SEM LEO 1550 MANUAL. I. Introduction. Generality

Size: px
Start display at page:

Download "SEM LEO 1550 MANUAL. I. Introduction. Generality"

Transcription

1 SEM LEO 1550 MANUAL RESERVATION POLICY: 2 booking slots maximum per day and per person (ie. 1h). 6 booking slots maximum per week and per person (ie. 3h). Reservation names must correspond to the operators. I. Introduction Generality The CMI SEM LEO 1550 is composed of one GEMINI column (cf. Figure 2), one process chamber with a motorized stage on 5 degree of freedoms (X, Y, Z, Tilt and Rotation) and one airlock. You can use 4 diferents holder according to the needs (cf. Figure 1). (1) for small sample(s) which are fixed with the help of a conductor double side adhesive (2) for cleaved sample(s) (3) for 150 mm wafer (4) for 100 mm wafer Figure 1: Different SEM holders 1

2 the Field Emission Gun (3), which provides the source of the electron beam the Condenser Lens (7), used only in special operating modes the Beam Booster, composed of Anode (5), Vacuum Tube (6), Apertures (8), Alignment Coils (9a, b, c), Stigmator (13), and Isolating Valve (15) The GEMINI Objective Lens (10,11) which focuses the electron beam onto the specimen (12), also containing the Deflecting System (14) Figure 2: Schematic view of the GEMINI column GEMINI column The GEMINI column s electron source is a Schottky Field Emission type made of tungsten and Zirconium (ZrO2). Two secondary electrons detectors have been installed: InLens: installed inside the column (high resolution detector). SE2: installed out of the column axis (better topography visualization). The chamber vaccum (few 10-7 mbar) is held thanks to a couple primary/turbomolecular pumps and the secondary vaccum (few mbar) is held thanck to an ionic pump. The GEMINI column advantages: We can use a low electronic acceleration from 0.5 kev to several kev according to the sample. So the charging effects dues to the insulators materials are limited. Both detectors are complementary and it s very easy to compare these performances. II. Access conditions 1. The CMI SEM LEO is reserved to the regular CMI users. 2. It is exclusively reserved to the control of processes which have been done with the CMI installation. 3. The maximal booking time per day is 1 hour (so 2 slots). 4. Saved pictures are available on STI network (EPFL local or VPN): "\\sti1files\cmi-transfert\z01-zeiss-leo". III. Basic operation 1. Login on the zone computer 2. Login on the SEM control computer 3. Place your sample(s) on the appropriate holder 4. Load your sample into the SEM airlock 5. Start the "Specimen Change via Airlock" routine (button 1 Figure 3) 6. Load the holder into the chamber (cf. Sample loading and unloading) 7. Set the SEM (cf. Sample observation) 8. Observe your sample(s) (cf. Sample observation) 9. Unload the holder into the airlock (cf. Sample loading and unloading) 2

3 nb Left Button (LB) Middle Button (MB) 1 Specimen Change via Airloock Resume exchange 2 Stage initialize Stage stop 3 Pixel averaging 1, 2, 3, 4 : the larger the number longer is the averaging 4 Full frame scan Reduced raster Split scan 5 Stigmation setting Brightness/Contrast Switch camera/detector Detector selection Magnification/Working distance (= Focal Distance) 6 Save picture (increase the picture number automatically) Print picture Figure 3: Software tools bar Continuous averaging 1, 2, 3 : bigger is the number longer is the averaging Scanning dialog Appertures dialog Detectors dialog Aperture alignment setting Auto-contrast and brightness Detectors dialog Detectors dialog Auto focus and stigmatisme Save Image dialog Printer dialog Sample loading and unloading Initial SEM state: The EHT is off and the airlock is at atmospheric pressure. Loading 1. Install your sample(s) on the appropriate holder 2. Initial airlock state, everything is closed 3. Open the number 1 (unlock the front door) 4. Unlock the black slider by pulling out the number 2 button. 3

4 5. Slide the door by holding it with the black part and lift at the same time to avoid vibration. Once you start to move the black part release the number 2 button. Continue to move the black slider until the number 2 locks again 6. Release 3 and open the door 7. Screw your holder on the transfer arm 8. Return the door parallel to the load lock 9. Release 2 and slide the door by holding it with the black part and lift it at the same time to avoid vibration on the sample till the number 2 is locked again 10. Lock it with number 1 (the door must be closed tightly to avoid leaks during pumping) 4

5 11. Click on the Specimen Change via Airloock button (cf. Figure 3) 12. Wait for the green light (= vacuum is good enough to open the door between the airlock and the main chamber) 13. Open the valve between the chamber and the airlock with number 5 and Insert your sample(s) onto the stage rail. The loading arm must stop at the beginning of the white tape. If you can go further it s because the holder has not loaded properly onto the stage (so retry to load it). 15. Unscrew the loading arm and pull it out from the chamber. Make sure the loading arm is properly unscrewed from the holder. 16. Close the airlock/chamber door (5) valve with and lock (6) 5

6 17. Click on the Resume Exchange button Unloading 1. Start by clicking on the Specimen Change via Airlock button (check before if the front door is properly closed, cf. Figure 3) 2. When the green light is lit open the valve between the airlock and the chamber 3. Insert the transfer arm 4. Screw on the holder 5. Extract the arm to the maximum 6. Close the valve between the airlock and the chamber 7. Click on the Resume Exchange button (cf. previous part step 17) 8. Unlock the door with the number 1 9. Wait untill the airlock reach the atmospheric pressure 10. Slide out the door by holding the black part and lift it at the same time to avoid vibration of the sample till the number 2 is locked again 11. Open the door with the number Enscrew your holder from the transfer arm 13. Close the airlock as you will pump it again before leaving 14. Close the door 15. Slide the door by holding the black part till the number 2 is blocked again 16. Lock it with number 1 Sample observation 1. Start the Extra High Tension (EHT) on the software bottom right corner (cf. Figures 4) 2. Figure 4: EHT On 6

