / Istruzioni d'uso Bedienungsanleitung / Operating Instructions / Instrucciones de uso / Mode d emploi / Istruzioni d'uso Pressluftgewehr LG30 LG30 Junior cal..177 (4,5mm) Marketing/Vertrieb Service/Technik Carl Walther

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual Denton Explorer 14 RF/DC Sputter Version 1.0 Page 1 of 11 Contents 1. Picture and Location 2. Process Capabilities 1. Cleanliness Standard 2. Available for Sputtering Materials

More information

Microscopy Cryogenic WORKSTATION. Performance by Design

Microscopy Cryogenic WORKSTATION. Performance by Design CRYO Microscopy Cryogenic WORKSTATION Optical cryostat for use in microscopy and spectroscopy Performance by Design RC102-CFM Microscopy Cryostat offers fast cooldown, high efficiency, excellent temperature

More information

Pressure Dump Valve Service Kit for Series 3000 Units

Pressure Dump Valve Service Kit for Series 3000 Units Instruction Sheet Pressure Dump Valve Service Kit for Series 000 Units. Overview The Nordson pressure dump valve is used to relieve hydraulic pressure instantly in Series 00, 400, 500, and 700 applicator

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual Branson IPC 3000 O 2 Asher Copyright 2014 by Hong Kong University of Science & Technology. All rights reserved. Page 1 Contents 1. Picture and Location 2. Process Capabilities

More information

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION BEST KNOWN METHODS Transpector XPR3 Gas Analysis System DESCRIPTION The Transpector XPR3 is a third-generation, quadrupole-based residual gas analyzer that operates at PVD process pressures and is the

More information

University of Minnesota Nano Fabrication Center

University of Minnesota Nano Fabrication Center Equipment name: Karl Suss Mask Aligners Badger name: MA6-P Revision number: 0 Model: MA6 Revisionist: Paul Kimani Location: Bay 4 PAN Date: September 17, 2015 1. Introduction The Karl Suss MA6-PAN is a

More information

RARS5000 AIR BODY SAW OWNER S OPERATING MANUAL

RARS5000 AIR BODY SAW OWNER S OPERATING MANUAL RARS5000 AIR BODY SAW OWNER S OPERATING MANUAL DESCRIPTION 1. No mar 2. No mar tip 3. Housing grip 4. Trigger 5. Air inlet 6. Air inlet plug 7. Plastic board Important! It is essential that you read the

More information

3190A NEO-ANGLE DOOR INSTALLATION INSTRUCTIONS. Series MODEL NO

3190A NEO-ANGLE DOOR INSTALLATION INSTRUCTIONS. Series MODEL NO NEO-ANGLE DOOR INSTALLATION INSTRUCTIONS Series 30A Please read these instructions carefully to familiarize yourself with the required tools, materials, and installation sequences. The Exploded Diagram

More information

! WARNING! IMPORTANT HPA AIR TANK SAFETY INSTRUCTION AND GUIDELINES ! WARNING! IMPORTANT SAFETY INSTRUCTION AND GUIDELINES

! WARNING! IMPORTANT HPA AIR TANK SAFETY INSTRUCTION AND GUIDELINES ! WARNING! IMPORTANT SAFETY INSTRUCTION AND GUIDELINES ! WARNING! IMPORTANT SAFETY INSTRUCTION AND GUIDELINES! WARNING! IMPORTANT HPA AIR TANK SAFETY INSTRUCTION AND GUIDELINES This Paintball Marker is NOT A TOY. Misuse can cause serious injury or death. It

More information

A3S Bellows Sealed Stop Valve Installation and Maintenance Instructions

A3S Bellows Sealed Stop Valve Installation and Maintenance Instructions 1326050/3 IM-P132-11 ST Issue 3 A3S Bellows Sealed Stop Valve Installation and Maintenance Instructions 1 General safety information 2 General product information 3 Installation 4 Commissioning 5 Operation

More information

5.1.3 Mechanical Hazards Drive assemblies have sufficient power to cause injury. Keep hands, fingers, clothing and tools clear of moving parts.

5.1.3 Mechanical Hazards Drive assemblies have sufficient power to cause injury. Keep hands, fingers, clothing and tools clear of moving parts. Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the PE4400. All users are expected to have read and understood this document. It is

More information

sigma User Manual Educational Microscopy

sigma User Manual Educational Microscopy R sigma User Manual Educational Microscopy To ensure proper use of this instrument as well as to avoid injury while operating instrument, understanding this manual completely before use is highly recommended.