7 Tip: Insulating Samples - When you click you first switch the EHT on the accelerating voltage will be set to 3 kev. If your sample is an insulator to avoid charging effects it may be better to reduce the acceleration voltage to around 1 kev. In the worst case if you need to observe a polymer this voltage can be reduce to 0.5 kev to avoid your structures collapsing too fast (basically to still have time to do the settings and pictures). It is often also helpful to reduce the current of the electron beam. Reducing the current reduces the number of electrons hitting your sample in a given time. To reduce the current you need to choose a smaller aperture. The default aperture is 30 µm try reducing this to 10 µm. When you reduce the acceleration voltage the SEM resolution drops and adjusting the alignment could be harder. When you reduce the aperture size the brightness and contrast will reduce and your image will become noisier. 2. Switch on the InLens by clicking on the appropriate button an the tools bar (cf. Figure 3) 3. Click on the Magnification and WD tools and reduce the Magnification to the minimum 4. Set the WD at 5 mm (double click on the WD value on the white data bar on the bottom of the screen) 5. Reduce the distance between the sample and the column by moving the stage on the Z axis untill the picture is sharp (cf. Stage movements) Tip: Sometimes it s a bit difficult to find structures on the sample surface to adjust the alignments. You can usually find some on the edge of the sample. We usually don t do the adjustments on important structures because by scanning the surface we affect it (especially with insulators and polymers). 7. Set the Brightness and the contrast (cf. Figure 3) 8. Press shift + F2 to recalibrat the deflection system hysteresis 9. Set the WD more accuracy with the mouse (now it should be between 3 and 7 mm). 10. To obtain sharp pictures you have to do a 3 steps sequence 2 or 3 times (depending on the final resolution needed): o Increase and decrease the WD and try to see the stigmation default. If you cannot see it zoom in and try again till you see it (depends on the final resolution need, cf. below). Figures 5 and 6: Effect of the stigmation default o o When you have highlighted it set the WD in such a way that you have set it just in the middle of the 2 privileged directions. Set the stigmatism thanks to the encoders or with the mouse in coarse mode (left mouse button on the appropriate icon, cf. cf. Figure 3) in order to have the better picture as possible Then open the Apertures dialog (SEM Control -> Apertures or medium mouse button on the appropriate icon, cf. Figure 3). Start the focus wobble and try to stop the picture movement with the Aperture Alignement setting. Again you can use the encoders or the mouse in coarse mode to perform it. 7

8 Figure 7: Apertures dialog Tip: As long as you don t change any fundamental parameters on the system like EHT, aperture or observation plane you don t have to do these settings again (except the WD). 11. Now you can zoom out and move where you want on your sample. 12. To reduce the noise we can use 2 different averagings: o Continuous averaging: Do an averaging with several frames with a short electron beam exposition time. It s particularlly useful with insulator or polymer to reduce the charging effects and the surface affectation (cf. Figure 3). o Pixel averaging: Do an averaging pixel per pixel by increasing its electron beam exposition time. It gives good results with conductor and semiconductor (cf. Figure 3). 13. When you are happy with the picture quality you can start the distance estimations with the annotation tools (Below the Data Zone). 14. To save picture go in File -> save image. The file name is composed of a prefix and a suffix. The prefix is write in the File name and the suffix is a number. When you have specify once the file name and the directory you can just press on the appropriate button on the tools bar (cf. Figure 3) to save a new picture by increasing the suffix number. Figure 8: Save image dialog 8

9 Stage movements With the remote control 2 joysticks: right to move X, Y, Rotation and left to move Z and Tilt. Figure 9: Remote control With the mouse Press Ctrl + tab on the keybord them you centred your sample where you click on the picture Press Ctrl + shift + tab on the keybord them you can drag a rectangle on the picture and the system will automatically zoom in and centred your sample into it. Keybord shortcut Ctrl + d : data zone display it/ hide it + (numerical keybord) : increase the scaning rate - (numerical keybord) : decrease the scaning rate tab : fine/coarse mode F3 : dialogs display them/hide them Ctrl + tab : you centred your sample where you click on the picture Ctrl + shift + tab : drag a rectangle on the picture and the system will automatically zoom in and centred your sample into it Shift + F2 : recalibrat the deflection system hysteresis Ctrl + A : annotation tools 9

10 IV. Photos galery Figure 10: Be careful with cleaved samples. Don t break it when you load it into the airlock Figure 11: Exchange position Figure 12: Initial sample position inside the chamber Figure 13: Deep silicon etching Figure 14: Dry etch of AlSi(1%) Figure 15: Deep fused silica etching (40 µm) Figure 16: Silicon submicrometric pillars 10

LEO 1525 FEG SEM Standard Operating Procedures Mar. 6, 2012 For additional assistance, please contact the facility manager.

LEO 1525 FEG SEM Standard Operating Procedures Mar. 6, 2012 For additional assistance, please contact the facility manager. LEO 1525 FEG SEM Standard Operating Procedures Mar. 6, 2012 For additional assistance, please contact the facility manager. Please contact in case of emergency: SEM microscopist: Mr. Eric Miller, 7-0789

More information

LEO SEM SOP Page 1 of 9 Revision 1.4 LEO 440 SEM SOP. Leica Leo Stereoscan 440i

LEO SEM SOP Page 1 of 9 Revision 1.4 LEO 440 SEM SOP. Leica Leo Stereoscan 440i LEO SEM SOP Page 1 of 9 LEO 440 SEM SOP Gun (Filament) Column Manual Valves Chamber Window Chamber Stage Movement Leica Leo Stereoscan 440i 1. Scope 1.1 This document provides the procedure for operating

More information

TYPICAL OPERATION PROCEDURES

TYPICAL OPERATION PROCEDURES JSM-7500F TYPICAL OPERATION PROCEDURES Reminders! Remember to sign in on the sign up sheet. WEAR GLOVES: always wear gloves when handling your samples, the sample holders and the specimen exchange rod.

More information

R I T. Title: Amray 1830 SEM Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 09/29/03 1 SCOPE 2 REFERENCE DOCUMENTS

R I T. Title: Amray 1830 SEM Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 09/29/03 1 SCOPE 2 REFERENCE DOCUMENTS Fabrication Laboratory Revision: A Rev Date: 09/29/03 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Amray 1830 SEM. All users

More information

Scanning Electron Microscope JEOL JSM F

Scanning Electron Microscope JEOL JSM F Scanning Electron Microscope JEOL JSM - 7600F How to Use This Manual You will be promised to obtain successful results if you follow this step by step manual gently and carefully. In order to that, you

More information

[needs to be discussed with faculty if any new material to be done]

[needs to be discussed with faculty if any new material to be done] Tool Name: EVO 18 Version 1 Authorized Users: Gayatri, Vishnu kant Materials Used: Sample holder has facility to mount 9 stubs: Si, Glass, Materials allowed : [needs to be discussed with faculty if any

More information

Login to ilab Kiosk. Revised 05/22/2018. Load your sample:

Login to ilab Kiosk. Revised 05/22/2018. Load your sample: Login to ilab Kiosk Load your sample: 1. Check: The analysis chamber pressure is

More information

Thermo K-Alpha XPS Standard Operating Procedure

Thermo K-Alpha XPS Standard Operating Procedure Thermo K-Alpha XPS Standard Operating Procedure Quick Guide Draft v.0.1 Procedure overview 1. Vent the loadlock 2. Secure your sample to the stage using clips, check the height of the final assembly. 3.

More information

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1 BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1 The front door of the main body of the instrument may be open when you arrive. Take the sample chamber and slide it into position with the

More information

Edge Isolation Tool. Standard Operating Procedure. Version 1.1. Date: Prepared by, Sandeep S S. Department of Electrical Engineering

Edge Isolation Tool. Standard Operating Procedure. Version 1.1. Date: Prepared by, Sandeep S S. Department of Electrical Engineering Edge Isolation Tool Standard Operating Procedure Version 1.1 Date: 20-9-2012 Prepared by, Sandeep S S Department of Electrical Engineering IIT Bombay This document is meant for internal circulation only

More information

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone:

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone: MJB4 Mask Aligner Operating Procedure Effective Date: 07/12/2012 Author(s): Jiong Hua Phone: 402-472-3773 Email: jhua2@unl.edu 1 1 Introduction 1.1 Key Words Karl Suss MJB4 Mask Aligner, Optical Lithography,

More information

Angstrom E-Beam Instructions. PROCESS CHECKS: the tooling factors of each metal; Ti/Au layer for wire bonding pull test.

Angstrom E-Beam Instructions. PROCESS CHECKS: the tooling factors of each metal; Ti/Au layer for wire bonding pull test. Angstrom E-Beam Instructions Tool Manager: Joe Palmer (jpalmer@princeton.edu) Office: 8-4706; Cell:609-751-1353 Backup: David Barth (dbarth@princeton.edu) Office: 8-4626; Cell: 610-405-8227 PROCESS CHECKS:

More information

Approved by BNC Managing Director Paul Lum Date: Approved by PI (s) /Super User (s): _Peter Hosemann/Andy Minor_Date:

Approved by BNC Managing Director Paul Lum Date: Approved by PI (s) /Super User (s): _Peter Hosemann/Andy Minor_Date: Approved by BNC Managing Director Paul Lum Date: Approved by PI (s) /Super User (s): _Peter Hosemann/Andy Minor_Date: Standard Operating Procedure BNC FEI Quanta 3D FEG/FIB ESEM Version 2011 Aug 30 I.

More information

Quintel Mask Aligner

Quintel Mask Aligner Quintel Mask Aligner (quintel) 1.0 1.0 Title Quintel Q4000 Mask Aligner 2.0 2.0 Purpose The Quintel Q4000 MA (Mask Aligner) is a top and bottom side contact lithography printer with the video-view split

More information

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012 JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012 UNIVERSITY OF TEXAS AT ARLINGTON Nanofabrication Research and Teaching Facility TABLE OF CONTENTS 1. Introduction....2 1.1 Scope of Work.....2

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual Branson IPC 3000 O 2 Asher Copyright 2014 by Hong Kong University of Science & Technology. All rights reserved. Page 1 Contents 1. Picture and Location 2. Process Capabilities

More information

PLASMA ENHANCED CHEMICAL VAPOR DEPOSITION (PECVD) SOP OXFORD PLASMALAB SYSTEM 100

PLASMA ENHANCED CHEMICAL VAPOR DEPOSITION (PECVD) SOP OXFORD PLASMALAB SYSTEM 100 PLASMA ENHANCED CHEMICAL VAPOR DEPOSITION (PECVD) SOP OXFORD PLASMALAB SYSTEM 100 June 2013 Interface Overview. The Oxford software is divided into 5 main screens. 1) Pump Control page 2) Recipe page 3)