More information

Operation of the mask aligner MJB-55

Operation of the mask aligner MJB-55 John Paul Adrian Glaubitz Operation of the mask aligner MJB-55 Department of Physics Faculty of Mathematics and Natural Sciences University of Oslo 1 Introduction The mask aligner is an essential tool

More information

Linkam Scientific Instruments USER GUIDE TST350. Temperature Controlled Tensile Stress Testing Stage

Linkam Scientific Instruments USER GUIDE TST350. Temperature Controlled Tensile Stress Testing Stage Linkam Scientific Instruments TST350 Temperature Controlled Tensile Stress Testing Stage USER GUIDE Contents Before Setting Up Your Equipment...3 Important Notice...4 Warranty...4 Technical Support...4

More information

IVIS Spectrum Whole Animal Imager

IVIS Spectrum Whole Animal Imager IVIS Spectrum Whole Animal Imager Kyle Marchuk Adam Fries Jordan Briscoe Taylor Shagam July 2018 Contents 1 Introduction 2 2 Hardware - Startup 4 3 Start-Up Software 5 4 Shut Down 11 1 1 Introduction The

More information

INSTRUCTION MANUAL. January 23, 2003, Revision 0

INSTRUCTION MANUAL. January 23, 2003, Revision 0 INSTRUCTION MANUAL Model 810A In-Vitro Test Apparatus for 310B Muscle Lever January 23, 2003, Revision 0 Copyright 2003 Aurora Scientific Inc. Aurora Scientific Inc. 360 Industrial Parkway S., Unit 4 Aurora,

More information

Eco Snow CO2 Clean System Operation

Eco Snow CO2 Clean System Operation Eco Snow CO2 Clean System Operation This equipment is to be operated only by individuals that have been trained and qualified by the MNFC staff. Do NOT use or attempt to use this tool unless you are qualified

More information

Approved by Principal Investigator Date: Approved by Super User: Date:

Approved by Principal Investigator Date: Approved by Super User: Date: Approved by Principal Investigator Date: Approved by Super User: Date: Standard Operating Procedure BNC OAI 200 Lithographic Mask Aligner (Aligner 3) Version 2011 June 2 I. Purpose This Standard Operating

More information

Approved by Principal Investigator Date: Approved by Super User: Date:

Approved by Principal Investigator Date: Approved by Super User: Date: Approved by Principal Investigator Date: Approved by Super User: Date: Standard Operating Procedure BNC OAI Lithographic Mask Aligner (Aligner 2) Version 2008 October 31 I. Purpose This Standard Operating

More information

Read this first. Zetasizer nano series Self installation and Quick start guide MRK825-02

Read this first. Zetasizer nano series Self installation and Quick start guide MRK825-02 ! Read this first Zetasizer nano series Self installation and Quick start guide I N S T R U M E N T S MRK825-02 Zetasizer Nano series Self installation and Quick start guide MAN0383 Issue 1.1 July 2007

More information

Variable Temperature Operation

Variable Temperature Operation Variable Temperature Operation AVIII 400 MHz Spectrometer November 11, 2012 The CBC NMR Laboratory University of Delaware AVIII 400 User Guide: Variable Temperature Operation 1 INTRODUCTION... 5 1.1 SAFETY...

More information

Quintel Mask Aligner

Quintel Mask Aligner Quintel Mask Aligner (quintel) 1.0 1.0 Title Quintel Q4000 Mask Aligner 2.0 2.0 Purpose The Quintel Q4000 MA (Mask Aligner) is a top and bottom side contact lithography printer with the video-view split

More information

USER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve

USER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve USER'S GUIDE Nanoflow Probe 3-4mm Red Stripe ptfe Sleeve To Injector ZU1XC Fused Silica 20µm 90µm TTF115 ptfe Sleeve Fused Silica 25µm 375µm Micromass UK Limited Floats Road Wythenshawe M23 9LZ Tel: +44

More information

Paddle Bar Replacement

Paddle Bar Replacement This procedure is to help facilitate the replacement of the 23 Paddle Bar Assembly on the ANKOM Dietary Fiber Analyzer. Note: The following items will be sent in a replacement package as part of the 23

More information