More information

r β tr t t s t r s t s tr tt r t r t r r r tt r e s tt e tt r e s tt r e r r e r 2 t ö t t r s s t t r r r ts r rr t

r β tr t t s t r s t s tr tt r t r t r r r tt r e s tt e tt r e s tt r e r r e r 2 t ö t t r s s t t r r r ts r rr t tr µs 2 r β t r tr t s r t t rst 1 2 s t s tr tt r t r t r r r tt r e s tt e r 2 st e tt r e r2 e r s tt r e r r e r 2 t ö t t r s s t t r r r ts r rr t t s e s r r tt rs r LaB 6 ss s r t s r t3 t r t3

More information

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Sensitivity: 1x10 11 A m 2 Applied DC fields: 0 T to 5 T Applied AC fields: 0 G to 3 G (zero-to-peak), 0.01 Hz to 1000 Hz Temperatures

More information

MANUAL FOR SPTS APS (DIELECTRICS ETCHER)

MANUAL FOR SPTS APS (DIELECTRICS ETCHER) MANUAL FOR SPTS APS (DIELECTRICS ETCHER) To be read first: SPTS APS is an etcher dedicated to dielectrics (SiO 2, Si 3 N 4, glass types...). Dielectrics etching in AMS200 is no more CMi standard. AMS200

More information

Angstrom Dielectric Sputterer Operation Manual

Angstrom Dielectric Sputterer Operation Manual Angstrom Dielectric Sputterer Operation Manual I. System overview The Angstrom Dielectric Sputterer (ADS) has a similar interface as the Angstrom metal sputterer. It has two screens, the process screen

More information

R I T. Title: STS ASE Semiconductor & Microsystems Fabrication Laboratory Revision: Original Rev Date: 01/21/ SCOPE 2 REFERENCE DOCUMENTS

R I T. Title: STS ASE Semiconductor & Microsystems Fabrication Laboratory Revision: Original Rev Date: 01/21/ SCOPE 2 REFERENCE DOCUMENTS Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the STS ASE. All users are expected to have read and understood this document. It is

More information

University of MN, Minnesota Nano Center Standard Operating Procedure

University of MN, Minnesota Nano Center Standard Operating Procedure Equipment name: STS Etcher Badger name: STS Revision number: 3 Model: 320 Revisionist: Paul Kimani Location: Bay 3 Date: 1 October 2013 A. Description The 320 is a manually loaded batch plasma etching

More information

Xactix XeF2 OPERATION MANUAL

Xactix XeF2 OPERATION MANUAL General Information The Xactix e-1 is a xenon difluoride (XeF 2) isotropic silicon etcher. XeF 2 is a vapor phase etch, which exhibits very high selectivity of silicon to photo-resist, silicon dioxide,

More information

KARL SUSS MJB3 MASK ALIGNER STANDARD OPERATING PROCEDURE

KARL SUSS MJB3 MASK ALIGNER STANDARD OPERATING PROCEDURE KARL SUSS MJB3 MASK ALIGNER STANDARD OPERATING PROCEDURE Purpose of this Instrument: This instrument is for patterning photosensitive polymers with UV light. Location: White Hall 410 Cleanroom Primary

More information

Standard Operating Procedure. For. PVD E-Beam

Standard Operating Procedure. For. PVD E-Beam P a g e 1 Standard Operating Procedure For PVD E-Beam P a g e 2 Introduction The PVD Electron-Beam Evaporator (E-Beam) thin film deposition machine uses a magnetically guided and collimated stream of electrons

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual AB-M Mask Aligner Version 1.1 Page 1 of 18 Contents 1. Picture and Location 2. Process Capabilities 2.1 Cleanliness Standard 2.2 Wafer Chuck Selection 2.3 Mask Holder Selection

More information

Oerlikon Sputtering Evaporator SOP

Oerlikon Sputtering Evaporator SOP Oerlikon Sputtering Evaporator SOP Short UNT Cleanroom 1. Taking out sample holder from Transport Chamber : Log in FOM to access the software Go to the software and log in with user1 and password user1

More information

Karl Suss MA6 Mask Aligner SOP

Karl Suss MA6 Mask Aligner SOP Page 1 of 11 Karl Suss MA6 Mask Aligner SOP Safety UV Exposure: The high energy light produced by the high pressure Mercury Xenon lamp can cause eye damage and skin burns. Be sure that the light guards

More information

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure PPMS Service Note 1099-412 In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure Introduction: Loss of electricity to the PPMS EverCool II should not cause damage to

More information

Wire Buddy Manual IOpath Software

Wire Buddy Manual IOpath Software Wire Buddy Manual IOpath Software Wire Buddy is a calculation reference tool for Electricians and Maintenance personnel. Wire Buddy was created for the Maintenance Electrician. There are many occasions

More information

Operation of the contact mask aligner Canon PPC 210 Projection Print Camera

Operation of the contact mask aligner Canon PPC 210 Projection Print Camera Operation of the contact mask aligner Canon PPC 210 Projection Print Camera Start-up Procedure: 1. Start the grey pump in the service corridor behind the aligner 2. Open the valve for the Nitrogen gas

More information

The SPI Sputter Coater Handbook

The SPI Sputter Coater Handbook The SPI Sputter Coater Handbook Coating of Specimens SPI-Module Sputter Coater with Etch Mode 1. Mount the specimens onto the SEM stub. Keep in mind that many adhesives have high vapor pressure solvents

More information

1)! DO NOT PROCEED BEYOND THIS MARK

1)! DO NOT PROCEED BEYOND THIS MARK Operating Instructions for X-ray Photoelectron Spectrometer: Physical Electronics Model 555 XPS/AES (John H. Thomas, III, Ph.D., Electron Spectroscopy) Sample Insertion: figure 1. Sample insertion rod

More information

Previous Release Notes

Previous Release Notes Release Notes Shearwater Desktop 3.1.5 Support for NERD 2. Previous Release Notes Version 3.1.4 Improved Bluetooth Reliability with the initial connection. Bug Notes: dded software workaround to allow

More information

OPERATION OF THE DIMPLER

OPERATION OF THE DIMPLER OPERATION OF THE DIMPLER After thinning your sample to ~80 μm you can now do a dimpling process to thin the center up to 10 μm. When you walk in and use the DIMPLER it should already be calibrated and

More information

ADDENDUM IMPORTANT DOCUMENT INVITATION TO BID ADDENDUM. OPENING DATE & TIME: January 3, 2018 at 3:00PM

ADDENDUM IMPORTANT DOCUMENT INVITATION TO BID ADDENDUM. OPENING DATE & TIME: January 3, 2018 at 3:00PM Procurement Services ADDENDUM IMPORTANT DOCUMENT INVITATION TO BID ADDENDUM ITB NUMBER: 1708MC OPENING DATE & TIME: January 3, 2018 at 3:00PM ITB TITLE: X-Ray Photoelectron Spectroscopy System ADDENDUM

More information

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION BEST KNOWN METHODS Transpector XPR3 Gas Analysis System DESCRIPTION The Transpector XPR3 is a third-generation, quadrupole-based residual gas analyzer that operates at PVD process pressures and is the

More information

Title: Xactix XeF2 Etcher Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 03/23/2016

Title: Xactix XeF2 Etcher Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 03/23/2016 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Xactix XeF2 Etcher. All users are expected to have read and understood this document.

More information

COBILT CA-800 Mask Aligner Equipment Operation

COBILT CA-800 Mask Aligner Equipment Operation COBILT CA-800 Mask Aligner Equipment Operation For the Micro-Electronics Laboratory At University of Notre Dame Department of Electrical Engineering This user manual is not be removed from room 247A. This

More information

IVIS Spectrum Whole Animal Imager

IVIS Spectrum Whole Animal Imager IVIS Spectrum Whole Animal Imager Kyle Marchuk Adam Fries Jordan Briscoe Taylor Shagam July 2018 Contents 1 Introduction 2 2 Hardware - Startup 4 3 Start-Up Software 5 4 Shut Down 11 1 1 Introduction The

More information

ClubHub. User s Guide

ClubHub. User s Guide ClubHub User s Guide Table of Contents Setup... Initial Club Setup...7 Changing Clubs...5 Settings...8 My Clubs... Turn On/Off Sounds...9 Play Round Mode...0 List View...8 Social Sharing...0 Viewing D

More information

Version 3.1.0: New Features/Improvements: Improved Bluetooth connection on Windows 10

Version 3.1.0: New Features/Improvements: Improved Bluetooth connection on Windows 10 Version 3.1.0: Improved Bluetooth connection on Windows 10 ***Important notice for Mac Users: Upgrading from Shearwater Desktop 3.0.8 to 3.1.0 will not cause issues. Upgrading from any 2.X.X to any 3.X.X

More information

MIMS PROCEDURES. Updated 10 Sept 2008

MIMS PROCEDURES. Updated 10 Sept 2008 Updated 10 Sept 2008 MIMS PROCEDURES A. OPERATING PROCEDURES...2 1. Tuning...2 Coarse Tuning...2 Fine Tuning...2 Implantation...2 2. Balancing Trolleys...3 3. Procedure for Bringing Up New Cesium Source...4

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual Branson IPC 3000 O 2 Asher Page 1 of 14 Contents 1 Picture and Location 2 Process Capabilities 2.1 Cleanliness Standard 2.2 Recipes 2.3 Performance of Branson IPC 3000 O 2 Asher

More information

David Cockayne Centre for Electron Microscopy.

David Cockayne Centre for Electron Microscopy. David Cockayne Centre for Electron Microscopy. Depart of Materials, University of Oxford JEOL JEM-2100 TEM USER GUIDE November 2014 JLH / NPY 1 Contents Control panels...4 Tracker Ball...7 Aperture controls....8

More information

OXFORD PLASMALAB 80PLUS (CLOEY)

OXFORD PLASMALAB 80PLUS (CLOEY) Arizona State University NanoFab OXFORD PLASMALAB 80PLUS (CLOEY) Rev D Table of Contents Contents Table of Contents...1 1. Purpose / Scope...2 2. Reference Documents...2 3. Equipment / Supplies / Material...2

More information

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI NanoSight NS300 Operation instructions Injection/flushing brief overview: 1. Do not exceed flow of 1 ml per 20 seconds. 2. Inject two 1 ml syringes with nano-pure or DI water. 3. If the water does not

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual Allwin21 AW610 RTP Page 1 of 18 Contents 1 Picture and Location 2 Process Capabilities 2.1 Cleanliness Standard 2.2 Recipes 2.3 Performance of Allwin21 AW610 RTP 3 Contact List

More information

Operation of the mask aligner MJB-55

Operation of the mask aligner MJB-55 John Paul Adrian Glaubitz Operation of the mask aligner MJB-55 Department of Physics Faculty of Mathematics and Natural Sciences University of Oslo 1 Introduction The mask aligner is an essential tool

More information

O P E R ATING INSTRUCTIONS FOR MODEL SPR-45 Automatic Screen and Stencil Printer

O P E R ATING INSTRUCTIONS FOR MODEL SPR-45 Automatic Screen and Stencil Printer O P E R ATING INSTRUCTIONS FOR MODEL SPR-45 Automatic Screen and Stencil Printer TABLE OF CONTENTS I. SPECIFICATIONS...3. II. SAFETY INSTRUCTIONS...4. III. INSTALLATION...5. IV. SET-UP...6. V. SYSTEM OPERATION...9.

More information

FireHawk M7 Interface Module Software Instructions OPERATION AND INSTRUCTIONS

FireHawk M7 Interface Module Software Instructions OPERATION AND INSTRUCTIONS FireHawk M7 Interface Module Software Instructions OPERATION AND INSTRUCTIONS WARNING THE WARRANTIES MADE BY MSA WITH RESPECT TO THE PRODUCT ARE VOIDED IF THE PRODUCT IS NOT USED AND MAINTAINED IN ACCORDANCE

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual ARC12M Sputter Copyright 11.2015 by Hong Kong University of Science & Technology. All rights reserved. Page 1 Contents 1. Picture and Location 2. Process Capabilities 2.1 Cleanliness

More information

Plasma-Therm PECVD. Operating Characteristics. Operating Instructions. Typical Processes. I. Loading. II. Operating

Plasma-Therm PECVD. Operating Characteristics. Operating Instructions. Typical Processes. I. Loading. II. Operating Plasma-Therm PECVD A PECVD (plasma enhanced chemical vapor deposition) reacts gases in a RF (radio frequency) induced plasma to deposit materials such as silicon dioxide and silicon nitride. This PECVD

More information

SHIMADZU LC-10/20 PUMP

SHIMADZU LC-10/20 PUMP SHIMADZU LC-10/20 PUMP Clarity Control Module ENG Code/Rev.: M091/70C Date: 24.10.2017 Phone: +420 251 013 400 DataApex Ltd. Fax: +420 251 013 401 Petrzilkova 2583/13 clarity@dataapex.com 158 00 Prague

More information

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008 UNITY 2 TM Air Server Series 2 Operators Manual Version 1.0 February 2008 1. Introduction to the Air Server Accessory for UNITY 2...2 1.1. Summary of Operation...2 2. Developing a UNITY 2-Air Server method

More information

Usage Policies Notebook for Karl Suss MA6 Mid / Deep UV Mask Aligner

Usage Policies Notebook for Karl Suss MA6 Mid / Deep UV Mask Aligner Usage Policies Notebook for Karl Suss MA6 Mid / Deep UV Mask Aligner Revision date September 2014 2 Emergency Plan for Karl Suss MA6 Aligner Standard Operating Procedures for Emergencies Contact information

More information

Inventory User Guide

Inventory User Guide Inventory User Guide User Guide ~ Table of Contents ~ Sign On/Select Facility Rates & Inventory Update Tee Times Load Tee Times AutoLoad Schedule Rate Fences Dashboards Revenue At Risk, Rounds & Revenue,

More information

AKTA pure 25 New Owner s Intro

AKTA pure 25 New Owner s Intro AKTA pure 25 New Owner s Intro The exercise below will give a quick demonstration of how easy and intuitive the AKTA pure 25 will be for you in demonstrating downstream processing to your students. Steps

More information

Figure SM1: Front panel of the multipatcher software graphic user interface (GUI) at the beginning of multipatcher operation.

Figure SM1: Front panel of the multipatcher software graphic user interface (GUI) at the beginning of multipatcher operation. APPENDIX 2. Multipatcher Software Setup and Operation. The multipatcher program is organized into four panels. There are controls that allow the user to specify various parameters into the system. The

More information

Golfshot : Golf GPS Android User Guide

Golfshot : Golf GPS Android User Guide Golfshot : Golf GPS Android User Guide Version 3.5 Scorecard Version 3.5 QUICK TIPS: New in 3.5: Track Menu Track Menu Press the Golf Ball icon in the upper left corner to slide down the Track Menu. Here

More information

Nanofabrication Facility: PECVD SOP Rev. 00, April 24

Nanofabrication Facility: PECVD SOP Rev. 00, April 24 Author: Charlie Yao & Mario Beaudoin Email: charlieyao@gmail.com; Beaudoin@physics.ubc.ca Phone: 604-822-1853(MB). Purpose This document outlines the standard operation for the Trion Plasma Enhanced Chemical

More information

Arizona State University NanoFab XACTIX ETCHER. Rev A

Arizona State University NanoFab XACTIX ETCHER. Rev A Arizona State University NanoFab XACTIX ETCHER Rev A Table of Contents Contents Table of Contents... 1 1. Purpose / Scope... 2 2. Reference Documents... 2 3. Equipment / Supplies / Material... 2 4. Safety...

More information

Chromat Calibration Updated October 27th, 2017

Chromat Calibration Updated October 27th, 2017 Chromat Calibration Updated October 27th, 2017 Calibrating the Chromatograph Performing the Chromat Calibration is highly recommended when starting a new well. The MLogger already has a default calibration

More information

AX5000 Operational Manual

AX5000 Operational Manual MIYACHI AMERICA CORPORATION The World Leader in Hermetic Sealing Systems AX5000 Operational Manual 0 Document #107-00092-001 Dec, 2013 AX5000 Operational Manual Miyachi America Corporation 1820 S. Myrtle

More information

7250 Sys. Ruska Multi-Range Pressure Calibration System. GE Sensing & Inspection Technologies. Features. Applications

7250 Sys. Ruska Multi-Range Pressure Calibration System. GE Sensing & Inspection Technologies. Features. Applications GE Sensing & Inspection Technologies 7250 Sys Ruska Multi-Range Pressure Calibration System Features Fully integrated multi-range pressure test and calibration systems Select either an eight range or the

More information

ALPHA TETRA Stencil Frame System

ALPHA TETRA Stencil Frame System SM814-2 ALPHA TETRA Stencil Frame System DESCRIPTION ALPHA TETRA is a revolutionary patented precision stencil mounting system that eliminates the need for a dedicated frame for each stencil, unlike traditional

More information

Quadair. Air Hockey Table Owners Manual. Assembly operation and care instuctions. Serial # Distributed By. Sales Person. Technical Service #

Quadair. Air Hockey Table Owners Manual. Assembly operation and care instuctions. Serial # Distributed By. Sales Person. Technical Service # Version 12.1.11 Quadair Air Hockey Table Owners Manual Assembly operation and care instuctions. Serial # Distributed By Sales Person Technical Service # QUADAIR PATENTED WORLDWIDE Forward First, we would

More information

Arizona State University Center for Solid State Electronic Research. Table of Contents. Issue: C Title: Oxford Plasmalab 80plus (Floey) Page 1 of 8

Arizona State University Center for Solid State Electronic Research. Table of Contents. Issue: C Title: Oxford Plasmalab 80plus (Floey) Page 1 of 8 Title: Oxford Plasmalab 80plus (Floey) Page 1 of 8 Table of Contents 1.0 Purpose/Scope... 2 2.0 Reference Documents... 2 3.0 Equipment/Supplies/Material... 2 4.0 Safety... 2 5.0 Set Up Procedures... 2

More information

FIG: 27.1 Tool String

FIG: 27.1 Tool String Bring up Radioactive Tracer service. Click Acquisition Box - Edit - Tool String Edit the tool string as necessary to reflect the tool string being run. This is important to insure proper offsets, filters,

More information

RM-80 respiration monitor

RM-80 respiration monitor RM-80 respiration monitor User Manual September 18, 2015 0025-003M 950 North Hague Avenue Columbus, Ohio 43204-2121 USA Sales: sales@colinst.com Service: service@colinst.com Phone: (614) 276-0861 Fax:

More information

TEST REPORT BEAMSCAN WITH HALCYON

TEST REPORT BEAMSCAN WITH HALCYON TEST REPORT BEAMSCAN WITH HALCYON UZ Leuven, November 20 th, 2017 + = Ideal BEAMSCAN tank dimensions to fit perfectly into the Halcyon bore Fast tank positioning using the Halcyon lasers, no electronics

More information

SOP for Karl Suss MJB3 #1 Mask Aligner

SOP for Karl Suss MJB3 #1 Mask Aligner SOP for Karl Suss MJB3 #1 Mask Aligner Rev. 5 (30/11/2016) Safety UV Exposure: The high-energy light produced by the high-pressure Mercury Xenon lamp can cause eye damage and skin burns. Be sure that the

More information

Boyle s Law: Pressure-Volume Relationship in Gases

Boyle s Law: Pressure-Volume Relationship in Gases Boyle s Law: Pressure-Volume Relationship in Gases The primary objective of this experiment is to determine the relationship between the pressure and volume of a confined gas. The gas we will use is air,

More information

Mass Spec will not Autotune

Mass Spec will not Autotune Mass Spec will not Autotune Applies to 5973A/N MSD What could be the problem? There could be several things that would cause your Mass Spec not to Autotune. The most common, easily corrected Autotune problems

More information

[MYLAPS INTEGRATION]

[MYLAPS INTEGRATION] 2018 The Race Director MyLaps Integration Manual [MYLAPS INTEGRATION] This document explains how to manage the results data between your MyLaps readers and Race Director using manual file transfers. Contents

More information

SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA

SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA Page 1 of 9 SSI Solaris 150 RTA The Solaris 150 RTA is a rapid thermal annealing system capable of handling sample sizes up to 100mm (4 diameter) or smaller. The system can anneal in N 2 and Forming gas

More information

STS DRIE User SOP Location: NRF Cleanroon, Dry Etch Bay

STS DRIE User SOP Location: NRF Cleanroon, Dry Etch Bay Page 1 of 20 STS DRIE User SOP Location: NRF Cleanroon, Dry Etch Bay Page 2 of 20 STS DRIE SOP The STS DRIE is an Inductively Coupled Plasma Process etcher designed and configured to etch deep high aspect

More information

SMTmax Corp. AE-3090D Automatic Stencil printer. User Manual

SMTmax Corp. AE-3090D Automatic Stencil printer. User Manual SMTmax Corp. AE-3090D Automatic Stencil printer User Manual SMTmax Corporation 5675 Kimball CT, Chino California 91710 www.smtmax.com AE-3090D Index 1. Introduction 2. Specifications 3. Installation guide

More information

MM3D C-MAP by Jeppesen FAQ

MM3D C-MAP by Jeppesen FAQ MM3D C-MAP by Jeppesen FAQ Contents 1. Is NavNet 3D Compatible with C-MAP MAX (or 4D) SD-Cards?... 2 2. What is the retail price of MM3D C-MAP by Jeppesen charts?... 2 3. When will the MM3D C-MAP by Jeppesen

More information

Using the Akta Prime plus October 22, 2012

Using the Akta Prime plus October 22, 2012 Some starting precautions: 1. Vacuum filter all buffers. Removes any large particles/debris that may clog your column De-gases the buffers 2. Clarify lysates first by centrifugation and then filtration

More information

Approved by Principal Investigator Date: Approved by Super User: Date:

Approved by Principal Investigator Date: Approved by Super User: Date: Approved by Principal Investigator Date: Approved by Super User: Date: Standard Operating Procedure BNC OAI 200 Lithographic Mask Aligner (Aligner 3) Version 2011 June 2 I. Purpose This Standard Operating

More information

Warnings: Notes: Revised: October 5, 2015

Warnings: Notes: Revised: October 5, 2015 Karl Suss MA6 Mask Aligner Standard Operating Procedure Faculty Supervisor: Prof. Robert White, Mechanical Engineering (x72210) Safety Office: Peter Nowak x73246 (Just dial this directly on any campus

More information

v2.3 USER MANUAL

v2.3 USER MANUAL v2.3 USER MANUAL www.foresightsports.com Table of Contents 03 04 05 09 12 17 20 21 Activation Getting Started Play Compete Improve Settings Update Manager Glossary 04 11 05 12 03 Activation FSX Activation

More information

SPLIT HOPKINSON PRESSURE BAR OPERATIONS MANUAL 5/9/2014

SPLIT HOPKINSON PRESSURE BAR OPERATIONS MANUAL 5/9/2014 B Calibration of Split Hopkinson Pressure Bar Setup: Calibration is performed by comparing the wave speed of the initial pulse versus the reflected pulse and calculating the Energy Ratio. This ratio gives

More information

Digital Vacuum Regulator

Digital Vacuum Regulator Temperature Control for Research and Industry Digital Vacuum Regulator User s Manual Model DVR-380 INDEX SECTION PAGE SAFETY NOTICES................................................. 3 1. QUICK OPERATING

More information

COMELEC C-30-S Parylene Coating System

COMELEC C-30-S Parylene Coating System COMELEC C-30-S Parylene Coating System Comelec C-30-S Parylene deposition system Introduction Parylene is a polymere deposited at room temperature in a vacuum chamber (few µb). Parylene coating is perfectly

More information

Presents the following instructional slide show:

Presents the following instructional slide show: Presents the following instructional slide show: Operating Procedures for the Sure Shot Acoustic Fluid Level System. Sure Shot Acoustic Fluid Level System Well head attachment and accessories Complete

More information

5.1.3 Mechanical Hazards Drive assemblies have sufficient power to cause injury. Keep hands, fingers, clothing and tools clear of moving parts.

5.1.3 Mechanical Hazards Drive assemblies have sufficient power to cause injury. Keep hands, fingers, clothing and tools clear of moving parts. Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the PE4400. All users are expected to have read and understood this document. It is

More information

Air Ball! Evaluation copy

Air Ball! Evaluation copy Air Ball! Computer 24 Do you ever wonder how the National Basketball Association (NBA) decides how much air should be in the basketballs used during a game? The NBA measures the pressure inside the ball

More information

BUILT TO WIN PAINTBALL LOADER SPIRE III MANUAL

BUILT TO WIN PAINTBALL LOADER SPIRE III MANUAL BUILT TO WIN PAINTBALL LOADER SPIRE III MANUAL SPIRE III SETUP Thank you for purchasing the Virtue Spire III paintball loader. This manual covers all aspects of your Spire III loader. It is recommended

More information

Read Before Operating!

Read Before Operating! Read Before Operating! IMPORTANT OPERATING INSTRUCTIONS THE DEUCE PITCHING MACHINE THROWS REAL REGULATION BALLS; HOWEVER, THE ACCURACY OF THE DEUCE DEPENDS ON THE QUALITY, HARDNESS AND TYPE OF BALLS YOU

More information

Unaxis ICP/RIE SOP Revision 8 09/30/16 Page 1 of 5. NRF Unaxis ICP/RIE Etch SOP

Unaxis ICP/RIE SOP Revision 8 09/30/16 Page 1 of 5. NRF Unaxis ICP/RIE Etch SOP Page 1 of 5 NRF Unaxis ICP/RIE Etch SOP Unaxis Shuttlelock Reactive Ion Etcher with Inductively Coupled Plasma Module. Etch Capabilities: SiO2, Si3N4, Al, dielectrics and other commonly used materials.

More information

Old Kings Oval Electronic Scoreboard. Scoreboard Operation and Advanced Scoring Instructions

Old Kings Oval Electronic Scoreboard. Scoreboard Operation and Advanced Scoring Instructions Old Kings Oval Electronic Scoreboard Scoreboard Operation and Advanced Scoring Instructions 1 Introduction... 3 Primary Contacts... 3 Solid Scoreboards 24/7 Support 0458 LED FIX (0458 533 349)Scoring a

More information

LAMINAR FLAME SPEED ANALYSIS USING PIV (PARTICLE IMAGE VELOCIMETRY)

LAMINAR FLAME SPEED ANALYSIS USING PIV (PARTICLE IMAGE VELOCIMETRY) LAMINAR FLAME SPEED ANALYSIS USING PIV (PARTICLE IMAGE VELOCIMETRY) OPERATIONAL PROCEDURES Venue: Project Room, Hopkinson Lab, CUED Experimental date: 25 March 2008 Prepared by: Cheng Tung Chong (ctc31@camacuk)

More information

Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300

Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300 Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300 The Thermo Scientific icap 6300 Inductively Coupled Plasma Optical Emission

More information

The MRL Furnaces USED FOR THIS MANUAL COVERS

The MRL Furnaces USED FOR THIS MANUAL COVERS The MRL Furnaces USED FOR 1. Making Silicon Oxide films from 10nm to 2um thickness 2. Annealing with temperatures from 30 800C and times from 5min to 2days 3. Diffusion with temperatures from 30 1100C

More information

Met One E-BAM Particulate Monitor

Met One E-BAM Particulate Monitor STANDARD OPERATING PROCEDURES Met One E-BAM Particulate Monitor AMBIENT AIR MONITORING PROGRAM for the 130 LIBERTY STREET DECONSTRUCTION PROJECT LOWER MANHATTAN DEVELOPMENT CORPORATION 1 Liberty Plaza

More information

SPUTTER STATION STANDARD OPERATING PROCEDURE

SPUTTER STATION STANDARD OPERATING PROCEDURE SPUTTER STATION STANDARD OPERATING PROCEDURE Purpose of this Instrument: This instrument is used for deposition of thin metal or oxide films. Source materials supplied by WVU Shared Research Facilities:

More information

Standard Operating Manual

Standard Operating Manual NANOELECTRONICS FABRICATION FACILITY (NFF), HKUST Standard Operating Manual STS ICP-DRIE Copyright 2014 by Hong Kong University of Science & Technology. All rights reserved. Page 1 NANOELECTRONICS FABRICATION

More information